Advertisement for orthosearch.org.uk
Results 241 - 254 of 254
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 103 - 111
1 Nov 2012
Vince KG

Seven stiff total knee arthroplasties are presented to illustrate the roles of: 1) manipulation under general anesthesia; 2) multiple concurrent diagnoses in addition to stiffness; 3) extra-articular pathology; 4) pain as part of the stiffness triad (pain and limits to flexion or extension); 5) component internal rotation; 6) multifactorial etiology; and 7) surgical exposure in this challenging clinical problem.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1339 - 1343
1 Oct 2014
Hamilton DF Burnett R Patton JT Howie CR Simpson AHRW

Instability is the reason for revision of a primary total knee replacement (TKR) in 20% of patients. To date, the diagnosis of instability has been based on the patient’s symptoms and a subjective clinical assessment. We assessed whether a measured standardised forced leg extension could be used to quantify instability.

A total of 25 patients (11 male/14 female, mean age 70 years; 49 to 85) who were to undergo a revision TKR for instability of a primary implant were assessed with a Nottingham rig pre-operatively and then at six and 26 weeks post-operatively. Output was quantified (in revolutions per minute (rpm)) by accelerating a stationary flywheel. A control group of 183 patients (71 male/112 female, mean age 69 years) who had undergone primary TKR were evaluated for comparison.

Pre-operatively, all 25 patients with instability exhibited a distinctive pattern of reduction in ‘mid-push’ speed. The mean reduction was 55 rpm (sd 33.2). Post-operatively, no patient exhibited this pattern and the reduction in ‘mid-push’ speed was 0 rpm. The change between pre- and post-operative assessment was significant (p < 0.001). No patients in the control group exhibited this pattern at any of the intervals assessed. The between-groups difference was also significant (p < 0.001). This suggests that a quantitative diagnostic test to assess the unstable primary TKR could be developed.

Cite this article: Bone Joint J 2014;96-B:1339–43.


Bone & Joint 360
Vol. 3, Issue 1 | Pages 17 - 20
1 Feb 2014

The February 2014 Knee Roundup360 looks at: whether sham surgery is as good as arthroscopic meniscectomy; distraction in knee osteoarthritis; whether trans-tibial tunnel placement increases the risk of graft failure in ACL surgery; whether joint replacements prevent cardiac events; the size of the pulmonary embolism problem; tranexamic acid and knee replacement haemostasis; matching the demand for knee replacement and follow-up; predicting the length of stay after knee replacement; and popliteal artery injury in TKR.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1697 - 1702
1 Dec 2013
Maroto MD Scolaro JA Henley MB Dunbar RP

Bicondylar tibial plateau fractures result from high-energy injuries. Fractures of the tibial plateau can involve the tibial tubercle, which represents a disruption to the extensor mechanism and logically must be stabilised. The purpose of this study was to identify the incidence of an independent tibial tubercle fracture in bicondylar tibial plateau fractures, and to report management strategies and potential complications. We retrospectively reviewed a prospectively collected orthopaedic trauma database for the period January 2003 to December 2008, and identified 392 bicondylar fractures of the tibial plateau, in which 85 tibial tubercle fractures (21.6%) were identified in 84 patients. There were 60 men and 24 women in our study group, with a mean age of 45.4 years (18 to 71). In 84 fractures open reduction and internal fixation was undertaken, either with screws alone (23 patients) or with a plate and screws (61 patients). The remaining patient was treated non-operatively. In all, 52 fractures were available for clinical and radiological assessment at a mean follow-up of 58.5 weeks (24 to 94). All fractures of the tibial tubercle united, but 24 of 54 fractures (46%) required a secondary procedure for their tibial plateau fracture. Four patients reported pain arising from prominent tubercle plates and screws, which in one patient required removal. Tibial tubercle fractures occurred in over one-fifth of the bicondylar tibial plateau fractures in our series. Fixation is necessary and can be reliably performed with screws alone or with a screw and plate, which restores the extensor mechanism and facilitates early knee flexion.

Cite this article: Bone Joint J 2013;95-B:1697–1702.


Bone & Joint 360
Vol. 1, Issue 6 | Pages 12 - 14
1 Dec 2012

The December 2012 Knee Roundup360 looks at: the demand for knee replacement; a Japanese knee outcome score; smoking and TKR; coronal alignment as a determinant of outcome in TKR; fixed flexion; MRI detected knee lesions; and lateral domed Oxford unicompartmental knee replacements.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 799 - 806
1 Jun 2010
Singh JA O’Byrne MM Colligan RC Lewallen DG

Seligman’s theory of causal attribution predicts that patients with a pessimistic explanatory style will have less favourable health outcomes. We identified 702 patients who had undergone 894 primary total knee replacements between 1993 and 2005, who responded to follow-up surveys at two (n = 783 knee replacements) and/or five years (n = 443 knee replacements) and had also completed the Minnesota Multiphasic Personality Inventory long before the joint replacement (median = 16.6 and 14.5 years for two- and five-year cohorts, respectively). Scores from the Minnesota Multiphasic Personality Inventory Optimism-Pessimism scale were used to categorise patients as pessimistic (t-score > 60) or non-pessimistic (t-score ≤ 60). Multivariate logistic regression models assessing the effect of pessimistic explanatory style on pain or improvement in knee function were adjusted for gender, age, distance from the place of treatment and depression score. Pessimists reported (a) significantly more moderate or severe pain at two years with odds ratio 2.21 (95% confidence interval (CI) 1.12 to 4.35; p = 0.02), but not at five years when the odds ratio was 1.21 (95% CI 0.51 to 2.83; p = 0.67); and (b) less improvement in knee function at two years when the odds ratio was 0.53 (95% CI 0.30 to 0.96; p = 0.04), but not at five years when the odds ratio was 1.26 (95% CI 0.57 to 2.77; p = 0.57).

No significant associations with moderate or severe limitation of activity were seen at two or five years. We conclude that a pessimistic explanatory style is associated with worse pain and functional outcomes two years after total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1009 - 1015
1 Aug 2012
Scott CEH Biant LC

Stems improve the mechanical stability of tibial components in total knee replacement (TKR), but come at a cost of stress shielding along their length. Their advantages include resistance to shear, reduced tibial lift-off and increased stability by reducing micromotion. Longer stems may have disadvantages including stress shielding along the length of the stem with associated reduction in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic fracture and end-of-stem pain. These features make long stems unattractive in the primary TKR setting, but often desirable in revision surgery with bone loss and instability. In the revision scenario, stems are beneficial in order to convey structural stability to the construct and protect the reconstruction of bony defects. Cemented and uncemented long stemmed implants have different roles depending on the nature of the bone loss involved.

This review discusses the biomechanics of the design of tibial components and stems to inform the selection of the component and the technique of implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 875 - 881
1 Jul 2012
Vanhegan IS Morgan-Jones R Barrett DS Haddad FS

This review summarises the opinions and conclusions reached from a symposium on infected total knee replacement (TKR) held at the British Association of Surgery of the Knee (BASK) annual meeting in 2011. The National Joint Registry for England and Wales reported 5082 revision TKRs in 2010, of which 1157 (23%) were caused by infection. The diagnosis of infection beyond the acute post-operative stage relies on the identification of the causative organism by aspiration and analysis of material obtained at arthroscopy. Ideal treatment then involves a two-stage surgical procedure with extensive debridement and washout, followed by antibiotics. An articulating or non-articulating drug-eluting cement spacer is used prior to implantation of the revision prosthesis, guided by the serum level of inflammatory markers. The use of a single-stage revision is gaining popularity and we would advocate its use in certain patients where the causative organism is known, no sinuses are present, the patient is not immunocompromised, and there is no radiological evidence of component loosening or osteitis.

It is our opinion that single-stage revision produces high-quality reproducible results and will soon achieve the same widespread acceptance as it does in infected hip arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1596 - 1599
1 Nov 2010
Song EK Seon JK Jeong MS

We describe a patient who developed a delayed-type hypersensitivity reaction to piperacillin/tazobactam in the cement beads and a spacer inserted at revision of total replacement of the left knee. We believe that this is the first report of such a problem. Our experience suggests that a delayed-type hypersensitivity reaction should be considered when a mixture of antibiotics such as piperacillin/tazobactam has been used in the bone cement, beads or spacer and the patient develops delayed symptoms of pain or painful paraesthesiae, fever, rash and abnormal laboratory findings in the absence of infection. The diagnosis was made when identical symptoms were induced by a provocation challenge test.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1299 - 1302
1 Oct 2006
Tillman R Kalra S Grimer R Carter S Abudu A

Peri- and sub-prosthetic fractures, or pathological fractures below an existing well-fixed femoral component, with or without an ipsilateral knee replacement, present a difficult surgical challenge.

We describe a simple solution, in which a custom-made prosthesis with a cylindrical design is cemented proximally to the stem of an existing, well-fixed femoral component. This effectively treats the fracture without sacrificing the good hip. We describe five patients with a mean age of 73 years (60 to 81) and a mean follow-up of 47 months (6 to 108).

The mean overlap of the prosthesis over the femoral component was 7.5 cm (5.5 to 10). There have been no mechanical failures, no new infections and no re-operations. We suggest that in highly selected cases, in which conventional fixation is not feasible, this technique offers a durable option and avoids the morbidity of a total femoral replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 331 - 334
1 Mar 2006
Haggis P Yates P Blakeway C Fick D Morgan DAF Holt M Wood D

Compartment syndrome is a rare complication of total knee arthroplasty that requires early recognition and prompt decompression in order to prevent long-term disability. We have found only one previous case report in the literature. We present a series of seven cases from four hospitals and five surgeons. Six of the cases resulted in the loss of at least one compartment, and one resulted in amputation. Four of the cases resulted in legal action.

We suggest that important risk factors contributing to the development of this condition include complex surgery, soft-tissue compromise, previous surgery, and possibly vascular disease. Delay in the diagnosis and hence delay in decompression was common in our series, and in five cases appeared to be related to the use of a postoperative epidural infusion for pain relief. The presence of associated neurological compromise may have also been a significant factor in the delay to diagnosis in two cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 796 - 799
1 Jun 2005
Lakdawala A Todo S Scott G

We investigated the changes in surface roughness of retrieved femoral components in 18 men and four women at revision knee surgery. The mean age at revision was 68.4 years and the mean period of implantation was for 55.6 months. Eighteen implants were retrieved for aseptic loosening and four for infection. The surface changes in the articulating areas were inspected visually and the roughness (Ra) analysed with a profilometer. Parallel scratching and burnishing were the two main forms of damage. The mean Ra measurements in the articulating areas showed no statistically significant difference when compared with those in a control area on either side of the patellar groove at the apex of the femoral flange. This suggests that it is not essential to revise a well-fixed and correctly aligned femoral component where the polished surface has become burnished or bears fine parallel scratches, if the revision is conducted solely for failure of the tibial component.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 49 - 53
1 Jan 2006
Tsumara N Yoshiya S Chin T Shiba R Kohso K Doita M

We undertook a prospective, randomised study in order to evaluate the efficacy of clamping the drains after intra-articular injection of saline with 1:500 000 adrenaline compared with post-operative blood salvage in reducing blood loss in 212 total knee arthroplasties. The mean post-operative drained blood volume after drain clamping was 352.1 ml compared to 662.3 ml after blood salvage (p < 0.0001). Allogenic blood transfusion was needed in one patient in the drain group and for three in the blood salvage group. Drain clamping with intra-articular injection of saline with adrenaline is more effective than post-operative autologous blood transfusion in reducing blood loss during total knee arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 656 - 663
1 May 2005
Toms AD McClelland D Chua L de Waal Malefijt M Verdonschot N Jones RS Kuiper J

Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray.

We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement.

Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The migration of short-stemmed trays correlated inversely with the density of the impacted groups. That of impaction-grafted tibial trays was in the range reported for uncemented primary trays. Movements of short-stemmed trays without cortical support were largest and sensitive to the degree of compaction of the graft. If support of the rim was sufficient or a long stem was used, impacted morsellised bone graft achieved adequate initial stability.