In this prospective study a total of 80 consecutive
Chinese patients with Crowe type I or II developmental dysplasia of
the hip were randomly assigned for hip resurfacing arthroplasty
(HRA) or total hip replacement (THR). Three patients assigned to HRA were converted to THR, and three
HRA patients and two THR patients were lost to follow-up. This left
a total of 34 patients (37 hips) who underwent HRA and 38 (39 hips)
who underwent THR. The mean follow-up was 59.4 months (52 to 70)
in the HRA group and 60.6 months (50 to 72) in the THR group. There was
no failure of the prosthesis in either group. Flexion of the hip
was significantly better after HRA, but there was no difference
in the mean post-operative Harris hip scores between the groups.
The mean size of the acetabular component in the HRA group was significantly
larger than in the THR group (49.5 mm vs 46.1 mm, p = 0.001). There was
no difference in the mean abduction angle of the acetabular component
between the two groups. Although the patients in this series had risk factors for failure
after HRA, such as low body weight, small femoral heads and dysplasia,
the clinical results of
Metal-on-metal hip resurfacing was performed for developmental dysplasia in 96 hips in 85 patients, 78 in women and 18 in men, with a mean age at the time of surgery of 43 years (14 to 65). These cases were matched for age, gender, operating surgeon and date of operation with a group of patients with primary osteoarthritis who had been treated by
We describe the experience with the first consecutive 230 Birmingham hip resurfacings at our centre. At a mean follow-up of three years (25 to 52 months) survivorship was 99.14% with revision in one patient for a loose acetabular component and one death from unrelated causes. One patient developed a fracture of the femoral neck at six weeks which united unremarkably after a period of non-weight-bearing. The Harris hip score improved from a mean of 62.54 (8 to 92) to 97.74 (61 to 100). The mean flexion improved from 91.52° (25 to 140) to 110.41° (80 to 145). Most patients (97%) considered the outcome to be good or excellent. Our preliminary experience with this implant is encouraging and the results are superior to the earlier generation of
In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.Aims
Methods
The October 2012 Hip &
Pelvis Roundup. 360. looks at: diagnosing the infected hip replacement; whether tranexamic acid has a low complication rate; the relationship between poor cementing technique and early failure of
When performing total hip replacement (THR) in high dislocated hips, the presence of soft-tissue contractures means that most surgeons prefer to use a femoral shortening osteotomy in order to avoid the risk of neurovascular damage. However, this technique will sacrifice femoral length and reduce the extent of any leg-length equalisation. We report our experience of 74 THRs performed between 2000 and 2008 in 65 patients with a high dislocated hip without a femoral shortening osteotomy. The mean age of the patients was 55 years (46 to 72) and the mean follow-up was 42 months (12 to 78). All implants were cementless except for one
We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7). The groups differed significantly from each other in Larsen grade, cementing of components and patellar
The Birmingham hip resurfacing metal-on-metal arthroplasty was introduced in 1997 and has shown promising short- to mid-term results. We used radiostereophotogrammetric analysis (RSA) to study the stability of 20
We assessed the use of woven carbon fibre pads to
In the native hip, the hip capsular ligaments tighten at the limits of range of hip motion and may provide a passive stabilizing force to protect the hip against edge loading. In this study we quantified the stabilizing force vectors generated by capsular ligaments at extreme range of motion (ROM), and examined their ability to prevent edge loading. Torque-rotation curves were obtained from nine cadaveric hips to define the rotational restraint contributions of the capsular ligaments in 36 positions. A ligament model was developed to determine the line-of-action and effective moment arms of the medial/lateral iliofemoral, ischiofemoral, and pubofemoral ligaments in all positions. The functioning ligament forces and stiffness were determined at 5 Nm rotational restraint. In each position, the contribution of engaged capsular ligaments to the joint reaction force was used to evaluate the net force vector generated by the capsule.Aims
Methods
Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), now limited to a few specialized centres. In this study, we provide the clinical results of 400 consecutive MoM HRAs implanted more than 20 years ago in one such centre. A total of 355 patients (400 hips) were treated with Conserve Plus HRA between November 1996 and November 2000. There were 96 female (27%) and 259 male patients (73%). Their mean age was 48.2 years (SD 10.9). The University of California, Los Angeles (UCLA) hip scores and 12-item Short Form Survey (SF-12) quality of life scores were reported. Survivorship was assessed using Kaplan-Meier analyses.Aims
Methods
The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up. A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.Aims
Methods
Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory pseudotumours, arising around these devices. They may cause soft-tissue destruction with severe symptoms and a poor outcome after revision surgery. The aim of this study was to determine the incidence of and risk factors for pseudotumours that are serious enough to require revision surgery. Since 1999, 1419 metal-on-metal hip resurfacings have been implanted by our group in 1224 patients; 1.8% of the patients had a revision for pseudotumour. In this series the Kaplan-Meier cumulative revision rate for pseudotumour increased progressively with time. At eight years, in all patients, it was 4% (95% confidence interval (CI) 2.2 to 5.8). Factors significantly associated with an increase in revision rate were female gender (p <
0.001), age under 40 (p = 0.003), small components (p = 0.003), and dysplasia (p = 0.019), whereas implant type was not (p = 0.156). These factors were inter-related, however, and on fitting a Cox proportional hazard model only gender (p = 0.002) and age (p = 0.024) had a significant independent influence on revision rate; size nearly reached significance (p = 0.08). Subdividing the cohort according to significant factors, we found that the revision rate for pseudotumours in men was 0.5% (95% CI 0 to 1.1) at eight years wheras in women over 40 years old it was 6% (95% CI 2.3 to 10.1) at eight years and in women under 40 years it was 13.1% at six years (95% CI 0 to 27) (p <
0.001). We recommend that
We sought to determine the degree of correlation between the condition of the patellar articular cartilage and patellofemoral symptoms and function in osteoarthritic patients undergoing total knee arthroplasty. The depth of the osteoarthritic lesion, as graded by the Outerbridge classification and its size and location were assessed to determine the condition of the patellar cartilage in 80 consecutive osteoarthritic knees undergoing total knee arthroplasty. The association between the condition of the cartilage and patellofemoral symptoms and function was investigated by correlation analysis. The depth and size of the lesion had a significant but weak correlation with anterior knee pain (r = −0.300 and −0.289; p = 0.007 and 0.009, respectively), whereas location had no significant association (p >
0.05). None had a significant association with patellofemoral functional parameters (chair-rising, stair-climbing, and quadriceps power) (p >
0.05). Our study indicates that patellofemoral symptoms and function are not completely determined by the condition of the cartilage. Caution should be taken when the symptoms and functional limitations are attributed to a lesion in the patellofemoral joint in making a decision regarding patellar
A multicentre, randomized, clinician-led, pragmatic, parallel-group orthopaedic trial of two surgical procedures was set up to obtain high-quality evidence of effectiveness. However, the trial faced recruitment challenges and struggled to maintain recruitment rates over 30%, although this is not unusual for surgical trials. We conducted a qualitative study with the aim of gathering information about recruitment practices to identify barriers to patient consent and participation to an orthopaedic trial. We collected 11 audio recordings of recruitment appointments and interviews of research team members (principal investigators and research nurses) from five hospitals involved in recruitment to an orthopaedic trial. We analyzed the qualitative data sets thematically with the aim of identifying aspects of informed consent and information provision that was either unclear, disrupted, or hindered trial recruitment.Aims
Methods
The outcome of tibial allograft reconstruction after resection of a tumour is inconsistent and has a high rate of failure. There are few reports on the use of tibial allografts in children with open growth plates. We performed 21 allograft reconstructions (16 osteoarticular, five intercalary) in 19 consecutive patients between seven and 17 years of age. Two had Ewing’s sarcoma, one an adamantinoma and 16 osteosarcoma, one with multifocal disease. Five patients have died; the other 14 were free from disease at the time of follow-up. Six surviving patients (eight allograft reconstructions) continue to have good or excellent function at a mean of 59 months (14 to 132). One patient has poor function at 31 months. The other seven patients have a good or excellent function after additional procedures including exchange of the allograft and
A 60-year-old man developed severe neuropathic pain and foot-drop in his left leg following
The aim of our study was to investigate whether placing of the femoral component of a hip resurfacing in valgus protected against spontaneous fracture of the femoral neck. We performed a hip resurfacing in 20 pairs of embalmed femora. The femoral component was implanted at the natural neck-shaft angle in the left femur and with a 10° valgus angle on the right. The bone mineral density of each femur was measured and CT was performed. Each femur was evaluated in a materials testing machine using increasing cyclical loads. In specimens with good bone quality, the 10° valgus placement of the femoral component had a protective effect against fractures of the femoral neck. An adverse effect was detected in osteoporotic specimens. When
Metal-on-metal (MOM) bearings for hip arthroplasty are increasing in popularity. Concern remains, however, regarding the potential toxicological effects of the metal ions which these bearings release. The serum levels of cobalt and chromium in 22 patients who had undergone MOM