Advertisement for orthosearch.org.uk
Results 261 - 280 of 384
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 958 - 962
1 Jul 2010
Wood PLR Karski MT Watmough P

We describe the early results of a prospective study of 100 total ankle replacements (96 patients) at a single centre using the Mobility Total Ankle Replacement. At final review, six patients had died and five ankles (5%) had been revised, two by fusion and three by exchange of components. All remaining patients were reviewed at a minimum of three years. The mean follow-up was 43 months (4 to 63). The three-year survival was 97% (95% confidence interval (CI) 91 to 99). The four-year survival was 93.6% (95% CI 84.7 to 97.4). The portion of bony interface that was visible on plain radiograph was divided into 15 zones and a radiolucent line or osteolytic cavity was seen in one zone in 14 ankles. It was not seen in more than one zone. In five ankles it was > 10 mm in width.

This study suggests that the early outcome of ankle replacement is comparable to that of other total joint replacements.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 405 - 408
1 Mar 2009
Monk AP Keys GW Murray DW

We describe a technique for the diagnosis of loosening of the femoral component of the Oxford Unicompartmental Knee Replacement using accurately aligned lateral radiographs in extension and flexion. If gaps are present between the component and cement on one radiograph and not on the other, the component is loose.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 881 - 886
1 Jul 2013
Bjorgul K Novicoff WN Andersen ST Ahlund OR Bunes A Wiig M Brevig K

A total of 397 hips were randomised to receive Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP) or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered polished femoral component (MS-30). There were 129 MoM hips in 123 patients (39 male and 84 female, mean age 63.3 years (40.7 to 72.9)), 137 MoP hips in 127 patients (39 male and 88 female, mean age 62.8 years (24.5 to 72.7)) and 131 CoP hips in 124 patients (51 male and 73 female, mean age 63.9 years (30.6 to 73.8)). All acetabular components were cemented Weber polyethylene components with the appropriate inlay for the MoM articulation. Clinical evaluation was undertaken using the Harris hip score (HHS) and radiological assessments were made at two, five and seven years. The HHS and radiological analysis were available for 341 hips after seven years. The MoM group had the lowest mean HHS (p = 0.124), a higher rate of revision (p < 0.001) and a higher incidence of radiolucent lines in unrevised hips (p < 0.001). In all, 12 revisions had been performed in 12 patients: eight in the MoM group (four for infection, four for aseptic loosening, three in the MoP group (one each of infection, dislocation and pain) and one in the CoP group (infection).

Our findings reveal no advantage to the MoM bearing and identified a higher revision rate and a greater incidence of radiolucent lines than with the other articulations. We recommend that patients with a 28 mm Metasul MoM bearing be followed carefully.

Cite this article: Bone Joint J 2013;95-B:881–6.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1178 - 1183
1 Sep 2013
Li H Mao Y Oni JK Dai K Zhu Z

In developmental dysplasia of the hip (DDH), a bone defect is often observed superior to the acetabulum after the reconstruction at the level of the true acetabulum during total hip replacement (THR). However, the essential amount of uncemented acetabular component coverage required for a satisfactory outcome remains controversial. The purpose of this study was to assess the stability and function of acetabular components with a lack of coverage > 30% (31% to 50%). A total of 760 DDH patients underwent THR with acetabular reconstruction at the level of the true floor. Lack of coverage above the acetabular component of > 30% occurred in 56 patients. Intra-operatively, autogenous morcellised bone grafts were used to fill the uncovered portion. Other than two screws inserted through the acetabular shell, no additional structural supports were used in these hips. In all, four patients were lost to follow-up. Therefore, 52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1 years (42 to 78) were available for this study at a mean of 4.8 years (3 to 7). There were no instances of prosthesis revision or marked loosening during the follow-up. The Harris hip score improved from a mean of 40.7 points (sd 12.2) pre-operatively to 91.1 (sd 5.0) at the last follow-up. Radiological analysis with medical imaging software allowed us to calculate the extent of the uncoverage in terms of the uncovered arc of the implant as viewed on the anteroposterior pelvic radiograph. From this we propose that up to 17 mm of lateral undercoverage in the presence of a stable initial implantation in the presence of bone autografting, with an inclination angle of the acetabular component between 40° and 55°, is acceptable. This represents undercoverage of ≤ 50%.

Cite this article: Bone Joint J 2013;95-B:1178–83.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1286 - 1292
1 Oct 2006
McLaughlin JR Lee KR

We studied a consecutive series of 285 uncemented total hip replacements in 260 patients using the Taperloc femoral component and the T-Tap acetabular component. The outcome of every hip was determined in both living and deceased patients. A complete clinical and radiological follow-up was obtained for 209 hips in 188 living patients, followed for a mean of 14.5 years (10 to 18.9). They were divided into two groups, obese and non-obese, as determined by their body mass index. There were 100 total hip replacements in 89 patients in the obese cohort (body mass index ≥ 30 kg/m2), and 109 in 99 non-obese (body mass index < 30 kg/m2) patients. A subgroup analysis of 31 patients of normal weight (body mass index 20 kg/m2 to 25 kg/m2) (33 hips) and 26 morbidly obese patients (body mass index ≥ 35 kg/m2) (30 hips) was also carried out.

In the obese group five femoral components (5%) were revised and one (1%) was loose by radiological criteria. Femoral cortical osteolysis was seen in eight hips (8%). The acetabular component was revised in 57 hips (57%) and a further 17 (17%) were loose. The mean Harris hip score improved from 52 (30 to 66) pre-operatively to 89 (49 to 100) at final follow-up. Peri-operative complications occurred in seven patients (7%).

In the non-obese group six (6%) femoral components were revised and one (1%) was loose. Femoral cortical osteolysis occurred in six hips (6%). The acetabular component was revised in 72 hips (66%) and a further 18 (17%) were loose. The mean Harris hip score increased from 53 (25 to 73) prior to surgery to 89 (53 to 100) at the time of each patient’s final follow-up radiograph.

No statistically significant difference was identified between the obese and non-obese patients with regards to clinical and radiological outcome or complications. The subgroup analysis of patients of normal weight and those who were morbidly obese showed no statistically significant difference in the rate of revision of either component.

Our findings suggest there is no evidence to support withholding total hip replacement from obese patients with arthritic hips on the grounds that their outcome will be less satisfactory than those who are not obese.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 922 - 928
1 Jul 2010
Bose VC Baruah BD

We performed 96 Birmingham resurfacing arthroplasties of the hip in 71 consecutive patients with avascular necrosis of the femoral head. A modified neck-capsule-preserving approach was used which is described in detail. The University of California, Los Angeles outcome score, the radiological parameters and survival rates were assessed. The mean follow-up was for 5.4 years (4.0 to 8.1). All the patients remained active with a mean University of California, Los Angeles activity score of 6.86 (6 to 9). Three hips failed, giving a cumulative survival rate of 95.4%. With failure of the femoral component as the endpoint, the cumulative survival rate was 98.0%. We also describe the combined abduction-valgus angle of the bearing couple, which is the sum of the inclination angle of the acetabular component and the stem-shaft angle, as an index of the optimum positioning of the components in the coronal plane.

Using a modified surgical technique, it is possible to preserve the femoral head in avascular necrosis by performing hip resurfacing in patients with good results.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1584 - 1590
1 Dec 2006
Hook S Moulder E Yates PJ Burston BJ Whitley E Bannister GC

We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up.

The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips.

All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure.

Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p < 0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p < 0.001).

The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis.

Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1210 - 1215
1 Sep 2005
Pospischill M Knahr K

We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1).

The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant were found. One patient needed revision surgery because of a late deep infection. In 11 hips (10.7%), the reason for revision was progressive wear of the polyethylene liner. Exchange of the acetabular component because of aseptic loosening without detectable liner wear was carried out in three hips (2.9%).

After 15 years the survivorship with aseptic loosening as the definition for failure was 95.6% for the acetabular component and 100% for the femoral component.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1500 - 1507
1 Nov 2013
Zaidi R Cro S Gurusamy K Sivanadarajah N Macgregor A Henricson A Goldberg A

We performed a systematic review and meta-analysis of modern total ankle replacements (TARs) to determine the survivorship, outcome, complications, radiological findings and range of movement, in patients with end-stage osteoarthritis (OA) of the ankle who undergo this procedure. We used the methodology of the Cochrane Collaboration, which uses risk of bias profiling to assess the quality of papers in favour of a domain-based approach. Continuous outcome scores were pooled across studies using the generic inverse variance method and the random-effects model was used to incorporate clinical and methodological heterogeneity. We included 58 papers (7942 TARs) with an interobserver reliability (Kappa) for selection, performance, attrition, detection and reporting bias of between 0.83 and 0.98. The overall survivorship was 89% at ten years with an annual failure rate of 1.2% (95% confidence interval (CI) 0.7 to 1.6). The mean American Orthopaedic Foot and Ankle Society score changed from 40 (95% CI 36 to 43) pre-operatively to 80 (95% CI 76 to 84) at a mean follow-up of 8.2 years (7 to 10) (p < 0.01). Radiolucencies were identified in up to 23% of TARs after a mean of 4.4 years (2.3 to 9.6). The mean total range of movement improved from 23° (95% CI 19 to 26) to 34° (95% CI 26 to 41) (p = 0.01).

Our study demonstrates that TAR has a positive impact on patients’ lives, with benefits lasting ten years, as judged by improvement in pain and function, as well as improved gait and increased range of movement. However, the quality of evidence is weak and fraught with biases and high quality randomised controlled trials are required to compare TAR with other forms of treatment such as fusion.

Cite this article: Bone Joint J 2013;95-B:1500–7.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 391 - 395
1 Mar 2013
Klatte TO Junghans K Al-Khateeb H Rueger JM Gehrke T Kendoff D Neumann J

There have been only a few small studies of patients with an infected shoulder replacement treated with a single-stage exchange procedure. We retrospectively reviewed 35 patients (19 men and 16 women) with a peri-prosthetic infection of the shoulder who were treated in this way. A total of 26 were available for clinical examination; three had died, two were lost to follow-up and four patients had undergone revision surgery. The mean follow-up time was 4.7 years (1.1 to 13.25), with an infection-free survival of 94%.

The organisms most commonly isolated intra-operatively were Staphylococcus epidermidis and Propionibacterium acnes; two patients developed a recurrent infection. Three different types of prosthesis were used: a hemiarthroplasty, a hemiarthroplasty with a bipolar head and reverse prosthesis. The mean Constant-Murley score at final follow-up was 43.3 (14 to 90) for patients with a hemiarthroplasty, 56 (40 to 88) for those with a hemiarthroplasty with a bipolar head and 61 (7 to 90) for those with a reverse prosthesis. The mean hospital stay was 10.6 days (5 to 29).

Single-stage exchange is a successful and practical treatment for patients with peri-prosthetic infection of the shoulder.

Cite this article: Bone Joint J 2013;95-B:391–5.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1540 - 1547
1 Nov 2010
Kim BS Knupp M Zwicky L Lee JW Hintermann B

We report the clinical and radiological outcome of total ankle replacement performed in conjunction with hindfoot fusion or in isolation. Between May 2003 and June 2008, 60 ankles were treated with total ankle replacement with either subtalar or triple fusion, and the results were compared with a control group of 288 ankles treated with total ankle replacement alone.

After the mean follow-up of 39.5 months (12 to 73), the ankles with hindfoot fusion showed significant improvement in the mean visual analogue score for pain (p < 0.001), the mean American Orthopaedic Foot and Ankle Society score (p < 0.001), and the mean of a modified version of this score (p < 0.001). The mean visual analogue pain score (p = 0.304) and mean modified American Orthopaedic Foot and Ankle Society score (p = 0.119) were not significantly different between the hindfoot fusion and the control groups. However, the hindfoot fusion group had a significantly lower mean range of movement (p = 0.009) and a higher rate of posterior focal osteolysis (p = 0.04). Both groups showed various complications (p = 0.131) and failure occurring at a similar rate (p = 0.685).

Subtalar or triple fusion is feasible and has minimal adverse effects on ankles treated with total ankle replacement up to midterm follow-up. The clinical outcome of total ankle replacement when combined with hindfoot fusion is comparable to that of ankle replacement alone. Thus, hindfoot fusion should be performed in conjunction with total ankle replacement when indicated.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1075 - 1082
1 Aug 2013
Choi GW Kim HJ Yeo ED Song SY

In a retrospective study we compared 32 HINTEGRA total ankle replacements (TARs) and 35 Mobility TARs performed between July 2005 and May 2010, with a minimum follow-up of two years. The mean follow-up for the HINTEGRA group was 53 months (24 to 76) and for the Mobility group was 34 months (24 to 45). All procedures were performed by a single surgeon.

There was no significant difference between the two groups with regard to the mean AOFAS score, visual analogue score for pain or range of movement of the ankle at the latest follow-up. Most radiological measurements did not differ significantly between the two groups. However, the most common grade of heterotopic ossification (HO) was grade 3 in the HINTEGRA group (10 of 13 TARs, 76.9%) and grade 2 in the Mobility group (four of seven TARs, 57.1%) (p = 0.025). Although HO was more frequent in the HINTEGRA group (40.6%) than in the Mobility group (20.0%), this was not statistically significant (p = 0.065).The difference in peri-operative complications between the two groups was not significant, but intra-operative medial malleolar fractures occurred in four (11.4%) in the Mobility group; four (12.5%) in the HINTEGRA group and one TAR (2.9%) in the Mobility group failed (p = 0.185).

Cite this article: Bone Joint J 2013;95-B:1075–82.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 643 - 648
1 May 2013
Wang J Hsu C Huang C Lin P Chen W

Structural allografts may be used to manage uncontained bone defects in revision total knee replacement (TKR). However, the availability of cadaver grafts is limited in some areas of Asia. The aim of this study was to evaluate the mid-term outcome of the use of femoral head allografts for the reconstruction of uncontained defects in revision TKR, focusing on complications related to the graft.

We retrospectively reviewed 28 patients (30 TKRs) with Anderson Orthopaedic Research Institute (AORI) type 3 bone defects, who underwent revision using femoral head allografts and stemmed components. The mean number of femoral heads used was 1.7 (1 to 3). The allograft–host junctions were packed with cancellous autograft.

At a mean follow-up of 76 months (38 to 136) the mean American Knee Society knee score improved from 37.2 (17 to 60) pre-operatively to 90 (83 to 100) (p < 0.001). The mean function score improved from 26.5 (0 to 50) pre-operatively to 81 (60 to 100) (p < 0.001). All the grafts healed to the host bone. The mean time to healing of the graft was 6.6 months (4 to 16). There have been no complications of collapse of the graft, nonunion, infection or implant loosening. No revision surgery was required.

The use of femoral head allografts in conjunction with a stemmed component and autogenous bone graft in revision TKR in patients with uncontained bone defects resulted in a high rate of healing of the graft with minimal complications and a satisfactory outcome. Longer follow-up is needed to observe the evolution of the graft.

Cite this article: Bone Joint J 2013;95-B:643–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 745 - 749
1 Jun 2009
Ritter MA

The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age > 70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p < 0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age < 55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure.

We attribute the success of the AGC implant to its relatively unconstrained articular geometry and the durability of a non-modular metal-backed tibial component with compression moulded polyethylene.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 609 - 615
1 May 2013
Cadossi M Chiarello E Savarino L Tedesco G Baldini N Faldini C Giannini S

We undertook a randomised controlled trial to compare bipolar hemiarthroplasty (HA) with a novel total hip replacement (THR) comprising a polycarbonate–urethane (PCU) acetabular component coupled with a large-diameter metal femoral head for the treatment of displaced fractures of the femoral neck in elderly patients. Functional outcome, assessed with the Harris hip score (HHS) at three months and then annually after surgery, was the primary endpoint. Rates of revision and complication were secondary endpoints.

Based on a power analysis, 96 consecutive patients aged > 70 years were randomised to receive either HA (49) or a PCU-THR (47). The mean follow-up was 30.1 months (23 to 50) and 28.6 months (22 to 52) for the HA and the PCU group, respectively.

The HHS showed no statistically significant difference between the groups at every follow-up. Higher pain was recorded in the PCU group at one and two years’ follow-up (p = 0.006 and p = 0.019, respectively). In the HA group no revision was performed. In the PCU-THR group six patients underwent revision and one patient is currently awaiting re-operation. The three-year survival rate of the PCU-THR group was 0.841 (95% confidence interval 0.680 to 0.926).

Based on our findings we do not recommend the use of the PCU acetabular component as part of the treatment of patients with fractures of the femoral neck.

Cite this article: Bone Joint J 2013;95-B:609–15.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 295 - 300
1 Mar 2013
Cawley DT Kelly N McGarry JP Shannon FJ

The optimum cementing technique for the tibial component in cemented primary total knee replacement (TKR) remains controversial. The technique of cementing, the volume of cement and the penetration are largely dependent on the operator, and hence large variations can occur. Clinical, experimental and computational studies have been performed, with conflicting results. Early implant migration is an indication of loosening. Aseptic loosening is the most common cause of failure in primary TKR and is the product of several factors. Sufficient penetration of cement has been shown to increase implant stability.

This review discusses the relevant literature regarding all aspects of the cementing of the tibial component at primary TKR.

Cite this article: Bone Joint J 2013;95-B:295–300.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1464 - 1473
1 Nov 2013
Vendittoli P Rivière C Roy AG Barry J Lusignan D Lavigne M

A total of 219 hips in 192 patients aged between 18 and 65 years were randomised to 28-mm metal-on-metal uncemented total hip replacements (THRs, 107 hips) or hybrid hip resurfacing (HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3) there was no significant difference between the THR and HR groups regarding rate of revision (4.0% (4 of 99) vs 5.8% (6 of 104), p = 0.569) or re-operation rates without revision (5.1% (5 of 99) vs 2.9% (3 of 104), p = 0.428). In the THR group one recurrent dislocation, two late deep infections and one peri-prosthetic fracture required revision, whereas in the HR group five patients underwent revision for femoral head loosening and one for adverse reaction to metal debris. The mean University of California, Los Angeles activity scores were significantly higher in HR (7.5 (sd 1.7) vs 6.9 (sd 1.7), p = 0.035), but similar mean Western Ontario and McMaster Universities Osteoarthritis Index scores were obtained (5.8 (sd 9.5) in HR vs 5.1 (sd 8.9) in THR, p = 0.615) at the last follow-up. Osteolysis was found in 30 of 81 THR patients (37.4%), mostly in the proximal femur, compared with two of 83 HR patients (2.4%) (p < 0.001). At five years the mean metal ion levels were < 2.5 μg/l for cobalt and chromium in both groups; only titanium was significantly higher in the HR group (p = 0.001). Although revision rates and functional scores were similar in both groups at mid-term, long-term survival analysis is necessary to determine whether one procedure is more advantageous than the other.

Cite this article: Bone Joint J 2013;95-B:1464–73.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 333 - 338
1 Mar 2013
Jack CM Molloy DO Walter WL Zicat BA Walter WK

The practice of removing a well-fixed cementless femoral component is associated with high morbidity. Ceramic bearing couples are low wearing and their use minimises the risk of subsequent further revision due to the production of wear debris. A total of 165 revision hip replacements were performed, in which a polyethylene-lined acetabular component was revised to a new acetabular component with a ceramic liner, while retaining the well-fixed femoral component. A titanium sleeve was placed over the used femoral trunnion, to which a ceramic head was added. There were 100 alumina and 65 Delta bearing couples inserted.

The mean Harris hip score improved significantly from 71.3 (9.0 to 100.0) pre-operatively to 91.0 (41.0 to 100.0) at a mean follow up of 4.8 years (2.1 to 12.5) (p < 0.001). No patients reported squeaking of the hip.

There were two fractures of the ceramic head, both in alumina bearings. No liners were seen to fracture. No fractures were observed in components made of Delta ceramic. At 8.3 years post-operatively the survival with any cause of failure as the endpoint was 96.6% (95% confidence interval (CI) 85.7 to 99.3) for the acetabular component and 94.0% (95% CI 82.1 to 98.4) for the femoral component.

The technique of revising the acetabular component in the presence of a well-fixed femoral component with a ceramic head placed on a titanium sleeve over the used trunnion is a useful adjunct in revision hip practice. The use of Delta ceramic is recommended.

Cite this article: Bone Joint J 2013;95-B:333–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 324 - 329
1 Mar 2008
Takeda M Higuchi H Kimura M Kobayashi Y Terauchi M Takagishi K

We prospectively examined the physical and imaging findings, including MRI, of 23 patients with spontaneous osteonecrosis of the knee after obtaining informed consent to acquire tissue specimens at surgery. There were four men and 19 women, with a mean age of 67.5 years (58 to 77). Plain radiographs were designated as stages 1, 2, 3 or 4 according to the classification of Koshino. Five knees were classified as stage 1, five as stage 2, seven as stage 3 and six as stage 4. The histological specimens were stained with haematoxylin and eosin and tetrachrome.

In the early stages of the condition, a subchondral fracture was noted in the absence of any features of osteonecrosis, whereas in advanced stages, osteonecrotic lesions were confined to the area distal to the site of the fracture which showed impaired healing. In such cases, formation of cartilage and fibrous tissue, occurred indicating delayed or nonunion. These findings strongly suggest that the histopathology at each stage of spontaneous osteonecrosis is characterised by different types of repair reaction for subchondral fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1311 - 1316
1 Oct 2008
Kim Y Kim J

The purpose of this study was to determine objectively the outcome of total knee replacement in patients with ankylosed knees.

There were 82 patients (99 knees) with ankylosed knees who underwent total knee replacement with a condylar constrained or a posterior stabilised prosthesis. Their mean age was 41.9 years (23 to 60) and the mean follow-up was for 8.9 years (6.6 to 14). Pre- and post-operative data included the Hospital for Special Surgery (HSS), the Knee Society (KS) and the Western Ontario and McMaster University Osteoarthritis index (WOMAC) scores.

The mean HSS, KS and WOMAC scores improved from 60, 53, and 79 pre-operatively to 81, 85, and 37 at follow-up. These improvements were statistically significant (p = 0.018, 0.001 and 0.014 respectively). The mean physical, social and emotional WOMAC scores also improved significantly (p = 0.032, p = 0.023 and p < 0.001 respectively). The mean satisfaction score was 8.5 (sd 1.5).

Total knee replacement gives good mid-term results in patients with ankylosed knees.