The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
Aims. The aim of this study was to identify predictors of return to
work (RTW) after revision lower limb arthroplasty in patients of
working age in the United Kingdom. Patients and Methods. We assessed 55 patients aged ≤ 65 years after revision total
hip arthroplasty (THA). There were 43 women and 12 men with a mean
age of 54 years (23 to 65). We also reviewed 30 patients after revision
total knee arthroplasty (TKA). There were 14 women and 16 men with
a mean age of 58 years (48 to 64). Preoperatively, age, gender,
body mass index, social deprivation, mode of failure, length of
primary implant survival, work status and nature, activity level
(University of California, Los Angeles (UCLA) score), and Oxford
Hip and Knee Scores were recorded. Postoperatively, RTW status,
Oxford Hip and Knee Scores, EuroQol-5D (EQ-5D), UCLA score, and
Work, Osteoarthritis and Joint-Replacement Questionnaire (WORQ)
scores were obtained. Univariate and multivariate analysis was performed. Results. Overall, 95% (52/55) of patients were working before their revision
THA. Afterwards, 33% (17/52) RTW by one year, 48% (25/52) had retired,
and 19% (10/52) were receiving welfare benefit. RTW was associated
with age, postoperative Oxford Hip Score,
Tissue responses to debris formed by abrasion of polymethylmethacrylate
(PMMA) spacers at two-stage revision arthroplasty for prosthetic
joint infection are not well described. We hypothesised that PMMA
debris induces immunomodulation in periprosthetic tissues. Samples of tissue were taken during 35 two-stage revision arthroplasties
(nine total hip and 26 total knee arthroplasties) in patients whose
mean age was 67 years (44 to 85). Fourier transform infrared microscopy
was used to confirm the presence of PMMA particles. Histomorphometry
was performed using Sudan Red and Haematoxylin-Eosin staining.
CD-68, CD-20, CD-11(c), CD-3 and IL-17 antibodies were used to immunophenotype
the inflammatory cells. All slides were scored semi-quantitatively
using the modified Willert scoring system.Aims
Patients and Methods
The aim of this study was to present data on 11 459 patients
who underwent total hip (THA), total knee (TKA) or unicompartmental
knee arthroplasty (UKA) between November 2002 and April 2014 with
aspirin as the primary agent for pharmacological thromboprophylaxis. We analysed the incidence of deep vein thrombosis (DVT) and pulmonary
embolism (PE) then compared the 90-day all-cause mortality with
the corresponding data in the National Joint Registry for England
and Wales (NJR). Aims
Patients and Methods
Objectives. Wear debris released from bearing surfaces has been shown to
provoke negative immune responses in the recipient. Excessive wear
has been linked to
The purpose of this study was to compare the
diagnostic accuracy for the detection of infection between the culture of
fluid obtained by sonication (SFC) and the culture of peri-implant
tissues (PITC) in patients with early and delayed implant failure,
and those with unsuspected and suspected septic failure. It was
hypothesised that SFC increases the diagnostic accuracy for infection
in delayed, but not early, implant failure, and in unsuspected septic
failure. The diagnostic accuracy for infection of all consecutive
implants (hardware or prostheses) that were removed for failure was
compared between SFC and PITC. This prospective study included 317
patients with a mean age of 62.7 years (9 to 97). The sensitivity
for detection of infection using SFC was higher than using PITC
in an overall comparison (89.9% versus 67%, respectively;
p <
0.001), in unsuspected septic failure (100% versus 48.5%,
respectively; p <
0.001), and in delayed implant failure (88% versus 58%,
respectively; p <
0.001). PITC sensitivity dropped significantly
in unsuspected compared with suspected septic failure (p = 0.007),
and in delayed compared with
We analysed data from the Oxford hip and knee questionnaires collected by the New Zealand Joint Registry at six months and five years after joint replacement, to determine if there was any relationship between the scores and the risk of early revision. Logistic regression of the six-month scores indicated that for every one-unit decrease in the Oxford score, the risk of revision within two years increased by 9.7% for total hip replacement (THR), 9.9% for total knee replacement (TKR) and 12.0% for unicompartmental knee replacement (UKR). Our findings showed that 70% of the revisions within two years for TKR and 67% for THR and UKR would have been captured by monitoring the lowest 22%, 28% and 28%, respectively, of the Oxford scores. When analysed using the Kalairajah classification a score of <
27 (poor) was associated with a risk of revision within two years of 7.6% for THR, 7.0% for TKR and 24.3% for UKR, compared with risks of 0.7%, 0.7% and 1.8%, respectively, for scores >
34 (good or excellent). Our study confirms that the Oxford hip and knee scores at six months are useful predictors of early revision after THR and TKR and we recommend their use for the monitoring of the outcome and potential failure in these patients.