Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims

The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer.

Materials and Methods

Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 455 - 460
1 Apr 2018
Mumith A Thomas M Shah Z Coathup M Blunn G

Increasing innovation in rapid prototyping (RP) and additive manufacturing (AM), also known as 3D printing, is bringing about major changes in translational surgical research.

This review describes the current position in the use of additive manufacturing in orthopaedic surgery.

Cite this article: Bone Joint J 2018;100-B:455-60.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 900 - 907
1 Nov 1997
McLaughlin JR Lee KR

We followed 138 patients (145 hips) who had had uncemented total hip arthroplasty using the Taperloc femoral component for a mean of ten years (8 to 12.5). No patient was lost to follow-up; 31 (31 hips) died before the minimum time of eight years for inclusion in the study, and 30 of these still had their femoral component in place. One well-fixed prosthesis had been exchanged at the time of acetabular revision. Of the remaining 114 hips, one femoral component required revision for aseptic loosening and one for sepsis. Three other well-fixed femoral components were removed during acetabular revision. Complete clinical and radiological follow-up was obtained in the 109 hips which had not had revision. Clinically, 94 (87%) were rated good or excellent, eight (7%) fair and seven (6%) poor. The average Harris hip score increased from 48 before operation to 88 at the time of the last follow-up. Radiologically, 103 hips (94%) had fixation by bone ingrowth, three (3%) showed stable fibrous ingrowth and three (3%) were unstable. Osteolysis of the femoral cortex was seen in seven hips (6%), with major lysis in only one. At a mean follow-up of ten years, the results of the Taperloc femoral component are comparable with those of modern techniques of cementing in primary total hip arthroplasty