The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups. This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups.Aims
Methods
The mobile bearing Oxford unicompartmental knee arthroplasty (OUKA) is recommended to be performed with the leg in the hanging leg (HL) position, and the thigh placed in a stirrup. This comparative cadaveric study assesses implant positioning and intraoperative kinematics of OUKA implanted either in the HL position or in the supine leg (SL) position. A total of 16 fresh-frozen knees in eight human cadavers, without macroscopic anatomical defects, were selected. The knees from each cadaver were randomized to have the OUKA implanted in the HL or SL position.Aims
Methods
The primary aim of this study was to determine the surgical team’s
learning curve for introducing robotic-arm assisted unicompartmental
knee arthroplasty (UKA) into routine surgical practice. The secondary
objective was to compare accuracy of implant positioning in conventional
jig-based UKA versus robotic-arm assisted UKA. This prospective single-surgeon cohort study included 60 consecutive
conventional jig-based UKAs compared with 60 consecutive robotic-arm
assisted UKAs for medial compartment knee osteoarthritis. Patients
undergoing conventional UKA and robotic-arm assisted UKA were well-matched
for baseline characteristics including a mean age of 65.5 years
(Aims
Patients and Methods
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase implant stability. This review discusses the relevant literature regarding all aspects
of the cementing of the tibial component at primary TKR. Cite this article:
The purpose of this study was to measure the
radiological parameters of femoral component alignment of the Oxford
Phase 3 unicompartmental knee replacement (UKR), and evaluate their
effect on clinical outcome. Multiple regression analysis was used
to examine the relative contributions of the radiological assessment
of femoral component alignment in 189 consecutive UKRs performed
by a single surgeon. The American Knee Society scores were compared
between groups, defined as being within or outside recommended tolerances
of the position of the femoral component. For the flexion/extension
position 21 UKRs (11.1%) lay outside the recommended limits, and for
posterior overhang of the femoral component nine (4.8%) lay outside
the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest
canal distance from the distal femoral entry point of the alignment
jig and coronal entry-point position had significant effects on
the flexion/extension position. Pre-operative HKA angle had a significant
influence on posterior overhang of the femoral component. However,
there was no significant difference in American Knee Society scores
relative to the position of the femoral component.
Patient-reported outcome measures (PROMs) are
increasingly being used to assess functional outcome and patient satisfaction.
They provide a framework for comparisons between surgical units,
and individual surgeons for benchmarking and financial remuneration.
Better performance may bring the reward of more customers as patients and
commissioners seek out high performers for their elective procedures.
Using National Joint Registry (NJR) data linked to PROMs we identified
22 691 primary total knee replacements (TKRs) undertaken for osteoarthritis
in England and Wales between August 2008 and February 2011, and
identified the surgical factors that influenced the improvements
in the Oxford knee score (OKS) and EuroQol-5D (EQ-5D) assessment
using multiple regression analysis. After correction for patient
factors the only surgical factors that influenced PROMs were implant
brand and hospital type (both p <
0.001). However, the effects
of surgical factors upon the PROMs were modest compared with patient
factors. For both the OKS and the EQ-5D the most important factors
influencing the improvement in PROMs were the corresponding pre-operative
score and the patient’s general health status. Despite having only
a small effect on PROMs, this study has shown that both implant
brand and hospital type do influence reported subjective functional
scores following TKR. In the current climate of financial austerity,
proposed performance-based remuneration and wider patient choice,
it would seem unwise to ignore these effects and the influence of
a range of additional patient factors.
This prospective study describes the outcome of the first 1000 phase 3 Oxford medial unicompartmental knee replacements (UKRs) implanted using a minimally invasive surgical approach for the recommended indications by two surgeons and followed up independently. The mean follow-up was 5.6 years (1 to 11) with 547 knees having a minimum follow-up of five years. At five years their mean Oxford knee score was 41.3 ( The incidence of implant-related re-operations was 2.9%; of these 29 re-operations two were revisions requiring revision knee replacement components with stems and wedges, 17 were conversions to a primary total knee replacement, six were open reductions for dislocation of the bearing, three were secondary lateral UKRs and one was revision of a tibial component. The most common reason for further surgical intervention was progression of arthritis in the lateral compartment (0.9%), followed by dislocation of the bearing (0.6%) and revision for unexplained pain (0.6%). If all implant-related re-operations are considered failures, the ten-year survival rate was 96% (95% confidence interval, 92.5 to 99.5). If only revisions requiring revision components are considered failures the ten-year survival rate is 99.8% (confidence interval 99 to 100). This is the largest published series of UKRs implanted through a minimally invasive surgical approach and with ten-year survival data. The survival rates are similar to those obtained with a standard open approach whereas the function is better. This demonstrates the effectiveness and safety of a minimally invasive surgical approach for implanting the Oxford UKR.
We retrospectively reviewed 35 cemented unicompartmental knee replacements performed for medial unicompartmental osteoarthritis of the knee in 31 patients ≤50 years old (mean 46, 31 to 49). Patients were assessed clinically and radiologically using the Knee Society scores at a mean follow-up of 9.7 years (5 to 16) and survival at 12 years was calculated. The mean Knee Society Function Score improved from 54 points (25 to 64) pre-operatively to 89 (80 to 100) post-operatively (p <
0.0001). Six knees required revision, four for polyethylene wear treated with an isolated exchange of the tibial insert, one for aseptic loosening and one for progression of osteoarthritis. The 12-year survival according to Kaplan-Meier was 80.6% with revision for any reason as the endpoint. Despite encouraging clinical results, polyethylene wear remains a major concern affecting the survival of unicompartmental knee replacement in patients younger than 50.
We performed a prospective, randomised study to compare the results and rates of complications of primary total knee replacement performed using a quadriceps-sparing technique or a standard arthrotomy in 120 patients who had bilateral total knee replacements carried out under the same anaesthetic. The clinical results, pain scales, surgical and hospital data, post-operative complications and radiological results were compared. No significant differences were found between the two groups with respect to the blood loss, knee score, function score, pain scale, range of movement or radiological findings. In contrast, the operating time (p = 0.0001) and the tourniquet time (p <
0.0001) were significantly longer in the quadriceps-sparing group, as was the rate of complications (p = 0.0468). We therefore recommend the use of a standard arthrotomy with the shortest possible skin incision for total knee replacement.