Beaulé PE Krismer M Mayrhofer P Wanner S Le Duff M Mattesich M Stoeckl B Amstutz HC Biedermann R

Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers.

The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively.

EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 604 - 607
1 May 2005
Stallard J


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 163 - 167
1 Feb 2006
Kalteis T Handel M Bäthis H Perlick L Tingart M Grifka J

In a prospective randomised clinical study acetabular components were implanted either freehand (n = 30) or using CT-based (n = 30) or imageless navigation (n = 30). The position of the component was determined post-operatively on CT scans of the pelvis.

Following conventional freehand placement of the acetabular component, only 14 of the 30 were within the safe zone as defined by Lewinnek et al (40° inclination sd 10°; 15° anteversion sd 10°). After computer-assisted navigation 25 of 30 acetabular components (CT-based) and 28 of 30 components (imageless) were positioned within this limit (overall p < 0.001). No significant differences were observed between CT-based and imageless navigation (p = 0.23); both showed a significant reduction in variation of the position of the acetabular component compared with conventional freehand arthroplasty (p < 0.001). The duration of the operation was increased by eight minutes with imageless and by 17 minutes with CT-based navigation.

Imageless navigation proved as reliable as that using CT in positioning the acetabular component.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 290 - 294
1 Mar 2006
Anderson GA


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 163 - 166
1 Feb 2005
Loughead JM Chesney D Holland JP McCaskie AW

Hip resurfacing is being performed more frequently in the United Kingdom. The possible benefits include more accurate restoration of leg length, femoral offset and femoral anteversion than occurs after total hip arthroplasty (THA).

We compared anteroposterior radiographs from 26 patients who had undergone hybrid THA (uncemented cup/cemented stem), with 28 who had undergone Birmingham Hip Resurfacing arthroplasty (BHR). We measured the femoral offset, femoral length, acetabular offset and acetabular height with reference to the normal contralateral hip. The data were analysed by paired t-tests.

There was a significant reduction in femoral offset (p = 0.0004) and increase in length (p = 0.001) in the BHR group. In the THA group, there was a significant reduction in acetabular offset (p = 0.0003), but femoral offset and overall hip length were restored accurately. We conclude that hip resurfacing does not restore hip mechanics as accurately as THA.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 965 - 968
1 Jul 2005
Stannard JP Singhania AK Lopez-Ben RR Anderson ER Farris RC Volgas DA McGwin GR Alonso JE

We report the incidence and location of deep-vein thrombosis in 312 patients who had sustained high-energy, skeletal trauma. They were investigated using magnetic resonance venography and Duplex ultrasound.

Despite thromboprophylaxis, 36 (11.5%) developed venous thromboembolic disease with an incidence of 10% in those with non-pelvic trauma and 12.2% in the group with pelvic trauma. Of patients who developed deep-vein thrombosis, 13 of 27 in the pelvic group (48%) and only one of nine in the non-pelvic group (11%) had a definite pelvic deep-vein thrombosis. When compared with magnetic resonance venography, ultrasound had a false-negative rate of 77% in diagnosing pelvic deep-vein thrombosis. Its value in the pelvis was limited, although it was more accurate than magnetic resonance venography in diagnosing clots in the lower limbs. Additional screening may be needed to detect pelvic deep-vein thrombosis in patients with pelvic or acetabular fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 928 - 933
1 Jul 2005
Berli BJ Schäfer D Morscher EW

The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement.

We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant in situ. Nine could only be interviewed by telephone. We included 85 patients with 88 hips in the clinical and radiological follow-up examinations.

None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed.

The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1605 - 1610
1 Dec 2005
Rathur HM Boulton AJM


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 544 - 547
1 Apr 2005
Katz DA Kim Y Millis MB

We treated eight dysplastic acetabula in six skeletally mature patients with Down’s syndrome by a modified Bernese periacetabular osteotomy. The mean age at the time of surgery was 16.5 years (12.8 to 28.5). Mean length of follow-up was five years (2 to 10.4).

Pre-operatively the mean (Tönnis) acetabular angle was 28°, the centre-edge angle was −9°, and the extrusion index was 60%; post-operatively they were 3°, 37°, and 17%, respectively. Two patients with post-operative (Tönnis) acetabular angles > 10° developed subluxation post-operatively and required secondary varus derotation femoral osteotomies. Another patient developed a late labral tear which was treated arthroscopically. All eight hips remain clinically stable, and are either asymptomatic or symptomatically improved.

These results suggest that the modified Bernese periacetabular osteotomy can be used successfully in the treatment of acetabular dysplasia in patients with Down’s syndrome.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 783 - 789
1 Jun 2006
Sen C Eralp L Gunes T Erdem M Ozden VE Kocaoglu M

In this retrospective study we evaluated the method of acute shortening and distraction osteogenesis for the treatment of tibial nonunion with bone loss in 17 patients with a mean age of 36 years (10 to 58). The mean bone loss was 5.6 cm (3 to 10). In infected cases, we performed the treatment in two stages. The mean follow-up time was 43.5 months (24 to 96). The mean time in external fixation was 8.0 months (4 to 13) and the mean external fixator index was 1.4 months/cm (1.1 to 1.8). There was no recurrence of infection. The bone evaluation results were excellent in 16 patients and good in one, while functional results were excellent in 15 and good in two. The complication rate was 1.2 per patient.

We conclude that acute shortening and distraction osteogenesis is a safe, reliable and successful method for the treatment of tibial nonunion with bone loss, with a shorter period of treatment and lower rate of complication.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 790 - 795
1 Jun 2006
Ahlmann ER Menendez LR Kermani C Gotha H

We reviewed retrospectively the results in 211 consecutive patients who had undergone limb salvage for bone neoplasia with endoprosthetic reconstruction of the proximal femur (96), distal femur (78), proximal tibia (30) and total femur (7). Their mean age was 50 years (11 to 86) and the mean follow-up period was 37.3 months (1 to 204). A total of 35 (16.6%) prostheses failed. Overall, implant survival was 78% (95% confidence interval (CI) 0.29 to 0.54) at five years, 60% (95% CI 0.93 to 2.35) at ten years and 60% (95% CI 1.27 to 3.88) at 15 years. Survivorship of the limb was 97.6% (95% CI 1.73 to 3.35) at ten years. The gender, age, diagnosis and location of the tumour were not prognostic variables for failure. Modular endoprosthetic replacement in the lower limb is a durable long-term reconstructive option, with the implants generally outlasting the patient.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 90 - 94
1 Jan 2006
Ramachandran M Birch R Eastwood DM

Between 1998 and 2002, 37 neuropathies in 32 patients with a displaced supracondylar fracture of the humerus who were referred to a nerve injury unit were identified. There were 19 boys and 13 girls with a mean age of 7.9 years (3.6 to 11.3). A retrospective review of these injuries was performed. The ulnar nerve was injured in 19, the median nerve in ten and the radial nerve in eight cases. Fourteen neuropathies were noted at the initial presentation and 23 were diagnosed after treatment of the fracture. After referral, exploration of the nerve was planned for 13 patients. Surgery was later cancelled in three because of clinical recovery. Six patients underwent neurolysis alone. Excision of neuroma and nerve grafting were performed in four. At follow-up, 26 patients had an excellent, five a good and one a fair outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 10 | Pages 1434 - 1438
1 Oct 2005
Eckardt H Ding M Lind M Hansen ES Christensen KS Hvid I

The re-establishment of vascularity is an early event in fracture healing; upregulation of angiogenesis may therefore promote the formation of bone. We have investigated the capacity of vascular endothelial growth factor (VEGF) to stimulate the formation of bone in an experimental atrophic nonunion model.

Three groups of eight rabbits underwent a standard nonunion operation. This was followed by interfragmentary deposition of 100 μg VEGF, carrier alone or autograft.

After seven weeks, torsional failure tests and callus size confirmed that VEGF-treated osteotomies had united whereas the carrier-treated osteotomies failed to unite. The biomechanical properties of the groups treated with VEGF and autograft were identical. There was no difference in bone blood flow.

We considered that VEGF stimulated the formation of competent bone in an environment deprived of its normal vascularisation and osteoprogenitor cell supply. It could be used to enhance the healing of fractures predisposed to nonunion.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 1012 - 1018
1 Jul 2005
Beck M Kalhor M Leunig M Ganz R

Recently, femoroacetabular impingement has been recognised as a cause of early osteoarthritis. There are two mechanisms of impingement: 1) cam impingement caused by a non-spherical head and 2) pincer impingement caused by excessive acetabular cover. We hypothesised that both mechanisms result in different patterns of articular damage. Of 302 analysed hips only 26 had an isolated cam and 16 an isolated pincer impingement. Cam impingement caused damage to the anterosuperior acetabular cartilage with separation between the labrum and cartilage. During flexion, the cartilage was sheared off the bone by the non-spherical femoral head while the labrum remained untouched. In pincer impingement, the cartilage damage was located circumferentially and included only a narrow strip. During movement the labrum is crushed between the acetabular rim and the femoral neck causing degeneration and ossification.

Both cam and pincer impingement lead to osteoarthritis of the hip. Labral damage indicates ongoing impingement and rarely occurs alone.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 416 - 420
1 Mar 2005
Bobyn JD Hacking SA Krygier JJ Harvey EJ Little DG Tanzer M

The effect of zoledronic acid on bone ingrowth was examined in an animal model in which porous tantalum implants were placed bilaterally within the ulnae of seven dogs. Zoledronic acid in saline was administered via a single post-operative intravenous injection at a dose of 0.1 mg/kg. The ulnae were harvested six weeks after surgery. Undecalcified transverse histological sections of the implant-bone interfaces were imaged with backscattered scanning electron microscopy and the percentage of available pore space that was filled with new bone was calculated. The mean extent of bone ingrowth was 6.6% for the control implants and 12.2% for the zoledronic acid-treated implants, an absolute difference of 5.6% (95% confidence interval, 1.2 to 10.1) and a relative difference of 85% which was statistically significant. Individual islands of new bone formation within the implant pores were similar in number in both groups but were 69% larger in the zoledronic acid-treated group. The bisphosphonate zoledronic acid should be further investigated for use in accelerating or enhancing the biological fixation of implants to bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1267 - 1271
1 Sep 2005
Allami MK Jamil W Fourie B Ashton V Gregg PJ

The Department of Health and the Public Health Laboratory Service established the Nosocomial Infection National Surveillance Scheme in order to standardise the collection of information about infections acquired in hospital in the United Kingdom and provide national data with which hospitals could measure their own performance. The definition of superficial incisional infection (skin and subcutaneous tissue), set by the Center for Disease Control (CDC), should meet at least one of the defined criteria which would confirm the diagnosis and determine the need for specific treatment.

We have assessed the interobserver reliability of the criteria for superficial incisional infection set by the CDC in our current practice. The incisional site of 50 patients who had an elective primary arthroplasty of the hip or knee was evaluated independently by two orthopaedic clinical research fellows and two orthopaedic ward sisters for the presence or absence of surgical-site infection. Interobserver reliability was assessed by comparison of the criteria for wound infection used by the four observers using kappa reliability coefficients. Our study demonstrated that some of the components of the current CDC criteria were unreliable and we recommend their revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 790 - 795
1 Jun 2005
Eren A Ömeroglu H Güven M Ugutmen E Altintas F

We treated 26 hips (24 consecutive patients) with residual dysplasia by a technique of incomplete triple pelvic osteotomy. The mean age of the patients was 21.6 years.

The mean values for the pre-operative centre-edge angle of Wiberg, the refined centre-edge angle, the acetabular angle of Sharp, the modified acetabular angle and femoral head lateralisation were 7.7°, −3.1°, 49.3°, 53.2° and 17.2 mm, respectively. After a mean follow-up of 3.3 years they were 27.0°, 13.0°, 38.9°, 44.3° and 15.9 mm, respectively (p < 0.05). The osteoarthritic grading changed adversely in one hip. The mean pre-operative and latest Harris hip scores were 74.9 and 93.0, respectively (p < 0.05).

This technique provides a stable osteotomy with maintenance of the posterior column which allows early mobilisation and minimal internal fixation. The technique is not complex and requires minimal blood transfusion. The use of an image intensifier is not necessary and harvesting of a subcristal bone graft avoids post-operative complications at the donor site.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 264 - 269
1 Feb 2006
Arora A Nadkarni B Dev G Chattopadhya D Jain AK Tuli SM Kumar S

We studied 51 patients with osteo-articular tuberculosis who were divided into two groups. Group I comprised 31 newly-diagnosed patients who were given first-line antituberculous treatment consisting of isoniazid, rifampicin, ethambutol and pyrazinamide. Group II (non-responders) consisted of 20 patients with a history of clinical non-responsiveness to supervised uninterrupted antituberculous treatment for a minimum of three months or a recurrence of a previous lesion which on clinical observation had healed. No patient in either group was HIV-positive. Group II were treated with an immunomodulation regime of intradermal BCG, oral levamisole and intramuscular diphtheria and tetanus vaccines as an adjunct for eight weeks in addition to antituberculous treatment. We gave antituberculous treatment for a total of 12 to 18 months in both groups and they were followed up for a mean of 30.2 months (24 to 49). A series of 20 healthy blood donors served as a control group.

Twenty-nine (93.6%) of the 31 patients in group I and 14 of the 20 (70%) in group II had a clinicoradiological healing response to treatment by five months.

The CD4 cell count in both groups was depressed at the time of enrolment, with a greater degree of depression in the group-II patients (686 cells/mm3 (sd 261) and 545 cells/mm3 (sd 137), respectively; p < 0.05). After treatment for three months both groups showed significant elevation of the CD4 cell count, reaching a level comparable with the control group. However, the mean CD4 cell count of group II (945 cells/mm3 (sd 343)) still remained lower than that of group I (1071 cells/mm3 (sd 290)), but the difference was not significant. Our study has shown encouraging results after immunomodulation and antituberculous treatment in non-responsive patients. The pattern of change in the CD4 cell count in response to treatment may be a reliable clinical indicator.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 191 - 195
1 Feb 2005
Mighell MA Dunham RC Rommel EA Frankle MA

We present six patients with chronic dislocation of the elbow who were treated by primary semiconstrained total elbow arthroplasty. All were women with a mean age of 65 years (51 to 76), the mean interval between dislocation and surgery was 17 weeks (5 to 52) and the mean follow-up 58 months (24 to 123).

The most dramatic improvement was in function. The mean American Shoulder and Elbow Surgeon score was 5.2 times better (p < 0.001) and the mean total range of movement increased from 33° to 121° (p < 0.001) after operation. Three patients developed wear of polyethylene. One required revision for a periprosthetic fracture, and another required a bushing exchange.

Primary semiconstrained elbow arthroplasty provides significant, predictable functional improvement. Potential solutions for wear of polyethylene include a different operative technique or design of implant. Despite the high incidence of such wear, total elbow arthroplasty should be considered as a viable treatment option for chronic dislocation of the elbow in elderly patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 950 - 954
1 Jul 2005
Khatri M Stirrat AN

We present the outcome of 47 Souter-Strathclyde replacements of the elbow with a mean follow-up of 82 months (12 to 129). The clinical results were assessed using a condition-specific outcome measure. The mean total score (maximum 100) before the operation was 47.21 and improved to 79.92 (p < 0.001). The mean pain score (maximum 50) improved from 21.41 to 46.70 (p < 0.001) and the mean functional component of the score (maximum 30) from 11.19 to 18.65 (p < 0.001). There was negligible change in the score for the range of movement although a significant improvement in mean flexion from 124° to 136° was noted (p < 0.001).

Revision surgery was required in four patients, for dislocation, wound dehiscence and early infection in one, late infection in two and aseptic loosening in one. The cumulative survival was 75% at nine years for all causes of failure and 97% at ten years for aseptic loosening alone. Our study demonstrates the value of the Souter-Strathclyde total elbow arthroplasty in providing relief from pain and functional improvement in rheumatoid patients.