Advertisement for orthosearch.org.uk
Results 181 - 200 of 222
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1561 - 1565
1 Dec 2009
Blakey CM Eswaramoorthy VK Hamilton LC Biant LC Field RE

We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips).

Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant.

This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 1 - 11
1 Jan 2011
Murray IR Amin AK White TO Robinson CM

Most proximal humeral fractures are stable injuries of the ageing population, and can be successfully treated non-operatively. The management of the smaller number of more complex displaced fractures is more controversial and new fixation techniques have greatly increased the range of fractures that may benefit from surgery.

This article explores current concepts in the classification and clinical aspects of these injuries, reviewing the indications, innovations and outcomes for the most common methods of treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 851 - 854
1 Jul 2009
Timperley AJ Whitehouse SL

Recently the National Patient Safety Agency in the United Kingdom published a report entitled “Mitigating surgical risk in patients undergoing hip arthroplasty for fractures of the proximal femur”. A total of 26 deaths had been reported to them when cement was used at hemiarthroplasty between October 2003 and October 2008. This paper considers the evidence for using cement fixation of a hemiarthroplasty in the treatment of hip fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 617 - 623
1 May 2010
McNamara I Deshpande S Porteous M

The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had radiolucent lines and two acetabular components migrated initially before stabilising.

Acetabular reconstruction in both primary and revision surgery using a 1:1 mixture of frozen, ground, irriadiated bone and Apapore 60 appears to be a reliable method of managing acetabular defects. Longer follow-up will be required to establish whether this technique is as effective as using fresh-frozen allograft.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1678 - 1678
1 Dec 2006
Schindler O


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1669 - 1677
1 Dec 2010
Thakar C Alsousou J Hamilton TW Willett K

We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p < 0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p < 0.01).

Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1137 - 1142
1 Sep 2008
Heilpern GNA Shah NN Fordyce MJF

We report the outcome at a minimum of five years of 110 consecutive metal-on-metal Birmingham Hip Resurfacing arthroplasties in 98 patients. The procedures were performed between October 1999 and June 2002 by one surgeon. All patients were followed up clinically and radiologically. The mean follow-up was 71 months (60 to 93). Revision of either component was defined as failure.

The mean Harris Hip score at follow-up was 96.4 (53 to 100). The mean Oxford hip score was 41.9 (16 to 57) pre-operatively and 15.4 (12 to 49) post-operatively (p < 0.001). The mean University of California Los Angeles activity score was 3.91 (1 to 10) pre-operatively and 7.5 (4 to 10) post-operatively (p < 0.001).

There were four failures giving a survival at five years of 96.3% (95% confidence interval 92.8 to 99.8). When applying a new method to estimate narrowing of the femoral neck we identified a 10% thinning of the femoral neck in 16 hips (14.5%), but the relevance of this finding to the long-term outcome remains unclear.

These good medium-term results from an independent centre confirm the original data from Birmingham.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 52 - 57
1 Jan 2009
Newman J Pydisetty RV Ackroyd C

Between 1989 and 1992 we had 102 knees suitable for unicompartmental knee replacement (UKR). They were randomised to receive either a St Georg Sled UKR or a Kinematic modular total knee replacement (TKR). The early results demonstrated that the UKR group had less complications and more rapid rehabilitation than the TKR group. At five years there were an equal number of failures in the two groups but the UKR group had more excellent results and a greater range of movement.

The cases were reviewed by a research nurse at 8, 10 and 12 years after operation. We report the outcome at 15 years follow-up. A total of 43 patients (45 knees) died with their prosthetic knees intact. Throughout the review period the Bristol knee scores of the UKR group have been better and at 15 years 15 (71.4%) of the surviving UKRs and 10 (52.6%) of the surviving TKRs had achieved an excellent score. The 15 years survivorship rate based on revision or failure for any reason was 24 (89.8%) for UKR and 19 (78.7%) for TKR. During the 15 years of the review four UKRs and six TKRs failed.

The better early results with UKR are maintained at 15 years with no greater failure rate. The median Bristol knee score of the UKR group was 91.1 at five years and 92 at 15 years, suggesting little functional deterioration in either the prosthesis or the remainder of the joint. These results justify the increased use of UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 393 - 395
1 Mar 2007
Briant-Evans TW Norton MR Fern ED

We describe two cases of fracture of Corin Taper-Fit stems used for cement-in-cement revision of congenital dysplasia of the hip. Both prostheses were implanted in patients in their 50s, with high offsets (+7.5 mm and +3.5 mm), one with a large diameter (48 mm) head and one with a constrained acetabular component. Fracture of the stems took place at nine months and three years post-operatively following low-demand activity. Both fractures occurred at the most medial of the two stem introducer holes in the neck of the prosthesis, a design feature that is unique to the Taper-Fit stem. We would urge caution in the use of these particular stems for cement-in-cement revisions.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 27 - 30
1 Jan 2008
Rajaratnam SS Jack C Tavakkolizadeh A George MD Fletcher RJ Hankins M Shepperd JAN

Between 1986 and 1991 we implanted 331 consecutive Furlong hydroxyapatite-coated femoral components of a total hip replacement in 291 patients. A cemented acetabular prosthesis was used in 217 hips and a hydroxyapatite-coated component in 114. We describe the long-term clinical and radiological survival of the femoral component at a mean follow-up of 17.5 years (15 to 21). Only two patients (0.68%) were lost to follow-up. With revision of the femoral component for any reason as the endpoint, the survival at a mean of 17 years was 97.4% (95% confidence interval 94.1 to 99.5), and with revision for aseptic loosening as the endpoint it was 100%. The survival at a maximum of 21 years with revision of the femoral component for any reason as the endpoint was 97.4% (95% confidence interval 81.0 or 99.5). These results compare favourably with the best long-term results of cemented or uncemented femoral components used in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 84 - 87
1 Jan 2008
Tillman RM Myers GJC Abudu AT Carter SR Grimer RJ

Pathological fractures due to metastasis with destruction of the acetabulum and central dislocation of the hip present a difficult surgical challenge. We describe a series using a single technique in which a stable and long-lasting reconstruction was obtained using standard primary hip replacement implants augmented by strong, fully-threaded steel rods with cement and steel mesh, where required.

Between 1997 and 2006, 19 patients with a mean age of 66 years (48 to 83) were treated using a modified Harrington technique. Acetabular destruction was graded as Harrington class II in six cases and class III in 13.

Reconstruction was achieved using three 6.5 mm rods inserted through a separate incision in the iliac crest followed by augmentation with cement and a conventional cemented Charnley or Exeter primary hip replacement. There were no peri-operative deaths. At the final follow-up (mean 25 months (5 to 110)) one rod had fractured and one construct required revision. Of the 18 patients who did not require revision, 13 had died. The mean time to death was 16 months (5 to 55). The mean follow-up of the five survivors was 31 months (18 to 47). There were no cases of dislocation, deep infection or injury to a nerve, the blood vessels or the bladder.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 486 - 489
1 Apr 2007
Bassi RS Simmons D Ali F Nuttall D Birch A Trail IA Stanley JK

The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function.

No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 668 - 673
1 May 2008
Karva AR Board TN Porter ML

We present a 25-year-old patient with juvenile rheumatoid arthritis and ankylosis of both hips and both knees treated by staged bilateral hip and knee arthroplasty. She was followed up for 18 months. We discuss the pre-operative planning, surgical details and post-operative rehabilitation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1594 - 1600
1 Dec 2009
Khan A Bunker TD Kitson JB

There are no long-term published results on the survival of a third-generation cemented total shoulder replacement. We describe a clinical and radiological study of the Aequalis total shoulder replacement for a minimum of ten years. Between September 1996 and May 1998, 39 consecutive patients underwent a primary cemented total shoulder replacement using this prosthesis. Data were collected prospectively on all patients each year, for a minimum of ten years, or until death or failure of the prosthesis. At a follow-up of at least ten years, 12 patients had died with the prosthesis intact and two had emigrated, leaving 25 available for clinical review. Of these, 13 had rheumatoid arthritis and 12 osteoarthritis. One refused radiological review leaving 24 with fresh radiographs.

Survivorship at ten years was 100% for the humeral component and 92% for the glenoid component. The incidence of lucent lines was low. No humeral component was thought to be at risk and only two glenoid components. The osteoarthritic group gained a mean 65° in forward flexion and their Constant score improved by a mean 41.4 points (13 to 55). The rheumatoid group gained a mean of 24° in flexion and their Constant score improved by 29.4 points. This difference may have been due to failure of the rotator cuff in 75% of the patients with rheumatoid arthritis.

Thus a third-generation total shoulder replacement gives an excellent result in patients with osteoarthritis and an intact rotator cuff. Patients with rheumatoid arthritis have a 75% risk of failure of the rotator cuff at ten years.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1000 - 1004
1 Aug 2008
Wraighte PJ Howard PW

Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998.

The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification.

The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively.

Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p < 0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 37 - 43
1 Jan 2009
Hassaballa M Mehendale S Poniatowski S Kalantzis G Smith E Learmonth ID

Loss of bone stock is a major problem in revision surgery of the hip. Impaction bone grafting of the femur is frequently used when dealing with deficient bone stock. In this retrospective study a consecutive series of 68 patients (69 hips) who had revision of a hip replacement with femoral impaction grafting were reviewed. Irradiated bone allograft was used in all hips. Radiological measurement of subsidence of the stem, incorporation of the graft and remodelling was carried out and showed incorporation of the graft in 26 of 69 hips (38%). However, there was no evidence of trabecular remodelling. Moderate subsidence of 5 mm to 10 mm occurred in ten hips (14.5%), and massive subsidence of > 10 mm in five (7.2%).

The results of this study are less favourable than those of others describing studies of revision of the femoral stem using impaction bone grafting. The absence of the characteristic changes of graft remodelling noted in other series raises the question as to whether irradiated bone graft may be a significant factor influencing the post-operative outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 143 - 150
1 Feb 2009
Toms AD Mandalia V Haigh R Hopwood B

The management of patients with a painful total knee replacement requires careful assessment and a stepwise approach in order to diagnose the underlying pathology accurately. The management should include a multidisciplinary approach to the patient’s pain as well as addressing the underlying aetiology. Pain should be treated with appropriate analgesia, according to the analgesic ladder of the World Health Organisation. Special measures should be taken to identify and to treat any neuropathic pain. There are a number of intrinsic and extrinsic causes of a painful knee replacement which should be identified and treated early. Patients with unexplained pain and without any recognised pathology should be treated conservatively since they may improve over a period of time and rarely do so after a revision operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 460 - 465
1 Apr 2008
Strickland JP Sperling JW Cofield RH

While frequently discussed as a standard treatment for the management of an infected shoulder replacement, there is little information on the outcome of two-stage re-implantation.

We examined the outcome of 17 consecutive patients (19 shoulders) who were treated between 1995 and 2004 with a two-stage re-implantation for the treatment of a deep-infection after shoulder replacement. All 19 shoulders were followed for a minimum of two years or until the time of further revision surgery. The mean clinical follow-up was for 35 months (24 to 80). The mean radiological follow-up was 27 months (7 to 80). There were two excellent results, four satisfactory and 13 unsatisfactory. In 12 of the 19 shoulders (63%) infection was considered to be eradicated. The mean pain score improved from 4.2 (3 to 5 (out of 5)) to 1.8 (1 to 4). The mean elevation improved from 42° (0° to 140°) to 89° (0° to 165°), mean external rotation from 30° (0° to 90°) to 43° (0° to 90°), and mean internal rotation from the sacrum to L5. There were 14 complications.

Our study suggests that two-stage re-implantation for an infected shoulder replacement is associated with a high rate of unsatisfactory results, marginal success at eradicating infection and a high complication rate.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1452 - 1456
1 Nov 2007
Roberts VI Esler CNA Harper WM

This is a 15-year follow-up observational study of 4390 patients with 4606 primary total knee replacements (TKRs) implanted in the Trent health region between 1990 and 1992. The operations were performed in 21 hospitals, including both district general and teaching hospitals, with 77 different surgeons as named consultant.

The main objective was to analyse the survival of the patients and of the prostheses, and to evaluate what impact different variables have on survival. In addition, the 1480 patients (33.7%) (1556 TKRs) alive at 15 years following operation were sent a self-administered questionnaire which examined their level of satisfaction, of pain, and their quality of life at 15 years. Completed responses were received from 912 TKRs (58.6%).

Three survival curves were constructed: a best-case scenario based on the patients entered into the life tables, another included failures not reported in the revision database, and a third worst-case scenario based on all patients lost to follow-up presumed to have had a failed primary TKR. In the best-case scenario survival at 15 years was 92.2%, and in the worst-case scenario was 81.1%. Survival was significantly increased in women and older patients (Mantel-Cox log-rank test, p < 0.005 and p < 0.001, respectively).

Revision as a result of infection was required in 40 TKRs (18.8%) representing 0.87% of the original cohort.

The limited information available from the questionnaire indicated that satisfaction was less frequent among men, patients with osteoarthritis and those who required revision (chi-squared test, p < 0.05, p < 0.05 and p < 0.0001, respectively). With regard to pain, older patients, females and patients who still had their primary replacement in place at 15 years, reported the least pain (chi-squared test for trends, p < 0.0005, p < 0.005 and p < 0.0001, respectively). The reported quality of life was not affected by any variable.