We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the risk of revision was related
to patient, surgeon and implant covariates. The overall rate of
revision at five years was 2.4% (99% confidence interval 2.02 to
2.79). In the final adjusted model, we found that the risk of revision
was significantly higher in patients receiving metal-on-metal (MoM:
hazard ratio (HR) 1.93, p <
0.001) and ceramic-on-ceramic bearings
(CoC: HR 1.55, p = 0.003) compared with the best performing bearing
(metal-on-polyethylene). The risk of revision was also greater for
smaller femoral stems (sizes 8 to 10: HR 1.82, p <
0.001) compared
with mid-range sizes. In a secondary analysis of only patients where body
mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly
increased the risk of revision (HR 1.55, p = 0.002). The influence
of the bearing on the risk of revision remained significant (MoM:
HR 2.19, p <
0.001; CoC: HR 2.09,
p = 0.001). The risk of revision was independent of age, gender,
head size and offset, shell, liner and stem type, and surgeon characteristics. We found significant differences in failure between bearing surfaces
and femoral stem size after adjustment for a range of covariates
in a large cohort of single-brand cementless THRs. In this study
of procedures performed since 2003, hard bearings had significantly
higher rates of revision, but we found no evidence that head size
had an effect. Patient characteristics, such as BMI and American
Society of Anesthesiologists grade, also influence the survival
of cementless components. Cite this article:
This was a retrospective analysis of the medium-
to long-term results of 46 TC3 Sigma revision total knee replacements
using long uncemented stems in press-fit mode. Clinical and radiological analysis took place pre-operatively,
at two years post-operatively, and at a mean follow-up of 8.5 years
(4 to 12). The mean pre-operative International Knee Society (IKS)
clinical score was 42 points (0 to 74), improving to 83.7 (52 to
100) by the final follow-up. The mean IKS score for function improved
from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up.
At the final follow-up 30 knees (65.2%) had an excellent result, seven
(15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor
result. There were two failures, one with anteroposterior instability
and one with aseptic loosening. The TC3 revision knee system, when used with press-fit for long
intramedullary
Peri-prosthetic infection is amongst the most
common causes of failure following total knee replacement (TKR).
In the presence of established infection, thorough joint debridement
and removal of all components is necessary following which new components
may be implanted. This can be performed in one or two stages; two-stage revision
with placement of an interim antibiotic-loaded spacer is regarded
by many to be the standard procedure for eradication of peri-prosthetic
joint infection. We present our experience of a consecutive series of 50 single-stage
revision TKRs for established deep infection performed between 1979
and 2010. There were 33 women and 17 men with a mean age at revision
of 66.8 years (42 to 84) and a mean follow-up of 10.5 years (2 to
24). The mean time between the primary TKR and the revision procedure
was 2.05 years (1 to 8). Only one patient required a further revision for recurrent infection,
representing a success rate of 98%. Nine patients required further
revision for aseptic loosening, according to microbiological testing
of biopsies taken at the subsequent surgery. Three other patients
developed a further septic episode but none required another revision. These results suggest that a single-stage revision can produce
comparable results to a two-stage revision. Single-stage revision
offers a reduction in costs as well as less morbidity and inconvenience
for patients. Cite this article:
Short uncemented humeral stems can preserve humeral
bone stock and facilitate revision surgery but may be prone to malalignment
or loosening, especially when used in reverse total shoulder replacement
(TSR). We undertook a retrospective review of 44 patients with a
mean age of 76 years (59 to 92) who underwent primary reverse TSR
with a short uncemented humeral stem. There were 29 females. The
indications for joint replacement included cuff tear associated
arthropathy (33), avascular necrosis (six), post-traumatic arthritis
(two), and inflammatory arthritis (three). At a mean follow-up of
27 months (24 to 40), pain was rated as mild or none in 43 shoulders
(97.7%). The mean active elevation improved from 54° Cite this article:
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
In skeletally immature patients, resection of
bone tumours and reconstruction of the lower limb often results
in leg-length discrepancy. The Stanmore non-invasive extendible
endoprosthesis, which uses electromagnetic induction, allows post-operative
lengthening without anaesthesia. Between 2002 and 2009, 55 children
with a mean age of 11.4 years (5 to 16) underwent reconstruction
with this prosthesis; ten patients (18.2%) died of disseminated
disease and one child underwent amputation due to infection. We
reviewed 44 patients after a mean follow-up of 41.2 months (22 to
104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to
30) and the Toronto Extremity Salvage score was 92.3% (55.2% to
99.0%). There was no local recurrence of tumour. Complications developed
in 16 patients (29.1%) and ten (18.2%) underwent revision. The mean length gained per patient was 38.6 mm (3.5 to 161.5),
requiring a mean of 11.3 extensions (1 to 40), and ten component
exchanges were performed in nine patients (16.4%) after attaining
the maximum lengthening capacity of the implant. There were 11 patients
(20%) who were skeletally mature at follow-up, ten of whom had equal
leg lengths and nine had a full range of movement of the hip and
knee. This is the largest reported series using non-invasive extendible
endoprostheses after excision of primary bone tumours in skeletally
immature patients. The technique produces a good functional outcome,
with prevention of limb-length discrepancy at skeletal maturity.
Interfacial defects between the cement mantle and a hip implant may arise from constrained shrinkage of the cement or from air introduced during insertion of the stem. Shrinkage-induced interfacial porosity consists of small pores randomly located around the stem, whereas introduced interfacial gaps are large, individual and less uniformly distributed areas of stem-cement separation. Using a validated CT-based technique, we investigated the extent, morphology and distribution of interfacial gaps for two types of stem, the Charnley-Kerboul and the Lubinus SPII, and for two techniques of implantation, line-to-line and undersized. The interfacial gaps were variable and involved a mean of 6.43% (. sd. 8.99) of the surface of the stem. Neither the type of implant nor the technique of implantation had a significant effect on the regions of the gaps, which occurred more often over the flat areas of the implant than along the corners of the stems, and were more common proximally than distally for Charnley-Kerboul
Femoral revision after cemented total hip replacement
(THR) might include technical difficulties, following essential cement
removal, which might lead to further loss of bone and consequently
inadequate fixation of the subsequent revision stem. . Femoral impaction allografting has been widely used in revision
surgery for the acetabulum, and subsequently for the femur. In combination
with a primary
We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up. Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain. Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated
We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p <
0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p <
0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.
Total hip replacement (THR) is a very common
procedure undertaken in up to 285 000 Americans each year. Patient
satisfaction with THR is very high, with improvements in general
health, quality of life, and function while at the same time very
cost effective. Although the majority of patients have a high degree
of satisfaction with their THR, 27% experience some discomfort,
and up to 6% experience severe chronic pain. Although it can be
difficult to diagnose the cause of the pain in these patients, this
clinical issue should be approached systematically and thoroughly.
A detailed history and clinical examination can often provide the
correct diagnosis and guide the appropriate selection of investigations, which
will then serve to confirm the clinical diagnosis made. Cite this article:
We reviewed the long-term results at ten to 12 years of 118 total hip replacements in 109 patients using a second-generation hemispherical cementless acetabular component (Reflection) designed to address the problem of backside wear. Five patients (five hips) died and six patients (seven hips) were lost to follow-up. The remaining 98 patients (106 hips) had a mean age of 62.9 years (34.0 to 86.2) A rate of revision for aseptic loosening of 0.9%, and predictable results were found with respect to radiological evidence of fixation, lack of pain, walking ability, range of movement and function. One component was revised for aseptic loosening, and of the 101 hips (95.2%) that did not have a revision, minor osteolytic lesions of the pelvis were seen in six (5.9%). Kaplan-Meier survival analysis for the total cohort of 118 hips revealed a 96.4% survival at both ten (95% confidence interval 90 to 98) and 12 years (95% confidence interval 86 to 98).
Bone loss in the proximal tibia and distal femur
is frequently encountered in revision knee replacement surgery.
The various options for dealing with this depend on the extent of
any bone loss. We present our results with the use of cementless
metaphyseal metal sleeves in 103 patients (104 knees) with a mean
follow-up of 43 months (30 to 65). At final follow-up, sleeves in
102 knees had good osseointegration. Two tibial sleeves were revised
for loosening, possibly due to infection. The average pre-operative Oxford Knee Score was 23 (11 to 36)
and this improved to 32 (15 to 46) post-operatively. These early
results encourage us to continue with the technique and monitor
the outcomes in the long term. Cite this article:
We undertook a randomised controlled trial to
compare the piriformis-sparing approach with the standard posterior approach
used for total hip replacement (THR). We recruited 100 patients
awaiting THR and randomly allocated them to either the piriformis-sparing
approach or the standard posterior approach. Pre- and post-operative
care programmes and rehabilitation regimes were identical for both
groups. Observers were blinded to the allocation throughout; patients
were blinded until the two-week assessment. Follow-up was at six
weeks, three months, one year and two years. In all 11 patients
died or were lost to follow-up. There was no significant difference between groups for any of
the functional outcomes. However, for patients in the piriformis-sparing
group there was a trend towards a better six-minute walk test at
two weeks and greater patient satisfaction at six weeks. The acetabular
components were less anteverted (p = 0.005) and had a lower mean
inclination angle (p = 0.02) in the piriformis-sparing group. However,
in both groups the mean component positions were within Lewinnek’s
safe zone. Surgeons perceived the piriformis-sparing approach to
be significantly more difficult than the standard approach (p =
0.03), particularly in obese patients. In conclusion, performing THR through a shorter incision involving
sparing piriformis is more difficult and only provides short-term
benefits compared with the standard posterior approach.
Despite the worldwide usage of the cemented Contemporary
acetabular component (Stryker), no published data are available
regarding its use in patients aged <
50 years. We undertook a
mid- to long-term follow-up study, including all consecutive patients
aged
<
50 years who underwent a primary total hip replacement using
the Contemporary acetabular component with the Exeter cemented stem
between January 1999 and January 2006. There were 152 hips in 126
patients, 61 men and 65 women, mean age at surgery 37.6 years (16
to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to
12.0). All clinical questionnaire scores, including Harris hip score,
Oxford hip score and several visual analogue scales, were found
to have improved. The eight year survivorship of all acetabular
components for the endpoints revision for any reason or revision
for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2
to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up
was complete for 146 implants. The eight year survival for the endpoint
radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving
implants were considered radiologically loose but were asymptomatic.
The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent
lines (n = 20, 13.9%) in the 144 surviving cups indicates a need
for continued observation in the second decade of follow-up in order
to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival
rate >
90% in young patients support the use of the Contemporary
acetabular component in this specific patient group. Cite this article:
The management of joint replacement in lysosomal storage diseases has not been well reported. We present three patients with progressive degenerative changes of the hips who required bilateral total hip replacement in early childhood. The stature of the patients make it essential to have access to appropriately scaled prostheses. Consideration has to be given to associated disorders of the skeleton which must be carefully screened to ensure safety in providing appropriate anaesthesia as well as ensuring that there is no cardiac abnormality. In one patient, a periprosthetic fracture was sustained in one hip in the early post-operative course requiring internal fixation. The patient made a full recovery and all six hips were clinically and radiologically satisfactory at mid-term review.