The aims of this study were to: 1) report on a cohort of skeletally mature patients with native hip and knee septic arthritis over a 14-year period; 2) to determine the rate of joint failure in patients who had experienced an episode of hip or knee septic arthritis; and 3) to assess the outcome following septic arthritis relative to the infecting organism, whether those patients infected by All microbiological samples from joint aspirations between March 2000 and December 2014 at our institution were reviewed in order to identify cases of culture-proven septic arthritis. Cases in children (aged < 16 years) and prosthetic joints were excluded. Data were abstracted on age at diagnosis, sex, joint affected (hip or knee), type of organisms isolated, cause of septic arthritis, comorbidities within the Charlson Comorbidity Index (CCI), details of treatment, and outcome.Aims
Methods
Breast cancer survivors have known risk factors that might influence the results of total hip arthroplasty (THA) or total knee arthroplasty (TKA). This study evaluated clinical outcomes of patients with breast cancer history after primary THA and TKA. Our total joint registry identified patients with breast cancer history undergoing primary THA (n = 423) and TKA (n = 540). Patients were matched 1:1 based upon age, sex, BMI, procedure (hip or knee), and surgical year to non-breast cancer controls. Mortality, implant survival, and complications were assessed via Kaplan-Meier methods. Clinical outcomes were evaluated via Harris Hip Scores (HHSs) or Knee Society Scores (KSSs). Mean follow-up was six years (2 to 15).Aims
Methods
Periprosthetic fracture is a significant complication of total
hip and knee arthroplasty. This study aimed to describe the survival
of patients sustaining periprosthetic femoral fractures and compare
this with that of the general population, as well as to identify
the factors that influence survival. A total of 151 patients (women: men 116:35, mean age 74.6 years,
standard deviation 11.5) that sustained a periprosthetic fracture
between January 2005 and October 2012 were retrospectively analysed.
Epidemiological data, comorbidities, type of surgical management,
type of implant, and mortality data were studied.Aims
Patients and Methods
We evaluated the duration of hospitalisation,
occurrence of infections, hip dislocations, revisions, and mortality following
primary hip and knee replacement in 857 patients with Parkinson’s
disease and compared them with 2571 matched control patients. The
data were collected from comprehensive nationwide Finnish health
registers. The mean follow-up was six years (1 to 13). The patients
with Parkinson’s disease had a longer mean length of stay (21 days
[1 to 365] Cite this article:
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article:
Peri-prosthetic fracture after joint replacement in the lower limb is associated with significant morbidity. The primary aim of this study was to investigate the incidence of peri-prosthetic fracture after total hip replacement (THR) and total knee replacement (TKR) over a ten-year period using a population-based linked dataset. Between 1 April 1997 and 31 March 2008, 52 136 primary THRs, 8726 revision THRs, 44 511 primary TKRs, and 3222 revision TKRs were performed. Five years post-operatively, the rate of fracture was 0.9% after primary THR, 4.2% after revision THR, 0.6% after primary TKR and 1.7% after revision TKR. Comparison of
As part of the government’s initiative to reduce waiting times for major joint surgery in Wales, the Cardiff and Vale NHS Trust sent 224 patients (258 knees) to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement. The Kinemax total knee replacement system was used in all cases. The cumulative survival rate at three years was 79.2% (95% confidence interval (CI) 69.2 to 86.8) using re-operation for any cause as an endpoint and 85.3% (95% CI 75.9 to 91.8) using aseptic revision as an endpoint. This is significantly worse than that recorded in the published literature. These poor results have resulted in a significant impact on our service.
We compared the radiological appearances and survival of four methods of fixation of a femoral stem in 538 hips after follow-up for five or ten years. The fixation groups were: 1) press-fit shot-blasted smooth Ti-Al-V stem; 2) press-fit shot-blasted proximally ridged stem; 3) proximal hydroxyapatite (HA) coating; and 4) cementing.
The Kinematic Stabilizer is a posterior-cruciate-substituting design of total knee replacement. We have reviewed 109 primary total knee replacements in 95 patients at a mean follow-up time of 12.7 years (10 to 14). We used