In order to address acetabular defects, porous metal revision acetabular components and augments have been developed, which require fixation to each other. The fixation technique that results in the smallest relative movement between the components, as well as its influence on the primary stability with the host bone, have not previously been determined. A total of 18 composite hemipelvises with a Paprosky IIB defect were implanted using a porous titanium 56 mm multihole acetabular component and 1 cm augment. Each acetabular component and augment was affixed to the bone using two screws, while the method of fixation between the acetabular component and augment varied for the three groups of six hemipelvises: group S, screw fixation only; group SC, screw plus cement fixation; group C, cement fixation only. The implanted hemipelvises were cyclically loaded to three different loading maxima (0.5 kN, 0.9 kN, and 1.8 kN).Objectives
Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
Periprosthetic joint infection (PJI) is a serious complication
of total hip arthroplasty (THA). Different bearing surface materials
have different surface properties and it has been suggested that
the choice of bearing surface may influence the risk of PJI after
THA. The objective of this meta-analysis was to compare the rate
of PJI between metal-on-polyethylene (MoP), ceramic-on-polyethylene
(CoP), and ceramic-on-ceramic (CoC) bearings. Electronic databases (Medline, Embase, Cochrane library, Web
of Science, and Cumulative Index of Nursing and Allied Health Literature)
were searched for comparative randomized and observational studies
that reported the incidence of PJI for different bearing surfaces.
Two investigators independently reviewed studies for eligibility, evaluated
risk of bias, and performed data extraction. Meta-analysis was performed
using the Mantel–Haenzel method and random-effects model in accordance
with methods of the Cochrane group.Aims
Patients and Methods
Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.Objectives
Methods
The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding.Aims
Patients and Methods
The cytotoxicity induced by cobalt ions (Co2+) and cobalt nanoparticles (Co-NPs) which released following the insertion of a total hip prosthesis, has been reported. However, little is known about the underlying mechanisms. In this study, we investigate the toxic effect of Co2+ and Co-NPs on liver cells, and explain further the potential mechanisms. Co-NPs were characterised for size, shape, elemental analysis, and hydrodynamic diameter, and were assessed by Transmission Electron Microscope, Scanning Electron Microscope, Energy Dispersive X-ray Spectroscopy and Dynamic Light Scattering. BRL-3A cells were used in this study. Cytotoxicity was evaluated by MTT and lactate dehydrogenase release assay. In order to clarify the potential mechanisms, reactive oxygen species, Bax/Bcl-2 mRNA expression, IL-8 mRNA expression and DNA damage were assessed on BRL-3A cells after Co2+ or Co-NPs treatment.Objectives
Methods
Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding ceramic fracture.
This study aims to report the risk of revision for fracture of modern
CoC bearings and identify factors that might influence this risk,
using data from the National Joint Registry (NJR) for England, Wales, Northern
Ireland and the Isle of Man. We analysed data on 223 362 bearings from 111 681 primary CoC
THAs and 182 linked revisions for bearing fracture recorded in the
NJR. We used implant codes to identify ceramic bearing composition
and generated Kaplan-Meier estimates for implant survivorship. Logistic
regression analyses were performed for implant size and patient specific
variables to determine any associated risks for revision.Aims
Patients and Methods
A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS
ion studies performed more than one year after surgery. The mean
age of the patients at the time of surgery was 51.7 years (29 to
70), with 38 women (68%) and 18 men (32%).Aims
Patients and Methods
Non-traumatic osteonecrosis of the femoral head
is a potentially devastating condition, the prevalence of which
is increasing. Many joint-preserving forms of treatment, both medical
and surgical, have been developed in an attempt to slow or reverse
its progression, as it usually affects young patients. However, it is important to evaluate the best evidence that is
available for the many forms of treatment considering the variation
in the demographics of the patients, the methodology and the outcomes
in the studies that have been published, so that it can be used
effectively. The purpose of this review, therefore, was to provide an up-to-date,
evidence-based guide to the management, both non-operative and operative,
of non-traumatic osteonecrosis of the femoral head. Cite this article:
Modular or custom-made femoral components have been preferred
for total hip arthroplasty (THA) in patients with a history of Perthes’
disease because of the distortion in the anatomy of the proximal
femur. However, it has not been established whether a monobloc cementless
stem will fit the distorted proximal femur or whether the results
of the procedure are satisfactory in this group of patients. We reviewed 68 consecutive patients who had undergone THA for
childhood Perthes’ disease between June 2003 and December 2008.
There were 35 men and 33 women with a mean age of 48 years (16 to
73) at the time of index arthroplasty. Their mean body mass index
was 24.4 (18.3 to 32.9). Of the 68 hips, 32 were classified as Stulberg
class III and 36 as class IV. The mean pre-operative shortening
of the affected leg was 17.2 mm (5 to 34). The minimum follow-up
was five years (mean 8.5 years; 5.2 to 10).Aims
Patients and Methods
The leading indication for revision total hip
arthroplasty (THA) remains aseptic loosening owing to wear. The younger,
more active patients currently undergoing THA present unprecedented
demands on the bearings. Ceramic-on-ceramic (CoC) bearings have
consistently shown the lowest rates of wear. The recent advances,
especially involving alumina/zirconia composite ceramic, have led
to substantial improvements and good results Alumina/zirconia composite ceramics are extremely hard, scratch
resistant and biocompatible. They offer a low co-efficient of friction
and superior lubrication and lower rates of wear compared with other
bearings. The major disadvantage is the risk of fracture of the
ceramic. The new composite ceramic has reduced the risk of fracture
of the femoral head to 0.002%. The risk of fracture of the liner
is slightly higher (0.02%). Assuming that the components are introduced without impingement,
CoC bearings have major advantages over other bearings. Owing to
the superior hardness, they produce less third body wear and are
less vulnerable to intra-operative damage. The improved tribology means that CoC bearings are an excellent
choice for young, active patients requiring THA. Cite this article: