The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection. From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.Aims
Methods
The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases. This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications.Aims
Methods
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
Limb salvage in bone tumour patients replaces the bone with massive segmental prostheses where achieving bone integration at the shoulder of the implant through extracortical bone growth has been shown to prevent loosening. This study investigates the effect of multidrug chemotherapy on extracortical bone growth and early radiological signs of aseptic loosening in patients with massive distal femoral prostheses. A retrospective radiological analysis was performed on adult patients with distal femoral arthroplasties. In all, 16 patients were included in the chemotherapy group with 18 patients in the non-chemotherapy control group. Annual radiographs were analyzed for three years postoperatively. Dimensions of the bony pedicle, osseointegration of the hydroxyapatite (HA) collar surface, bone resorption at the implant shoulder, and radiolucent line (RLL) formation around the cemented component were analyzed.Aims
Methods
The consensus is that bipolar hemiarthroplasty (BHA) in allograft-prosthesis composite (APC) reconstruction of the proximal femur following primary tumour resection provides more stability than total hip arthroplasty (THA). However, no comparative study has been performed. In this study, we have compared the outcome and complication rates of these two methods. In a retrospective study, 57 patients who underwent APC reconstruction of proximal femur following the primary tumour resection, either using BHA (29) or THA (28), were included. Functional outcome was assessed using the Musculoskeletal Tumour Society (MSTS) scoring system and Harris Hip Score (HHS). Postoperative complications of the two techniques were also compared.Aims
Methods
We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.Aims
Methods
The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children. Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one.Aims
Patients and Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
Aims. Following the resection of an extensive amount of bone in the
treatment of a tumour, the residual segment may be insufficient
to accept a standard length intramedullary
We determined the efficacy of a devitalised autograft
(n = 13) and allograft (n = 16) cortical strut bone graft combined
with long-stem endoprosthetic reconstruction in the treatment of
massive tumours of the lower limb. A total of 29 patients (18 men:11
women, mean age 20.1 years (12 to 45) with a ratio of length of
resection to that of the whole prosthesis of >
50% were treated
between May 2003 and May 2012. The mean follow-up was 47 months
(15 to 132). The stem of the prosthesis was introduced through bone
graft struts filled with cement, then cemented into the residual
bone. Bone healing was achieved in 23 patients (86%). The mean Musculoskeletal
Tumour Society functional score was 85% (57 to 97). The five-year
survival rate of the endoprostheses was 81% (95% confidence intervals
67.3 to 92.3). The mean length of devitalised autografts and allografts
was 8.6 cm (5 to 15), which increased the ratio of the the length
of the stem of the prosthesis to that of the whole length of the
prosthesis from a theoretical 35% to an actual 55%. Cortical strut bone grafting and long-stem endoprosthetic reconstruction
is an option for treating massive segmental defects following resection
of a tumour in the lower limb. Patients can regain good function
with a low incidence of aseptic loosening. The strut graft and the
residual bone together serve as a satisfactory bony environment
for a revision prosthesis, if required, once union is achieved. Cite this article:
Giant cell tumour is the most common aggressive
benign tumour of the musculoskeletal system and has a high rate of
local recurrence. When it occurs in proximity to the hip, reconstruction
of the joint is a challenge. Options for reconstruction after wide
resection include the use of a megaprosthesis or an allograft-prosthesis
composite. We performed a clinical and radiological study to evaluate
the functional results of a proximal femoral allograft-prosthesis
composite in the treatment of proximal femoral giant cell tumour
after wide resection. This was an observational study, between 2006
and 2012, of 18 patients with a mean age of 32 years (28 to 42)
and a mean follow-up of 54 months (18 to 79). We achieved excellent
outcomes using Harris Hip Score in 13 patients and a good outcome
in five. All allografts united. There were no complications such
as infection, failure, fracture or resorption of the graft, or recurrent
tumour. Resection and reconstruction of giant cell tumours with
proximal femoral allograft–prosthesis composite is a better option
than using a prosthesis considering preservation of bone stock and excellent
restoration of function. A good result requires demanding bone banking techniques, effective
measures to prevent infection and stability at the allograft-host
junction. Cite this article:
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
In skeletally immature patients, resection of
bone tumours and reconstruction of the lower limb often results
in leg-length discrepancy. The Stanmore non-invasive extendible
endoprosthesis, which uses electromagnetic induction, allows post-operative
lengthening without anaesthesia. Between 2002 and 2009, 55 children
with a mean age of 11.4 years (5 to 16) underwent reconstruction
with this prosthesis; ten patients (18.2%) died of disseminated
disease and one child underwent amputation due to infection. We
reviewed 44 patients after a mean follow-up of 41.2 months (22 to
104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to
30) and the Toronto Extremity Salvage score was 92.3% (55.2% to
99.0%). There was no local recurrence of tumour. Complications developed
in 16 patients (29.1%) and ten (18.2%) underwent revision. The mean length gained per patient was 38.6 mm (3.5 to 161.5),
requiring a mean of 11.3 extensions (1 to 40), and ten component
exchanges were performed in nine patients (16.4%) after attaining
the maximum lengthening capacity of the implant. There were 11 patients
(20%) who were skeletally mature at follow-up, ten of whom had equal
leg lengths and nine had a full range of movement of the hip and
knee. This is the largest reported series using non-invasive extendible
endoprostheses after excision of primary bone tumours in skeletally
immature patients. The technique produces a good functional outcome,
with prevention of limb-length discrepancy at skeletal maturity.
The best method of reconstruction after resection of malignant tumours of the tibial diaphysis is unknown. In the absence of any long-term studies analysing the results of intercalary endoprosthetic replacement, we present a retrospective review of 18 patients who underwent limb salvage using a tibial diaphyseal endoprosthetic replacement following excision of a malignant bone tumour. There were ten men and eight women with a mean age of 42.5 years (16 to 76). Mean follow-up was 58.5 months (20 to 141) for all patients and 69.3 months (20 to 141) for the 12 patients still alive. Cumulative patient survival was 59% (95% confidence interval (CI) 32 to 84) at five years. Implant survival was 63% (95% CI 35 to 90) at ten years. Four patients required revision to a proximal tibial replacement at a mean follow-up of 29 months (10 to 54). Complications included metastases in five patients, aseptic loosening in four, peri-prosthetic fracture in two, infection in one and local recurrence in one. The mean Musculoskeletal Tumor Society score and the mean Toronto Extremity Salvage Score were 23 (17 to 28) and 74% (53 to 91), respectively. Although rates of complication and revision were high, custom-made tibial diaphyseal replacement following resection of malignant bone tumours enables early return to function and provides an attractive alternative to other surgical options, without apparent compromise of patient survival.
This paper describes the preliminary results of a proximal tibial endoprosthesis which spares the knee joint and enables retention of the natural articulation by replacing part of the tibial metaphysis and diaphysis. In eight patients who had a primary malignant bone tumour of the proximal tibia, the distal stem, which had a hydroxyapatite-coated collar to improve fixation, was cemented into the medullary canal. The proximal end had hydroxyapatite-coated extracortical plates which were secured to the remaining proximal tibial metaphysis using cortical screws. The mean age of the patients at operation was 28.9 years (8 to 43) and the mean follow-up was for 35 months (4 to 48). The mean Musculoskeletal Tumour Society score was 79% (57% to 90%), the mean Oxford Knee score was 40 points of 48 (36 to 46) and the mean knee flexion was 112° (100° to 120°). In one patient, revision to a below-knee amputation through the prosthesis was required because of recurrence of the tumour. Another patient sustained a periprosthetic fracture which healed with a painful malunion. This was revised to a further endoprosthesis which replaced the knee. In the remaining six patients the prosthesis allowed preservation of the knee joint with good function and no early evidence of loosening. Further follow-up is required to assess the longevity of these prostheses.
Massive endoprostheses using a
We have made a retrospective study of 1001 custom-made prostheses used as replacements after surgery for bone tumours. There were 493 distal femoral, 263 proximal femoral and 245 proximal tibial prostheses. Aseptic loosening was shown to be the principal mode of failure of the implants, and 71 patients had revision for aseptic loosening of a