We explored the literature surrounding whether
allergy and hypersensitivity has a clinical basis for implant selection
in total knee arthroplasty (TKA). In error, the terms hypersensitivity
and allergy are often used synonymously. Although a relationship
is present, we could not find any evidence of implant failure due
to allergy. There is however increasing basic science that suggests
a link between loosening and metal ion production. This is not an
allergic response but is a potential problem. With a lack of evidence
logically there can be no justification to use ‘hypoallergenic’
implants in patients who have pre-existing skin sensitivity to the
metals used in TKA. Cite this article:
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium
modular revision stems and present an argument in favour of the
more selective use of modular designs. Cite this article:
The continual cycle of bone formation and resorption
is carried out by osteoblasts, osteocytes, and osteoclasts under
the direction of the bone-signaling pathway. In certain situations
the host cycle of bone repair is insufficient and requires the assistance
of bone grafts and their substitutes. The fundamental properties
of a bone graft are osteoconduction, osteoinduction, osteogenesis,
and structural support. Options for bone grafting include autogenous
and allograft bone and the various isolated or combined substitutes
of calcium sulphate, calcium phosphate, tricalcium phosphate, and
coralline hydroxyapatite. Not all bone grafts will have the same
properties. As a result, understanding the requirements of the clinical
situation and specific properties of the various types of bone grafts
is necessary to identify the ideal graft. We present a review of
the bone repair process and properties of bone grafts and their
substitutes to help guide the clinician in the decision making process. Cite this article:
The ‘jumbo’ acetabular component is now commonly
used in acetabular revision surgery where there is extensive bone
loss. It offers high surface contact, permits weight bearing over
a large area of the pelvis, the need for bone grafting is reduced
and it is usually possible to restore centre of rotation of the
hip. Disadvantages of its use include a technique in which bone
structure may not be restored, a risk of excessive posterior bone
loss during reaming, an obligation to employ screw fixation, limited
bone ingrowth with late failure and high hip centre, leading to increased
risk of dislocation. Contraindications include unaddressed pelvic
dissociation, inability to implant the component with a rim fit,
and an inability to achieve screw fixation. Use in acetabulae with
<
50% bone stock has also been questioned. Published results
have been encouraging in the first decade, with late failures predominantly because
of polyethylene wear and aseptic loosening. Dislocation is the most
common complication of jumbo acetabular revisions, with an incidence
of approximately 10%, and often mandates revision. Based on published results,
a hemispherical component with an enhanced porous coating, highly
cross-linked polyethylene, and a large femoral head appears to represent
the optimum tribology for jumbo acetabular revisions. Cite this article:
The long-term functional outcome of total hip arthroplasty (THA)
performed by trainees is not known. A multicentre retrospective
study of 879 THAs was undertaken to investigate any differences
in outcome between those performed by trainee surgeons and consultants. A total of 879 patients with a mean age of 69.5 years (37 to
94) were included in the study; 584 THAs (66.4%) were undertaken
by consultants, 138 (15.7%) by junior trainees and 148 (16.8%) by
senior trainees. Patients were scored using the Harris Hip Score
(HHS) pre-operatively and at one, three, five, seven and ten years
post-operatively. Surgical outcome, complications and survival were
compared between groups. The effect of supervision was determined
by comparing supervised and unsupervised trainees. A primary univariate
analysis was used to select variables for inclusion in multivariate
analysis. Aims
Patients and Methods
Given the increasing number of total hip arthroplasty
procedures being performed annually, it is imperative that orthopaedic
surgeons understand factors responsible for instability. In order
to treat this potentially complex problem, we recommend correctly
classifying the type of instability present based on component position, abductor
function, impingement, and polyethylene wear. Correct classification
allows the treating surgeon to choose the appropriate revision option
that ultimately will allow for the best potential outcome. Cite this article:
The goals of this study were to define the risk factors, nature,
chronology, and treatment strategies adopted for periprosthetic
femoral fractures in 32 644 primary total hip arthroplasties (THAs). There were 564 intra-operative fractures (1.7%); 529 during uncemented
stem placement (3.0%) and 35 during cemented stem placement (0.23%).
Intra-operative fractures were more common in females and patients
over 65 years (p <
0.001). The majority occurred during placement
of the femoral component (60%), and involved the calcar (69%). There
were 557 post-operative fractures (20-year probability: 3.5%; 95%
confidence interval (CI) 3.2 to 3.9); 335 fractures after placement
of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to
9.1) and 222 after placement of a cemented stem (20-year probability:
2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture
within 30 days after an uncemented stem was ten times higher than
a cemented stem. The most common post-operative fracture type was
a Vancouver AG (32%; n = 135), with 67% occurring after
a fall. In all, 36% (n = 152) were treated with revision arthroplasty. Aim and Methods
Results
The purpose of this study was to validate the diagnosis of periprosthetic
joint infection (PJI) in the Danish Hip Arthroplasty Register (DHR). We identified a cohort of patients from the DHR who had undergone
primary total hip arthroplasty (THA) since 1 January 2005 and followed
them until first-time revision, death, emigration or until 31 December
2012. Revision for PJI, as registered in the DHR, was validated against
a benchmark which included information from microbiology databases,
prescription registers, clinical biochemistry registers and clinical
records. We estimated the sensitivity, specificity, positive predictive
value (PPV) and negative predictive value (NPV) for PJI in the DHR
alone and in the DHR when combined with microbiology databases.Aims
Patients and Methods
We aimed to assess the comparability of data in joint replacement
registries and identify ways of improving the comparisons between
registries and the overall monitoring of joint replacement surgery. We conducted a review of registries that are full members of
the International Society of Arthroplasty Registries with publicly
available annual reports in English. Of the six registries which
were included, we compared the reporting of: mean age, definitions
for revision and re-operation, reasons for revision, the approach
to analysing revisions, and patient-reported outcome measures (PROMs)
for primary and revision total hip arthroplasty (THA) and hip resurfacing
arthroplasty (HRA).Aims
Materials and Methods
Bactericidal levels of antibiotics are difficult
to achieve in infected total joint arthroplasty when intravenous antibiotics
or antibiotic-loaded cement spacers are used, but intra-articular
(IA) delivery of antibiotics has been effective in several studies.
This paper describes a protocol for IA delivery of antibiotics in
infected knee arthroplasty, and summarises the results of a pharmacokinetic
study and two clinical follow-up studies of especially difficult
groups: methicillin-resistant Cite this article:
Recently, the use of metal-on-metal articulations
in total hip arthroplasty (THA) has led to an increase in adverse
events owing to local soft-tissue reactions from metal ions and
wear debris. While the majority of these implants perform well,
it has been increasingly recognised that a small proportion of patients
may develop complications secondary to systemic cobalt toxicity
when these implants fail. However, distinguishing true toxicity
from benign elevations in cobalt ion levels can be challenging. The purpose of this two part series is to review the use of cobalt
alloys in THA and to highlight the following related topics of interest:
mechanisms of cobalt ion release and their measurement, definitions
of pathological cobalt ion levels, and the pathophysiology, risk factors
and treatment of cobalt toxicity. Historically, these metal-on-metal
arthroplasties are composed of a chromium-cobalt articulation. The release of cobalt is due to the mechanical and oxidative
stresses placed on the prosthetic joint. It exerts its pathological
effects through direct cellular toxicity. This manuscript will highlight the pathophysiology of cobalt
toxicity in patients with metal-on-metal hip arthroplasties. Take home message: Patients with new or evolving hip symptoms
with a prior history of THA warrant orthopaedic surgical evaluation.
Increased awareness of the range of systemic symptoms associated
with cobalt toxicity, coupled with prompt orthopaedic intervention, may
forestall the development of further complications. Cite this article:
Infection is a leading indication for revision
arthroplasty. Established criteria used to diagnose prosthetic joint infection
(PJI) include a range of laboratory tests. Leucocyte esterase (LE)
is widely used on a colorimetric reagent strip for the diagnosis
of urinary tract infections. This inexpensive test may be used for
the diagnosis or exclusion of PJI. Aspirates from 30 total hip arthroplasties
(THAs) and 79 knee arthroplasties (KA) were analysed for LE activity. Semi-quantitative
reagent strip readings of 15, 70, 125 and 500 white blood cells
(WBC) were validated against a manual synovial white cell count
(WCC). A receiver operating characteristic (ROC) curve was constructed
to determine the optimal cut-off point for the semi-quantitative
results. Based on established criteria, six THAs and 15 KAs were
classified as infected. The optimal cut-off point for the diagnosis
of PJI was 97 WBC. The closest semi-quantitative reading for a positive
result was 125 WBC, achieving a sensitivity of 81% and a specificity
of 93%. The positive and negative predictive values of the LE test
strip were 74% and 95% respectively. The LE reagent strip had a high specificity and negative predictive
value. A negative result may exclude PJI and negate the need for
further diagnostic tests. Cite this article:
Vascular injuries during total hip arthroplasty
(THA) are rare but when they occur, have serious consequences. These have
traditionally been managed with open exploration and repair, but
more recently there has been a trend towards percutaneous endovascular
management. We performed a systematic review of the literature to assess
if this change in trend has led to an improvement in the overall
reported rates of morbidity and mortality during the last 22 years
in comparison with the reviews of the literature published previously. We found a total of 61 articles describing 138 vascular injuries
in 124 patients. Injuries because of a laceration were the most
prevalent (n = 51, 44%) and the most common presenting feature,
when recorded, was bleeding (n = 41, 53.3%). Delay in diagnosis
was associated with the type of vascular lesion (p <
0.001) and
the clinical presentation (p = 0.002). Open exploration and repair was the most common form of management,
however percutaneous endovascular intervention was used in one third
of the injuries and more constantly during the last 13 years. The main overall reported complications included death (n = 9,
7.3%), amputation (n = 2, 1.6%), and persistent ischaemia (n = 9,
7.3%). When compared with previous reviews there was a similar rate
of mortality but lower rates of amputation and permanent disability,
especially in patients managed by endovascular strategies. Cite this article:
We present the results of 62 consecutive acetabular
revisions using impaction bone grafting and a cemented polyethylene
acetabular component in 58 patients (13 men and 45 women) after
a mean follow-up of 27 years (25 to 30). All patients were prospectively
followed. The mean age at revision was 59.2 years (23 to 82). We performed Kaplan–Meier (KM) analysis and also a Competing
Risk (CR) analysis because with long-term follow-up, the presence
of a competing event (i.e. death) prevents the occurrence of the
endpoint of re-revision. A total of 48 patients (52 hips) had died or had been re-revised
at final review in March 2011. None of the deaths were related to
the surgery. The mean Harris hip score of the ten surviving hips
in ten patients was 76 points (45 to 99). The KM survivorship at 25 years for the endpoint ‘re-revision
for any reason’ was 58.0% (95% confidence interval (CI) 38 to 73)
and for ‘re-revision for aseptic loosening’ 72.1% (95% CI 51 to
85). With the CR analysis we calculated the KM analysis overestimates
the failure rate with respectively 74% and 93% for these endpoints.
The current study shows that acetabular impaction bone grafting
revisions provide good clinical results at over 25 years. Cite this article:
We report patient-reported outcomes and complications
associated with contemporary periacetabular osteotomy (PAO) surgery
in treating symptomatic acetabular dysplasia and compare these outcomes
with total hip arthroplasty (THA) in patients with similar demographic
details. Two consecutive cohorts included patients between aged
18 to 40 years who had undergone either PAO (100 hips; 24 male,
76 female) or THA (55 hips; 18 male, 37 female). At a mean follow-up
of 5.9 years (2 to 13), there was significant improvement in the
modified Harris hip pain (p <
0.001, PAO and p <
0.001, THA),
function (p <
0.001, PAO and p = 0.001, THA), and total scores
(p <
0.001, PAO and p <
0.001, THA) within each cohort. There
were no significant differences in the clinical outcome scores between
the groups. Complication rates were low and similar in each cohort
(p = 0.68). Similar to THA, contemporary PAO surgery is a clinically
effective procedure that improves function and activity levels,
provides pain relief and is associated with an acceptable complication
rate. Cite this article:
A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
The aim of our study was to describe the characteristics,
treatment, and outcomes of patients with periprosthetic joint infection
(PJI) and normal inflammatory markers after total knee arthroplasty
(TKA) and total hip arthroplasty (THA). In total 538 TKAs and 414 THAs underwent surgical treatment for
PJI and met the inclusion criteria. Pre-operative erythrocyte sedimentation
rate (ESR) and C-reactive protein level (CRP) were reviewed to identify
the seronegative cohort. An age- and gender-matched cohort was identified
from the remaining patients for comparison. Overall, 4% of confirmed
infections were seronegative (21 TKA and 17 THA). Of those who underwent
pre-operative aspiration, cultures were positive in 76% of TKAs
(n = 13) and 64% of THAs (n = 7). Cell count and differential were
suggestive of infection in 85% of TKA (n = 11) and all THA aspirates
(n = 5). The most common organism was coagulase-negative Cite this article: