Advertisement for orthosearch.org.uk
Results 1 - 20 of 25
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 458 - 462
1 Apr 2020
Limberg AK Tibbo ME Pagnano MW Perry KI Hanssen AD Abdel MP

Aims. Varus-valgus constrained (VVC) implants are often used during revision total knee arthroplasty (TKA) to gain coronal plane stability. However, the increased mechanical torque applied to the bone-cement interface theoretically increases the risk of aseptic loosening. We assessed mid-term survivorship, complications, and clinical outcomes of a fixed-bearing VVC device in revision TKAs. Methods. A total of 416 consecutive revision TKAs (398 patients) were performed at our institution using a single fixed-bearing VVC TKA from 2007 to 2015. Mean age was 64 years (33 to 88) with 50% male (199). Index revision TKA diagnoses were: instability (n = 122, 29%), aseptic loosening (n = 105, 25%), and prosthetic joint infection (PJI) (n = 97, 23%). All devices were cemented on the epiphyseal surfaces. Femoral stems were used in 97% (n = 402) of cases, tibial stems in 95% (n = 394) of cases; all were cemented. In total, 93% (n = 389) of cases required a stemmed femoral and tibial component. Femoral cones were used in 29%, and tibial cones in 40%. Survivorship was assessed via competing risk analysis; clinical outcomes were determined using Knee Society Scores (KSSs) and range of movement (ROM). Mean follow-up was four years (2 to 10). Results. The five-year cumulative incidence of subsequent revision for aseptic loosening and instability were 2% (95% confidence interval (CI) 0.2 to 3, number at risk = 154) and 4% (95% CI 2 to 6, number at risk = 153), respectively. The five-year cumulative incidence of any subsequent revision was 14% (95% CI 10 to 18, number at risk = 150). Reasons for subsequent revision included PJI (n = 23, of whom 12 had previous PJI), instability (n = 13), and aseptic loosening (n = 11). The use of this implant without stems was found to be a significant risk factor for subsequent revision (hazard ratio (HR) 7.58 (95% CI 3.98 to 16.03); p = 0.007). KSS improved from 46 preoperatively to 81 at latest follow-up (p < 0.001). ROM improved from 96° prerevision to 108° at latest follow-up (p = 0.016). Conclusion. The cumulative incidence of subsequent revision for aseptic loosening and instability was very low at five years with this fixed-bearing VVC implant in revision TKAs. Routine use of cemented and stemmed components with targeted use of metaphyseal cones likely contributed to this low rate of aseptic loosening. Cite this article: Bone Joint J 2020;102-B(4):458–462


Bone & Joint Open
Vol. 4, Issue 9 | Pages 682 - 688
6 Sep 2023
Hampton M Balachandar V Charalambous CP Sutton PM

Aims

Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA.

Methods

A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1279 - 1285
1 Dec 2023
Baker JF Nadar AC Jouflas AC Smith LS Sachdeva S Yakkanti MR Malkani AL

Aims

The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

Methods

This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1286 - 1293
1 Dec 2023
Yang H Cheon J Jung D Seon J

Aims

Fungal periprosthetic joint infections (PJIs) are rare, but their diagnosis and treatment are highly challenging. The purpose of this study was to investigate the clinical outcomes of patients with fungal PJIs treated with two-stage exchange knee arthroplasty combined with prolonged antifungal therapy.

Methods

We reviewed our institutional joint arthroplasty database and identified 41 patients diagnosed with fungal PJIs and treated with two-stage exchange arthroplasty after primary total knee arthroplasty (TKA) between January 2001 and December 2020, and compared them with those who had non-fungal PJIs during the same period. After propensity score matching based on age, sex, BMI, American Society of Anesthesiologists grade, and Charlson Comorbidity Index, 40 patients in each group were successfully matched. The surgical and antimicrobial treatment, patient demographic and clinical characteristics, recurrent infections, survival rates, and relevant risk factors that affected joint survivorship were analyzed. We defined treatment success as a well-functioning arthroplasty without any signs of a PJI, and without antimicrobial suppression, at a minimum follow-up of two years from the time of reimplantation.


Bone & Joint Open
Vol. 5, Issue 5 | Pages 401 - 410
20 May 2024
Bayoumi T Burger JA van der List JP Sierevelt IN Spekenbrink-Spooren A Pearle AD Kerkhoffs GMMJ Zuiderbaan HA

Aims

The primary objective of this registry-based study was to compare patient-reported outcomes of cementless and cemented medial unicompartmental knee arthroplasty (UKA) during the first postoperative year. The secondary objective was to assess one- and three-year implant survival of both fixation techniques.

Methods

We analyzed 10,862 cementless and 7,917 cemented UKA cases enrolled in the Dutch Arthroplasty Registry, operated between 2017 and 2021. Pre- to postoperative change in outcomes at six and 12 months’ follow-up were compared using mixed model analyses. Kaplan-Meier and Cox regression models were applied to quantify differences in implant survival. Adjustments were made for patient-specific variables and annual hospital volume.


Bone & Joint Research
Vol. 11, Issue 2 | Pages 91 - 101
1 Feb 2022
Munford MJ Stoddart JC Liddle AD Cobb JP Jeffers JRT

Aims

Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia.

Methods

In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee.


Bone & Joint Research
Vol. 10, Issue 8 | Pages 467 - 473
2 Aug 2021
Rodríguez-Collell JR Mifsut D Ruiz-Sauri A Rodríguez-Pino L González-Soler EM Valverde-Navarro AA

Aims

The main objective of this study is to analyze the penetration of bone cement in four different full cementation techniques of the tibial tray.

Methods

In order to determine the best tibial tray cementation technique, we applied cement to 40 cryopreserved donor tibiae by four different techniques: 1) double-layer cementation of the tibial component and tibial bone with bone restrictor; 2) metallic cementation of the tibial component without bone restrictor; 3) bone cementation of the tibia with bone restrictor; and 4) superficial bone cementation of the tibia and metallic keel cementation of the tibial component without bone restrictor. We performed CT exams of all 40 subjects, and measured cement layer thickness at both levels of the resected surface of the epiphysis and the endomedular metaphyseal level.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 51 - 58
1 Jun 2021
Yang J Heckmann ND Nahhas CR Salzano MB Ruzich GP Jacobs JJ Paprosky WG Rosenberg AG Nam D

Aims

Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design.

Methods

All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical tibial baseplate and shortened keel. Patient demographic details, Knee Society Scores (KSS), component alignment, and the presence of radiolucent lines at final follow-up were recorded. Kaplan-Meier analyses were performed to estimate survivorship.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 116 - 122
1 Jun 2020
Bedard NA Cates RA Lewallen DG Sierra RJ Hanssen AD Berry DJ Abdel MP

Aims

Metaphyseal cones with cemented stems are frequently used in revision total knee arthroplasty (TKA). However, if the diaphysis has been previously violated, the resultant sclerotic canal can impair cemented stem fixation, which is vital for bone ingrowth into the cone, and long-term fixation. We report the outcomes of our solution to this problem, in which impaction grafting and a cemented stem in the diaphysis is combined with an uncemented metaphyseal cone, for revision TKA in patients with severely compromised bone.

Methods

A metaphyseal cone was combined with diaphyseal impaction grafting and cemented stems for 35 revision TKAs. There were two patients with follow-up of less than two years who were excluded, leaving 33 procedures in 32 patients in the study. The mean age of the patients at the time of revision TKA was 67 years (32 to 87); 20 (60%) were male. Patients had undergone a mean of four (1 to 13) previous knee arthroplasty procedures. The indications for revision were aseptic loosening (80%) and two-stage reimplantation for prosthetic joint infection (PJI; 20%). The mean follow-up was four years (2 to 11).


Bone & Joint Open
Vol. 2, Issue 1 | Pages 48 - 57
19 Jan 2021
Asokan A Plastow R Kayani B Radhakrishnan GT Magan AA Haddad FS

Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes.

Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty.

This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice.

Cite this article: Bone Jt Open 2021;2(1):48–57.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 107 - 115
1 Jun 2020
Tetreault MW Perry KI Pagnano MW Hanssen AD Abdel MP

Aims

Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system.

Methods

We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1087 - 1092
1 Sep 2019
Garceau S Warschawski Y Dahduli O Alshaygy I Wolfstadt J Backstein D

Aims

The aim of this study was to assess the effects of transferring patients to a specialized arthroplasty centre between the first and second stages (interstage) of prosthetic joint infection (PJI) of the knee.

Patients and Methods

A search of our institutional database was performed to identify patients having undergone two-stage revision total knee arthroplasty (TKA) for PJI. Two cohorts were created: continuous care (CC) and transferred care (TC). Baseline characteristics and outcomes were collected and compared between cohorts.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 108 - 114
1 Jul 2019
Ji G Xu R Niu Y Li N Ivashkiv L Bostrom MPG Greenblatt MB Yang X

Aims

It is increasingly appreciated that coordinated regulation of angiogenesis and osteogenesis is needed for bone formation. How this regulation is achieved during peri-implant bone healing, such as osseointegration, is largely unclear. This study examined the relationship between angiogenesis and osteogenesis in a unique model of osseointegration of a mouse tibial implant by pharmacologically blocking the vascular endothelial growth factor (VEGF) pathway.

Materials and Methods

An implant was inserted into the right tibia of 16-week-old female C57BL/6 mice (n = 38). Mice received anti-VEGF receptor-1 (VEGFR-1) antibody (25 mg/kg) and VEGF receptor-2 (VEGFR-2) antibody (25 mg/kg; n = 19) or an isotype control antibody (n = 19). Flow cytometric (n = 4/group) and immunofluorescent (n = 3/group) analyses were performed at two weeks post-implantation to detect the distribution and density of CD31hiEMCNhi endothelium. RNA sequencing analysis was performed using sorted CD31hiEMCNhi endothelial cells (n = 2/group). Osteoblast lineage cells expressing osterix (OSX) and osteopontin (OPN) were also detected with immunofluorescence. Mechanical pull-out testing (n = 12/group) was used at four weeks post-implantation to determine the strength of the bone-implant interface. After pull-out testing, the tissue attached to the implant surface was harvested. Whole mount immunofluorescent staining of OSX and OPN was performed to determine the amount of osteoblast lineage cells.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1331 - 1347
1 Nov 2019
Jameson SS Asaad A Diament M Kasim A Bigirumurame T Baker P Mason J Partington P Reed M

Aims

Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients.

Materials and Methods

Data from the National Joint Registry (NJR) of England and Wales were obtained for all primary cemented TKAs between March 2003 and July 2016. Patient, implant, and surgical variables were analyzed. Cox proportional hazards models were used to assess the influence of ALBC on risk of revision. Body mass index (BMI) data were available in a subset of patients.


Aims

The objective of this study was to compare early postoperative functional outcomes and time to hospital discharge between conventional jig-based total knee arthroplasty (TKA) and robotic-arm assisted TKA.

Patients and Methods

This prospective cohort study included 40 consecutive patients undergoing conventional jig-based TKA followed by 40 consecutive patients receiving robotic-arm assisted TKA. All surgical procedures were performed by a single surgeon using the medial parapatellar approach with identical implant designs and standardized postoperative inpatient rehabilitation. Inpatient functional outcomes and time to hospital discharge were collected in all study patients.


The Bone & Joint Journal
Vol. 98-B, Issue 10_Supple_B | Pages 41 - 47
1 Oct 2016
Lisowski LA Meijer LI Bekerom MPJVD Pilot P Lisowski AE

Aims

The interest in unicompartmental knee arthroplasty (UKA) for medial osteoarthritis has increased rapidly but the long-term follow-up of the Oxford UKAs has yet to be analysed in non-designer centres. We have examined our ten- to 15-year clinical and radiological follow-up data for the Oxford Phase III UKAs.

Patients and Methods

Between January 1999 and January 2005 a total of 138 consecutive Oxford Phase III arthroplasties were performed by a single surgeon in 129 patients for medial compartment osteoarthritis (71 right and 67 left knees, mean age 72.0 years (47 to 91), mean body mass index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were prospectively recorded and obtained at intervals. Of the 129 patients, 32 patients (32 knees) died, ten patients (12 knees) were not able to take part in the final clinical and radiological assessment due to physical and mental conditions, but via telephone interview it was confirmed that none of these ten patients (12 knees) had a revision of the knee arthroplasty. One patient (two knees) was lost to follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1311 - 1316
1 Oct 2008
Kim Y Kim J

The purpose of this study was to determine objectively the outcome of total knee replacement in patients with ankylosed knees.

There were 82 patients (99 knees) with ankylosed knees who underwent total knee replacement with a condylar constrained or a posterior stabilised prosthesis. Their mean age was 41.9 years (23 to 60) and the mean follow-up was for 8.9 years (6.6 to 14). Pre- and post-operative data included the Hospital for Special Surgery (HSS), the Knee Society (KS) and the Western Ontario and McMaster University Osteoarthritis index (WOMAC) scores.

The mean HSS, KS and WOMAC scores improved from 60, 53, and 79 pre-operatively to 81, 85, and 37 at follow-up. These improvements were statistically significant (p = 0.018, 0.001 and 0.014 respectively). The mean physical, social and emotional WOMAC scores also improved significantly (p = 0.032, p = 0.023 and p < 0.001 respectively). The mean satisfaction score was 8.5 (sd 1.5).

Total knee replacement gives good mid-term results in patients with ankylosed knees.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 794 - 798
1 Jun 2010
Minoda Y Ikebuchi M Kobayashi A Iwaki H Inori F Nakamura H

Bone mineral density (BMD) around the femoral component has been reported to decrease after total knee replacement (TKR) because of stress shielding. Our aim was to determine whether a cemented mobile-bearing component reduced the post-operative loss of BMD. In our study 28 knees receiving a cemented fixed-bearing TKR were matched with 28 receiving a cemented mobile-bearing TKR. They underwent dual-energy x-ray absorptiometry, pre-operatively and at three weeks and at three, six, 12, 18 and 24 months post-operatively. The patients were not taking medication to improve the BMD.

The pre-operative differences in the BMD of the femoral neck, wrist, lumbar spine and knee in the two groups were not significant. The BMD of the femur decreased postoperatively in the fixed-bearing group, but not the mobile-bearing group. The difference in the post-operative change in the BMD in the two groups was statistically significant (p < 0.05) at 18 and 24 months.

Our findings show that a cemented mobile-bearing TKR has a favourable effect on the BMD of the distal femur after TKR in the short term. Further study is required to determine the long-term effects.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 896 - 902
1 Jul 2009
Gulati A Chau R Pandit HG Gray H Price AJ Dodd CAF Murray DW

Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement tibial components have been referred to as physiological radiolucencies. These should be distinguished from pathological radiolucencies, which are poorly defined, wide and progressive, and associated with loosening and infection. We studied the incidence and clinical significance of tibial radiolucent lines in 161 Oxford unicondylar knee replacements five years after surgery. All the radiographs were aligned with fluoroscopic control to obtain views parallel to the tibial tray to reveal the tibial bone-implant interface.

We found that 49 knees (30%) had complete, 52 (32%) had partial and 60 (37%) had no radiolucent lines. There was no relationship between the incidence of radiolucent lines and patient factors such as gender, body mass index and activity, or operative factors including the status of the anterior cruciate ligament and residual varus deformity. Nor was any statistical relationship established between the presence of radiolucent lines and clinical outcome, particularly pain, assessed by the Oxford Knee score and the American Knee Society score.

We conclude that radiolucent lines are common after Oxford unicompartmental knee replacement but that their aetiology remains unclear. Radiolucent lines were not a source of adverse symptoms or pain. Therefore, when attempting to identify a source of postoperative pain after Oxford unicompartmental knee replacement the presence of a physiological radiolucency should be ignored.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 351 - 356
1 Mar 2009
Parratte S Argenson JA Pearce O Pauly V Auquier P Aubaniac J

We retrospectively reviewed 35 cemented unicompartmental knee replacements performed for medial unicompartmental osteoarthritis of the knee in 31 patients ≤50 years old (mean 46, 31 to 49). Patients were assessed clinically and radiologically using the Knee Society scores at a mean follow-up of 9.7 years (5 to 16) and survival at 12 years was calculated. The mean Knee Society Function Score improved from 54 points (25 to 64) pre-operatively to 89 (80 to 100) post-operatively (p < 0.0001). Six knees required revision, four for polyethylene wear treated with an isolated exchange of the tibial insert, one for aseptic loosening and one for progression of osteoarthritis.

The 12-year survival according to Kaplan-Meier was 80.6% with revision for any reason as the endpoint. Despite encouraging clinical results, polyethylene wear remains a major concern affecting the survival of unicompartmental knee replacement in patients younger than 50.