The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
This study aimed to develop and validate a fully automated system that quantifies proximal femoral bone mineral density (BMD) from CT images. The study analyzed 978 pairs of hip CT and dual-energy X-ray absorptiometry (DXA) measurements of the proximal femur (DXA-BMD) collected from three institutions. From the CT images, the femur and a calibration phantom were automatically segmented using previously trained deep-learning models. The Hounsfield units of each voxel were converted into density (mg/cm3). Then, a deep-learning model trained by manual landmark selection of 315 cases was developed to select the landmarks at the proximal femur to rotate the CT volume to the neutral position. Finally, the CT volume of the femur was projected onto the coronal plane, and the areal BMD of the proximal femur (CT-aBMD) was quantified. CT-aBMD correlated to DXA-BMD, and a receiver operating characteristic (ROC) analysis quantified the accuracy in diagnosing osteoporosis.Aims
Methods
Our objective was describing an algorithm to identify and prevent vascular injury in patients with intrapelvic components. Patients were defined as at risk to vascular injuries when components or cement migrated 5 mm or more beyond the ilioischial line in any of the pelvic incidences (anteroposterior and Judet view). In those patients, a serial investigation was initiated by a CT angiography, followed by a vascular surgeon evaluation. The investigation proceeded if necessary. The main goal was to assure a safe tissue plane between the hardware and the vessels.Aims
Methods
Acetabular edge-loading was a cause of increased wear rates in metal-on-metal hip arthroplasties, ultimately contributing to their failure. Although such wear patterns have been regularly reported in retrieval analyses, this study aimed to determine their in vivo location and investigate their relationship with acetabular component positioning. 3D CT imaging was combined with a recently validated method of mapping bearing surface wear in retrieved hip implants. The asymmetrical stabilizing fins of Birmingham hip replacements (BHRs) allowed the co-registration of their acetabular wear maps and their computational models, segmented from CT scans. The in vivo location of edge-wear was measured within a standardized coordinate system, defined using the anterior pelvic plane.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
The aims of this study were to develop an in vivo model of periprosthetic joint infection (PJI) in cemented hip hemiarthroplasty, and to monitor infection and biofilm formation in real-time. Sprague-Dawley rats underwent cemented hip hemiarthroplasty via the posterior approach with pre- and postoperative gait assessments. Infection with Aims
Methods
Uncemented metal acetabular components show good osseointegration, but material stiffness causes stress shielding and retroacetabular bone loss. Cemented monoblock polyethylene components load more physiologically; however, the cement bone interface can suffer fibrous encapsulation and loosening. It was hypothesized that an uncemented titanium-sintered monoblock polyethylene component may offer the optimum combination of osseointegration and anatomical loading. A total of 38 patients were prospectively enrolled and received an uncemented monoblock polyethylene acetabular (pressfit) component. This single cohort was then retrospectively compared with previously reported randomized cohorts of cemented monoblock (cemented) and trabecular metal (trabecular) acetabular implants. The primary outcome measure was periprosthetic bone density using dual-energy x-ray absorptiometry over two years. Secondary outcomes included radiological and clinical analysis.Aims
Methods
We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.Aims
Methods
This study aimed to determine if macrophages can attach and directly affect the oxide layers of 316L stainless steel, titanium alloy (Ti6Al4V), and cobalt-chromium-molybdenum alloy (CoCrMo) by releasing components of these alloys. Murine peritoneal macrophages were cultured and placed on stainless steel, CoCrMo, and Ti6Al4V discs into a 96-well plate. Cells were activated with interferon gamma and lipopolysaccharide. Macrophages on stainless steel discs produced significantly more nitric oxide (NO) compared to their control counterparts after eight to ten days and remained elevated for the duration of the experiment.Aims
Methods
Using a simple classification method, we aimed to estimate the collapse rate due to osteonecrosis of the femoral head (ONFH) in order to develop treatment guidelines for joint-preserving surgeries. We retrospectively analyzed 505 hips from 310 patients (141 men, 169 women; mean age 45.5 years Objectives
Methods
The purpose of this study was to evaluate the biological fixation of a 3D printed porous implant, with and without different hydroxyapatite (HA) coatings, in a canine model. A canine transcortical model was used to evaluate the characteristics of bone ingrowth of Ti6Al4V cylindrical implants fabricated using laser rapid manufacturing (LRM). At four and 12 weeks post-implantation, we performed histological analysis and mechanical push-out testing on three groups of implants: a HA-free control (LRM), LRM with precipitated HA (LRM-PA), and LRM with plasma-sprayed HA (LRM-PSHA).Aims
Materials and Methods
Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell ( This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m2 and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (Aims
Patients and Methods
Few reconstructive techniques are available for patients requiring
complex acetabular revisions such as those involving Paprosky type
2C, 3A and 3B deficiencies and pelvic discontinuity. Our aim was
to describe the development of the patient specific Triflange acetabular
component for use in these patients, the surgical technique and
mid-term results. We include a description of the pre-operative
CT scanning, the construction of a model, operative planning, and
surgical technique. All implants were coated with porous plasma
spray and hydroxyapatite if desired. A multicentre, retrospective review of 95 complex acetabular
reconstructions in 94 patients was performed. A total of 61 (64.2%)
were female. The mean age of the patients was 66 (38 to 85). The
mean body mass index was 29 kg/m2 (18 to 51). Outcome
was reported using the Harris Hip Score (HHS), complications, failures
and survival.Aims
Patients and Methods
Until now, there has been no consensus as to whether stemmed
acetabular components are appropriate for use in patients undergoing
revision total hip arthroplasty (THA) who have major acetabular
defects or pelvic discontinuity. We wished to address this deficiency
in the literature. We carried out a retrospective study of 35 patients (six men
and 29 women) with a mean age of 68 years (37 to 87), with major
acetabular defects who underwent revision THA between 2000 and 2012.Aim
Patients and Methods
This study reports the clinical outcome of reconstruction
of deficient abductor muscles following revision total hip arthroplasty
(THA), using a fresh–frozen allograft of the extensor mechanism
of the knee. A retrospective analysis was conducted of 11 consecutive
patients with a severe limp because of abductor deficiency which
was confirmed on MRI scans. The mean age of the patients (three
men and eight women) was 66.7 years (52 to 84), with a mean follow-up
of 33 months (24 to 41). Following surgery, two patients had no limp, seven had a mild
limp, and two had a persistent severe limp (p = 0.004). The mean
power of the abductors improved on the Medical Research Council
scale from 2.15 to 3.8 (p <
0.001). Pre-operatively, all patients
required a stick or walking frame; post-operatively, four patients
were able to walk without an aid. Overall, nine patients had severe
or moderate pain pre-operatively; ten patients had no or mild pain
post-operatively. At final review, the Harris hip score was good in five patients,
fair in two and poor in four. We conclude that using an extensor mechanism allograft is relatively
effective in the treatment of chronic abductor deficiency of the
hip after THA when techniques such as local tissue transfer are
not possible. Longer-term follow-up is necessary before the technique can be
broadly applied. Cite this article:
Conventional cemented acetabular components are
reported to have a high rate of failure when implanted into previously
irradiated bone. We recommend the use of a cemented reconstruction
with the addition of an acetabular reinforcement cross to improve
fixation. We reviewed a cohort of 45 patients (49 hips) who had undergone
irradiation of the pelvis and a cemented total hip arthroplasty
(THA) with an acetabular reinforcement cross. All hips had received
a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour
or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3:
45 to 60; mean: 49.57, 32 to 72). The mean follow-up after THA was 51 months (17 to 137). The cumulative
probability of revision of the acetabular component for a mechanical
reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60
months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One
hip was revised for mechanical failure and three for infection. Cemented acetabular components with a reinforcement cross provide
good medium-term fixation after pelvic irradiation. These patients
are at a higher risk of developing infection of their THA. Cite this article:
A pelvic discontinuity occurs when the superior
and inferior parts of the hemi-pelvis are no longer connected, which
is difficult to manage when associated with a failed total hip replacement.
Chronic pelvic discontinuity is found in 0.9% to 2.1% of hip revision
cases with risk factors including severe pelvic bone loss, female
gender, prior pelvic radiation and rheumatoid arthritis. Common
treatment options include: pelvic plating with allograft, cage reconstruction,
custom triflange implants, and porous tantalum implants with modular augments.
The optimal technique is dependent upon the degree of the discontinuity,
the amount of available bone stock and the likelihood of achieving
stable healing between the two segments. A method of treating pelvic
discontinuity using porous tantalum components with a distraction
technique that achieves both initial stability and subsequent long-term
biological fixation is described. Cite this article:
An atypical femoral fracture (AFF), with a transverse
fracture radiologically through the lateral cortex is a rare but serious
condition. In order to improve our ability to identify patients
with this condition, we retrospectively surveyed all subtrochanteric,
peri-implant and diaphyseal femoral fractures in patients aged ≥ 65
years who underwent surgical treatment at our hospital between 2004
and 2011. We describe the incidence of atypical fractures and their characteristics,
with observational data including a review of the hospital and general
practitioner records. Clinical outcomes were evaluated using the
Harris hip score (HHS) and the timed up-and-go (TUG) test. Atypical fractures only occurred in women with an incidence of
9.8 per 100 000 person-years. The incidence in those who were treated
with bisphosphonates was 79.0 per 100 000 person-years; eight of
17 fractures occurred around
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
Total hip replacement (THR) is a very common
procedure undertaken in up to 285 000 Americans each year. Patient
satisfaction with THR is very high, with improvements in general
health, quality of life, and function while at the same time very
cost effective. Although the majority of patients have a high degree
of satisfaction with their THR, 27% experience some discomfort,
and up to 6% experience severe chronic pain. Although it can be
difficult to diagnose the cause of the pain in these patients, this
clinical issue should be approached systematically and thoroughly.
A detailed history and clinical examination can often provide the
correct diagnosis and guide the appropriate selection of investigations, which
will then serve to confirm the clinical diagnosis made. Cite this article: