Advertisement for orthosearch.org.uk
Results 1 - 6 of 6
Results per page:
Bone & Joint Open
Vol. 5, Issue 6 | Pages 457 - 463
2 Jun 2024
Coviello M Abate A Maccagnano G Ippolito F Nappi V Abbaticchio AM Caiaffa E Caiaffa V

Aims

Proximal femur fractures treatment can involve anterograde nailing with a single or double cephalic screw. An undesirable failure for this fixation is screw cut-out. In a single-screw nail, a tip-apex distance (TAD) greater than 25 mm has been associated with an increased risk of cut-out. The aim of the study was to examine the role of TAD as a risk factor in a cephalic double-screw nail.

Methods

A retrospective study was conducted on 112 patients treated for intertrochanteric femur fracture with a double proximal screw nail (Endovis BA2; EBA2) from January to September 2021. The analyzed variables were age, sex, BMI, comorbidities, fracture type, side, time of surgery, quality of reduction, pre-existing therapy with bisphosphonate for osteoporosis, screw placement in two different views, and TAD. The last follow-up was at 12 months. Logistic regression was used to study the potential factors of screw cut-out, and receiver operating characteristic curve to identify the threshold value.


Bone & Joint Open
Vol. 4, Issue 8 | Pages 584 - 593
15 Aug 2023
Sainio H Rämö L Reito A Silvasti-Lundell M Lindahl J

Aims

Several previously identified patient-, injury-, and treatment-related factors are associated with the development of nonunion in distal femur fractures. However, the predictive value of these factors is not well defined. We aimed to assess the predictive ability of previously identified risk factors in the development of nonunion leading to secondary surgery in distal femur fractures.

Methods

We conducted a retrospective cohort study of adult patients with traumatic distal femur fracture treated with lateral locking plate between 2009 and 2018. The patients who underwent secondary surgery due to fracture healing problem or plate failure were considered having nonunion. Background knowledge of risk factors of distal femur fracture nonunion based on previous literature was used to form an initial set of variables. A logistic regression model was used with previously identified patient- and injury-related variables (age, sex, BMI, diabetes, smoking, periprosthetic fracture, open fracture, trauma energy, fracture zone length, fracture comminution, medial side comminution) in the first analysis and with treatment-related variables (different surgeon-controlled factors, e.g. plate length, screw placement, and proximal fixation) in the second analysis to predict the nonunion leading to secondary surgery in distal femur fractures.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 114 - 121
1 Jan 2014
Pekmezci M McDonald E Buckley J Kandemir U

We investigated a new intramedullary locking nail that allows the distal interlocking screws to be locked to the nail. We compared fixation using this new implant with fixation using either a conventional nail or a locking plate in a laboratory simulation of an osteoporotic fracture of the distal femur. A total of 15 human cadaver femora were used to simulate an AO 33-A3 fracture pattern. Paired specimens compared fixation using either a locking or non-locking retrograde nail, and using either a locking retrograde nail or a locking plate. The constructs underwent cyclical loading to simulate single-leg stance up to 125 000 cycles. Axial and torsional stiffness and displacement, cycles to failure and modes of failure were recorded for each specimen. When compared with locking plate constructs, locking nail constructs had significantly longer mean fatigue life (75 800 cycles (sd 33 900) vs 12 800 cycles (sd 6100); p = 0.007) and mean axial stiffness (220 N/mm (sd 80) vs 70 N/mm (sd 18); p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (sd 0.9) vs 5.1 Nm/° (sd 1.5); p = 0.008). In addition, in the nail group the mode of failure was either cut-out of the distal screws or breakage of nails, and in the locking plate group breakage of the plate was always the mode of failure. Locking nail constructs had significantly longer mean fatigue life than non-locking nail constructs (78 900 cycles (sd 25 600) vs 52 400 cycles (sd 22 500); p = 0.04).

The new locking retrograde femoral nail showed better stiffness and fatigue life than locking plates, and superior fatigue life to non-locking nails, which may be advantageous in elderly patients.

Cite this article: Bone Joint J 2014;96-B:114–21.


The Bone & Joint Journal
Vol. 95-B, Issue 6 | Pages 825 - 830
1 Jun 2013
Abram SGF Pollard TCB Andrade AJMD

The Gamma nail is frequently used in unstable peri-trochanteric hip fractures. We hypothesised that mechanical failure of the Gamma nail was associated with inadequate proximal three-point fixation. We identified a consecutive series of 299 Gamma nails implanted in 299 patients over a five-year period, 223 of whom fulfilled our inclusion criteria for investigation. The series included 61 men and 162 women with a mean age of 81 years (20 to 101). Their fractures were classified according to the Modified Evans’ classification and the quality of fracture reduction was graded. The technical adequacy of three points of proximal fixation was recorded from intra-operative fluoroscopic images, and technical inadequacy for each point was defined. All patients were followed to final follow-up and mechanical failures were identified. A multivariate statistical analysis was performed, adjusting for confounders. A total of 16 failures (7.2%) were identified. The position of the lag screw relative to the lateral cortex was the most important point of proximal fixation, and when inadequate the failure rate was 25.8% (eight of 31: odds ratio 7.5 (95% confidence interval 2.5 to 22.7), p < 0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral fractures is rare (< 1%) when three-point proximal fixation is achieved. However, when proximal fixation is inadequate, failure rates increase. The strongest predictor of failure is positioning the lateral end of the lag screw short of the lateral cortex. Adherence to simple technical points minimises the risk of fixation failure in this vulnerable patient group. Cite this article: Bone Joint J 2013;95-B:825–30


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 391 - 397
1 Mar 2012
Parker MJ Bowers TR Pryor GA

In a randomised trial involving 598 patients with 600 trochanteric fractures of the hip, the fractures were treated with either a sliding hip screw (n = 300) or a Targon PF intramedullary nail (n = 300). The mean age of the patients was 82 years (26 to 104). All surviving patients were reviewed at one year with functional outcome assessed by a research nurse blinded to the treatment used. The intramedullary nail was found to have a slightly increased mean operative time (46 minutes (sd 12.3) versus 49 minutes (sd 12.7), p < 0.001) and an increased mean radiological screening time (0.3 minutes (sd 0.2) versus 0.5 minutes (sd 0.3), p <  0.001). Operative difficulties were more common with the intramedullary nail. There was no statistically significant difference between implants for wound healing complications (p = 1), or need for post-operative blood transfusion (p = 1), and medical complications were similarly distributed in both groups. There was a tendency to fewer revisions of fixation or conversion to an arthroplasty in the nail group, although the difference was not statistically significant (nine versus three cases, p = 0.14). The extent of shortening, loss of hip flexion, mortality and degree of residual pain were similar in both groups. The recovery of mobility was superior for those treated with the intramedullary nails (p = 0.01 at one year from injury).

In summary, both implants produced comparable results but there was a tendency to better return of mobility for those treated with the intramedullary nail.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 387 - 392
1 Mar 2011
Robinson CM Murray IR

Fractures and nonunions of the proximal humerus are increasingly treated by open reduction and internal fixation. The extended deltopectoral approach remains the most widely used for this purpose. However, it provides only limited exposure of the lateral and posterior aspects of the proximal humerus. We have previously described the alternative extended deltoid-splitting approach. In this paper we outline variations and extensions of this technique that we have developed in the management of further patients with these fractures.