Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset femoral stems, and occurs within a central concentric
zone of the femoral head bore taper. Further investigation is required
to determine the effect of increased head size, and variations in
head–neck taper design. Cite this article:
Dislocation is a common and well-studied complication after total hip replacement. However, subluxation, which we define as a clinically recognised episode of incomplete movement of the femoral head outside the acetabulum with spontaneous reduction, has not been studied previously. Out of a total of 2521 hip replacements performed over 12 years by one surgeon, 30 patients experienced subluxations which occurred in 31 arthroplasties. Data were collected prospectively with a minimum follow-up of two years. Subluxation occurred significantly more frequently after revision than after primary hip replacement, and resolved in 19 of 31 cases (61.3%). In six of the 31 hips (19.4%) the patient subsequently dislocated the affected hip, and in six hips (19.4%) intermittent subluxation continued. Four patients had a revision operation for instability, three for recurrent dislocation and one for recurrent subluxation. Clinical and radiological comparisons with a matched group of stable total hips showed no correlation with demographic or radiological parameters. Patients with subluxing hips reported significantly more concern that their hip would dislocate, more often changed their behaviour to prevent instability and had lower postoperative Harris hip scores than patients with stable replacements.
Reported rates of dislocation in hip hemiarthroplasty
(HA) for the treatment of intra-capsular fractures of the hip, range
between 1% and 10%. HA is frequently performed through a direct
lateral surgical approach. The aim of this study is to determine
the contribution of the anterior capsule to the stability of a cemented
HA through a direct lateral approach. A total of five whole-body cadavers were thawed at room temperature,
providing ten hip joints for investigation. A Thompson HA was cemented
in place via a direct lateral approach. The cadavers were then positioned
supine, both knee joints were disarticulated and a digital torque
wrench was attached to the femur using a circular frame with three
half pins. The wrench applied an external rotation force with the
hip in extension to allow the hip to dislocate anteriorly. Each
hip was dislocated twice; once with a capsular repair and once without
repairing the capsule. Stratified sampling ensured the order in
which this was performed was alternated for the paired hips on each
cadaver. Comparing peak torque force in hips with the capsule repaired
and peak torque force in hips without repair of the capsule, revealed
a significant difference between the ‘capsule repaired’ (mean 22.96
Nm, standard deviation ( Cite this article:
The term developmental dysplasia of the hip (DDH)
describes a spectrum of disorders that results in abnormal development
of the hip joint. If not treated successfully in childhood, these
patients may go on to develop hip symptoms and/or secondary osteoarthritis
in adulthood. In this review we describe the altered anatomy encountered
in adults with DDH along with the management options, and the challenges
associated with hip arthroscopy, osteotomies and arthroplasty for
the treatment of DDH in young adults. Cite this article:
The December 2014 Hip &
Pelvis Roundup360 looks at: Sports and total hips; topical tranexamic acid and blood conservation in hip replacement; blind spots and biases in hip research; no recurrence in cam lesions at two years; to drain or not to drain?; sonication and diagnosis of implant associated infection; and biomarkers and periprosthetic infection
We compared the accuracy of the growth remaining
method of assessing leg-length discrepancy (LLD) with the straight-line
graph method, the multiplier method and their variants. We retrospectively
reviewed the records of 44 patients treated by percutaneous epiphysiodesis
for LLD. All were followed up until maturity. We used the modified Green–Anderson
growth-remaining method (Method 1) to plan the timing of epiphysiodesis.
Then we presumed that the other four methods described below were
used pre-operatively for calculating the timing of epiphysiodesis. We
then assumed that these four methods were used pre-operatively.
Method 2 was the original Green–Anderson growth-remaining method;
Method 3, Paley’s multiplier method using bone age; Method 4, Paley’s
multiplier method using chronological age; and Method 5, Moseley’s
straight-line graph method. We compared ‘Expected LLD at maturity
with surgery’ with ‘Final LLD at maturity with surgery’ for each
method. Statistical analysis revealed that ‘Expected LLD at maturity
with surgery’ was significantly different from ‘Final LLD at maturity
with surgery’. Method 2 was the most accurate. There was a significant
correlation between ‘Expected LLD at maturity with surgery’ and
‘Final LLD at maturity with surgery’, the greatest correlation being
with Method 2. Generally all the methods generated an overcorrected
value. No method generates the precise ‘Expected LLD at maturity
with surgery’. It is essential that an analysis of the pattern of
growth is taken into account when predicting final LLD. As many
additional data as possible are required. Cite this article:
We report our experience of using a computer
navigation system to aid resection of malignant musculoskeletal tumours
of the pelvis and limbs and, where appropriate, their subsequent
reconstruction. We also highlight circumstances in which navigation
should be used with caution. We resected a musculoskeletal tumour from 18 patients (15 male,
three female, mean age of 30 years (13 to 75) using commercially
available computer navigation software (Orthomap 3D) and assessed
its impact on the accuracy of our surgery. Of nine pelvic tumours,
three had a biological reconstruction with extracorporeal irradiation,
four underwent endoprosthetic replacement (EPR) and two required
no bony reconstruction. There were eight tumours of the bones of
the limbs. Four diaphyseal tumours underwent biological reconstruction.
Two patients with a sarcoma of the proximal femur and two with a
sarcoma of the proximal humerus underwent extra-articular resection
and, where appropriate, EPR. One soft-tissue sarcoma of the adductor
compartment which involved the femur was resected and reconstructed
using an EPR. Computer navigation was used to aid reconstruction
in eight patients. Histological examination of the resected specimens revealed tumour-free
margins in all patients. Post-operative radiographs and CT showed
that the resection and reconstruction had been carried out as planned
in all patients where navigation was used. In two patients, computer
navigation had to be abandoned and the operation was completed under
CT and radiological control. The use of computer navigation in musculoskeletal oncology allows
accurate identification of the local anatomy and can define the
extent of the tumour and proposed resection margins. Furthermore,
it helps in reconstruction of limb length, rotation and overall
alignment after resection of an appendicular tumour. Cite this article:
Achieving arthrodesis of the ankle can be difficult
in the presence of infection, deformity, poor soft tissues and bone loss.
We present a series of 48 patients with complex ankle pathology,
treated with the Ilizarov technique. Infection was present in 30
patients and 30 had significant deformity before surgery. Outcome
was assessed clinically and with patient-reported outcome measures
(Modified American Orthopaedic Foot and Ankle Society (MAOFAS) scale and
the Short-Form (SF-36)). Arthrodesis was achieved in 40 patients with the Ilizarov technique
alone and in six further patients with additional surgery. Infection
was eradicated in all patients at a mean follow-up of 46.6 months
(13 to 162). Successful arthrodesis was less likely in those with
comorbidities and in tibiocalcaneal fusion compared with tibiotalar
fusion. These patients had poor general health scores compared with the
normal population before surgery. The mean MAOFAS score improved
significantly from 24.3 (0 to 90) pre-operatively to 56.2 (30 to
90) post-operatively, but there was only a modest improvement in
general health; the mean SF-36 improved from 44.8 (19 to 66) to
50.1 (21 to 76). There was a major benefit in terms of pain relief. Arthrodesis using the Ilizarov technique is an effective treatment
for complex ankle pathology, with good clinical outcomes and eradication
of infection. However, even after successful arthrodesis general
health scores remain limited. Cite this article:
This review examines the future of total hip arthroplasty, aiming to avoid past mistakes
We report the outcome of 39 patients who underwent
a modified Pauwels’ intertrochanteric osteotomy for nonunion of
a femoral neck fracture following failed osteosynthesis. There were
31 men and eight women with a mean age of 47.2 years (34 to 59).
By Pauwels’ classification, there were 11 Type II fractures and
28 Type III fractures. The mean follow-up was 7.9 years (2 to 19).
In the 11 patients whose initial treatment had been osteotomy, union was
achieved in nine (81.8%). In 28 patients whose initial treatment
had been with a lag screw or a dynamic hip screw, union was achieved
in 27 (96.4%). Limb lengths were equalised in 14 of 16 patients
(87.5%) with pre-operative shortening. The mean neck-shaft angle
improved significantly from 100.5° (80° to 120°) to 131.6° (120°
to 155°) (p = 0.004). The mean modified Harris hip score was 85.6
points (70 to 97) and the mean modified Merle d’Aubigné score was
14.3 (11 to 18). Good to excellent functional outcomes were achieved
in 32 patients (88.8%). A modified Pauwels’ intertrochanteric osteotomy
is a reliable method of treating ununited fractures of the femoral neck
following failed osteosynthesis: coxa vara and shortening can also
simultaneously be addressed. Cite this article
The purpose of this study was to define immediate post-operative ‘quality’ in total hip replacements and to study prospectively the occurrence of failure based on these definitions of quality. The evaluation and assessment of failure were based on ten radiological and clinical criteria. The cumulative summation (CUSUM) test was used to study 200 procedures over a one-year period. Technical criteria defined failure in 17 cases (8.5%), those related to the femoral component in nine (4.5%), the acetabular component in 32 (16%) and those relating to discharge from hospital in five (2.5%). Overall, the procedure was considered to have failed in 57 of the 200 total hip replacements (28.5%). The use of a new design of acetabular component was associated with more failures. For the CUSUM test, the level of adequate performance was set at a rate of failure of 20% and the level of inadequate performance set at a failure rate of 40%; no alarm was raised by the test, indicating that there was no evidence of inadequate performance. The use of a continuous monitoring statistical method is useful to ensure that the quality of total hip replacement is maintained, especially as newer implants are introduced.
We compared the rate of revision for instability
after total hip replacement (THR) when lipped and non-lipped acetabular
liners were used. We hypothesised that the use of a lipped liner
in a modular uncemented acetabular component reduces the risk of
revision for instability after primary THR. Using data from the
New Zealand Joint Registry, we found that the use of a lipped liner
was associated with a significantly decreased rate of revision for instability
and for all other indications. Adjusting for the size of the femoral
head, the surgical approach and the age and gender of the patient,
this difference remained strongly significant (p <
0.001). We conclude that evidence from the New Zealand registry suggests
that the use of lipped liners with modular uncemented acetabular
components is associated with a decreased rate of revision for instability
after primary THR. Cite this article:
Despite a lack of long-term follow-up, there
is an increasing trend towards using femoral heads of large diameter
in total hip replacement (THR), partly because of the perceived
advantage of lower rates of dislocation. However, increasing the
size of the femoral head is not the only way to reduce the rate
of dislocation; optimal alignment of the components and repair of
the posterior capsule could achieve a similar effect. In this prospective study of 512 cemented unilateral THRs (Male:Female
230:282) performed between 2004 and 2011, we aimed to determine
the rate of dislocation in patients who received a 22 mm head on
a 9/10 Morse taper through a posterior approach with capsular repair
and using the transverse acetabular ligament (TAL) as a guide for the
alignment of the acetabular component. The mean age of the patients
at operation was 67 years (35 to 89). The mean follow-up was 2.8
years (0.5 to 6.6). Pre- and post-operative assessment included
Oxford hip, Short Form-12 and modified University of California
Los Angeles and Merle D’Aubigne scores. The angles of inclination
and anteversion of the acetabular components were measured using
radiological software. There were four dislocations (0.78%), all
of which were anterior. In conclusion, THR with a 22 mm diameter head performed through
a posterior approach with capsular repair and using the TAL as a
guide for the alignment of the acetabular component was associated
with a low rate of dislocation. Cite this article:
We assessed leg length discrepancy and hip function in 90 patients undergoing primary total hip arthroplasty before surgery and at three and 12 months after. Function was measured using the Oxford hip score (OHS). After surgery the mean OHS improved by 26 points after three months and by 30 points after 12. After operation 56 (62%) limbs were long by a mean of 9 mm and this was perceived by 24 (43%) patients after three months and by 18 (33%) after 12. The mean OHS in patients who perceived true lengthening was 27% worse than the rest of the population after three months and 18% worse after 12. In 55 (98%) patients, lengthening occurred in the femoral component. Appropriate placement of the femoral component could significantly reduce a patient’s perception of discrepancy of length.
The aim of this study was to evaluate the functional
and oncological outcome of extracorporeally irradiated autografts
used to reconstruct the pelvis after a P1/2 internal hemipelvectomy. The study included 18 patients with a primary malignant bone
tumour of the pelvis. There were 13 males and five females with
a mean age of 24.8 years (8 to 62). Of these, seven had an osteogenic
sarcoma, six a Ewing’s sarcoma, and five a chondrosarcoma. At a
mean follow-up of 51.6 months (4 to 185), nine patients had died
with metastatic disease while nine were free from disease. Local
recurrence occurred in three patients all of whom eventually died of
their disease. Deep infection occurred in three patients and required
removal of their graft in two while the third underwent a hindquarter
amputation for extensive flap necrosis. The mean Musculoskeletal Tumor Society functional score of the
16 patients who could be followed-up for at least 12 months was
77% (50 to 90). Those 15 patients who completed the Toronto Extremity
Salvage Score questionnaire had a mean score of 71% (53 to 85). Extracorporeal irradiation and re-implantation of bone is a valid
method of reconstruction after an internal hemipelvectomy. It has
an acceptable morbidity and a functional outcome that compares favourably
with other available reconstructive techniques. Cite this article:
The aim of this study was to determine whether
an osteoplasty of the femoral neck performed at the same time as an
intertrochanteric Imhäuser osteotomy led to an improved functional
outcome or increased morbidity. A total of 20 hips in 19 patients
(12 left, 8 right, 13 male, 6 female), who underwent an Imhäuser
intertrochanteric osteotomy following a slipped capital femoral
epiphysis were assessed over an eight-year period. A total of 13
hips in 13 patients had an osteoplasty of the femoral neck at the
same time. The remaining six patients (seven hips) had intertrochanteric
osteotomy alone. The mean age was 15.3 years (13 to 20) with a mean
follow-up of 57.8 months (15 to 117); 19 of the slips were severe
(Southwick grade III) and one was moderate (grade II), with a mean
slip angle of 65.3° (50° to 80°); 17 of the slips were stable and
three unstable at initial presentation. The mean Non-Arthritic Hip Scores
(NAHS) in patients who underwent osteoplasty was 91.7 (76.3 to 100)
and the mean NAHS in patients who did not undergo osteoplasty was
76.6 (41.3 to 100) (p = 0.056). Two patients required a subsequent
arthroplasty and neither of these patients had an osteoplasty. No
hips developed osteonecrosis or chondrolysis, and there was no increase
in complications related to the osteoplasty. We recommend that for
patients with a slipped upper femoral epiphysis undergoing an intertrochanteric
osteotomy, the addition of an osteoplasty of the femoral neck should
be considered. Cite this article:
We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone regeneration and subsidence were assessed for 101
hips. The mean Harris Hip Score was 88.2 (45.8 to 100) after five years
and there was an increase of the mean Barnett and Nordin-Score,
a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7).
Five stems had to be revised (4.3%), three (2.9%) showed subsidence,
five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%)
a periprosthetic infection. At the latest follow-up, the survival with revision of the stem
as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%)
and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic
fractures were not observed. We report excellent results with respect to subsidence, the risk
of fracture, and loosening after femoral revision using a modular
curved revision stem with distal cone-in-cone fixation. A successful
outcome depends on careful pre-operative planning and the use of
a transfemoral approach when the anatomy is distorted or a fracture
is imminent, or residual cement or a partially-secured existing
stem cannot be removed. The shortest appropriate stem should, in
our opinion, be used and secured with >
3 cm fixation at the femoral
isthmus, and distal interlocking screws should be used for additional
stability when this goal cannot be realised. Cite this article:
We have compared the biomechanical nature of the reconstruction of the hip in conventional total hip arthroplasty (THA) and surface replacement arthroplasty (SRA) in a randomised study involving 120 patients undergoing unilateral primary hip replacement. The contralateral hip was used as a control. Post-operatively, the femoral offset was significantly increased with THA (mean 5.1 mm; −2.8 to 11.6) and decreased with SRA (mean −3.3 mm; −8.9 to 8.2). Femoral offset was restored within Restoration of the normal proximal femoral anatomy was more precise with SRA. The enhanced stability afforded by the use of a large-diameter femoral head avoided over-lengthening of the limb or increased offset to improve soft-tissue tension as occurs sometimes in THA. In a subgroup of patients with significant pre-operative deformity, restoration of the normal hip anatomy with lower pre-operative femoral offset or significant shortening of the leg was still possible with SRA.
Endoprosthetic reconstruction following resection of 31 tumours of the proximal femur in 30 patients was performed using a Wagner SL femoral revision stem. The mean follow-up was 25.6 months (0.6 to 130.0). Of the 28 patients with a metastasis, 27 died within a mean follow-up period of 18.1 months (0.6 to 56.3) after the operation, and the remaining patient was excluded from the study 44.4 months post-operatively when the stem was removed. The two patients with primary bone tumours were still alive at the latest follow-up of 81.0 and 130.0 months, respectively. One stem only was removed for suspected low-grade infection 44.4 months post-operatively. The worst-case survival rate with removal of the stem for any cause and/or loss to follow-up was 80.0% (95% confidence interval 44.9 to 100) at 130.0 months. The mean Karnofsky index increased from 44.2% (20% to 70%) pre-operatively to 59.7% (0% to 100%) post-operatively, and the mean Merle d’Aubigné score improved from 4.5 (0 to 15) to 12.0 (0 to 18). The mean post-operative Musculoskeletal Tumour Society score was 62.4% (3.3% to 100%). The Wagner SL femoral revision stem offers an alternative to special tumour prostheses for the treatment of primary and secondary tumours of the proximal femur. The mid-term results are very promising, but long-term experience is necessary.
We report a case of spontaneous physeal growth arrest of the distal femur in a nine-year-old child with Ewing’s sarcoma of the proximal femur treated with chemotherapy and endoprosthetic replacement. Owing to the extent of disuse osteoporosis at the time of surgery, the entire intramedullary canal up to the distal femoral physis was filled with cement. Three years later, the femur remained at its pre-operative length of 19 cm. Pre-operative calculations of further growth failed to account for the growth arrest, and the initial expandable growing prosthesis inserted has been revised to a longer one in order to address the leg-length discrepancy. To our knowledge, this is the only reported case of distal femoral physeal growth arrest following cemented endoprosthetic replacement of the proximal femur.