We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still in
situ and functional in 42 of 53 elbows (79%) at a mean of
94.5 months (26 to 266) after surgery. In 31 of these 42 elbows
(74%) the Mayo Elbow Performance Score was good or excellent. Overall,
of the 53 elbows, 18 (34%) required re-operation, ten (19%) for
loosening. A classification system was developed to identify those
not suitable for revision by this technique, and using this we have
showed that successful
Intraspinal
While frequently discussed as a standard treatment for the management of an infected shoulder replacement, there is little information on the outcome of two-stage
This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection. Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was
We present the outcome of 47 Souter-Strathclyde replacements of the elbow with a mean follow-up of 82 months (12 to 129). The clinical results were assessed using a condition-specific outcome measure. The mean total score (maximum 100) before the operation was 47.21 and improved to 79.92 (p <
0.001). The mean pain score (maximum 50) improved from 21.41 to 46.70 (p <
0.001) and the mean functional component of the score (maximum 30) from 11.19 to 18.65 (p <
0.001). There was negligible change in the score for the range of movement although a significant improvement in mean flexion from 124° to 136° was noted (p <
0.001). Revision surgery was required in four patients, for dislocation, wound dehiscence and early infection in one, late infection in two and aseptic loosening in one. The cumulative survival was 75% at nine years for all causes of failure and 97% at ten years for aseptic loosening alone. Our study demonstrates the value of the Souter-Strathclyde total elbow arthroplasty in providing relief from pain and functional improvement in rheumatoid patients.
We performed a prospective study in 186 patients with a minimum follow-up of two years in order to attempt to predict successful major