header advert
Results 1 - 5 of 5
Results per page:
The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 455 - 460
1 Apr 2018
Mumith A Thomas M Shah Z Coathup M Blunn G

Increasing innovation in rapid prototyping (RP) and additive manufacturing (AM), also known as 3D printing, is bringing about major changes in translational surgical research.

This review describes the current position in the use of additive manufacturing in orthopaedic surgery.

Cite this article: Bone Joint J 2018;100-B:455-60.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


Bone & Joint Research
Vol. 5, Issue 3 | Pages 73 - 79
1 Mar 2016
Anwander H Cron GO Rakhra K Beaule PE

Objectives

Hips with metal-on-metal total hip arthroplasty (MoM THA) have a high rate of adverse local tissue reactions (ALTR), often associated with hypersensitivity reactions. Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) measures tissue perfusion with the parameter Ktrans (volume transfer constant of contrast agent). Our purpose was 1) to evaluate the feasibility of DCE-MRI in patients with THA and 2) to compare DCE-MRI in patients with MoM bearings with metal-on-polyethylene (MoP) bearings, hypothesising that the perfusion index Ktrans in hips with MoM THA is higher than in hips with MoP THA.

Methods

In this pilot study, 16 patients with primary THA were recruited (eight MoM, eight MoP). DCE-MRI of the hip was performed at 1.5 Tesla (T). For each patient, Ktrans was computed voxel-by-voxel in all tissue lateral to the bladder. The mean Ktrans for all voxels was then calculated. These values were compared with respect to implant type and gender, and further correlated with clinical parameters.


Bone & Joint Research
Vol. 3, Issue 3 | Pages 60 - 68
1 Mar 2014
Langton DJ Sidaginamale RP Holland JP Deehan D Joyce TJ Nargol AVF Meek RD Lord JK

Objectives

Wear debris released from bearing surfaces has been shown to provoke negative immune responses in the recipient. Excessive wear has been linked to early failure of prostheses. Analysis using coordinate measuring machines (CMMs) can provide estimates of total volumetric material loss of explanted prostheses and can help to understand device failure. The accuracy of volumetric testing has been debated, with some investigators stating that only protocols involving hundreds of thousands of measurement points are sufficient. We looked to examine this assumption and to apply the findings to the clinical arena.

Methods

We examined the effects on the calculated material loss from a ceramic femoral head when different CMM scanning parameters were used. Calculated wear volumes were compared with gold standard gravimetric tests in a blinded study.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 799 - 807
1 Jun 2007
Warwick D Friedman RJ Agnelli G Gil-Garay E Johnson K FitzGerald G Turibio FM

Patients who have undergone total hip or knee replacement (THR and TKR, respectively) are at high risk of venous thromboembolism. We aimed to determine the time courses of both the incidence of venous thromboembolism and effective prophylaxis. Patients with elective primary THR and TKR were enrolled in the multi-national Global Orthopaedic Registry. Data on the incidence of venous thromboembolism and prophylaxis were collected from 6639 THR and 8326 TKR patients.

The cumulative incidence of venous thromboembolism within three months of surgery was 1.7% in the THR and 2.3% in the TKR patients. The mean times to venous thromboembolism were 21.5 days (sd 22.5) for THR, and 9.7 days (sd 14.1) for TKR. It occurred after the median time to discharge in 75% of the THR and 57% of the TKA patients who developed venous thromboembolism. Of those who received recommended forms of prophylaxis, approximately one-quarter (26% of THR and 27% of TKR patients) were not receiving it seven days after surgery, the minimum duration recommended at the time of the study.

The risk of venous thromboembolism extends beyond the usual period of hospitalisation, while the duration of prophylaxis is often shorter than this. Practices should be re-assessed to ensure that patients receive appropriate durations of prophylaxis.