In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any
Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined. Results. There were 158,677 individuals (42.4%) with an implant containing CoCr. There were 47,963 deaths, 27,332 heart outcomes, 35,720 cancers, and 22,025 neurodegenerative disorders. There was no evidence of an association between patients with CoCr implants and higher rates of any of the outcomes. Conclusion. CoCr-containing THAs did not have an increased risk of all-cause mortality, or clinically meaningful heart outcomes, cancer, or neurodegenerative disorders into the second decade post-implantation. Our findings will help reassure clinicians and the increasing number of patients receiving primary THA worldwide that the use of CoCr-containing implants is not associated with significant
Infection of a total hip replacement (THR) requires component removal and thorough local debridement. Usually, long-term antibiotic treatment in conjunction with a two-stage revision is required. This may take several months. One-stage revision using antibiotic-loaded cement has not gained widespread use, although the clinical and economic advantages are obvious. Allograft bone may be impregnated with high levels of antibiotics, and in revision of infected THR, act as a carrier providing a sustained high local concentration. We performed 37 one-stage revision of infected THRs, without the use of cement. There were three hips which required further revision because of recurrent infection, the remaining 34 hips (92%) stayed free from infection and stable at a mean follow-up of 4.4 years (2 to 8). No
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface. Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer. The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no
We have investigated the effect of using tranexamic
acid (TXA) during peri-acetabular osteotomy (PAO) on peri-operative
blood loss and blood transfusion requirements. In addition we analysed
whether the use of TXA was associated with an increased risk of
venous thromboembolism (VTE) following this procedure. A consecutive series of 96 PAOs, performed by a single surgeon,
were reviewed. A total of 48 patients received TXA and 48 did not.
The TXA group received a continuous infusion of TXA at a rate of
10 mg/kg/h. The primary outcome measure was the requirement for
blood transfusion. Secondary outcomes included total blood loss,
the decrease in the level of haemoglobin in the blood, the length
of hospital stay, and the complications of this treatment. The mean rate of transfusion was significantly lower in the TXA
group (62.5% vs 12.5%, p <
0.001). The mean blood
loss was also significantly reduced in the TXA group (1.9 L (standard
deviation (. sd. ) 0.9) vs 1.5 L (. sd. 0.7),
p <
0.01). No post-operative episodes of VTE were identified
in either group. . The use of TXA reduced the blood loss and the rate of transfusion
after PAO significantly, without
Perthes’ disease (PD) is a childhood hip disorder that can affect the quality of life in adulthood due to femoral head deformity and osteoarthritis. There is very little data on how PD patients function as adults, especially from the patients’ perspective. The purpose of this study was to collect treatment history, demographic details, the University of California, Los Angeles activity score (UCLA), the 36-Item Short Form survey (SF-36) score, and the Hip disability and Osteoarthritis Outcome score (HOOS) of adults who had PD using a web-based survey method and to compare their outcomes to the outcomes from an age- and sex-matched normative population. The English REDCap-based survey was made available on a PD study group website. The survey included childhood and adult PD history, UCLA, SF-36, and HOOS. Of the 1,182 participants who completed the survey, the 921 participants who did not have a total hip arthroplasty are the focus of this study. The mean age at survey was 38 years (SD 12) and the mean duration from age at PD onset to survey participation was 30.8 years (SD 12.6).Aims
Methods
Refobacin Bone Cement R and Palacos Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.Aims
Methods
Metal-on-metal bearings are being increasingly used in young patients. The potential
The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes. We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).Aims
Methods
Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane. Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.Aims
Methods
To evaluate the effect of ultrasound-targeted simvastatin-loaded microbubble destruction (UTMD In vitro, OA chondrocytes were treated with ultrasound (US), US-targeted microbubble destruction (UTMD), simvastatin (SV), and UTMDAims
Methods
Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
Postoperative delirium (POD) and postoperative cognitive decline (POCD) are common surgical complications. In the UK, the Best Practice Tariff incentivizes the screening of delirium in patients with hip fracture. Further, a National Hip Fracture Database (NHFD) performance indicator is the reduction in the incidence of POD. To aid in its recognition, we sought to determine factors associated with POD and POCD in patients with hip fractures. We interrogated the NHFD data on patients presenting with hip fractures to our institution from 2016 to 2018. POD was determined using the 4AT score, as recommended by the NHFD and UK Department of Health. POCD was defined as a decline in Abbreviated Mental Test Score (AMTS) of two or greater. Using logistic regression, we adjusted for covariates to identify factors associated with POD and POCD.Aims
Methods
The practice of overlapping surgery has been increasing in the delivery of orthopaedic surgery, aiming to provide efficient, high-quality care. However, there are concerns about the safety of this practice. The purpose of this study was to examine the safety and efficacy of a model of partially overlapping surgery that we termed ‘swing room’ in the practice of primary total hip (THA) and knee arthroplasty (TKA). A retrospective review of prospectively collected data was carried out on patients who underwent primary THA and TKA between 2006 and 2017 in two academic centres. Cases were stratified as partially overlapping (swing room), in which the surgeon is in one operating room (OR) while the next patient is being prepared in another, or nonoverlapping surgery. The demographic details of the patients which were collected included operating time, length of stay (LOS), postoperative complications within six weeks of the procedure, unplanned hospital readmissions, and unplanned reoperations. Fisher's exact, Wilcoxon rank-sum tests, chi-squared tests, and logistic regression analysis were used for statistical analysis.Aims
Methods
In previous studies, we identified multiple factors influencing the survivorship of hip resurfacing arthroplasties (HRAs), such as initial anatomical conditions and surgical technique. In addition, the University of California, Los Angeles (UCLA) activity score presents a ceiling effect, so a better quantification of activity is important to determine which activities may be advisable or detrimental to the recovered patient. We aimed to determine the effect of specific groups of sporting activities on the survivorship free of aseptic failure of a large series of HRA. A total of 661 patients (806 hips) representing 77% of a consecutive series of patients treated with metal-on-metal hybrid HRA answered a survey to determine the types and amounts of sporting activities they regularly participated in. There were 462 male patients (70%) and 199 female patients (30%). Their mean age at the time of surgery was 51.9 years (14 to 78). Their mean body mass index (BMI) was 26.5 kg/m2 (16.7 to 46.5). Activities were regrouped into 17 categories based on general analogies between these activities. Scores for typical frequency and duration of the sessions were used to quantify the patients’ overall time spent engaging in sporting activities. Impact and cycle scores were computed. Multivariable models were used.Aims
Patients and Methods
Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results. We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (Aims
Patients and Methods
The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction. Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study.Aims
Patients and Methods