The MAGnetic Expansion Control (MAGEC) system
is used increasingly in the management of early-onset scoliosis.
Good results have been published, but there have been recent reports
identifying implant failures that may be associated with significant
metallosis surrounding the implants. This article aims to present
the current knowledge regarding the performance of this implant,
and the potential implications and strategies that may be employed
to identify and limit any problems. We urge surgeons to apply caution to patient and construct selection;
engage in prospective patient registration using a spine registry;
ensure close clinical monitoring until growth has ceased; and send
all explanted MAGEC rods for independent analysis. The MAGEC system may be a good instrumentation system for the
treatment of early-onset scoliosis. However, it is innovative and
like all new technology, especially when deployed in a paediatric
population, robust systems to assess long-term outcome are required
to ensure that patient safety is maintained. Cite this article:
Failure of fixation is a common problem in the treatment of osteoporotic fractures around the hip. The reinforcement of bone stock or of fixation of the implant may be a solution. Our study assesses the existing evidence for the use of bone substitutes in the management of these fractures in osteoporotic patients. Relevant publications were retrieved through Medline research and further scrutinised. Of 411 studies identified, 22 met the inclusion criteria, comprising 12 experimental and ten clinical reports. The clinical studies were evaluated with regard to their level of evidence. Only four were prospective and randomised. Polymethylmethacrylate and calcium-phosphate cements increased the primary stability of the implant-bone construct in all experimental and clinical studies, although there was considerable variation in the design of the studies. In randomised, controlled studies, augmentation of intracapsular fractures of the neck of the femur with calcium-phosphate cement was associated with poor long-term results. There was a lack of data on the long-term outcome for trochanteric fractures. Because there were only a few, randomised, controlled studies, there is currently poor evidence for the use of bone cement in the treatment of fractures of the hip.