Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory pseudotumours, arising around these devices. They may cause soft-tissue destruction with severe symptoms and a poor outcome after revision surgery. The aim of this study was to determine the incidence of and risk factors for pseudotumours that are serious enough to require revision surgery. Since 1999, 1419 metal-on-metal hip resurfacings have been implanted by our group in 1224 patients; 1.8% of the patients had a revision for pseudotumour. In this series the Kaplan-Meier cumulative revision rate for pseudotumour increased progressively with time. At eight years, in all patients, it was 4% (95% confidence interval (CI) 2.2 to 5.8). Factors significantly associated with an increase in revision rate were female gender (p <
0.001), age under 40 (p = 0.003), small components (p = 0.003), and dysplasia (p = 0.019), whereas implant type was not (p = 0.156). These factors were inter-related, however, and on fitting a Cox proportional hazard model only gender (p = 0.002) and age (p = 0.024) had a significant independent influence on revision rate; size nearly reached significance (p = 0.08). Subdividing the cohort according to significant factors, we found that the revision rate for pseudotumours in men was 0.5% (95% CI 0 to 1.1) at eight years wheras in women over 40 years old it was 6% (95% CI 2.3 to 10.1) at eight years and in women under 40 years it was 13.1% at six years (95% CI 0 to 27) (p <
0.001). We recommend that resurfacings are undertaken with caution in women, particularly those under 40 years of age but they remain a good option in young men. Further work is required to understand the aetiology of pseudotumours so that this complication can be avoided.
Total joint arthroplasty (TJA) is commonly performed in elderly
patients. Frailty, an aggregate expression of vulnerability, becomes
increasingly common with advanced age, and independently predicts
adverse outcomes and the use of resources after a variety of non-cardiac
surgical procedures. Our aim was to assess the impact of frailty
on outcomes after TJA. We analysed the impact of pre-operative frailty on death and
the use of resources after elective TJA in a population-based cohort
study using linked administrative data from Ontario, Canada.Aims
Patients and Methods
Juvenile idiopathic arthritis (JIA) is a chronic disease of childhood; it causes joint damage which may require surgical intervention, often in the young adult. The aim of this study was to describe the long-term outcome and survival of hip replacement in a group of adult patients with JIA and to determine predictors of survival for the prosthesis. In this retrospective comparative study patients were identified from the database of a regional specialist adult JIA clinic. This documented a series of 47 hip replacements performed in 25 adult patients with JIA. Surgery was performed at a mean age of 27 years (11 to 47), with a mean follow-up of 19 years (2 to 36). The mean Western Ontario and McMaster Universities osteoarthritis index questionnaire (WOMAC) score at the last follow-up was 53 (19 to 96) and the mean Health Assessment Questionnaire score was 2.25 (0 to 3). The mean pain component of the WOMAC score (60 (20 to 100)) was significantly higher than the mean functional component score (46 (0 to 97)) (p = 0.02). Kaplan-Meier survival analysis revealed a survival probability of 46.6% (95% confidence interval 37.5 to 55.7) at 19 years, with a trend towards enhanced survival with the use of a cemented acetabular component and a cementless femoral component. This was not, however, statistically significant (acetabular component, p = 0.76, femoral component, p = 0.45). Cox’s proportional hazards regression analysis showed an implant survival rate of 54.9% at 19 years at the mean of covariates. Survival of the prosthesis was significantly poorer (p = 0.001) in patients who had been taking long-term corticosteroids and significantly better (p = 0.02) in patients on methotrexate.
To investigate the longevity of uncemented fixation of a femoral
component in total hip arthroplasty (THA) in patients with Dorr
type C proximal femoral morphology. A total of 350 consecutive uncemented THA in 320 patients were
performed between 1983 and 1987, by a single surgeon using the Taperloc
femoral component. The 63 patients (68 hips) with Dorr type C proximal
femoral morphology were the focus of this review. The mean age of
the patients was 69 years (24 to 88) and mean follow-up was 16.6
years (ten to 29). Survival analysis included eight patients (eight
hips) who died without undergoing revision surgery prior to obtaining
ten years follow-up. All 55 surviving patients (60 hips) were available
for clinical assessment and radiographic review. As a comparator
group, the survival and implant fixation in the remaining 282 THAs
(257 patients) with Dorr type A and B morphology were evaluated.
The mean age of these patients was 52 years (20 to 82).Aims
Patients and Methods
The practice of removing a well-fixed cementless
femoral component is associated with high morbidity. Ceramic bearing
couples are low wearing and their use minimises the risk of subsequent
further revision due to the production of wear debris. A total of
165 revision hip replacements were performed, in which a polyethylene-lined acetabular
component was revised to a new acetabular component with a ceramic
liner, while retaining the well-fixed femoral component. A titanium
sleeve was placed over the used femoral trunnion, to which a ceramic
head was added. There were 100 alumina and 65 Delta bearing couples
inserted. The mean Harris hip score improved significantly from 71.3 (9.0
to 100.0) pre-operatively to 91.0 (41.0 to 100.0) at a mean follow
up of 4.8 years (2.1 to 12.5) (p <
0.001). No patients reported
squeaking of the hip. There were two fractures of the ceramic head, both in alumina
bearings. No liners were seen to fracture. No fractures were observed
in components made of Delta ceramic. At 8.3 years post-operatively
the survival with any cause of failure as the endpoint was 96.6%
(95% confidence interval (CI) 85.7 to 99.3) for the acetabular component and
94.0% (95% CI 82.1 to 98.4) for the femoral component. The technique of revising the acetabular component in the presence
of a well-fixed femoral component with a ceramic head placed on
a titanium sleeve over the used trunnion is a useful adjunct in
revision hip practice. The use of Delta ceramic is recommended. Cite this article:
Hip resurfacing arthroplasty (HRA) is an alternative to conventional
total hip arthroplasty for patients with osteonecrosis (ON) of the
femoral head. Our aim was to report the long-term outcome of HRA,
which is not currently known. Long-term survivorship, clinical scores and radiographic results
for 82 patients (99 hips) treated with HRA for ON over a period
of 18 years were reviewed retrospectively. The mean age of the 67
men and 15 women at the time of surgery was 40.8 years (14 to 64).
Patients were resurfaced regardless of the size of the osteonecrotic
lesion.Aims
Patients and Methods
The inherent challenges of total hip replacement
(THR) in children include the choice of implant for the often atypical
anatomical morphology, its fixation to an immature growing skeleton
and the bearing surface employed to achieve a successful long-term
result. We report the medium-term results of 52 consecutive uncemented
THRs undertaken in
35 paediatric patients with juvenile idiopathic arthritis. The mean
age at the time of surgery was 14.4 years (10 to 16). The median
follow-up was 10.5 years (6 to 15). During the study period 13 THRs
underwent revision surgery. With revision as an endpoint, subgroup
analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs
and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This
resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral
component and 62% (95% CI 41.0 to 78.0) of the acetabular component.
Revision of the acetabular component for wear and osteolysis were
the most common reasons for failure accounting for 11 of the 13
revisions. The success seen in patients with a ceramic-on-ceramic articulation
seems to indicate that this implant strategy has the potential to
make a major difference to the long-term outcome in this difficult
group of patients.
We report the survival, radiological and functional outcomes of a single surgeon series of his first 144 consecutive Birmingham hip resurfacing procedures (130 patients) at a minimum of ten years. There were ten revisions during this time. Although no patients were lost to follow-up some did not complete the scoring assessment or undergo radiological assessment at ten years. The ten-year survival for male patients was 98.0% (95% confidence interval 95.2 to 100). The ten-year survival for the total cohort with aseptic revision as the endpoint was 95.5% (95% confidence interval 91.8 to 99.0) and including revisions for sepsis was 93.5% (95% confidence interval 89.2 to 97.6). The median modified Oxford hip score at ten years was 4.2% (interquartile range 0 to 19) and the median University of California, Los Angeles score was 7.0 (interquartile range 5.0 to 8.0). This study confirms the midterm reports that metal-on-metal hip resurfacing using the Birmingham Hip provides a durable alternative to total hip replacement, particularly in younger male patients wishing to maintain a high level of function, with low risk of revision for at least ten years.
We retrospectively compared the outcome after
the treatment of giant cell tumours of bone either with curettage alone
or with adjuvant cementation. Between 1975 and 2008, 330 patients
with a giant cell tumour were treated primarily by intralesional
curettage, with 84 (25%) receiving adjuvant bone cement in the cavity.
The local recurrence rate for curettage alone was 29.7% (73 of 246)
compared with 14.3% (12 of 84) for curettage and cementation (p
= 0.001). On multivariate analysis both the stage of disease and
use of cement were independent significant factors associated with
local recurrence. The use of cement was associated with a higher
risk of the subsequent need for joint replacement. In patients without
local recurrence, 18.1% (13 of 72) of those with cement needed a
subsequent joint replacement compared to 2.3% (4 of 173) of those
without cement (p = 0.001). In patients who developed local recurrence,
75.0% (9 of 12) of those with previous cementation required a joint
replacement, compared with 45.2% (33 of 73) of those without cement
(p = 0.044).
We investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the distal femur after resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database, review of the design of the implant and case records. In total, 335 patients underwent a distal femoral replacement, 162 having a fixed-hinge design and 173 a rotating-hinge. The median age of the patients was 24 years (interquartile range 17 to 48). A total of 192 patients remained alive with a mean follow-up of 12 years (5 to 30). The risk of revision for any reason was 17% at five years, 33% at ten years and 58% at 20 years. Aseptic loosening was the main reason for revision of the fixed-hinge knees while infection and fracture of the stem were the most common for the rotating-hinge implant. The risk of revision for aseptic loosening was 35% at ten years with the fixed-hinge knee, which has, however, been replaced by the rotating-hinge knee with a hydroxyapatite collar. The overall risk of revision for any reason fell by 52% when the rotating-hinge implant was used. Improvements in the design of distal femoral endoprostheses have significantly decreased the need for revision operations, but infection remains a serious problem. We believe that a cemented, rotating-hinge prosthesis with a hydroxyapatite collar offers the best chance of long-term survival of the prosthesis.
We report on the long-term results of 163 bicruciate-retaining
Hermes 2C total knee replacements in 130 patients at a mean follow-up
of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament
had a partially degenerative appearance it was preserved as long
as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The
description and surgical technique of this minimally constrained
prosthesis were published in 1983 and the ten-year clinical results
in 1999. A total of 12% of the knees (20 of 163) in this study were revised
because of wear of the polyethylene tibial insert. Excellent stability
was achieved and the incidence of aseptic component loosening was
4.3% (seven of 163). The survival rate using revision for any reason
as the endpoint was 82% (95% confidence interval 76.2 to 88.0). Although this series included a relatively small number of replacements,
it demonstrated that the anterior cruciate ligament, even when partially
degenerated at the time of TKR, remained functional and provided
adequate stability at a long-term follow-up. Cite this article:
We evaluated the outcome of 41 consecutive Charnley
low-friction arthroplasties (LFAs) performed by a single surgeon
in 28 patients aged ≤ 35 years at operation between 23 and 36 years
previously. There were 20 women and eight men with a mean age of
32 years (23 to 35) at surgery. Two patients (three hips) were lost
to follow-up at 12 and 17 years post-operatively, respectively,
and one patient (one hip) died at 13 years post-operatively. These patients
were excluded from the final evaluation. The survival rate of the
acetabular components was 92.7% (95% confidence interval (CI) 88.7
to 96.7) at ten years, 67.1% (95% CI 59.75 to 74.45) at 20 years
and 53.2% (95% CI 45.3 to 61.1) at 25 years. For the femoral component
the survival was 95.1% (95% CI 91.8 to 98.5) at ten years, 77.1%
(95% CI 73.9 to 80.3) at 20 years and 68.2% (95% CI 60.7 to 75.8)
at 25 years. The results indicate that the Charnley LFA remains
a reasonable choice in the treatment of young patients and can serve
for comparison with newer techniques and implants. Cite this article:
Survival analysis is an important tool for assessing the outcome of total joint replacement. The Kaplan-Meier method is used to estimate the incidence of revision of a prosthesis over time, but does not account appropriately for competing events which preclude revision. In the presence of competing death, this method will lead to statistical bias and the curve will lose its interpretability. A valid comparison of survival results between studies using the method is impossible without accounting for different rates of competing events. An alternative and easily applicable approach, the cumulative incidence of competing risk, is proposed. Using three simulated data sets and realistic data from a cohort of 406 consecutive cementless total hip prostheses, followed up for a minimum of ten years, both approaches were compared and the magnitude of potential bias was highlighted. The Kaplan-Meier method overestimated the incidence of revision by almost 4% (60% relative difference) in the simulations and more than 1% (31.3% relative difference) in the realistic data set. The cumulative incidence of competing risk approach allows for appropriate accounting of competing risk and, as such, offers an improved ability to compare survival results across studies.
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
We evaluated the impact of pre-coating the tibial
component with polymethylmethacrylate (PMMA) on implant survival
in a cohort of 16 548 primary NexGen total knee replacements (TKRs)
in 14 113 patients. In 13 835 TKRs a pre-coated tray was used while
in 2713 TKRs the non-pre-coated version of the same tray was used.
All the TKRs were performed between 2001 and 2009 and were cemented.
TKRs implanted with a pre-coated tibial component had a lower cumulative
survival than those with a non-pre-coated tibial component (p =
0.01). After adjusting for diagnosis, age, gender, body mass index,
American Society of Anesthesiologists grade, femoral coupling design, surgeon
volume and hospital volume, pre-coating was an independent risk
factor for all-cause aseptic revision (hazard ratio 2.75, p = 0.006).
Revision for aseptic loosening was uncommon for both pre-coated
and non-pre-coated trays (rates of 0.12% and 0%, respectively).
Pre-coating with PMMA does not appear to be protective of revision
for this tibial tray design at short-term follow-up. Cite this article:
We undertook a randomised controlled trial to
compare bipolar hemiarthroplasty (HA) with a novel total hip replacement
(THR) comprising a polycarbonate–urethane (PCU) acetabular component
coupled with a large-diameter metal femoral head for the treatment
of displaced fractures of the femoral neck in elderly patients. Functional
outcome, assessed with the Harris hip score (HHS) at three months
and then annually after surgery, was the primary endpoint. Rates
of revision and complication were secondary endpoints. Based on a power analysis, 96 consecutive patients aged >
70
years were randomised to receive either HA (49) or a PCU-THR (47).
The mean follow-up was 30.1 months (23 to 50) and 28.6 months (22
to 52) for the HA and the PCU group, respectively. The HHS showed no statistically significant difference between
the groups at every follow-up. Higher pain was recorded in the PCU
group at one and two years’ follow-up
(p = 0.006 and p = 0.019, respectively). In the HA group no revision
was performed. In the PCU-THR group six patients underwent revision
and one patient is currently awaiting
re-operation. The three-year survival rate of the PCU-THR group
was 0.841 (95% confidence interval 0.680 to 0.926). Based on our findings we do not recommend the use of the PCU
acetabular component as part of the treatment of patients with fractures
of the femoral neck. Cite this article:
In a randomised controlled pragmatic trial we
investigated whether local infiltration analgesia would result in earlier
readiness for discharge from hospital after total knee replacement
(TKR) than patient-controlled epidural analgesia (PCEA) plus femoral
nerve block. A total of 45 patients with a mean age of 65 years
(49 to 81) received a local infiltration with a peri-articular injection
of bupivacaine, morphine and methylprednisolone, as well as adjuvant
analgesics. In 45 PCEA+femoral nerve blockade patients with a mean
age of 67 years (50 to 84), analgesia included a bupivacaine nerve
block, bupivacaine/hydromorphone PCEA, and adjuvant analgesics.
The mean time until ready for discharge was 3.2 days (1 to 14) in
the local infiltration group and 3.2 days (1.8 to 7.0) in the PCEA+femoral
nerve blockade group. The mean pain scores for patients receiving
local infiltration were higher when walking (p = 0.0084), but there
were no statistically significant differences at rest. The mean
opioid consumption was higher in those receiving local infiltration. The choice between these two analgesic pathways should not be
made on the basis of time to discharge after surgery. Most secondary
outcomes were similar, but PCEA+femoral nerve blockade patients
had lower pain scores when walking and during continuous passive
movement. If PCEA+femoral nerve blockade is not readily available, local
infiltration provides similar length of stay and similar pain scores
at rest following TKR. Cite this article:
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
We treated 98 patients with peri-acetabular tumours by resection and reconstruction with a custom-made pelvic endoprosthesis. The overall survival of the patients was 67% at five years, 54% at ten years and 51% at 30 years. One or more complications occurred in 58.1% of patients (54), of which infection was the most common, affecting 30% (28 patients). The rate of local recurrence was 31% (29 patients) after a mean follow-up of 71 months (11 to 147). Dislocation occurred in 20% of patients (19). Before 1996 the rate was 40.5% (17 patients) but this was reduced to 3.9% (two patients) with the introduction of a larger femoral head. There were six cases of palsy of the femoral nerve with recovery in only two. Revision or excision arthroplasty was performed in 23.7% of patients (22), principally for uncontrolled infection or aseptic loosening. Higher rates of death, infection and revision occurred in men. This method of treatment is still associated with high morbidity. Patients should be carefully selected and informed of this pre-operatively.
We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips). Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant. This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.