Advertisement for orthosearch.org.uk
Results 1 - 17 of 17
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 484 - 490
1 Apr 2019
Nandra R Matharu G Stevenson J Parry M Grimer R Jeys L

Aims

The aim of this study was to investigate the local recurrence rate at an extended follow-up in patients following navigated resection of primary pelvic and sacral tumours.

Patients and Methods

This prospective cohort study comprised 23 consecutive patients (nine female, 14 male) who underwent resection of a primary pelvic or sacral tumour, using computer navigation, between 2010 and 2012. The mean age of the patients at the time of presentation was 51 years (10 to 77). The rates of local recurrence and mortality were calculated using the Kaplan–Meier method.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 194 - 199
1 Feb 2012
Hoffart H Langenstein E Vasak N

The aim of this prospective single-centre study was to assess the difference in clinical outcome between total knee replacement (TKR) using computerised navigation and that of conventional TKR. We hypothesised that navigation would give a better result at every stage within the first five years. A total of 195 patients (195 knees) with a mean age of 70.0 years (39 to 89) were allocated alternately into two treatment groups, which used either conventional instrumentation (group A, 97 knees) or a navigation system (group B, 98 knees). After five years, complete clinical scores were available for 121 patients (62%). A total of 18 patients were lost to follow-up. Compared with conventional surgery, navigated TKR resulted in a better mean Knee Society score (p = 0.008). The difference in mean Knee Society scores over time between the two groups was not constant (p = 0.006), which suggests that these groups differed in their response to surgery with time. No significant difference in the frequency of malalignment was seen between the two groups.

In summary, computerised navigation resulted in a better functional outcome at five years than conventional techniques. Given the similarity in mechanical alignment between the two groups, rotational alignment may prove to be a better method of identifying differences in clinical outcome after navigated surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 477 - 480
1 Apr 2007
Spencer JM Chauhan SK Sloan K Taylor A Beaver RJ

We previously compared the component alignment in total knee replacement using a computer-navigated technique with a conventional jig-based method. We randomly allocated 71 patients to undergo either computer-navigated or conventional replacement. An improved alignment was seen in the computer-navigated group. The patients were then followed up post-operatively for two years, using the Knee Society score, the Short Form-36 health survey, the Western Ontario and McMaster Universities osteoarthritis index, the Bartlett Patellar pain questionnaire and the Oxford knee score, to assess functional outcome. At two years post-operatively 60 patients were available for assessment, 30 in each group and 62 patients completed a postal survey. No patient in either group had undergone revision. All variables were analysed for differences between the groups either by Student’s t-test or the Mann-Whitney U test. Differences between the two groups did not reach significance for any of the outcome measures at any time point. At two years postoperatively, the frequency of mild to severe anterior pain was not significantly different (p = 0.818), varying between 44% (14) for the computer-navigated group, and 47% (14) for the conventionally-replaced group. The Bartlett Patellar score and the Oxford knee score were also not significantly different (t-test p = 0.161 and p = 0.607, respectively). The clinical outcome of the patients with a computer-navigated knee replacement appears to be no different to that of a more conventional jig-based technique at two years post-operatively, despite the better alignment achieved with computer-navigated surgery


Bone & Joint 360
Vol. 13, Issue 2 | Pages 20 - 23
1 Apr 2024

The April 2024 Knee Roundup. 360. looks at: Challenging the status quo: re-evaluating the impact of obesity on unicompartmental knee arthroplasty outcomes; Timing matters: the link between ACL reconstruction delays and cartilage damage; Custom fit or off the shelf: evaluating patient outcomes in tailored versus standard knee replacements; Revolutionizing knee replacement: a comparative study on robotic-assisted and computer-navigated techniques; Pre-existing knee osteoarthritis and severe joint depression are associated with the need for total knee arthroplasty after tibial plateau fracture in patients aged over 60 years; Modern digital therapies?; A matched study on fracture rates following knee replacement surgeries;


Bone & Joint 360
Vol. 13, Issue 4 | Pages 16 - 19
2 Aug 2024

The August 2024 Knee Roundup360 looks at: Calcification’s role in knee osteoarthritis: implications for surgical decision-making; Lower complication rates and shorter lengths of hospital stay with technology-assisted total knee arthroplasty; Revision surgery: the hidden burden on surgeons; Are preoperative weight loss interventions worthwhile?; Total knee arthroplasty with or without prior bariatric surgery: a systematic review and meta-analysis; Aspirin triumphs in knee arthroplasty: a decade of evidence; Efficacy of DAIR in unicompartmental knee arthroplasty: a glimpse from Oxford.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 115 - 120
1 Mar 2024
Ricotti RG Flevas DA Sokrab R Vigdorchik JM Mayman DJ Jerabek SA Sculco TP Sculco PK

Aims

Periprosthetic femoral fracture (PPF) is a major complication following total hip arthroplasty (THA). Uncemented femoral components are widely preferred in primary THA, but are associated with higher PPF risk than cemented components. Collared components have reduced PPF rates following uncemented primary THA compared to collarless components, while maintaining similar prosthetic designs. The purpose of this study was to analyze PPF rate between collarless and collared component designs in a consecutive cohort of posterior approach THAs performed by two high-volume surgeons.

Methods

This retrospective series included 1,888 uncemented primary THAs using the posterior approach performed by two surgeons (PKS, JMV) from January 2016 to December 2022. Both surgeons switched from collarless to collared components in mid-2020, which was the only change in surgical practice. Data related to component design, PPF rate, and requirement for revision surgery were collected. A total of 1,123 patients (59.5%) received a collarless femoral component and 765 (40.5%) received a collared component. PPFs were identified using medical records and radiological imaging. Fracture rates between collared and collarless components were analyzed. Power analysis confirmed 80% power of the sample to detect a significant difference in PPF rates, and a Fisher’s exact test was performed to determine an association between collared and collarless component use on PPF rates.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1642 - 1645
1 Nov 2021
Kayani B Giebaly D Haddad FS


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1561 - 1570
1 Oct 2021
Blyth MJG Banger MS Doonan J Jones BG MacLean AD Rowe PJ

Aims

The aim of this study was to compare the clinical outcomes of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) during the first six weeks and at one year postoperatively.

Methods

A per protocol analysis of 76 patients, 43 of whom underwent TKA and 34 of whom underwent bi-UKA, was performed from a prospective, single-centre, randomized controlled trial. Diaries kept by the patients recorded pain, function, and the use of analgesics daily throughout the first week and weekly between the second and sixth weeks. Patient-reported outcome measures (PROMs) were compared preoperatively, and at three months and one year postoperatively. Data were also compared longitudinally and a subgroup analysis was conducted, stratified by preoperative PROM status.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 397 - 404
1 Jun 2021
Begum FA Kayani B Magan AA Chang JS Haddad FS

Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship.


Bone & Joint Research
Vol. 8, Issue 10 | Pages 438 - 442
1 Oct 2019
Kayani B Haddad FS


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1217 - 1222
1 Sep 2011
Bonner TJ Eardley WGP Patterson P Gregg PJ

Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR.

A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by > 3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47).

We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1457 - 1461
1 Nov 2012
Krishnan SP Dawood A Richards R Henckel J Hart AJ

Improvements in the surgical technique of total knee replacement (TKR) are continually being sought. There has recently been interest in three-dimensional (3D) pre-operative planning using magnetic resonance imaging (MRI) and CT. The 3D images are increasingly used for the production of patient-specific models, surgical guides and custom-made implants for TKR.

The users of patient-specific instrumentation (PSI) claim that they allow the optimum balance of technology and conventional surgery by reducing the complexity of conventional alignment and sizing tools. In this way the advantages of accuracy and precision claimed by computer navigation techniques are achieved without the disadvantages of additional intra-operative inventory, new skills or surgical time.

This review describes the terminology used in this area and debates the advantages and disadvantages of PSI.


Bone & Joint 360
Vol. 6, Issue 5 | Pages 12 - 16
1 Oct 2017


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1417 - 1424
1 Oct 2013
Jeys L Matharu GS Nandra RS Grimer RJ

We hypothesised that the use of computer navigation-assisted surgery for pelvic and sacral tumours would reduce the risk of an intralesional margin. We reviewed 31 patients (18 men and 13 women) with a mean age of 52.9 years (13.5 to 77.2) in whom computer navigation-assisted surgery had been carried out for a bone tumour of the pelvis or sacrum. There were 23 primary malignant bone tumours, four metastatic tumours and four locally advanced primary tumours of the rectum. The registration error when using computer navigation was <  1 mm in each case. There were no complications related to the navigation, which allowed the preservation of sacral nerve roots (n = 13), resection of otherwise inoperable disease (n = 4) and the avoidance of hindquarter amputation (n = 3). The intralesional resection rate for primary tumours of the pelvis and sacrum was 8.7% (n = 2): clear bone resection margins were achieved in all cases. At a mean follow-up of 13.1 months (3 to 34) three patients (13%) had developed a local recurrence. The mean time alive from diagnosis was 16.8 months (4 to 48).

Computer navigation-assisted surgery is safe and has reduced our intralesional resection rate for primary tumours of the pelvis and sacrum. We recommend this technique as being worthy of further consideration for this group of patients.

Cite this article: Bone Joint J 2013;95-B:1417–24.


Bone & Joint Research
Vol. 4, Issue 1 | Pages 6 - 10
1 Jan 2015
Goudie ST Deakin AH Deep K

Objectives

Acetabular component orientation in total hip arthroplasty (THA) influences results. Intra-operatively, the natural arthritic acetabulum is often used as a reference to position the acetabular component. Detailed information regarding its orientation is therefore essential. The aim of this study was to identify the acetabular inclination and anteversion in arthritic hips.

Methods

Acetabular inclination and anteversion in 65 symptomatic arthritic hips requiring THA were measured using a computer navigation system. All patients were Caucasian with primary osteoarthritis (29 men, 36 women). The mean age was 68 years (SD 8). Mean inclination was 50.5° (SD 7.8) in men and 52.1° (SD 6.7) in women. Mean anteversion was 8.3° (SD 8.7) in men and 14.4° (SD 11.6) in women.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 609 - 618
1 May 2014
Gøthesen Ø Espehaug B Havelin LI Petursson G Hallan G Strøm E Dyrhovden G Furnes O

We performed a randomised controlled trial comparing computer-assisted surgery (CAS) with conventional surgery (CONV) in total knee replacement (TKR). Between 2009 and 2011 a total of 192 patients with a mean age of 68 years (55 to 85) with osteoarthritis or arthritic disease of the knee were recruited from four Norwegian hospitals. At three months follow-up, functional results were marginally better for the CAS group. Mean differences (MD) in favour of CAS were found for the Knee Society function score (MD: 5.9, 95% confidence interval (CI) 0.3 to 11.4, p = 0.039), the Knee Injury and Osteoarthritis Outcome Score (KOOS) subscales for ‘pain’ (MD: 7.7, 95% CI 1.7 to 13.6, p = 0.012), ‘sports’ (MD: 13.5, 95% CI 5.6 to 21.4, p = 0.001) and ‘quality of life’ (MD: 7.2, 95% CI 0.1 to 14.3, p = 0.046). At one-year follow-up, differences favouring CAS were found for KOOS ‘sports’ (MD: 11.0, 95% CI 3.0 to 19.0, p = 0.007) and KOOS ‘symptoms’ (MD: 6.7, 95% CI 0.5 to 13.0, p = 0.035). The use of CAS resulted in fewer outliers in frontal alignment (> 3° malalignment), both for the entire TKR (37.9% vs 17.9%, p = 0.042) and for the tibial component separately (28.4% vs 6.3%, p = 0.002). Tibial slope was better achieved with CAS (58.9% vs 26.3%, p < 0.001). Operation time was 20 minutes longer with CAS. In conclusion, functional results were, statistically, marginally in favour of CAS. Also, CAS was more predictable than CONV for mechanical alignment and positioning of the prosthesis. However, the long-term outcomes must be further investigated.

Cite this article: Bone Joint J 2014; 96-B:609–18.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1296 - 1299
1 Oct 2011
Lang JE Mannava S Floyd AJ Goddard MS Smith BP Mofidi A M. Seyler T Jinnah RH

Robots have been used in surgery since the late 1980s. Orthopaedic surgery began to incorporate robotic technology in 1992, with the introduction of ROBODOC, for the planning and performance of total hip replacement. The use of robotic systems has subsequently increased, with promising short-term radiological outcomes when compared with traditional orthopaedic procedures. Robotic systems can be classified into two categories: autonomous and haptic (or surgeon-guided). Passive surgery systems, which represent a third type of technology, have also been adopted recently by orthopaedic surgeons.

While autonomous systems have fallen out of favour, tactile systems with technological improvements have become widely used. Specifically, the use of tactile and passive robotic systems in unicompartmental knee replacement (UKR) has addressed some of the historical mechanisms of failure of non-robotic UKR. These systems assist with increasing the accuracy of the alignment of the components and produce more consistent ligament balance. Short-term improvements in clinical and radiological outcomes have increased the popularity of robot-assisted UKR.

Robot-assisted orthopaedic surgery has the potential for improving surgical outcomes. We discuss the different types of robotic systems available for use in orthopaedics and consider the indication, contraindications and limitations of these technologies.