Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article:
Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis. At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p <
0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate. There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.