Vogt B Lueckingsmeier M Gosheger G Laufer A Toporowski G Antfang C Roedl R Frommer A

Aims

As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach.

Methods

A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30).


Bone & Joint Open
Vol. 4, Issue 7 | Pages 539 - 550
21 Jul 2023
Banducci E Al Muderis M Lu W Bested SR

Aims

Safety concerns surrounding osseointegration are a significant barrier to replacing socket prosthesis as the standard of care following limb amputation. While implanted osseointegrated prostheses traditionally occur in two stages, a one-stage approach has emerged. Currently, there is no existing comparison of the outcomes of these different approaches. To address safety concerns, this study sought to determine whether a one-stage osseointegration procedure is associated with fewer adverse events than the two-staged approach.

Methods

A comprehensive electronic search and quantitative data analysis from eligible studies were performed. Inclusion criteria were adults with a limb amputation managed with a one- or two-stage osseointegration procedure with follow-up reporting of complications.


Bone & Joint Open
Vol. 4, Issue 3 | Pages 146 - 157
7 Mar 2023
Camilleri-Brennan J James S McDaid C Adamson J Jones K O'Carroll G Akhter Z Eltayeb M Sharma H

Aims

Chronic osteomyelitis (COM) of the lower limb in adults can be surgically managed by either limb reconstruction or amputation. This scoping review aims to map the outcomes used in studies surgically managing COM in order to aid future development of a core outcome set.

Methods

A total of 11 databases were searched. A subset of studies published between 1 October 2020 and 1 January 2011 from a larger review mapping research on limb reconstruction and limb amputation for the management of lower limb COM were eligible. All outcomes were extracted and recorded verbatim. Outcomes were grouped and categorized as per the revised Williamson and Clarke taxonomy.


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 88 - 96
1 Jan 2023
Vogt B Rupp C Gosheger G Eveslage M Laufer A Toporowski G Roedl R Frommer A

Aims. Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. Methods. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages. Results. Osteolysis next to the telescopic junction was observed in 31/48 segments (65%) lengthened with the STRYDE nail before implant removal compared to 1/91 segment (1%) in the PRECICE cohort. In the STRYDE cohort, osteolysis initially increased, but decreased or resolved in almost all lengthened segments (86%) after implant removal. Implant failure was observed in 9/48 STRYDE (19%) and in 8/92 PRECICE nails (9%). Breakage of the distal locking bolts was found in 5/48 STRYDE nails (10%) compared to none in the PRECICE cohort. Treatment-associated pain was generally recorded as mild and found in 30/48 patients (63%) and 39/92 (42%) in the STRYDE and PRECICE cohorts, respectively. Temporary range of motion (ROM) limitations under distraction were registered in 17/48 (35%) segments treated with the STRYDE and 35/92 segments (38%) treated with the PRECICE nail. Conclusion. Osteolysis and periosteal reaction, implant breakage, and pain during lengthening and consolidation is more likely in patients treated with the STRYDE nail compared to the PRECICE nail. Temporary ROM limitations during lengthening occurred independent of the applied device. Implant-related osseous alterations seem to remodel after implant removal. Cite this article: Bone Joint J 2023;105-B(1):88–96


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims. This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported. Methods. This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed. Results. Adjacent to the nail’s telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail’s telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion. Conclusion. Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article: Bone Joint Res 2021;10(7):425–436


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1168 - 1172
1 Jun 2021
Iliadis AD Wright J Stoddart MT Goodier WD Calder P

Aims

The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device.

Methods

This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1561 - 1565
1 Nov 2014
Park JW Kim YS Yoon JO Kim JS Chang JS Kim JM Chun JM Jeon IH

Non-tuberculous mycobacterial (NTM) infection of the musculoskeletal tissue is a rare disease. An early and accurate diagnosis is often difficult because of the indolent clinical course and difficulty of isolating pathogens. Our goal was to determine the clinical features of musculoskeletal NTM infection and to present the treatment outcomes. A total of 29 patients (nine females, 20 males between 34 and 85 years old, mean age 61.7 years; 34 to 85) with NTM infection of the musculoskeletal system between 1998 to 2011 were identified and their treatment retrospectively analysed. Microbiological studies demonstrated NTM in 29 patients: the isolates were Mycobacterium intracellulare in six patients, M. fortuitum in three, M. abscessus in two and M. marinum in one. In the remaining patients we failed to identify the species. The involved sites were the hand/wrist in nine patients the knee in five patients, spine in four patients, foot in two patients, elbow in two patients, shoulder in one, ankle in two patients, leg in three patients and multiple in one patient. The mean interval between the appearance of symptoms and diagnosis was 20.8 months (1.5 to 180). All patients underwent surgical treatment and antimicrobial medication according to our protocol for chronic musculoskeletal infection: 20 patients had NTM-specific medication and nine had conventional antimicrobial therapy. At the final follow-up 22 patients were cured, three failed to respond to treatment and four were lost to follow-up. Identifying these diseases due the initial non-specific presentation can be difficult. Treatment consists of surgical intervention and adequate antimicrobial therapy, which can result in satisfactory outcomes.

Cite this article: Bone Joint J 2014;96-B:1561–5.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 807 - 811
1 Jun 2006
Roche SJ Fitzgerald D O’Rourke A McCabe JP

This prospective five-year study analyses the impact of methicillin-resistant Staphylococcus aureus (MRSA) on an Irish orthopaedic unit. We identified 318 cases of MRSA, representing 0.76% of all admissions (41 971). A total of 240 (76%) cases were colonised with MRSA, while 120 (37.7%) were infected. Patients were admitted from home (218; 68.6%), nursing homes (72; 22.6%) and other hospitals (28; 8.8%). A total of 115 cases (36.6%) were colonised or infected on admission. Many patients were both colonised and infected at some stage. The length of hospital stay was almost trebled because of the presence of MRSA infection.

Encouragingly, overall infection rates have not risen significantly over the five years of the study despite increased prevalence of MRSA. However, the financial burden of MRSA is increasing, highlighting the need for progress in understanding how to control this resistant pathogen more effectively.