Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up.Aims
Methods
Calcium sulphate has traditionally been used as a filler of dead space arising during surgery. Various complications have been described following the use of Stimulan bio-absorbable calcium sulphate beads. This study is a prospective observational study to assess the safety profile of these beads when used in revision arthroplasty, comparing the complication rates with those reported in the literature. A total of 755 patients who underwent 456 revision total knee arthroplasties (TKA) and 299 revision total hip arthroplasties (THA), with a mean follow-up of 35 months (0 to 78) were included in the study.Aims
Methods
We describe the use of a protocol of irrigation and debridement
(I&
D) with retention of the implant for the treatment of periprosthetic
infection of a total elbow arthroplasty (TEA). This may be an attractive
alternative to staged re-implantation. Between 1990 and 2010, 23 consecutive patients were treated in
this way. Three were lost to follow-up leaving 20 patients (21 TEAs)
in the study. There were six men and 14 women. Their mean age was
58 years (23 to 76). The protocol involved: component unlinking,
irrigation and debridement (I&
D), and the introduction of antibiotic
laden cement beads; organism-specific intravenous antibiotics; repeat
I&
D and re-linkage of the implant if appropriate; long-term
oral antibiotic therapy. Aims
Patients and Methods
The aim of this study was to establish what happens to patients
in the long term after endoprosthetic replacement for a primary
malignant tumour of bone. We conducted a retrospective analysis of a prospectively maintained
database to identify all patients who had undergone an endoprosthetic
replacement more than 25 years ago and who were still alive. Their
outcomes were investigated with reference to their complications and
need for further surgery. A total of 230 patients were identified.
Their mean age at diagnosis was 20.7 years (five to 62). The most
common diagnosis was osteosarcoma (132). The most common site was
the distal femur (102). Aims
Patients and Methods
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
Given the increasing number of total hip arthroplasty
procedures being performed annually, it is imperative that orthopaedic
surgeons understand factors responsible for instability. In order
to treat this potentially complex problem, we recommend correctly
classifying the type of instability present based on component position, abductor
function, impingement, and polyethylene wear. Correct classification
allows the treating surgeon to choose the appropriate revision option
that ultimately will allow for the best potential outcome. Cite this article:
Objectives. We wanted to investigate regional variations in the organisms
reported to be causing
The October 2015 Knee Roundup360 looks at: Allergy and outcome in arthroplasty; Physiotherapy and drains not such a bad combination?; Another nail in the coffin for arthroscopists?; Graft precondition hocus pocus; Extended dose steroids in knee arthritis?; Indolent
Periprosthetic joint infection (PJI) is a devastating
complication for patients and results in greatly increased costs
of care for both healthcare providers and patients. More than 15
500 revision hip and knee procedures were recorded in England, Wales
and Northern Ireland in 2013, with infection accounting for 13%
of revision hip and 23% of revision knee procedures. We report our experience of using antibiotic eluting absorbable
calcium sulphate beads in 15 patients (eight men and seven women
with a mean age of 64.8 years; 41 to 83) as part of a treatment
protocol for PJI in revision arthroplasty. The mean follow-up was 16 months (12 to 22). We report the outcomes
and complications, highlighting the risk of hypercalcaemia which
occurred in three patients. We recommend that serum levels of calcium be routinely sought
following the implantation of absorbable calcium sulphate beads
in orthopaedic surgery. Cite this article:
Charcot neuro-osteoarthropathy (CN) of the midfoot
presents a major reconstructive challenge for the foot and ankle
surgeon. The Synthes 6 mm Midfoot Fusion Bolt is both designed and
recommended for patients who have a deformity of the medial column
of the foot due to CN. We present the results from the first nine
patients (ten feet) on which we attempted to perform fusion of the
medial column using this bolt. Six feet had concurrent hindfoot fusion
using a retrograde nail. Satisfactory correction of deformity of
the medial column was achieved in all patients. The mean correction
of calcaneal pitch was from 6°
(-15° to +18°) pre-operatively to 16° (7° to 23°) post-operatively;
the mean Meary angle from 26° (3° to 46°) to 1° (1° to 2°); and
the mean talometatarsal angle on dorsoplantar radiographs from 27°
(1° to 48°) to 1° (1° to 3°). However, in all but two feet, at least one joint failed to fuse.
The bolt migrated in six feet, all of which showed progressive radiographic
osteolysis, which was considered to indicate loosening. Four of
these feet have undergone a revision procedure, with good radiological
evidence of fusion. The medial column bolt provided satisfactory correction
of the deformity but failed to provide adequate fixation for fusion
in CN deformities in the foot. In its present form, we cannot recommend the routine use of this
bolt. Cite this article:
Implant-associated infection is a major source
of morbidity in orthopaedic surgery. There has been extensive research
into the development of materials that prevent biofilm formation,
and hence, reduce the risk of infection. Silver nanoparticle technology
is receiving much interest in the field of orthopaedics for its
antimicrobial properties, and the results of studies to date are
encouraging. Antimicrobial effects have been seen when silver nanoparticles are
used in trauma implants, tumour prostheses, bone cement, and also
when combined with hydroxyapatite coatings. Although there are promising
results with Cite this article:
A retrospective review was performed of patients
undergoing primary cementless total knee replacement (TKR) using
porous tantalum performed by a group of surgical trainees. Clinical
and radiological follow-up involved 79 females and 26 males encompassing
115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up
was 7 years (2 to 11). Tibial and patellar components were porous
tantalum monoblock implants, and femoral components were posterior
stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological
assessments were made for implant positioning, alignment, radiolucencies,
lysis, and loosening. There was 95.7% survival of implants. There
was no radiological evidence of loosening and no osteolysis found.
No revisions were performed for aseptic loosening. Average tibial
component alignment was 1.4° of varus (4°of valgus to 9° varus),
and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean
femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral
alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar
tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient
satisfaction with improvement in pain was 91%. Cementless TKR incorporating
porous tantalum yielded good clinical and radiological outcomes
at a mean of follow-up of seven-years. Cite this article:
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
The treatment of infected nonunions is difficult.
Antibiotic cement-coated (ACC) rods provide stability as well as delivering
antibiotics. We conducted a review of 110 infected nonunions treated
with ACC rods. Patients were divided into two groups: group A (67
patients) with an infected arthrodesis, and group B (43 patients)
with an infected nonunion in a long bone. In group A, infected arthrodesis,
the success rate after the first procedure was 38/67 (57%), 29/67
(43%) required further surgery for either control of infection or
non-union. At last follow-up, five patients required amputation,
representing a limb salvage rate of 62/67 (93%) overall. In all,
29/67 (43%) presented with a bone defect with a mean size of 6.78
cm (2 to 25). Of those with a bone defect, 13/29 (45%) required
further surgery and had a mean size of defect of 7.2 cm (3.5 to
25). The cultures were negative in 17/67 (26%) and the most common
organism cultured was methicillin-resistant Cite this article:
The treatment of peri-prosthetic joint infection
(PJI) of the ankle is not standardised. It is not clear whether
an algorithm developed for hip and knee PJI can be used in the management
of PJI of the ankle. We evaluated the outcome, at two or more years
post-operatively, in 34 patients with PJI of the ankle, identified
from a cohort of 511 patients who had undergone total ankle replacement.
Their median age was 62.1 years (53.3 to 68.2), and 20 patients
were women. Infection was exogenous in 28 (82.4%) and haematogenous
in six (17.6%); 19 (55.9%) were acute infections and 15 (44.1%)
chronic. Staphylococci were the cause of 24 infections (70.6%).
Surgery with retention of one or both components was undertaken
in 21 patients (61.8%), both components were replaced in ten (29.4%),
and arthrodesis was undertaken in three (8.8%). An infection-free
outcome with satisfactory function of the ankle was obtained in
23 patients (67.6%). The best rate of cure followed the exchange
of both components (9/10, 90%). In the 21 patients in whom one or
both components were retained, four had a relapse of the same infecting organism
and three had an infection with another organism. Hence the rate
of cure was 66.7% (14 of 21). In these 21 patients, we compared
the treatment given to an algorithm developed for the treatment
of PJI of the knee and hip. In 17 (80.9%) patients, treatment was
not according to the algorithm. Most (11 of 17) had only one criterion against
retention of one or both components. In all, ten of 11 patients
with severe soft-tissue compromise as a single criterion had a relapse-free
survival. We propose that the treatment concept for PJI of the ankle
requires adaptation of the grading of quality of the soft tissues. Cite this article
Fungal
Total hip replacement (THR) still is a rare intervention
in many African countries. In Burkina Faso it is not performed on
a regular basis. A visiting programme for THR was started in a district
hospital with no previous relevant experience. In this paper we
present an analysis of the surgical technical problems and peri-operative
complications of 152 THRs in 136 patients and three bipolar hemiarthroplasties
in three patients undertaken in this new programme with limited
orthopaedic equipment. There were 86 male and 53 female patients
with a mean age of 49 years (21 to 78). We identified 77 intra-operative
technical problems in 51 operations. There were 24 peri-operative complications
in 21 patients, 17 of which were bony in nature. So far, ten revision
THRs have been performed in nine patients. Regular analysis of the technical problems and complications
was used to improve quality, and we identified patient selection
adapted to the local circumstances as important to avoid complications.
Our reflections on the problems encountered in initiating such a
programme may be of help to other teams planning similar projects. Cite this article: