Aims. We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre. Methods. Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year. Results. Eight patients (ten ankles) died during follow-up, but none required revision. Of the surviving 106 patients (108 ankles: rheumatoid arthritis (RA), n = 15; osteoarthritis (OA), n = 93), 38 were women and 68 were men, with a mean age of 68.2 years (48 to 86) at the time of surgery. Mean follow-up was 5.1 years (2.1 to 9.0). A total of ten implants failed (8.5%), thus requiring revision. The
Aims. We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants. Patients and Methods. Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year. Results. A total of 84 patients (87 ankles) were alive by the end of this study. Of the surviving 84 patients (87 ankles; rheumatoid arthritis (RA), n = 40; OA, n = 47), 45 were women and 39 were men, with a mean age of 54 years (18 to 72 years) at the time of surgery. A total of 32 implants failed (16%), requiring revision surgery. The mean time to revision was 80 months (2 to 257). The
The purpose of this study was to assess the success rate and functional outcomes of bone grafting for periprosthetic bone cysts following total ankle arthroplasty (TAA). Additionally, we evaluated the rate of graft incorporation and identified associated predisposing factors using CT scan. We reviewed a total of 37 ankles (34 patients) that had undergone bone grafting for periprosthetic bone cysts. A CT scan was performed one year after bone grafting to check the status of graft incorporation. For accurate analysis of cyst volumes and their postoperative changes, 3D-reconstructed CT scan processed with 3D software was used. For functional outcomes, variables such as the Ankle Osteoarthritis Scale score and the visual analogue scale for pain were measured.Aims
Methods
Despite the increasing numbers of ankle arthroplasties, there are limited studies on their survival and comparisons between different implants. The primary aim of this study was to determine the failure rates of primary ankle arthroplasties commonly used in the UK. A data linkage study combined National Joint Registry (NJR) data and NHS Digital data. The primary outcome of failure was defined as the removal or exchange of any components of the implanted device. Life tables and Kaplan-Meier survival charts were used to illustrate survivorship. Cox proportional hazards regression models were fitted to compare failure rates between 1 April 2010 and 31 December 2018.Aims
Methods
Implant failure has become more common as the number of primary total ankle arthroplasties (TAAs) performed has increased. Although revision arthroplasty has gained attention for functional preservation, the long-term results remain unclear. This study aimed to assess the long-term outcomes of revision TAA using a mobile-bearing prosthesis in a considerably large cohort; the risk factors for failure were also determined. This single-centre retrospective cohort study included 116 patients (117 ankles) who underwent revision TAA for failed primary TAA between July 2000 and March 2010. Survival analysis and risk factor assessment were performed, and clinical performance and patient satisfaction were evaluated preoperatively and at last follow-up.Aims
Methods
The aim of this study was to capture 12-month outcomes from a representative multicentre cohort of patients undergoing total ankle arthroplasty (TAA), describe the pattern of patient-reported outcome measures (PROMs) at 12 months, and identify predictors of these outcome measures. Patients listed for a primary TAA at 19 NHS hospitals between February 2016 and October 2017 were eligible. PROMs data were collected preoperatively and at six and 12 months including: Manchester-Oxford Foot and Ankle Questionnaire (MOXFQ (foot and ankle)) and the EuroQol five-dimension five-level questionnaire (EQ-5D-5L). Radiological pre- and postoperative data included Kellgren-Lawrence score and implant position measurement. This was supplemented by data from the National Joint Registry through record linkage to determine: American Society of Anesthesiologists (ASA) grade at index procedure; indication for surgery, index ankle previous fracture; tibial hind foot alignment; additional surgery at the time of TAA; and implant type. Multivariate regression models assessed outcomes, and the relationship between MOXFQ and EQ-5D-5L outcomes, with patient characteristics.Aims
Methods
This study reports updates the previously published two-year clinical, functional, and radiological results of a group of patients who underwent transfibular total ankle arthroplasty (TAA), with follow-up extended to a minimum of five years. We prospectively evaluated 89 patients who underwent transfibular TAA for end-stage osteoarthritis. Patients’ clinical and radiological examinations were collected pre- and postoperatively at six months and then annually for up to five years of follow-up. Three patients were lost at the final follow-up with a total of 86 patients at the final follow-up.Aims
Methods
Preoperative talar valgus deformity ≥ 15° is considered a contraindication for total ankle arthroplasty (TAA). We compared operative procedures and clinical outcomes of TAA in patients with talar valgus deformity ≥ 15° and < 15°. A matched cohort of patients similar for demographics and components used but differing in preoperative coronal-plane tibiotalar valgus deformity ≥ 15° (valgus, n = 50; 52% male, mean age 65.8 years (SD 10.3), mean body mass index (BMI) 29.4 (SD 5.2)) or < 15° (control, n = 50; 58% male, mean age 65.6 years (SD 9.8), mean BMI 28.7 (SD 4.2)), underwent TAA by one surgeon. Preoperative and postoperative radiographs, Ankle Osteoarthritis Scale (AOS) pain and disability and 36-item Short Form Health Survey (SF-36) version 2 scores were collected prospectively. Ancillary procedures, secondary procedures, and complications were recorded.Aims
Methods
To assess the effect of age on clinical outcome and revision rates in patients who underwent total ankle arthroplasty (TAA) for end-stage ankle osteoarthritis (OA). A consecutive series of 811 ankles (789 patients) that underwent TAA between May 2003 and December 2013 were enrolled. The influence of age on clinical outcome, including the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot score, and pain according to the visual analogue scale (VAS) was assessed. In addition, the risk for revision surgery that includes soft tissue procedures, periarticular arthrodeses/osteotomies, ankle joint debridement, and/or inlay exchange (defined as minor revision), as well as the risk for revision surgery necessitating the exchange of any of the metallic components or removal of implant followed by ankle/hindfoot fusion (defined as major revision) was calculated.Aims
Methods
Arthroplasty for end-stage hallux rigidus (HR) is controversial. Arthrodesis remains the gold standard for surgical treatment, although is not without its complications, with rates of up to 10% for nonunion, 14% for reoperation and 10% for metatarsalgia. The aim of this study was to analyze the outcome of a double-stemmed silastic implant (Wright-Medical, Memphis, Tennessee, USA) for patients with end-stage HR. We conducted a retrospective review of 108 consecutive implants in 76 patients, between January 2005 and December 2016, with a minimum follow-up of two years. The mean age of the patients at the time of surgery was 61.6 years (42 to 84). There were 104 females and four males. Clinical, radiological, patient reported outcome measures (PROMS) data, a visual analogue score (VAS) for pain, and satisfaction scores were collected.Aims
Methods
Total ankle arthroplasty (TAA) surgery is complex and attracts a wide variety of complications. The literature lacks consistency in reporting adverse events and complications. The aim of this article is to provide a comprehensive analysis of each of these complications from a literature review, and to compare them with rates from our Unit, to aid clinicians with the process of informed consent. A total of 278 consecutive total ankle arthroplasties (251 patients), performed by four surgeons over a six-year period in Wrightington Hospital (Wigan, United Kingdom) were prospectively reviewed. There were 143 men and 108 women with a mean age of 64 years (41 to 86). The data were recorded on each follow-up visit. Any complications either during initial hospital stay or subsequently reported on follow-ups were recorded, investigated, monitored, and treated as warranted. Literature search included the studies reporting the outcomes and complications of TAA implants.Aims
Patients and Methods
To examine the mid-term outcome and cost utility of the BioPro
metallic hemiarthroplasty for the treatment of hallux rigidius. We reviewed 97 consecutive BioPro metallic hemiarthroplasties
performed in 80 patients for end-stage hallux rigidus, with a minimum
follow-up of five years. There were 19 men and 61 women; their mean
age was 55 years (22 to 74). No patient was lost to follow-up.Aims
Patients and Methods
Few reports compare the contribution of the talonavicular articulation
to overall range of movement in the sagittal plane after total ankle
arthroplasty (TAA) and tibiotalar arthrodesis. The purpose of this
study was to assess changes in ROM and functional outcomes following
tibiotalar arthrodesis and TAA. Patients who underwent isolated tibiotalar arthrodesis or TAA
with greater than two-year follow-up were enrolled in the study.
Overall arc of movement and talonavicular movement in the sagittal
plane were assessed with weight-bearing lateral maximum dorsiflexion
and plantarflexion radiographs. All patients completed Short Form-12
version 2.0 questionnaires, visual analogue scale for pain (VAS)
scores, and the Foot and Ankle Ability Measure (FAAM).Aims
Patients and Methods
The purpose of this study was to evaluate the
change in sagittal tibiotalar alignment after total ankle arthroplasty (TAA)
for osteoarthritis and to investigate factors affecting the restoration
of alignment. This retrospective study included 119 patients (120 ankles) who
underwent three component TAA using the Hintegra prosthesis. A total
of 63 ankles had anterior displacement of the talus before surgery
(group A), 49 had alignment in the normal range (group B), and eight
had posterior displacement of the talus (group C). Ankles in group
A were further sub-divided into those in whom normal alignment was
restored following TAA (41 ankles) and those with persistent displacement
(22 ankles). Radiographic and clinical results were assessed. Pre-operatively, the alignment in group A was significantly more
varus than that in group B, and the posterior slope of the tibial
plafond was greater (p <
0.01 in both cases). The posterior slope
of the tibial component was strongly associated with restoration
of alignment: ankles in which the alignment was restored had significantly
less posterior slope (p <
0.001). An anteriorly translated talus was restored to a normal position
after TAA in most patients. We suggest that surgeons performing
TAA using the Hintegra prosthesis should aim to insert the tibial
component at close to 90° relative to the axis of the tibia, hence
reducing posterior soft-tissue tension and allowing restoration
of normal tibiotalar alignment following surgery. Cite this article:
We compared the clinical and radiographic results
of total ankle replacement (TAR) performed in non-diabetic and diabetic
patients. We identified 173 patients who underwent unilateral TAR
between 2004 and 2011 with a minimum of two years’ follow-up. There
were 88 male (50.9%) and 85 female (49.1%) patients with a mean
age of 66 years ( The mean AOS and AOFAS scores were significantly better in the
non-diabetic group (p = 0.018 and p = 0.038, respectively). In all,
nine TARs (21%) in the diabetic group had clinical failure at a
mean follow-up of five years (24 to 109), which was significantly
higher than the rate of failure of 15 (11.6%) in the non-diabetic
group (p = 0.004). The uncontrolled diabetic subgroup had a significantly
poorer outcome than the non-diabetic group (p = 0.02), and a higher
rate of delayed wound healing. The incidence of early-onset osteolysis was higher in the diabetic
group than in the non-diabetic group (p = 0.02). These results suggest
that diabetes mellitus, especially with poor glycaemic control,
negatively affects the short- to mid-term outcome after TAR. Cite this article:
Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities >
10 mm in diameter considered to be ‘marked’. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous coating. Care should be taken with ankle arthroplasty until more is known about the reasons for these severe osteolyses.