The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 865 - 874
1 Jul 2012
Mills LA Simpson AHRW

This review is aimed at clinicians appraising preclinical trauma studies and researchers investigating compromised bone healing or novel treatments for fractures. It categorises the clinical scenarios of poor healing of fractures and attempts to match them with the appropriate animal models in the literature.

We performed an extensive literature search of animal models of long bone fracture repair/nonunion and grouped the resulting studies according to the clinical scenario they were attempting to reflect; we then scrutinised them for their reliability and accuracy in reproducing that clinical scenario.

Models for normal fracture repair (primary and secondary), delayed union, nonunion (atrophic and hypertrophic), segmental defects and fractures at risk of impaired healing were identified. Their accuracy in reflecting the clinical scenario ranged greatly and the reliability of reproducing the scenario ranged from 100% to 40%.

It is vital to know the limitations and success of each model when considering its application.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1079 - 1084
1 Aug 2010
Muirhead-Allwood S Sandiford N Skinner JA Hua J Kabir C Walker PS

We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99).

These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1015 - 1018
1 Aug 2007
Wroblewski BM Siney PD Fleming PA

We studied survival to 38 years after Charnley low-friction arthroplasty of the hip. We used revision as an end-point, while adopting a policy of regular follow-up and early revision for radiological changes alone if indicated.

Between November 1962 and June 2005, 22 066 primary low-friction arthroplasties (17 409 patients) had been performed at Wrightington Hospital by more than 330 surgeons. By June 2006, 1001 (4.5%) hips had been revised and 1490 patients (2662 hips, 12%) had died. At 31 years, where a minimum of 40 hips were still attending follow-up, survival with revision for infection as an endpoint was 95%, for dislocation 98%, for a fractured stem 88.6%, for a loose stem 72.5% and for a loose acetabular component 53.7%. Wear and loosening of the ultra-high-molecular-weight polyethylene acetabular component were the main long-term problems.

We conclude that regular follow-up after hip replacement is essential and that all operative findings should be recorded at revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1069 - 1073
1 Aug 2009
Hamid N Loeffler BJ Braddy W Kellam JF Cohen BE Bosse MJ

The purpose of this study was to compare the clinical and radiological outcome of patients with intact, broken and removed syndesmosis screws after Weber B or C ankle fracture with an associated injury to the syndesmosis. We hypothesised that there would be no difference. Of a possible 142 patients who fulfilled our inclusion criteria, 52 returned for clinical and radiological assessment at least one year after surgery. Of these, 27 had intact syndesmosis screws, ten had broken screws, and 15 had undergone elective removal of the screw. The mean American Orthopaedic Foot and Ankle Society ankle/hindfoot score was 83.07 (sd 13.59) in the intact screw group, 92.40 (sd 12.69) in the broken screw group, and 85.80 (sd 11.33) in the removed screw group (p = 0.0466).

There was no difference in clinical outcome of patients with intact or removed syndesmotic screws. Paradoxically, patients with a broken syndesmosis screw had the best clinical outcome. Our data do not support the removal of intact or broken syndesmosis screws, and we caution against attributing post-operative ankle pain to breakage of the syndesmosis screw.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 572 - 579
1 May 2011
Haddad FS Thakrar RR Hart AJ Skinner JA Nargol AVF Nolan JF Gill HS Murray DW Blom AW Case CP

Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 720 - 724
1 Jun 2009
Ollivere B Darrah C Brankin RC Donnell ST Walton NP

We report the follow-up at 12 years of the use of the Elite Plus total hip replacement (THR). We have previously reported the results at a mean of 6.4 years. Of the 217 patients (234 THRs), 83 had died and nine had been lost to follow-up. The patients were reviewed radiologically and clinically using the Oxford hip score.

Of the 234 THRs, 19 (8.1%) had required a revision by the final follow-up in all but one for aseptic loosening. Survivorship analysis for revision showed a survival of 93.9% (95% confidence interval (CI) 89.2 to 96.5) at ten years, and of 88.0% (95% CI 81.8 to 92.3) at 12 years. At the final follow-up survival analysis showed that 37% (95% CI 37.3 to 44.7) of the prostheses had either failed radiologically or had been revised. Patients with a radiologically loose femoral component had a significantly poorer Oxford hip score than those with a well-fixed component (p = 0.03). Radiological loosening at 6.4 years was predictive of failure at 12 years.

Medium-term radiographs and clinical scores should be included in the surveillance of THR to give an early indication of the performance of specific implants.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 304 - 309
1 Mar 2009
Kerboull L Hamadouche M Kerboull M

We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30.

There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 344 - 350
1 Mar 2009
Luyckx T Didden K Vandenneucker H Labey L Innocenti B Bellemans J

The purpose of this study was to test the hypothesis that patella alta leads to a less favourable situation in terms of patellofemoral contact force, contact area and contact pressure than the normal patellar position, and thereby gives rise to anterior knee pain.

A dynamic knee simulator system based on the Oxford rig and allowing six degrees of freedom was adapted in order to simulate and record the dynamic loads during a knee squat from 30° to 120° flexion under physiological conditions. Five different configurations were studied, with variable predetermined patellar heights.

The patellofemoral contact force increased with increasing knee flexion until contact occurred between the quadriceps tendon and the femoral trochlea, inducing load sharing. Patella alta caused a delay of this contact until deeper flexion. As a consequence, the maximal patellofemoral contact force and contact pressure increased significantly with increasing patellar height (p < 0.01). Patella alta was associated with the highest maximal patellofemoral contact force and contact pressure. When averaged across all flexion angles, a normal patellar position was associated with the lowest contact pressures.

Our results indicate that there is a biomechanical reason for anterior knee pain in patients with patella alta.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 170 - 175
1 Feb 2009
Young L Duckett S Dunn A

We describe the survivorship of the Exeter femoral component in a District General Hospital. Between 1994 and 1996, 230 Exeter Universal cemented femoral components were implanted in 215 patients who were reviewed at a mean of 11.2 years (10 to 13). We used one acetabular implant, the Elite Ogee component, in 218 of the 230 hips.

During the period of this study 76 patients (79 hips) died. Of the remaining 139 patients (151 hips), 121 were able to attend for radiological analysis at a minimum of ten years. One patient was lost to follow-up.

No femoral component was revised for aseptic loosening. Three hips were revised for deep infection and six acetabular components required revision, four for loosening and two for recurrent dislocation. Taking the ‘worst-case scenario’ including the one patient lost to follow-up, the overall survival rate was 94.4% at 13 years.

Our results confirm excellent medium-term results for the Exeter Universal femoral component, implanted in a general setting. The excellent survival of this femoral component, when used in combination with the Ogee acetabular component, suggests that this is a successful pairing.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1508 - 1512
1 Nov 2006
Wimsey S Pickard R Shaw G

Magnification of anteroposterior radiographs of the pelvis is variable. To improve the accuracy of templating, reliable and radiographer-friendly methods of scaling are necessary. We assessed two methods of scaling digital radiographs of the pelvis: placing a coin of known diameter in the plane of interest between the patient’s thighs, and using a caliper to measure the bony width of the pelvis. A total of 39 patients who had recently undergone hemiarthroplasty of the hip or total hip replacement were enrolled in the study. The accuracy of the methods was assessed by comparing the actual diameter of the head of the prosthesis with the measured on-screen value. The coin method was within a mean of 1.12% (0% to 2.38%) of the actual measurement, the caliper group within 6.99% (0% to 16.67%). The coin method was significantly more accurate (p < 0.001). It was also reliable and radiographer friendly. We recommend it as the method of choice for scaling radiographs of the pelvis before hip surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 893 - 898
1 Jul 2008
Levy O Relwani J Zaman T Even T Venkateswaran B Copeland S

The aim of this study was to define the microcirculation of the normal rotator cuff during arthroscopic surgery and investigate whether it is altered in diseased cuff tissue.

Blood flow was measured intra-operatively by laser Doppler flowmetry. We investigated six different zones of each rotator cuff during the arthroscopic examination of 56 consecutive patients undergoing investigation for impingement, cuff tears or instability; there were 336 measurements overall.

The mean laser Doppler flowmetry flux was significantly higher at the edges of the tear in torn cuffs (43.1, 95% confidence interval (CI) 37.8 to 48.4) compared with normal cuffs (32.8, 95% CI 27.4 to 38.1; p = 0.0089). It was significantly lower across all anatomical locations in cuffs with impingement (25.4, 95% CI 22.4 to 28.5) compared with normal cuffs (p = 0.0196), and significantly lower in cuffs with impingement compared with torn cuffs (p < 0.0001).

Laser Doppler flowmetry analysis of the rotator cuff blood supply indicated a significant difference between the vascularity of the normal and the pathological rotator cuff. We were unable to demonstrate a functional hypoperfusion area or so-called ‘critical zone’ in the normal cuff. The measured flux decreases with advancing impingement, but there is a substantial increase at the edges of rotator cuff tears. This might reflect an attempt at repair.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 925 - 927
1 Jul 2006
Kalra KP Dhar SB Shetty G Dhariwal Q

We studied 15 patients with healed tuberculosis of the spine and a resultant kyphosis. We selected only those with no neurological deficit and performed a wedge resection of the vertebra using a transpedicular approach. The wedge was removed from the apex of the deformity. For those with a neurological deficit, we chose the conventional anterior debridement and decompression with 360° circumferential fusion. At a mean follow-up of 26.8 months (8 to 46) the outcome was good with an increase in the mean Oswestry Disability Index from 56.26 (48 to 62) pre-operatively to 11.2 (6 to 16) at the latest follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 342 - 348
1 Mar 2010
El Masri F Kerboull L Kerboull M Courpied JP Hamadouche M

We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component.

At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 405 - 410
1 Apr 2008
Dattani R Patnaik S Kantak A Srikanth B Selvan TP

The management of injury to the distal tibiofibular syndesmosis remains controversial in the treatment of ankle fractures. Operative fixation usually involves the insertion of a metallic diastasis screw. There are a variety of options for the position and characterisation of the screw, the type of cortical fixation, and whether the screw should be removed prior to weight-bearing. This paper reviews the relevant anatomy, the clinical and radiological diagnosis and the mechanism of trauma and alternative methods of treatment for injuries to the syndesmosis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1178 - 1181
1 Sep 2005
Harrison WJ


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 124 - 130
1 Jan 2009
Deuel CR Jamali AA Stover SM Hazelwood SJ

Bone surface strains were measured in cadaver femora during loading prior to and after resurfacing of the hip and total hip replacement using an uncemented, tapered femoral component. In vitro loading simulated the single-leg stance phase during walking. Strains were measured on the medial and the lateral sides of the proximal aspect and the mid-diaphysis of the femur. Bone surface strains following femoral resurfacing were similar to those in the native femur, except for proximal shear strains, which were significantly less than those in the native femur. Proximomedial strains following total hip replacement were significantly less than those in the native and the resurfaced femur.

These results are consistent with previous clinical evidence of bone loss after total hip replacement, and provide support for claims of bone preservation after resurfacing arthroplasty of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 859 - 864
1 Jul 2009
Gwynne-Jones DP Garneti N Wainwright C Matheson JA King R

We reviewed the results at nine to 13 years of 125 total hip replacements in 113 patients using the monoblock uncemented Morscher press-fit acetabular component. The mean age at the time of operation was 56.9 years (36 to 74). The mean clinical follow-up was 11 years (9.7 to 13.5) and the mean radiological follow-up was 9.4 years (7.7 to 13.1). Three hips were revised, one immediately for instability, one for excessive wear and one for deep infection.

No revisions were required for aseptic loosening. A total of eight hips (7.0%) had osteolytic lesions greater than 1 cm, in four around the acetabular component (3.5%). One required bone grafting behind a well-fixed implant. The mean wear rate was 0.11 mm/year (0.06 to 0.78) and was significantly higher in components with a steeper abduction angle.

Kaplan-Meier survival curves at 13 years showed survival of 96.8% (95% confidence interval 90.2 to 99.0) for revision for any cause and of 95.7% (95% confidence interval 88.6 to 98.4) for any acetabular re-operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 400 - 404
1 Mar 2008
Johansson HR Skripitz R Aspenberg P

We have examined the deterioration of implant fixation after withdrawal of parathyroid hormone (PTH) in rats. First, the pull-out force for stainless-steel screws in the proximal tibia was measured at different times after withdrawal. The stimulatory effect of PTH on fixation was lost after 16 days. We then studied whether bisphosphonates could block this withdrawal effect. Mechanical and histomorphometric measurements were conducted for five weeks after implantation. Subcutaneous injections were given daily. Specimens treated with either PTH or saline during the first two weeks showed no difference in the mechanical or histological results (pull-out force 76 N vs 81 N; bone volume density 19% vs 20%). Treatment with PTH for two weeks followed by pamidronate almost doubled the pull-out force (152 N; p < 0.001) and the bone volume density (37%; ANOVA, p < 0.001). Pamidronate alone did not have this effect (89 N and 25%, respectively). Thus, the deterioration can be blocked by bisphosphonates. The clinical implications are discussed.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 928 - 933
1 Jul 2005
Berli BJ Schäfer D Morscher EW

The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement.

We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant in situ. Nine could only be interviewed by telephone. We included 85 patients with 88 hips in the clinical and radiological follow-up examinations.

None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed.

The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence.