Advertisement for orthosearch.org.uk
Results 1 - 20 of 384
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 896 - 902
1 Jul 2009
Gulati A Chau R Pandit HG Gray H Price AJ Dodd CAF Murray DW

Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement tibial components have been referred to as physiological radiolucencies. These should be distinguished from pathological radiolucencies, which are poorly defined, wide and progressive, and associated with loosening and infection. We studied the incidence and clinical significance of tibial radiolucent lines in 161 Oxford unicondylar knee replacements five years after surgery. All the radiographs were aligned with fluoroscopic control to obtain views parallel to the tibial tray to reveal the tibial bone-implant interface. We found that 49 knees (30%) had complete, 52 (32%) had partial and 60 (37%) had no radiolucent lines. There was no relationship between the incidence of radiolucent lines and patient factors such as gender, body mass index and activity, or operative factors including the status of the anterior cruciate ligament and residual varus deformity. Nor was any statistical relationship established between the presence of radiolucent lines and clinical outcome, particularly pain, assessed by the Oxford Knee score and the American Knee Society score. We conclude that radiolucent lines are common after Oxford unicompartmental knee replacement but that their aetiology remains unclear. Radiolucent lines were not a source of adverse symptoms or pain. Therefore, when attempting to identify a source of postoperative pain after Oxford unicompartmental knee replacement the presence of a physiological radiolucency should be ignored


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 185 - 189
1 Feb 2009
Pandit H Jenkins C Beard DJ Gallagher J Price AJ Dodd CAF Goodfellow JW Murray DW

We randomised 62 knees to receive either cemented or cementless versions of the Oxford unicompartmental knee replacement. The implants used in both arms of the study were similar, except that the cementless components were coated with porous titanium and hydroxyapatite. The tibial interfaces were studied with fluoroscopically-aligned radiographs. At one year there was no difference in clinical outcome between the two groups. Narrow radiolucent lines were seen at the bone-implant interfaces in 75% of cemented tibial components. These were partial in 43%, and complete in 32%. In the cementless implants, partial radiolucencies were seen in 7% and complete radiolucencies in none. These differences are statistically significant (p < 0.0001) and imply satisfactory bone ingrowth into the cementless implants


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 83 - 90
1 Jan 2022
Batten TJ Gallacher S Evans JP Harding RJ Kitson J Smith CD Thomas WJ

Aims. The use and variety of stemless humeral components in anatomical total shoulder arthroplasty (TSA) have proliferated since their advent in 2004. Early outcomes are reassuring but independent mid-term results are scarce. This independent study reports a consecutive series of 143 Eclipse stemless shoulder prostheses with a minimum five-year (5 to 10) follow-up. Methods. Outcomes of 143 procedures undertaken for all indications in 131 patients were reviewed, with subset analysis of those for osteoarthritis (OA) (n = 99). The primary outcome was the Oxford Shoulder Score (OSS) at a minimum of five years. Secondary outcomes were ranges of motion and radiological analysis of humeral radiolucency, rotator cuff failure, and glenoid loosening. Results. Mean OSS at mean follow-up of 6.67 years (5.0 to 10.74) was 40.12 (9 to 48), with no statistically significant difference between those implanted for a non-OA indication and those for OA (p = 0.056) or time-dependent deterioration between two years and five years (p = 0.206). Ranges of motion significantly improved compared with preoperative findings and were maintained between two and five years with a mean external rotation of 38° (SD 18.1, 0 to 100) and forward elevation of 152° (SD 29.9, 90 to 180). Of those components with radiographs suitable for analysis (n = 83), 23 (28%) were found to have a least one humeral radiolucent line, which were predominantly incomplete, less than 2 mm, and in a single anatomical zone. No humeral components were loose. A radiolucent line was present around 22 (15%) of glenoid components, and 15 (10%) of components had failed. Rotator cuff failure was found in 21 (15%) components. The mean time to either glenoid or rotator cuff failure was greater than three years following implantation. Survivorship was 96.4% (95% CI 91.6 to 98.5, number at risk 128) at five years, and 94.3% (95% CI 88.2 to 97.3, number at risk 76) at seven years, both of which compare favourably with best results taken from available registries. Conclusion. Functional and radiological outcomes of the Eclipse stemless TSA are excellent, with no loose humeral components at minimum five-year follow-up. The presence of radiolucent lines is of interest and requires long-term observation but does not impact on the clinical results. Of the eight revisions required, this was predominantly for glenoid and rotator cuff failure. Cite this article: Bone Joint J 2022;104-B(1):83–90


Bone & Joint Open
Vol. 5, Issue 12 | Pages 1092 - 1100
8 Dec 2024
Fraser E Spence S Farhan-Alanie OM Doonan J Mahendra A Gupta S

Aims. Limb salvage surgery (LSS) is the primary treatment option for primary bone malignancy. It involves the removal of bone and tissue, followed by reconstruction with endoprosthetic replacements (EPRs) to prevent amputation. Trabecular metal (TM) collars have been developed to encourage bone ingrowth (osseointegration (OI)) into EPRs. The primary aim of this study was to assess whether OI occurs when TM collars are used in EPRs for tumour. Methods. A total of 124 patients from July 2010 to August 2021 who underwent an EPR for tumour under the West of Scotland orthopaedic oncology team were identified. Overall, 81 patients (65%) met the inclusion criteria, and two consultants independently analyzed radiographs at three and 12 months, as well as the last radiograph, using a modified version of the Stanford Radiological Assessment System. Results. OI of the TM collar occurred in approximately 65% of patients at last radiograph. The percentage of patients with OI at three months (65.4%) reflected the 12-month (65%) and long-term (64.4%) follow-up. The median amount of OI across all radiographs was one at all three timepoints, with only five cases (11.1%) showing OI in all four zones at last radiograph. Radiolucency at the bone:collar junction was present in 23 cases (28.4%) at three months, but only four (6.7%) showed progression of this at 12 months. The interobserver reliability was found to be highly reliable in all parameters (p < 0.001). Conclusion. OI occurs in approximately 65% of TM collars, and is similar at three months, 12 months, and last radiograph. The extent of OI at the bone:collar junction was found to have decreased at longer-term follow-up. Furthermore, radiolucency at the bone-collar impact junction does occur in some patients but only a low number will show radiolucency progression at longer-term follow-up. Cite this article: Bone Jt Open 2024;5(12):1092–1100


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 760 - 767
1 Jul 2019
Galea VP Rojanasopondist P Laursen M Muratoglu OK Malchau H Bragdon C

Aims. Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years. Patients and Methods. A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45). Radiostereometric analysis (RSA) was used to measure polyethylene wear and component migration. Differences in wear were assessed while adjusting for body mass index, activity level, acetabular inclination, anteversion, and head size. Plain radiographs were assessed for radiolucency and patient-reported outcome measures (PROMs) were administered at each follow-up. Results. In total, 73 patients (82%) completed the seven-year visit. Mean seven-year linear proximal penetration was -0.07 mm (. sd. 0.16) and 0.00 mm (. sd. 0.22) for the VEPE and ModXLPE cohorts, respectively (p = 0.116). PROMs (p = 0.310 to 0.807) and radiolucency incidence (p = 0.330) were not different between the polyethylene cohorts. The mean proximal shell migration rate was 0.04 mm per year (. sd. 0.09). At seven years, patients with radiolucency (34%) demonstrated greater migration (mean difference: 0.6 mm (. sd. 0.2); p < 0.001). PROMs were lower for patients with radiolucency and greater proximal migration (p = 0.009 to p = 0.045). No implants were revised for aseptic loosening. Conclusion. This is the first randomized controlled trial to report seven-year RSA results for VEPE. All wear rates were below the previously reported osteolysis threshold (0.1 mm per year). PTC shells demonstrated acceptable primary stability through seven years, as indicated by low migration and lack of aseptic loosening. However, patients with acetabular radiolucency were associated with higher shell migration and lower PROM scores. Cite this article: Bone Joint J 2019;101-B:760–767


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims. We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. Methods. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes. Results. At two years, 44 patients in each group were available for RSA assessment. The median total two-year proximal head penetration was -0.02 mm (IQR -0.09 to 0.07; p = 0.548) for the largest possible head and -0.01 mm (IQR -0.07 to 0.10; p = 0.525) for 32 mm heads. Their difference was not statistically significant (p = 0.323). Neither group demonstrated a bedding-in period. The median steady-state volumetric wear rates were 6.1 mm. 3. /year (IQR -59 to 57) and 3.5 mm. 3. /year (-21 to 34) respectively, and did not differ between the groups (p = 0.848). There were no statistically significant differences in periacetabular radiolucencies or patient-reported outcomes. Conclusion. The use of the largest possible metal head did not increase vitamin E-doped cross-linked polyethylene wear compared with 32 mm heads at two years. Linear wear was negligible and volumetric wear rates were very low in both head size groups. There was a tendency towards higher values of volumetric wear in large heads that warrants longer-term evaluation before any definite conclusions about the association between head size and wear can be drawn. Cite this article: Bone Joint J 2021;103-B(7):1206–1214


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 82 - 88
1 May 2024
Villa JM Rajschmir K Hosseinzadeh S Manrique-Succar J Grieco P Higuera-Rueda CA Riesgo AM

Aims. Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation. Methods. A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707). Results. Out of the 33 cases analyzed, six (18.2%) constructs required revision, with four revisions due to uncontrolled infection, one for dislocation, and one for aseptic loosening. Among the 27 non-revised constructs, only one showed wider radiolucencies compared to immediate postoperative radiographs, indicating potential loosening. Patients who underwent revision (n = 6) were significantly younger and had a higher BMI compared to those with non-revised constructs (p = 0.016 and p = 0.026, respectively). Sex, race, ethnicity, American Society of Anesthesiologists grade, infection status (patients with postoperative PJI diagnosis (septic) vs patients without such diagnosis (aseptic)), and mean follow-up did not significantly differ between revised and non-revised groups. Conclusion. The HiRISC technique may serve as a feasible short-term (about one to two years) alternative in patients with large acetabular defects, particularly in cases of PJI. Longer follow-up is necessary to establish the long-term survival of this technique. Cite this article: Bone Joint J 2024;106-B(5 Supple B):82–88


Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves modular cementless, tapered, revision femoral components in conjunction with distal interlocking screws. The aim of this study was to analyze the rate of reoperations and complications and union of the fracture, subsidence of the stem, mortality, and the clinical outcomes in these patients. Methods. A total of 46 femoral components in patients with Vancouver B3 fractures (23 with Paprosky type IIIA, 19 with type IIIB, and four with type IV defects) in 46 patients were revised with a transfemoral approach using a modular, tapered, cementless revision Revitan curved femoral component with distal cone-in-cone fixation and prospectively followed for a mean of 48.8 months (SD 23.9; 24 to 112). The mean age of the patients was 80.4 years (66 to 100). Additional distal interlocking was also used in 23 fractures in which distal cone-in-cone fixation in the isthmus was < 3 cm. Results. One patient (2.2%) died during the first postoperative year. After six months, 43 patients (93.5%) had osseous, and three had fibrous consolidation of the fracture and the bony flap, 42 (91.3%) had bony ingrowth and four had stable fibrous fixation of the stem. No patient had radiolucency around the interlocking screws and no screw broke. One patient had non-progressive subsidence and two had a dislocation. The mean Harris Hip Score increased from of 57.8 points (SD 7.9) three months postoperatively to 76.1 points (SD 10.7) 24 months postoperatively. Conclusion. The 2° tapered, fluted revision femoral component with distal cone-in-cone-fixation, combined with additional distal interlocking in patients with bony deficiency at the isthmus, led to reproducibly good results in patients with a Vancouver B3 fracture and Paprosky type IIIA, IIIB, and IV defects with regard to union of the fracture, subsidence or loosening of the stem, and clinical outcomes. Cite this article: Bone Joint J 2024;106-B(4):344–351


Bone & Joint Open
Vol. 3, Issue 8 | Pages 648 - 655
1 Aug 2022
Yeung CM Bhashyam AR Groot OQ Merchan N Newman ET Raskin KA Lozano-Calderón SA

Aims. Due to their radiolucency and favourable mechanical properties, carbon fibre nails may be a preferable alternative to titanium nails for oncology patients. We aim to compare the surgical characteristics and short-term results of patients who underwent intramedullary fixation with either a titanium or carbon fibre nail for pathological long-bone fracture. Methods. This single tertiary-institutional, retrospectively matched case-control study included 72 patients who underwent prophylactic or therapeutic fixation for pathological fracture of the humerus, femur, or tibia with either a titanium (control group, n = 36) or carbon fibre (case group, n = 36) intramedullary nail between 2016 to 2020. Patients were excluded if intramedullary fixation was combined with any other surgical procedure/fixation method. Outcomes included operating time, blood loss, fluoroscopic time, and complications. Fisher’s exact test and Mann-Whitney U test were used for categorical and continuous outcomes, respectively. Results. Patients receiving carbon nails as compared to those receiving titanium nails had higher blood loss (median 150 ml (interquartile range (IQR) 100 to 250) vs 100 ml (IQR 50 to 150); p = 0.042) and longer fluoroscopic time (median 150 seconds (IQR 114 to 182) vs 94 seconds (IQR 58 to 124); p = 0.001). Implant complications occurred in seven patients (19%) in the titanium group versus one patient (3%) in the carbon fibre group (p = 0.055). There were no notable differences between groups with regard to operating time, surgical wound infection, or survival. Conclusion. This pilot study demonstrates a non-inferior surgical and short-term clinical profile supporting further consideration of carbon fibre nails for pathological fracture fixation in orthopaedic oncology patients. Given enhanced accommodation of imaging methods important for oncological surveillance and radiation therapy planning, as well as high tolerances to fatigue stress, carbon fibre implants possess important oncological advantages over titanium implants that merit further prospective investigation. Level of evidence: III, Retrospective study. Cite this article: Bone Jt Open 2022;3(8):648–655


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 472 - 478
1 Apr 2022
Maccario C Paoli T Romano F D’Ambrosi R Indino C Federico UG

Aims. This study reports updates the previously published two-year clinical, functional, and radiological results of a group of patients who underwent transfibular total ankle arthroplasty (TAA), with follow-up extended to a minimum of five years. Methods. We prospectively evaluated 89 patients who underwent transfibular TAA for end-stage osteoarthritis. Patients’ clinical and radiological examinations were collected pre- and postoperatively at six months and then annually for up to five years of follow-up. Three patients were lost at the final follow-up with a total of 86 patients at the final follow-up. Results. A total of 86 patients were evaluated at a mean follow-up of 65.4 months (60 to 90). At five-year follow-up, statistically significant improvements (p < 0.001) were found in the mean American Orthopaedic Foot & Ankle Society Ankle Hindfoot Score (from 33.8 (SD 14.3) to 86.1 (SD 8.8)), visual analogue scale for pain (from 8.5 (SD 1.7) to 1.5 (SD 1.2)), Short Form-12 Physical and Mental Component Scores (from 29.9 (SD 6.7) and 43.3 (SD 8.6) to 47.3 (SD 7.5) and 52.2 (SD 8.0), respectively), and mean ankle dorsiflexion and plantarflexion (from 6.2° (SD 5.5°) and 9.6° (SD 5.8°) to 23.9° (SD 7.7°) and 16.9° (SD 7.2°), respectively). Radiologically, the implants maintained neutral alignment without subsidence. Tibial or talar radiolucency was found in eight patients, but none of these patients was symptomatic. At five-year follow up, 97.7% of implants (95% confidence interval 91.2 to 99.4) were free from revision or removal with 84 implants at risk. We recorded two cases (2.3%) of failure for septic loosening. Conclusion. Transfibular TAA is safe and effective with a high survival rate at mid-term follow-up and satisfactory clinical and radiological results. Further studies are required to determine the long-term performance of these implants. Cite this article: Bone Joint J 2022;104-B(4):472–478


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 185 - 191
1 Feb 2015
Kendrick BJL Kaptein BL Valstar ER Gill HS Jackson WFM Dodd CAF Price AJ Murray DW

The most common reasons for revision of unicompartmental knee arthroplasty (UKA) are loosening and pain. Cementless components may reduce the revision rate. The aim of this study was to compare the fixation and clinical outcome of cementless and cemented Oxford UKAs. . A total of 43 patients were randomised to receive either a cemented or a cementless Oxford UKA and were followed for two years with radiostereometric analysis (RSA), radiographs aligned with the bone–implant interfaces and clinical scores. The femoral components migrated significantly during the first year (mean 0.2 mm) but not during the second. There was no significant difference in the extent of migration between cemented and cementless femoral components in either the first or the second year. In the first year the cementless tibial components subsided significantly more than the cemented components (mean 0.28 mm (. sd. 0.17) vs. 0.09 mm (. sd. 0.19 mm)). In the second year, although there was a small amount of subsidence (mean 0.05 mm) there was no significant difference (p = 0.92) between cemented and cementless tibial components. There were no femoral radiolucencies. Tibial radiolucencies were narrow (< 1 mm) and were significantly (p = 0.02) less common with cementless (6 of 21) than cemented (13 of 21) components at two years. There were no complete radiolucencies with cementless components, whereas five of 21 (24%) cemented components had complete radiolucencies. The clinical scores at two years were not significantly different (p = 0.20). . As second-year migration is predictive of subsequent loosening, and as radiolucency is suggestive of reduced implant–bone contact, these data suggest that fixation of the cementless components is at least as good as, if not better than, that of cemented devices. Cite this article: Bone Joint J 2015; 97-B:185–91


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 181 - 187
1 Feb 2013
Liddle AD Pandit H O’Brien S Doran E Penny ID Hooper GJ Burn PJ Dodd CAF Beverland DE Maxwell AR Murray DW

The Cementless Oxford Unicompartmental Knee Replacement (OUKR) was developed to address problems related to cementation, and has been demonstrated in a randomised study to have similar clinical outcomes with fewer radiolucencies than observed with the cemented device. However, before its widespread use it is necessary to clarify contraindications and assess the complications. This requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless OUKRs in 881 patients at a minimum follow-up of one year. All patients had radiological assessment aligned to the bone–implant interfaces and clinical scores. Analysis was performed at a mean of 38.2 months (19 to 88) following surgery. A total of 17 patients died (comprising 19 knees (1.9%)), none as a result of surgery; there were no tibial or femoral loosenings. A total of 19 knees (1.9%) had significant implant-related complications or required revision. Implant survival at six years was 97.2%, and there was a partial radiolucency at the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies. There was no significant increase in complication rate compared with cemented fixation (p = 0.87), and no specific contraindications to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients with end-stage anteromedial osteoarthritis of the knee, with radiological evidence of improved fixation compared with previous reports using cemented fixation. Cite this article: Bone Joint J 2013;95-B:181–7


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1158 - 1166
14 Sep 2020
Kaptein BL den Hollander P Thomassen B Fiocco M Nelissen RGHH

Aims. The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). Methods. A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively. Results. The overall migration (mean maximum total point motion (MPTM)) at two years was comparable: mean 1.13 mm (95% confidence interval (CI), 0.97 to 1.30) for the ATTUNE and 1.16 mm (95% CI, 0.99 to 1.35) for the PFC-sigma. At two years, the mean backward tilting was -0.43° (95% CI, -0.65 to -0.21) for the ATTUNE and 0.08° (95% CI -0.16 to 0.31), for the PFC-sigma. Overall migration between the first and second postoperative year was negligible for both components. The clinical outcomes and PROMs improved compared with preoperative scores and were not different between groups. Radiolucencies at the implant-cement interface were mainly seen below the medial baseplate: 17% in the ATTUNE and 3% in the PFC-sigma at two weeks, and at two years 42% and 9% respectively (p = 0.001). Conclusion. In the first two postoperative years the initial version of the ATTUNE tibial component was not inferior with respect to overall migration, although it showed relatively more backwards tilting and radiolucent lines at the implant-cement interface than the PFC-sigma. The version of the ATTUNE tibial component examined in this study has subsequently undergone modification by the manufacturer. Level of Evidence: 1 (randomized controlled clinical trial). Cite this article: Bone Joint J 2020;102-B(9):1158–1166


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 458 - 464
1 Apr 2017
Abrahams JM Kim YS Callary SA De Ieso C Costi K Howie DW Solomon LB

Aims. This study aimed to determine the diagnostic performance of radiographic criteria to detect aseptic acetabular loosening after revision total hip arthroplasty (THA). Secondary aims were to determine the predictive values of different thresholds of migration and to determine the predictive values of radiolucency criteria. Patients and Methods. Acetabular component migration to re-revision was measured retrospectively using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements (Sutherland method) in two groups: Group A, 52 components (48 patients) found not loose at re-revision and Group B, 42 components (36 patients) found loose at re-revision between 1980 and 2015. The presence and extent of radiolucent lines was also assessed. Results. Using EBRA, both proximal translation and sagittal rotation were excellent diagnostic tests for detecting aseptic loosening. The area under the receiver operating characteristic (ROC) curves was 0.94 and 0.93, respectively. The thresholds of 2.5 mm proximal translation or 2° sagittal rotation (EBRA) in combination with radiolucency criteria had a sensitivity of 93% and specificity of 88% to detect aseptic loosening. The sensitivity, specificity, positive predictive value and negative predictive value (NPV) of radiolucency criteria were 41%, 100%, 100% and 68% respectively. Manual measurements of both proximal translation and sagittal rotation were very good diagnostic tests. The area under the ROC curve was 0.86 and 0.92 respectively. However, manual measurements had a decreased specificity compared with EBRA. Radiolucency criteria had a poor sensitivity and NPV of 41% and 68% respectively. Conclusion. This study shows that EBRA and manual migration measurements can be used as accurate diagnostic tools to detect aseptic loosening of cementless acetabular components used at revision THA. Radiolucency criteria should not be used in isolation to exclude loosening of cementless acetabular components used at revision THA given their poor sensitivity and NPV. Cite this article: Bone Joint J 2017;99-B:458–64


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1240 - 1246
1 Sep 2011
Melis B DeFranco M Lädermann A Molé D Favard L Nérot C Maynou C Walch G

Radiological changes and differences between cemented and uncemented components of Grammont reverse shoulder arthroplasties (DePuy) were analysed at a mean follow-up of 9.6 years (8 to 12). Of 122 reverse shoulder arthroplasties implanted in five shoulder centres between 1993 and 2000, a total of 68 (65 patients) were available for study. The indications for reversed shoulder arthroplasty were cuff tear arthropathy in 48 shoulders, revision of shoulder prostheses of various types in 11 and massive cuff tear in nine. The development of scapular notching, bony scapular spur formation, heterotopic ossification, glenoid and humeral radiolucencies, stem subsidence, radiological signs of stress shielding and resorption of the tuberosities were assessed on standardised true anteroposterior and axillary radiographs. A scapular notch was observed in 60 shoulders (88%) and was associated with the superolateral approach (p = 0.009). Glenoid radiolucency was present in 11 (16%), bony scapular spur and/or ossifications in 51 (75%), and subsidence of the stem and humeral radiolucency in more than three zones were present in three (8.8%) and in four (11.8%) of 34 cemented components, respectively, and in one (2.9%) and two (5.9%) of 34 uncemented components, respectively. Radiological signs of stress shielding were significantly more frequent with uncemented components (p < 0.001), as was resorption of the greater (p < 0.001) and lesser tuberosities (p = 0.009)


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 793 - 798
1 Jun 2012
Bianchi A Martinelli N Sartorelli E Malerba F

The Bologna–Oxford (BOX) total ankle replacement (TAR) was developed with the aim of achieving satisfactory pain-free movement of the ankle. To date, only one single multicentre study has reported its clinical results. The aim of this study was to conduct an independent review of its mid-term results. We retrospectively reviewed a total of 60 prospectively followed patients in whom 62 BOX TARs had been implanted between 2004 and 2008. We used the American Orthopedic Foot and Ankle Society (AOFAS) score to assess the clinical results. Standardised radiographs taken at the time of final follow-up were analysed by two observers. The overall survival was 91.9% at a mean follow-up of 42.5 months (24 to 71). The mean AOFAS score had improved from 35.1 points (. sd. 16.6; 4 to 73) pre-operatively to 78.0 (. sd. 10.7; 57 to 100) at final follow-up (p < 0.01). Tibial radiolucencies < 2 mm in width were seen around 16 TARs. Talar radiolucencies < 2 mm were seen around four TARs. A total of 47 patients (78.3%) were very satisfied or satisfied with the outcome. Five patients required revision for functional limitation or continuing pain


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1592 - 1599
1 Dec 2018
Galea VP Connelly JW Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen MB Muratoglu O Bragdon C Malchau H

Aims. The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure. Patients and Methods. A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery. Results. We observed similar bedding in one year postoperatively and wear two years postoperatively between the two types of liner. However, there was significantly more penetration of the femoral head in the ModXLPE cohort compared with the VEPE cohort five years postoperatively (p < 0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (β = 0.22, p = 0.010) and metal (vs ceramic) femoral head (β = 0.21, p = 0.013). There was no association between increased wear and the development of radiolucency (p = 0.866) or PROMs. No patient had evidence of osteolysis. Conclusion. Five years postoperatively, patients with VEPE (vs ModXLPE) and ceramic (vs metal) femoral heads had decreased wear. The rates of wear for both liners were very low and have not led to any osteolysis or implant failure due to aseptic loosening


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1336 - 1344
1 Oct 2018
Powell AJ Crua E Chong BC Gordon R McAuslan A Pitto RP Clatworthy MG

Aims. This study compares the PFC total knee arthroplasty (TKA) system in a prospective randomized control trial (RCT) of the mobile-bearing rotating-platform (RP) TKA against the fixed-bearing (FB) TKA. This is the largest RCT with the longest follow-up where cruciate-retaining PFC total knee arthroplasties are compared in a non-bilateral TKA study. Patients and Methods. A total of 167 patients (190 knees with 23 bilateral cases), were recruited prospectively and randomly assigned, with 91 knees receiving the RP and 99 knees receiving FB. The mean age was 65.5 years (48 to 82), the mean body mass index (BMI) was 29.7 kg/m. 2. (20 to 52) and 73 patients were female. The Knee Society Score (KSS), Knee Society Functional Score (KSFS), Oxford Knee Score (OKS), Western Ontario and McMaster Universities Arthritis Index (WOMAC), and 12-Item Short-Form Health Survey Physical and Mental Component Scores (SF-12 PCS, SF-12 MCS) were gathered and recorded preoperatively, at five-years’ follow-up, and at ten years’ follow-up. Additionally, Knee Injury and Osteoarthritis Outcome Scores (KOOS) were collected at five- and ten-year follow-ups. The prevalence of radiolucent lines (RL) on radiographs and implant survival were recorded at five- and ten-year follow-ups. Results. At the ten-year follow-up, the RP group (n = 39) had a statistically significant superior score in the OKS (p = 0.001), WOMAC (p = 0.023), SF-12 PCS (p = 0.019), KOOS Activities of Daily Living (ADL) (p = 0.010), and KOOS Sport and Recreation (Sport/Rec) (p = 0.006) compared with the FB group (n = 46). The OKS, SF-12 PCS, and KOOS Sport/Rec at ten years had mean scores above the minimal clinically important difference (MCID) threshold. There was no significant difference in prevalence of radiolucency between groups at five-years’ follow-up (p = 0.449), nor at ten-years’ follow-up (p = 0.08). Implant survival rate at 14 years postoperative was 95.2 (95% CI 90.7 to 99.8) and 94.7 (95% CI 86.8 to 100.0) for the RP and FB TKAs, respectively. Conclusion. At ten-year follow-up, the mobile-bearing knee joint arthroplasty had statistically and clinically relevant superior OKS, SF-12 PCS, and KOOS (Sport/Rec) than the fixed-bearing platform. No difference was seen in prevalence of radiolucent lines. There was a greater than 94% implant survival rate for both cohorts at 14 years. Cite this article: Bone Joint J 2018;100-B:1336–44


Bone & Joint Research
Vol. 8, Issue 10 | Pages 443 - 450
1 Oct 2019
Treacy RBC Holland JP Daniel J Ziaee H McMinn DJW

Objectives. Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. Methods. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (. sd. 11; 21 to 82), and 73% of implantations were in female patients. All patients were individually counselled, and a detailed informed consent was obtained prior to operation. Primary resurfacing was carried out in 85 hips, and three cases involved revision of previous MoM HRA. Clinical, radiological, and Oxford Hip Score (OHS) assessments were studied, along with implant survival. Results. There was no loss to follow-up and no actual or impending revision or reoperation. Median OHS increased from 24 (interquartile range (IQR) 20 to 28) preoperatively to 48 (IQR 46 to 48) at the latest follow-up (48 being the best possible score). Radiographs showed one patient had a head-neck junction lucency. No other radiolucency, osteolysis, component migration, or femoral neck thinning was noted. Conclusion. The results in this small consecutive cohort suggest that metal-on-monobloc-XLPE HRA is successful in the short term and merits further investigation as a conservative alternative to the current accepted standard of stemmed THA. However, we would stress that survival data with longer-term follow-up are needed prior to widespread adoption. Cite this article: R. B. C. Treacy, J. P. Holland, J. Daniel, H. Ziaee, D. J. W. McMinn. Preliminary report of clinical experience with metal-on-highly-crosslinked-polyethylene hip resurfacing. Bone Joint Res 2019;8:443–450. DOI: 10.1302/2046-3758.810.BJR-2019-0060.R1


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 313 - 319
1 Mar 2016
Baker P Rankin K Naisby S Agni N Brewster N Holland J

Aims . The highly cross-linked polyethylene Exeter RimFit flanged cemented acetabular component was introduced in the United Kingdom in 2010. This study aimed to examine the rates of emergence of radiolucent lines observed when the Rimfit acetabular component was implanted at total hip arthroplasty (THA) using two different techniques: firstly, the ‘rimcutter’ technique in which the flange sits on a pre-prepared acetabular rim; and secondly, the ‘trimmed flange’ technique in which the flange is trimmed and the acetabular component is seated inside the rim of the acetabulum. . Patients and Methods . The radiographs of 150 THAs (75 ‘rimcutter’, 75 ‘trimmed flange’) involving this component were evaluated to assess for radiolucencies at the cement/bone interface by three observers. . Results . Rimfit acetabular components implanted using the rimcutter technique had significantly higher rates of radiolucency than those introduced using the ‘trimmed flange’ technique one year post-operatively (one zone: 63/75 (84%) vs 17/75 (23%); two zones 42/75 (56%) vs 0/75 (0%); all three zones 17/75 (23%) vs 0/75 (0%):(all p < 0.001). . Conclusion . On the basis of these findings, we have stopped using the ‘rimcutter’ technique when implanting the Rimfit acetabular component and have reverted to the ‘trimmed flange’ technique. Take home message: Surgeons should be vigilant of the performance of the Rimfit acetabular component when used alongside the rim cutter device due to an observed higher rate of progressive radiolucencies with this combination of component / technique. Cite this article: Bone Joint J 2016;98-B:313–19