Vol. 5, Issue 11 | Pages 971 - 976
5 Nov 2024
Baker G Hill J O'Neill F McChesney J Stevenson M Beverland D

Aims

In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes.

Methods

A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 548 - 554
1 Jun 2024
Ohyama Y Minoda Y Masuda S Sugama R Ohta Y Nakamura H

Aims

The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group).

Methods

The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1045 - 1051
1 Oct 2023
Turgeon TR Righolt CH Burnell CD Gascoyne TC Hedden DR Bohm ER

Aims

The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components.

Methods

Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).


Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal stress-shielding was seen with the CFP stem, suggesting diaphyseal fixation, and questioning its femoral neck-sparing properties in the long term. Cite this article: Bone Joint J 2022;104-B(5):581–588


Bone & Joint Research
Vol. 11, Issue 3 | Pages 180 - 188
1 Mar 2022
Rajpura A Asle SG Ait Si Selmi T Board T

Aims

Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane.

Methods

Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. Methods. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis. Results. Results from mixed model analyses revealed a lower mean retroversion for line-to-line (0.72° (95% confidence interval (CI) 0.38° to 1.07°; p < 0.001), but no significant differences in subsidence between the techniques (-0.15 mm (95% CI -0.53 to 0.227; p = 0.429) at 24 months. Radiolucent lines measuring < 2 mm wide were found in three and five arthroplasties cemented by the standard and line-to-line method, respectively. Conclusion. The cemented Corail stem with a force-closed design seems to settle earlier and better with the line-to-line cementing method, although for subsidence the difference was not significant. However, the lower rate of migration into retroversion may reduce the wear and cement deformation, contributing to good long-term fixation and implant survival. Cite this article: Bone Joint J 2022;104-B(1):19–26


Bone & Joint Open
Vol. 2, Issue 9 | Pages 696 - 704
1 Sep 2021
Malhotra R Gautam D Gupta S Eachempati KK

Aims

Total hip arthroplasty (THA) in patients with post-polio residual paralysis (PPRP) is challenging. Despite relief in pain after THA, pre-existing muscle imbalance and altered gait may cause persistence of difficulty in walking. The associated soft tissue contractures not only imbalances the pelvis, but also poses the risk of dislocation, accelerated polyethylene liner wear, and early loosening.

Methods

In all, ten hips in ten patients with PPRP with fixed pelvic obliquity who underwent THA as per an algorithmic approach in two centres from January 2014 to March 2018 were followed-up for a minimum of two years (2 to 6). All patients required one or more additional soft tissue procedures in a pre-determined sequence to correct the pelvic obliquity. All were invited for the latest clinical and radiological assessment.


Bone & Joint Open
Vol. 2, Issue 5 | Pages 293 - 300
3 May 2021
Lewis PM Khan FJ Feathers JR Lewis MH Morris KH Waddell JP

Aims

“Get It Right First Time” (GIRFT) and NHS England’s Best Practice Tariff (BPT) have published directives advising that patients over the ages of 65 (GIRFT) and 69 years (BPT) receiving total hip arthroplasty (THA) should receive cemented implants and have brought in financial penalties if this policy is not observed. Despite this, worldwide, uncemented component use has increased, a situation described as a ‘paradox’. GIRFT and BPT do, however, acknowledge more data are required to support this edict with current policies based on the National Joint Registry survivorship and implant costs.

Methods

This study compares THA outcomes for over 1,000 uncemented Corail/Pinnacle constructs used in all age groups/patient frailty, under one surgeon, with identical pre- and postoperative pathways over a nine-year period with mean follow-up of five years and two months (range: nine months to nine years and nine months). Implant information, survivorship, and regular postoperative Oxford Hip Scores (OHS) were collected and two comparisons undertaken: a comparison of those aged over 65 years with those 65 and under and a second comparison of those aged 70 years and over with those aged under 70.


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1654 - 1661
1 Dec 2020
Perelgut ME Polus JS Lanting BA Teeter MG

Aims

The direct anterior (DA) approach has been associated with rapid patient recovery after total hip arthroplasty (THA) but may be associated with more frequent femoral complications including implant loosening. The objective of this study was to determine whether the addition of a collar to the femoral stem affects implant migration, patient activity, and patient function following primary THA using the DA approach.

Methods

Patients were randomized to either a collared (n = 23) or collarless (n = 26) cementless femoral stem implanted using the DA approach. Canal fill ratio (CFR) was measured on the first postoperative radiographs. Patients underwent a supine radiostereometric analysis (RSA) exam postoperatively on the day of surgery and at two, four, six, 12, 26, and 52 weeks postoperatively. Patient-reported outcome measures (Western Ontario and McMaster Universities Osteoarthritis (WOMAC) Index, the 12-item Short Form Health Survey Mental and Physical Score, and University of California, Los Angeles (UCLA) Activity Score) were measured preoperatively and at each post-surgery clinic visit. Activity and function were also measured as the weekly average step count recorded by an activity tracker, and an instrumented timed up-and-go (TUG) test in clinic, respectively.


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 779 - 786
1 Jul 2019
Lamb JN Baetz J Messer-Hannemann P Adekanmbi I van Duren BH Redmond A West RM Morlock MM Pandit HG

Aims

The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model.

Materials and Methods

A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an in vitro model using paired fresh frozen cadaveric femora.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 536 - 539
1 May 2019
Cassidy RS O hEireamhoin S Beverland DE

Aims

The aim of this retrospective audit was to determine the route of referral or presentation of patients requiring revision following primary total hip arthroplasty (THA).

Patients and Methods

A total of 4802 patients were implanted with an Orthopaedic Data Evaluation Panel (ODEP) 10A* cementless implant (Corail/Pinnacle) between 2005 and 2015; 80 patients with a mean age of 67.8 years (sd 10.8) underwent a subsequent revision. The primary outcome measure was route of referral for revision.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1559 - 1564
1 Dec 2018
McConnell JS Saunders PRJ Young SK

Aims. Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. Patients and Methods. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (. sd. 12.3) and median body mass index (BMI) of 29 kg/m. 2. (interquartile range (IQR) 26 to 32). The recordings were analyzed to identify the frequencies of the sounds produced during impaction of the femoral broach. Results. The emergence of a low-frequency band of sound in the 1 kHz range, during the final femoral broaching, was a strong predictor of a well-sized implant stem. The frequency was related to femoral length, supporting our hypothesis that the sound arose from the bone itself. Conclusion. The low-frequency sound generated during femoral broaching can be monitored spectrographically, its frequency can be predicted from femoral length, and it is a good predictor of appropriate stem sizing


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1010 - 1017
1 Aug 2018
Jain S Magra M Dube B Veysi VT Whitwell GS Aderinto JB Emerton ME Stone MH Pandit HG

Aims

This study aimed to evaluate implant survival of reverse hybrid total hip arthroplasty (THA) at medium-term follow-up.

Patients and Methods

A consecutive series of 1082 THAs in 982 patients with mean follow-up of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21 to 94). Of these, 194 (17.9%) were in patients under 60 years, 663 (61.3%) in female patients and 348 (32.2%) performed by a trainee. Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%). Survival analysis was performed and subgroups compared using log rank tests.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 33 - 41
1 Jan 2018
Matharu GS Hunt LP Murray DW Howard P Pandit HG Blom AW Bolland B Judge A

Aims

The aim of this study was to determine whether the rates of revision for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle components varied according to the year of the initial operation, and compare these with the rates of revision for other designs of MoM THA.

Patients and Methods

Data from the National Joint Registry for England and Wales included 36 mm MoM THAs with Pinnacle acetabular components which were undertaken between 2003 and 2012 with follow-up for at least five years (n = 10 776) and a control group of other MoM THAs (n = 13 817). The effect of the year of the primary operation on all-cause rates of revision was assessed using Cox regression and interrupted time-series analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1435 - 1441
1 Nov 2017
Buttaro MA Oñativia JI Slullitel PA Andreoli M Comba F Zanotti G Piccaluga F

Aims. The Corail stem has good long-term results. After four years of using this stem, we have detected a small group of patients who have presented with symptomatic metaphyseal debonding. The aim of this study was to quantify the incidence of this complication, to delineate the characteristics of patients presenting with this complication and to compare these patients with asymptomatic controls to determine any important predisposing factors. Patients and Methods. Of 855 Corail collarless cementless stems implanted for osteoarthritis, 18 presented with symptomatic metaphyseal debonding. A control group of 74 randomly selected patients was assembled. Clinical and radiological parameters were measured and a logistic regression model was created to evaluate factors associated with metaphyseal debonding. Results. The prevalence of this complication was 2.1% in our series. In the multivariable model, the presence of a Dorr B-type proximal femur was associated with metaphyseal debonding (odds ratio (OR) 10.73, 95% confidence interval (CI) 2.31 to 49.97, p = 0.002), as was a body mass index > 25 kg/m. 2. (OR 6.85, 95% CI 1.06 to 44.28, p = 0.04). Smaller stems and the use of a polyethylene acetabular liner appeared to be protective when compared with metal and ceramic setting hard-on-hard bearings. Conclusion. We have described an uncommon but important mode of failure of the Corail stem. Surgeons should be aware of this phenomenon; overweight patients with Dorr B-type femurs and in whom hard bearings are used appear to be particularly at risk. Cite this article: Bone Joint J 2017;99-B:1435–41


The Bone & Joint Journal
Vol. 99-B, Issue 9 | Pages 1147 - 1152
1 Sep 2017
Lainiala O Reito A Jämsä P Eskelinen A

Aims

To determine whether there is any association between glomerular filtration rate (GFR) and blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasty.

Patients and Methods

We identified 179 patients with a unilateral 36 mm diameter head as part of a stemmed Summit-Pinnacle MoM hip arthroplasty. GFR was calculated using the Modification of Diet in Renal Disease formula.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 445 - 450
1 Apr 2017
Marsh AG Nisar A El Refai M Patil S Meek RMD

Aims

The purpose of this study was to evaluate whether an innovative templating technique could predict the need for acetabular augmentation during primary total hip arthroplasty for patients with dysplastic hips.

Patients and Methods

We developed a simple templating technique to estimate acetabular component coverage at total hip arthroplasty, the True Cup: False Cup (TC:FC) ratio. We reviewed all patients with dysplastic hips who underwent primary total hip arthroplasty between 2005 and 2012. Traditional radiological methods of assessing the degree of acetabular dysplasia (Sharp’s angle, Tönnis angle, centre-edge angle) as well as the TC:FC ratio were measured from the pre-operative radiographs. A comparison of augmented and non-augmented hips was undertaken to determine any difference in pre-operative radiological indices between the two cohorts. The intra- and inter-observer reliability for all radiological indices used in the study were also calculated.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 741 - 748
1 Jun 2015
Bonnin MP Neto CC Aitsiselmi T Murphy CG Bossard N Roche S

The aim of this study was to investigate the relationship between the geometry of the proximal femur and the incidence of intra-operative fracture during uncemented total hip arthroplasty (THA). We studied the pre-operative CT scans of 100 patients undergoing THA with an uncemented femoral component. We measured the anteroposterior and mediolateral dimensions at the level of division of the femoral neck to calculate the aspect ratio of the femur. Wide variations in the shape of the femur were observed, from round, to very narrow elliptic. The femurs of women were narrower than those of men (p < 0.0001) and small femurs were also narrower than large ones. Patients with an intra-operative fracture of the calcar had smaller and narrower femurs than those without a fracture (p < 0.05) and the implanted Corail stems were smaller in those with a fracture (mean size 9 vs 12, p < 0.0001). The variability of the shape of the femoral neck at the level of division contributes to the understanding of the causation of intra-operative fractures in uncemented THA. Cite this article: Bone Joint J 2015;97-B:741–8


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 762 - 770
1 Jun 2015
Pennington MW Grieve R van der Meulen JH

There is little evidence on the cost effectiveness of different brands of hip prostheses. We compared lifetime cost effectiveness of frequently used brands within types of prosthesis including cemented (Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed) and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy). We used data from three linked English national databases to estimate the lifetime risk of revision, quality-adjusted life years (QALYs) and cost.

For women with osteoarthritis aged 70 years, the Exeter V40 Elite Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0 QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision risk, 9.22 QALYs), the most commonly used brand, and assuming a willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is most cost effective, with an incremental net monetary benefit of £876. Differences in cost effectiveness between the hybrid CPT Trilogy and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc Exceed were small, and a cautious interpretation is required, given the limitations of the available information.

However, it is unlikely that cemented brands are among the most cost effective. Similar patterns of results were observed for men and other ages. The gain in quality of life after total hip arthroplasty, rather than the risk of revision, was the main driver of cost effectiveness.

Cite this article: Bone Joint J 2015;97-B:762–70.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (sd 1.71) vs 0.32 mm (sd 0.45), p = 0.66; rotation 0.96° (sd 1.49) vs 1.41° (sd 2.95), p = 0.88; and total migration 1.09 mm (sd 1.74) vs 0.73 mm (sd 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem.