There is little information available at present regarding the mechanisms of failure of modern metallic radial head implants. Between 1998 and 2008, 44 consecutive patients (47 elbows) underwent removal of a failed metallic radial head replacement. In 13 patients (13 elbows) the initial operation had been undertaken within one week of a fracture of the radial head, at one to six weeks in seven patients (seven elbows) and more than six weeks (mean of 2.5 years (2 to 65 months)) in 22 patients (25 elbows). In the remaining two elbows the replacement was inserted for non-traumatic reasons. The most common indication for further surgery was painful loosening (31 elbows). Revision was undertaken for stiffness in 18
We assessed the long-term results of 58 Souter-Strathclyde total elbow replacements in 49 patients with rheumatoid arthritis. The mean length of follow-up was 9.5 years (0.7 to 16.7). The mean pre-operative Mayo Elbow Performance Score was 30 (15 to 80) and at final follow-up was 82 (60 to 95). A total of 13 elbows (22.4%) were revised, ten (17.2%) for aseptic loosening, one (1.7%) for instability, one (1.7%) for secondary loosening after fracture, and one elbow (1.7%) was removed because of deep infection. The Kaplan-Meier survival rate was 70% and 53% at ten and 16 years, respectively. Failure of the ulnar component was found to be the main problem in relation to the loosening. Anterior transposition of the ulnar nerve had no influence on ulnar nerve paresthaesiae in these patients.
Excision is not a suitable treatment for all comminuted fractures of the radial head. In
We have treated 16 patients with recurrent complex