We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine. We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The acetabular inclination angle was strongly positively correlated with the rate of wear. Femoral version was weakly negatively correlated with the rate of wear. Application of a threshold of >
5 μm/year for the rate of wear in order to separate the revisions into low or high wearing groups showed that more high wearing components were implanted outside Lewinnek’s safe zone, but that this was mainly due to the inclination of the acetabular component, which was the only parameter that significantly differed between the groups. We were unable to show that excess version of the acetabular component alone or combined with femoral version was associated with an increase in the rate of wear based on our assessment of version using CT.
There is evidence that high levels of physical
activity following arthroplasty of the hip or knee can lead to early revision.
However, the term ‘highly active’ is not well defined. A validated ankle accelerometer was used to quantify activity
in 13 patients, who had undergone a total of 20 arthroplasties of
the lower limbs and who had active lifestyles. The assessments were
taken at a mean of 8.7 years post-operatively (1.8 to 15.8). The mean gait cycles per day was 8273 (5964 to 12 557), which
extrapolates to 3.0 million cycles per year (cpy) (2.2 to 4.6).
The mean percentage of time spent in high activity mode was 4.3%,
or about one hour per day. The mean percentage of cycles in high
activity was 40%. Based on these data, we propose the following definitions of
high activity: >
3 million cpy; one hour per day in high activity
mode; 40% of cycles in high activity mode. Extrapolating the sample of activity over the time since operation,
the mean cycles per arthroplasty was 25.2 million, with a maximum
of 44.1 million. No joint has been revised, or shows evidence of
impending failure. Cite this article:
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher acetabular component inclination angles correlated
with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002).
Patients with increased ion levels had lower Harris hip scores (p
= 0.038). In conclusion, in well-functioning BHRs the metal ion levels
decreased significantly at ten years. An increase >
2.5 µg/l was
associated with poor function. Cite this article:
Metal-on-metal hip resurfacing was performed for developmental dysplasia in 96 hips in 85 patients, 78 in women and 18 in men, with a mean age at the time of surgery of 43 years (14 to 65). These cases were matched for age, gender, operating surgeon and date of operation with a group of patients with primary osteoarthritis who had been treated by resurfacing, to provide a control group of 96 hips (93 patients). A clinical and radiological follow-up study was performed. The dysplasia group were followed for a mean of 4.4 years (2.0 to 8.5) and the osteoarthritis group for a mean of 4.5 years (2.2 to 9.4). Of the dysplasia cases, 17 (18%) were classified as Crowe grade III or IV. There were five (5.2%) revisions in the dysplasia group and none in the osteoarthritic patients. Four of the failures were due to acetabular loosening and the other sustained a fracture of the neck of femur. There was a significant difference in survival between the two groups (p = 0.02). The five-year survival was 96.7% (95% confidence interval 90.0 to 100) for the dysplasia group and 100% (95% confidence interval 100 to 100) for the osteoarthritic group. There was no significant difference in the median Oxford hip score between the two groups at any time during the study. The medium-term results of metal-on-metal hip resurfacing in all grades of developmental dysplasia are encouraging, although they are significantly worse than in a group of matched patients with osteoarthritis treated in the same manner.
We have reviewed 42 patients who had revision of metal-on-metal resurfacing procedures, mostly because of problems with the acetabular component. The revisions were carried out a mean of 26.2 months (1 to 76) after the initial operation and most of the patients (30) were female. Malpositioning of the acetabular component resulted in 27 revisions, mostly because of excessive abduction (mean 69.9°; 56° to 98°) or insufficient or excessive anteversion. Seven patients had more than one reason for revision. The mean increase in the diameter of the component was 1.8 mm (0 to 4) when exchange was needed. Malpositioning of the components was associated with metallosis and a high level of serum ions. The results of revision of the femoral component to a component with a modular head were excellent, but four patients had dislocation after revision and four required a further revision.
We carried out metal artefact-reduction MRI, three-dimensional CT measurement of the position of the component and inductively-coupled plasma mass spectrometry analysis of cobalt and chromium levels in whole blood on 26 patients with unexplained pain following metal-on-metal resurfacing arthroplasty. MRI showed periprosthetic lesions around 16 hips, with 14 collections of fluid and two soft-tissue masses. The lesions were seen in both men and women and in symptomatic and asymptomatic hips. Using three-dimensional CT, the median inclination of the acetabular component was found to be 55° and its positioning was outside the Lewinnek safe zone in 13 of 16 cases. Using inductively-coupled plasma mass spectrometry, the levels of blood metal ions tended to be higher in painful compared with well-functioning metal-on-metal hips. These three clinically useful investigations can help to determine the cause of failure of the implant, predict the need for future revision and aid the choice of revision prostheses.
This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
Hips with metal-on-metal total hip arthroplasty (MoM THA) have a high rate of adverse local tissue reactions (ALTR), often associated with hypersensitivity reactions. Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) measures tissue perfusion with the parameter Ktrans (volume transfer constant of contrast agent). Our purpose was 1) to evaluate the feasibility of DCE-MRI in patients with THA and 2) to compare DCE-MRI in patients with MoM bearings with metal-on-polyethylene (MoP) bearings, hypothesising that the perfusion index Ktrans in hips with MoM THA is higher than in hips with MoP THA. In this pilot study, 16 patients with primary THA were recruited (eight MoM, eight MoP). DCE-MRI of the hip was performed at 1.5 Tesla (T). For each patient, Ktrans was computed voxel-by-voxel in all tissue lateral to the bladder. The mean Ktrans for all voxels was then calculated. These values were compared with respect to implant type and gender, and further correlated with clinical parameters.Objectives
Methods
The December 2012 Hip &
Pelvis Roundup360 looks at: swimming against the tide with resurfacing; hip impingement surgery; the relationship between obesity and co-morbidities and joint replacement infection; cemented hips; cross-linked polyethylene notching; whether cement is necessary in oncological arthroplasty; and how total hip replacement may result in weight gain.
The October 2012 Hip &
Pelvis Roundup360 looks at: diagnosing the infected hip replacement; whether tranexamic acid has a low complication rate; the relationship between poor cementing technique and early failure of resurfacing; debridement and retention for the infected replacement; triple-tapered stems and bone mineral density; how early discharge can be bad for your sleep; an updated QFracture algorithm to predict the risk of an osteoporotic fracture; and local infiltration analgesia and total hip replacement.
Hip simulators have been used for ten years to determine the tribological performance of large-head metal-on-metal devices using traditional test conditions. However, the hip simulator protocols were originally developed to test metal-on-polyethylene devices. We have used patient activity data to develop a more physiologically relevant test protocol for metal-on-metal devices. This includes stop/start motion, a more appropriate walking frequency, and alternating kinetic and kinematic profiles. There has been considerable discussion about the effect of heat treatments on the wear of metal-on-metal cobalt chromium molybdenum (CoCrMo) devices. Clinical studies have shown a higher rate of wear, levels of metal ions and rates of failure for the heat-treated metal compared to the as-cast metal CoCrMo devices. However, hip simulator studies in vitro under traditional testing conditions have thus far not been able to demonstrate a difference between the wear performance of these implants. Using a physiologically relevant test protocol, we have shown that heat treatment of metal-on-metal CoCrMo devices adversely affects their wear performance and generates significantly higher wear rates and levels of metal ions than in as-cast metal implants.
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.
Surface hip replacement (SHR) is generally used
in younger, active patients as an alternative conventional total
hip replacement in part because of the ability to preserve femoral
bone. This major benefit of surface replacement will only hold true
if revision procedures of SHRs are found to provide good clinical
results. A retrospective review of SHR revisions between 2007 and 2012
was presented, and the type of revision and aetiologies were recorded.
There were 55 SHR revisions, of which 27 were in women. At a mean
follow-up of 2.3 years (0.72 to 6.4), the mean post-operative Harris
hip score (HHS) was 94.8 (66 to 100). Overall 23 were revised for mechanical
reasons, nine for impingement, 13 for metallosis, nine for unexplained
pain and one for sepsis. Of the type of revision surgery performed,
14 were femoral-only revisions; four were acetabular-only revisions,
and 37 were complete revisions. We did not find that clinical scores were significantly different
between gender or different types of revisions. However, the mean
post-operative HHS was significantly lower in patients revised for
unexplained pain compared with patients revised for mechanical reasons
(86.9 (66 to 100) Based on the overall clinical results, we believe that revision
of SHR can have good or excellent results and warrants a continued
use of the procedure in selected patients. Close monitoring of these
patients facilitates early intervention, as we believe that tissue
damage may be related to the duration of an ongoing problem. There
should be a low threshold to revise a surface replacement if there
is component malposition, rising metal ion levels, or evidence of
soft-tissue abnormalities. Cite this article:
Recently, the use of metal-on-metal articulations
in total hip arthroplasty (THA) has led to an increase in adverse
events owing to local soft-tissue reactions from metal ions and
wear debris. While the majority of these implants perform well,
it has been increasingly recognised that a small proportion of patients
may develop complications secondary to systemic cobalt toxicity
when these implants fail. However, distinguishing true toxicity
from benign elevations in cobalt ion levels can be challenging. The purpose of this two part series is to review the use of cobalt
alloys in THA and to highlight the following related topics of interest:
mechanisms of cobalt ion release and their measurement, definitions
of pathological cobalt ion levels, and the pathophysiology, risk factors
and treatment of cobalt toxicity. Historically, these metal-on-metal
arthroplasties are composed of a chromium-cobalt articulation. The release of cobalt is due to the mechanical and oxidative
stresses placed on the prosthetic joint. It exerts its pathological
effects through direct cellular toxicity. This manuscript will highlight the pathophysiology of cobalt
toxicity in patients with metal-on-metal hip arthroplasties. Take home message: Patients with new or evolving hip symptoms
with a prior history of THA warrant orthopaedic surgical evaluation.
Increased awareness of the range of systemic symptoms associated
with cobalt toxicity, coupled with prompt orthopaedic intervention, may
forestall the development of further complications. Cite this article:
T-cells are considered to play an important role in the inflammatory response causing arthroplasty failure. The study objectives were to investigate the composition and distribution of CD4+ T-cell phenotypes in the peripheral blood (PB) and synovial fluid (SF) of patients undergoing revision surgery for failed metal-on-metal (MoM) and metal-on-polyethylene (MoP) hip arthroplasties, and in patients awaiting total hip arthroplasty. In this prospective case-control study, PB and SF were obtained from 22 patients (23 hips) undergoing revision of MoM (n = 14) and MoP (n = 9) hip arthroplasties, with eight controls provided from primary hip osteoarthritis cases awaiting arthroplasty. Lymphocyte subtypes in samples were analysed using flow cytometry.Objectives
Methods
The August 2012 Hip &
Pelvis Roundup360 looks at: whether cemented hip replacement might be bad for your health; highly cross-linked polyethylene; iHOT-33 - a new hip outcome measure; hamstring injuries; total hip replacement; stemmed metal-on-metal THR; bipolar hemiarthroplasty, neuromuscular disease and dislocation; the high risk of secondary hemiarthroplasty; and whether we have to repair the labrum after all?
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
The aim of this cadaver study was to identify
the change in position of the sciatic nerve during arthroplasty
using the posterior surgical approach to the hip. We investigated
the position of the nerve during this procedure by dissecting 11
formalin-treated cadavers (22 hips: 12 male, ten female). The distance
between the sciatic nerve and the femoral neck was measured before
and after dislocation of the hip, and in positions used during the
preparation of the femur. The nerve moves closer to the femoral
neck when the hip is flexed to >
30° and internally rotated to 90° (90°
IR). The mean distance between the nerve and femoral neck was 43.1
mm (standard deviation ( This study demonstrates that the sciatic nerve becomes closer
to the operative field during hip arthroplasty using the posterior
approach with progressive flexion of the hip. Cite this article:
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.