Advertisement for orthosearch.org.uk
Results 241 - 260 of 1999
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1610 - 1617
1 Dec 2014
Lainiala O Eskelinen A Elo P Puolakka T Korhonen J Moilanen T

We conducted a retrospective study to assess the prevalence of adverse reactions to metal debris (ARMD) in patients operated on at our institution with metal-on-metal (MoM) total hip replacements with 36 mm heads using a Pinnacle acetabular shell. A total of 326 patients (150 males, 175 hips; 176 females, 203 hips) with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5 years (0.1 to 10.8) participating in our in-depth modern MoM follow-up programme were included in the study, which involved recording whole blood cobalt and chromium ion measurements, Oxford hip scores (OHS) and plain radiographs of the hip and targeted cross-sectional imaging. Elevated blood metal ion levels (> 5 parts per billion) were seen in 32 (16.1%) of the 199 patients who underwent unilateral replacement. At 23 months after the start of our modern MoM follow-up programme, 29 new cases of ARMD had been revealed. Hence, the nine-year survival of this cohort declined from 96% (95% CI 95 to 98) with the old surveillance routine to 86% (95% CI 82 to 90) following the new protocol. Although ARMD may not be as common in 36 mm MoM THRs as in those with larger heads, these results support the Medicines and Healthcare Products Regulatory Agency guidelines on regular reviews and further investigations, and emphasise the need for specific a follow-up programme for patients with MoM THRs.

Cite this article: Bone Joint J 2014; 96-B:1610–17.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 3 | Pages 447 - 451
1 Apr 2002
Eggli S z’Brun S Gerber C Ganz R

In this prospective, randomised study, we have compared the wear rate of cemented, acetabular polyethylene cups articulating with either a 22 mm or a 32 mm cobalt-chromium head. We evaluated 89 patients who had a total of 484 radiographs. The mean follow-up period was 71.4 months (SD 29.1). All the radiographs were digitised and electronically measured.

The linear wear rate was significantly higher during the first two years and decreased after this period to a constant value. We suggest that this is partly due to a ‘run-in’ process caused by irregularities between surfaces of the cup and head and an initial plastic deformation of the polyethylene. The mean volumetric wear was 120.3 mm3/year for the 32 mm head, which was significantly higher than the 41.5 mm3/year for the 22 mm heads. The mean linear wear rate was not significantly different. We were, however, unable to find radiological signs of osteolysis in the patients who had higher wear rates.


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 5 | Pages 794 - 796
1 Sep 1990
Richards R Evans G Egan J Shearer

We compared, under laboratory conditions, the resistance to cutting out of the AO dynamic hip screw and the Pugh sliding nail. The mean load at cut out, adjusted for bone strength, was 70% greater for the Pugh sliding nail. The reasons for this difference are discussed.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 148 - 148
1 Jan 2004
COHEN J


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 5 | Pages 631 - 635
1 Jul 2002
Norton MR Yarlagadda R Anderson GH

We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at five years. All hips which failed did so because of aseptic loosening with progressive osteolysis or radiolucencies. We therefore recommend early and regular review of all patients with this combination of implants and early revision surgery in order to avoid massive bone loss.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 1 | Pages 113 - 115
1 Jan 2002
Mohammad S Port A Montgomery RJ

We describe a 15-year-old boy with a posterior dislocation of the hip, fracture of the posterior column of the acetabulum and separation of the femoral capital epiphysis. To our knowledge no previous case in a child has been reported. Such high-energy injuries are extremely rare, and a poor outcome is expected.

We advocate early referral to a specialised tertiary centre, and the use of a modification of Delbet’s classification to reflect the complexity and displacement which may occur with this injury.


The Journal of Bone & Joint Surgery British Volume
Vol. 69-B, Issue 5 | Pages 761 - 764
1 Nov 1987
Phillips T

The purpose of this study was to establish if the Bateman prosthesis functions as a bipolar device moving primarily at the inner metal-on-polyethylene bearing as originally proposed, or as a unipolar hemiarthroplasty moving at the outer metal-on-cartilage surface as has recently been suggested. One hundred hips were examined at one year follow-up; 78 were examined again at two to four years. The replacement was performed for arthritis in 76 hips and for femoral neck fracture in 24. Movement was assessed both with and without weight-bearing. In 80% of the arthritis group the prosthesis functioned as a bipolar hip replacement with movement occurring primarily at the inner metal-on-polyethylene surface. By contrast, in 75% of the fracture group the prosthesis functioned largely as a unipolar device with movement occurring primarily at the outer metal-on-cartilage surface. In all cases examined serially the movement pattern was the same at two to four years as it has been at one year. Clearly, the action of the prosthesis depends on the condition of the acetabular cartilage.


Bone & Joint Research
Vol. 11, Issue 12 | Pages 881 - 889
1 Dec 2022
Gómez-Barrena E Padilla-Eguiluz N López-Marfil M Ruiz de la Reina R

Aims. Successful cell therapy in hip osteonecrosis (ON) may help to avoid ON progression or total hip arthroplasty (THA), but the achieved bone regeneration is unclear. The aim of this study was to evaluate amount and location of bone regeneration obtained after surgical injection of expanded autologous mesenchymal stromal cells from the bone marrow (BM-hMSCs). Methods. A total of 20 patients with small and medium-size symptomatic stage II femoral head ON treated with 140 million BM-hMSCs through percutaneous forage in the EudraCT 2012-002010-39 clinical trial were retrospectively evaluated through preoperative and postoperative (three and 12 months) MRI. Then, 3D reconstruction of the original lesion and the observed postoperative residual damage after bone regeneration were analyzed and compared per group based on treatment efficacy. Results. The mean preoperative lesion volume was 18.7% (SD 10.2%) of the femoral head. This reduced to 11.6% (SD 7.5%) after three months (p = 0.015) and 3.7% (SD 3%) after one year (p < 0.001). Bone regeneration in healed cases represented a mean 81.2% (SD 13.8%) of the initial lesion volume at one year. Non-healed cases (n = 1 stage progression; n = 3 THAs) still showed bone regeneration but this did not effectively decrease the ON volume. A lesion size under mean 10% (SD 6%) of the femoral head at three months predicted no ON stage progression at one year. Regeneration in the lateral femoral head (C2 under Japanese Investigation Committee (JCI) classification) and in the central and posterior regions of the head was predominant in cases without ON progression. Conclusion. Bone regeneration was observed in osteonecrotic femoral heads three months after expanded autologous BM-hMSC injection, and the volume and location of regeneration indicated the success of the therapy. Cite this article: Bone Joint Res 2022;11(12):881–889


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1052 - 1059
1 Oct 2023
El-Sahoury JAN Kjærgaard K Ovesen O Hofbauer C Overgaard S Ding M

Aims. The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). Methods. A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size. Results. A total of 116 patients participated, of whom 77 were male. The median ages of the vE-PE 32 mm and 36 mm groups were 65 (interquartile range (IQR) 57 to 67) and 63 years (IQR 56 to 66), respectively, and of the XLPE 32 mm and 36 mm groups were 64 (IQR 58 to 66) and 61 years (IQR 54 to 66), respectively. Mean total head penetration was significantly lower into vE-PE acetabular liner groups than into XLPE acetabular liner groups (-0.219 mm (95% confidence interval -0.348 to -0.090); p = 0.001). There were no differences in wear according to head size, acetabular component migration, or PROMs, except for UCLA. There were no cases of aseptic loosening or failures requiring revision at long-term follow-up. Conclusion. Significantly lower wear was observed in vE-PE acetabular liners than in XLPE acetabular liners. No difference in wear was observed between different head size or PROMs except for the UCLA at ten years. Cite this article: Bone Joint J 2023;105-B(10):1052–1059


Bone & Joint Open
Vol. 4, Issue 11 | Pages 839 - 845
6 Nov 2023
Callary SA Sharma DK D’Apollonio TM Campbell DG

Aims. Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA. Methods. We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point. Results. The median proximal, 2D, and 3D wear rates calculated between one and ten years were all less than 0.005 mm/year, with no patient recording a proximal wear rate of more than 0.021 mm/year. Importantly, there was no increase in the wear rate between five and ten years. Conclusion. The very low wear rate of X3 XLPE liners with larger articulations remains encouraging for the future clinical performance of this material. Cite this article: Bone Jt Open 2023;4(11):839–845


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1032 - 1038
1 Sep 2022
García-Rey E Cruz-Pardos A Saldaña L

Aims. A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Methods. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package. Results. In the Durasul group, no patients underwent revision due to loosening or showed radiological evidence of osteolysis. In the Sulene group, four patients (four hips) were revised due to femoral component loosening. The 20-year cumulative failure incidence in the presence of the competing event of death for revision surgery was 4.5% (95% confidence interval (CI) 0.8 to 13.6) in the Durasul group, and 8.9% (95% CI 2.8 to 19.5) in the Sulene group. The mean wear one year after surgery was 0.09 mm (SD 0.007) in the Durasul group and 0.24 (SD 0.015) in the Sulene group (p < 0.001). From one to 20 years after surgery, the mean total penetration was 0.32 mm (SD 0.045) in the Durasul group and 1.07 mm (SD 0.13) in the Sulene group (p < 0.001). Mean femoral head penetration at 20 years was approximately 70.0% less in the Durasul group than the Sulene group. Conclusion. The significant reduction in femoral head penetration obtained with the Durasul compared with Sulene in uncemented THA resulted in lower osteolysis and revision rates after 20 years. Cite this article: Bone Joint J 2022;104-B(9):1032–1038


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1604 - 1610
1 Oct 2021
Takaoka Y Goto K Tamura J Okuzu Y Kawai T Kuroda Y Orita K Matsuda S

Aims. We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive. Methods. We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score. Results. The mean follow-up period was 13.0 years (10.0 to 16.3). Femoral head autografting was performed on 135 hips (72.2%). One acetabular component was retrieved because of deep infection. No revision was performed for the aseptic acetabular loosening. The Kaplan-Meier survival curve for the primary and secondary endpoints were 98.2% (95% confidence interval (CI) 88.6% to 99.8%) and 79.3% (95% CI 72.8% to 84.6%), respectively. RLLs were detected in 38 hips (21.2%), at a mean of 1.7 years (1 month to 6 years) postoperatively. None of the RLLs were progressive, and the presence of RLLs did not show a significant association with the survival and clinical score. RLLs were more frequently observed in hips without femoral head autografts than in those with autografts. Conclusion. The use of HXLPE cemented acetabular components in total hip arthroplasty demonstrated excellent clinical outcomes after ten years, and no RLLs were progressive, and their presence did not affect the outcome. Femoral head autografting did not negatively impact the acetabular component survival or the appearance of RLLs. Cite this article: Bone Joint J 2021;103-B(10):1604–1610


Bone & Joint 360
Vol. 11, Issue 6 | Pages 15 - 18
1 Dec 2022

The December 2022 Hip & Pelvis Roundup. 360. looks at: Fix and replace: simultaneous fracture fixation and hip arthroplasty for acetabular fractures in older patients; Is the revision rate for femoral neck fracture lower for total hip arthroplasty than for hemiarthroplasty?; Femoral periprosthetic fractures: data from the COMPOSE cohort study; Dual-mobility cups and fracture of the femur; What’s the deal with outcomes for hip and knee arthroplasty outcomes internationally?; Osteochondral lesions of the femoral head: is costal cartilage the answer?


Bone & Joint 360
Vol. 12, Issue 2 | Pages 13 - 16
1 Apr 2023

The April 2023 Hip & Pelvis Roundup. 360. looks at: Do technical errors determine outcomes of operatively managed femoral neck fractures in younger adults?; Single-stage or two-stage revision for hip prosthetic joint infection (INFORM); Fixation better than revision in type B periprosthetic fractures of taper slip stems; Can you maximize femoral head size at the expense of liner thickness?; Plasma D-dimer for periprosthetic joint infection?; How important is in vivo oxidation?; Total hip arthroplasty for HIV patients with osteonecrosis


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 568 - 574
1 May 2023
Kobayashi H Ito N Nakai Y Katoh H Okajima K Zhang L Tsuda Y Tanaka S

Aims. The aim of this study was to report the patterns of symptoms and insufficiency fractures in patients with tumour-induced osteomalacia (TIO) to allow the early diagnosis of this rare condition. Methods. The study included 33 patients with TIO who were treated between January 2000 and June 2022. The causative tumour was detected in all patients. We investigated the symptoms and evaluated the radiological patterns of insufficiency fractures of the rib, spine, and limbs. Results. The mean age of the patients was 57 years (24 to 87), and the mean duration of pain from onset to time of presentation was 3.9 years (0.75 to 23). The primary symptoms were low back pain (ten patients), chest wall pain (eight patients), and hip pain (eight patients). There were symptoms at more sites at the time of presentation compared with that at the time of the onset of symptoms. Bone scans showed the uptake of tracer in the rib (100%), thoracic and lumbar vertebrae (83%), proximal femur (62%), distal femur (66%), and proximal tibia (72%). Plain radiographs or MRI scans identified femoral neck fractures in 14 patients, subchondral insufficiency fractures of the femoral head and knee in ten and six patients, respectively, distal femoral fractures in nine patients, and proximal tibial fractures in 12 patients. Thoracic or lumbar vertebral fractures were identified in 23 of 29 patients (79.3%) when using any imaging study, and a biconcave deformity was the most common type of fracture. Conclusion. Insufficiency fractures in patients with TIO caused spinal pain, chest wall pain, and periarticular pain in the lower limbs. Vertebral fractures tended to be biconcave deformities, and periarticular fractures of the hips and knees included subchondral insufficiency fractures and epiphyseal or metaphyseal fractures. In patients with a tumour, the presence of one or more of these symptoms and an insufficiency fracture should suggest the diagnosis of TIO. Cite this article: Bone Joint J 2023;105-B(5):568–574


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 193 - 199
1 Feb 2022
Wang Q Wang H A G Xiao T Kang P

Aims. This study aimed to use intraoperative free electromyography to examine how the placement of a retractor at different positions along the anterior acetabular wall may affect the femoral nerve during total hip arthroplasty (THA) when undertaken using the direct anterior approach (THA-DAA). Methods. Intraoperative free electromyography was performed during primary THA-DAA in 82 patients (94 hips). The highest position of the anterior acetabular wall was defined as the “12 o’clock” position (middle position) when the patient was in supine position. After exposure of the acetabulum, a retractor was sequentially placed at the ten, 11, 12, one, and two o’clock positions (right hip; from superior to inferior positions). Action potentials in the femoral nerve were monitored with each placement, and the incidence of positive reactions (defined as explosive, frequent, or continuous action potentials, indicating that the nerve was being compressed) were recorded as the primary outcome. Secondary outcomes included the incidence of positive reactions caused by removing the femoral head, and by placing a retractor during femoral exposure; and the incidence of femoral nerve palsy, as detected using manual testing of the strength of the quadriceps muscle. Results. Positive reactions were significantly less frequent when the retractor was placed at the ten (15/94; 16.0%), 11 (12/94; 12.8%), or 12 o’clock positions (19/94; 20.2%), than at the one (37/94; 39.4%) or two o’clock positions (39/94; 41.5%) (p < 0.050). Positive reactions also occurred when the femoral head was removed (28/94; 29.8%), and when a retractor was placed around the proximal femur (34/94; 36.2%) or medial femur (27/94; 28.7%) during femoral exposure. After surgery, no patient had reduced strength in the quadriceps muscle. Conclusion. Placing the anterior acetabular retractor at the one or two o’clock positions (right hip; inferior positions) during THA-DAA can increase the rate of electromyographic signal changes in the femoral nerve. Thus, placing a retractor in these positions may increased the risk of the development of a femoral nerve palsy. Cite this article: Bone Joint J 2022;104-B(2):193–199


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1815 - 1820
1 Dec 2021
Huhnstock S Wiig O Merckoll E Svenningsen S Terjesen T

Aims. The aim of this study was to assess the prognostic value of the modified three-group Stulberg classification, which is based on the sphericity of the femoral head, in patients with Perthes’ disease. Methods. A total of 88 patients were followed from the time of diagnosis until a mean follow-up of 21 years. Anteroposterior pelvic and frog-leg lateral radiographs were obtained at diagnosis and at follow-up of one, five, and 21 years. At the five- and 21-year follow-up, the femoral heads were classified using a modified three-group Stulberg classification (round, ovoid, or flat femoral head). Further radiological endpoints at long-term follow-up were osteoarthritis (OA) of the hip and the requirement for total hip arthroplasty (THA). Results. There were 71 males (81%) and 17 females. A total of 13 patients had bilateral Perthes’ disease; thus 101 hips were analyzed. At five-year follow-up, 37 hips were round, 38 ovoid, and 26 flat. At that time, 66 hips (65%) were healed and 91 (90%) were skeletally immature. At long-term follow-up, when the mean age of the patients was 28 years (24 to 34), 20 hips had an unsatisfactory outcome (seven had OA and 13 had required THA). There was a strongly significant association between the modified Stulberg classification applied atfive-year follow-up and an unsatisfactory outcome at long-term follow-up (p < 0.001). Between the five- and 21-year follow-up, 67 hips (76%) stayed in their respective modified Stulberg group, indicating a strongly significant association between the Stulberg classifications at these follow-ups (p < 0.001). Conclusion. The modified Stulberg classification is a strong predictor of long-term radiological outcome in patients with Perthes’ disease. It can be applied at the healing stage, which is usually reached five years after the diagnosis is made and before skeletal maturity. Cite this article: Bone Joint J 2021;103-B(12):1815–1820


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 134 - 141
1 Jan 2022
Cnudde PHJ Nåtman J Hailer NP Rogmark C

Aims. The aim of this study was to investigate the potentially increased risk of dislocation in patients with neurological disease who sustain a femoral neck fracture, as it is unclear whether they should undergo total hip arthroplasty (THA) or hemiarthroplasty (HA). A secondary aim was to investgate whether dual-mobility components confer a reduced risk of dislocation in these patients. Methods. We undertook a longitudinal cohort study linking the Swedish Hip Arthroplasty Register with the National Patient Register, including patients with a neurological disease presenting with a femoral neck fracture and treated with HA, a conventional THA (cTHA) with femoral head size of ≤ 32 mm, or a dual-mobility component THA (DMC-THA) between 2005 and 2014. The dislocation rate at one- and three-year revision, reoperation, and mortality rates were recorded. Cox multivariate regression models were fitted to calculate adjusted hazard ratios (HRs). Results. A total of 9,638 patients with a neurological disease who also underwent unilateral arthroplasty for a femoral neck fracture were included in the study. The one-year dislocation rate was 3.7% after HA, 8.8% after cTHA < 32 mm), 5.9% after cTHA (= 32 mm), and 2.7% after DMC-THA. A higher risk of dislocation was associated with cTHA (< 32 mm) compared with HA (HR 1.90 (95% confidence interval (CI) 1.26 to 2.86); p = 0.002). There was no difference in the risk of dislocation with DMC-THA (HR 0.68 (95% CI 0.26 to 1.84); p = 0.451) or cTHA (= 32 mm) (HR 1.54 (95% CI 0.94 to 2.51); p = 0.083). There were no differences in the rate of reoperation and revision-free survival between the different types of prosthesis and sizes of femoral head. Conclusion. Patients with a neurological disease who sustain a femoral neck fracture have similar rates of dislocation after undergoing HA or DMC-THA. Most patients with a neurological disease are not eligible for THA and should thus undergo HA, whereas those eligible for THA could benefit from a DMC-THA. Cite this article: Bone Joint J 2022;104-B(1):134–141


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 8 - 11
1 Jan 2022
Wright-Chisem J Elbuluk AM Mayman DJ Jerabek SA Sculco PK Vigdorchik JM

Dislocation following total hip arthroplasty (THA) is a well-known and potentially devastating complication. Clinicians have used many strategies in attempts to prevent dislocation since the introduction of THA. While the importance of postoperative care cannot be ignored, particular emphasis has been placed on preoperative planning in the prevention of dislocation. The strategies have progressed from more traditional approaches, including modular implants, the size of the femoral head, and augmentation of the offset, to newer concepts, including patient-specific component positioning combined with computer navigation, robotics, and the use of dual-mobility implants. As clinicians continue to pursue improved outcomes and reduced complications, these concepts will lay the foundation for future innovation in THA and ultimately improved outcomes. Cite this article: Bone Joint J 2022;104-B(1):8–11


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 359 - 367
1 Mar 2022
Deere K Matharu GS Ben-Shlomo Y Wilkinson JM Blom AW Sayers A Whitehouse MR

Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined. Results. There were 158,677 individuals (42.4%) with an implant containing CoCr. There were 47,963 deaths, 27,332 heart outcomes, 35,720 cancers, and 22,025 neurodegenerative disorders. There was no evidence of an association between patients with CoCr implants and higher rates of any of the outcomes. Conclusion. CoCr-containing THAs did not have an increased risk of all-cause mortality, or clinically meaningful heart outcomes, cancer, or neurodegenerative disorders into the second decade post-implantation. Our findings will help reassure clinicians and the increasing number of patients receiving primary THA worldwide that the use of CoCr-containing implants is not associated with significant adverse systemic effects. Cite this article: Bone Joint J 2022;104-B(3):359–367