Advertisement for orthosearch.org.uk
Results 1 - 20 of 61
Results per page:
The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1052 - 1059
1 Oct 2023
El-Sahoury JAN Kjærgaard K Ovesen O Hofbauer C Overgaard S Ding M

Aims

The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA).

Methods

A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 768 - 774
1 Jul 2023
Wooster BM Kennedy NI Dugdale EM Sierra RJ Perry KI Berry DJ Abdel MP

Aims

Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis.

Methods

We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 833 - 843
1 Jul 2022
Kayani B Baawa-Ameyaw J Fontalis A Tahmassebi J Wardle N Middleton R Stephen A Hutchinson J Haddad FS

Aims. This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Methods. Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up. Results. Patients in Group C were associated with increased mean liner wear rates compared to patients in Group A (0.133 mm/yr (SD 0.21) vs 0.031 mm/yr (SD 0.07), respectively; p < 0.001) and Group B (0.133 mm/yr (SD 0.21) vs 0.022 mm/yr (SD 0.05), respectively; p < 0.001) at ten years’ follow-up. Patients in Group C were also associated with increased risk of osteolysis and aseptic loosening requiring revision surgery, compared with patients in Group A (7/133 vs 0/133, respectively; p = 0.007) and Group B (7/133 vs 0/135, respectively; p = 0.007). There was a non-statistically significant trend towards increased mean liner wear rates in Group A compared with Group B (0.031 mm/yr (SD 0.07) vs 0.022 mm/yr (SD 0.05), respectively; p = 0.128). All three groups were statistically comparable preoperatively and at ten years’ follow-up when measuring normalized Western Ontario and McMaster Universities Osteoarthritis Index (p = 0.410), 36-Item Short Form Health Survey (p = 0.465 mental, p = 0.713 physical), and pain scale scores (p = 0.451). Conclusion. The use of UHMWPE was associated with progressively increased annual liner wear rates after THA compared to XLPE. At ten years’ follow-up, the group receiving UHMWPE demonstrated an increased incidence of osteolysis and aseptic loosening requiring revision surgery compared to XLPE. Femoral heads composed of OxZr were associated with trend towards reduced wear rates compared to CoCr, but this did not reach statistical significance and did not translate to any differences in osteolysis, functional outcomes, or revision surgery between the two femoral head components. Cite this article: Bone Joint J 2022;104-B(7):833–843


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 852 - 858
1 Jul 2022
Grothe T Günther K Hartmann A Blum S Haselhoff R Goronzy J

Aims

Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors.

Methods

Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 820 - 825
1 Jul 2022
Dhawan R Baré JV Shimmin A

Aims

Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a DMA.

Methods

A multicentre study was performed using data from 227 patients undergoing primary total hip arthroplasty (THA), enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameter had a DMA inserted at the time of their surgery. The mean age was 76 years (22 to 93) and 63% were female (n = 145). At a mean of 14 months (5 to 31) postoperatively, the AOANJRR was analyzed for follow-up information. Reasons for revision and types of revision were identified.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 359 - 367
1 Mar 2022
Deere K Matharu GS Ben-Shlomo Y Wilkinson JM Blom AW Sayers A Whitehouse MR

Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined. Results. There were 158,677 individuals (42.4%) with an implant containing CoCr. There were 47,963 deaths, 27,332 heart outcomes, 35,720 cancers, and 22,025 neurodegenerative disorders. There was no evidence of an association between patients with CoCr implants and higher rates of any of the outcomes. Conclusion. CoCr-containing THAs did not have an increased risk of all-cause mortality, or clinically meaningful heart outcomes, cancer, or neurodegenerative disorders into the second decade post-implantation. Our findings will help reassure clinicians and the increasing number of patients receiving primary THA worldwide that the use of CoCr-containing implants is not associated with significant adverse systemic effects. Cite this article: Bone Joint J 2022;104-B(3):359–367


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1604 - 1610
1 Oct 2021
Takaoka Y Goto K Tamura J Okuzu Y Kawai T Kuroda Y Orita K Matsuda S

Aims

We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive.

Methods

We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims

We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice.

Methods

Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 20 - 26
1 Jul 2020
Romero J Wach A Silberberg S Chiu Y Westrich G Wright TM Padgett DE

Aims. This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. Methods. For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current. Results. The radiological review identified that 32 of 551 MDM liners (5.8%) were malseated. Malseating was noted in all of the three different cup designs. The incidence of malseating was significantly higher in low-volume MDM surgeons than high-volume MDM surgeons (p < 0.001). Pristine well-seated liners showed significantly lower fretting current values at all peak loads greater than 800 N (p < 0.044). Malseated liner-shell couples had lower fretting onset loads at 2,400 N. Conclusion. MDM malseating remains an issue that can occur in at least one in 20 patients at a high-volume arthroplasty centre. The onset of fretting and increased fretting current throughout loading cycles suggests susceptibility to corrosion when this occurs. These results support the hypothesis that malseated liners may be at risk for fretting corrosion. Clinicians should be aware of this phenomenon. Cite this article: Bone Joint J 2020;102-B(7 Supple B):20–26


Bone & Joint Research
Vol. 8, Issue 10 | Pages 443 - 450
1 Oct 2019
Treacy RBC Holland JP Daniel J Ziaee H McMinn DJW

Objectives. Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. Methods. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (. sd. 11; 21 to 82), and 73% of implantations were in female patients. All patients were individually counselled, and a detailed informed consent was obtained prior to operation. Primary resurfacing was carried out in 85 hips, and three cases involved revision of previous MoM HRA. Clinical, radiological, and Oxford Hip Score (OHS) assessments were studied, along with implant survival. Results. There was no loss to follow-up and no actual or impending revision or reoperation. Median OHS increased from 24 (interquartile range (IQR) 20 to 28) preoperatively to 48 (IQR 46 to 48) at the latest follow-up (48 being the best possible score). Radiographs showed one patient had a head-neck junction lucency. No other radiolucency, osteolysis, component migration, or femoral neck thinning was noted. Conclusion. The results in this small consecutive cohort suggest that metal-on-monobloc-XLPE HRA is successful in the short term and merits further investigation as a conservative alternative to the current accepted standard of stemmed THA. However, we would stress that survival data with longer-term follow-up are needed prior to widespread adoption. Cite this article: R. B. C. Treacy, J. P. Holland, J. Daniel, H. Ziaee, D. J. W. McMinn. Preliminary report of clinical experience with metal-on-highly-crosslinked-polyethylene hip resurfacing. Bone Joint Res 2019;8:443–450. DOI: 10.1302/2046-3758.810.BJR-2019-0060.R1


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 378 - 385
1 Apr 2019
García-Rey E Carbonell-Escobar R Cordero-Ampuero J García-Cimbrelo E

Aims

We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up.

Patients and Methods

Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1002 - 1009
1 Aug 2018
Westerman RW Whitehouse SL Hubble MJW Timperley AJ Howell JR Wilson MJ

Aims

The aim of this study was to report the initial results of the Exeter V40 stem, which became available in 2000.

Patients and Methods

A total of 540 total hip arthroplasties (THAs) were performed in our unit using this stem between December 2000 and May 2002. Our routine protocol is to review patients postoperatively and at one, five, and ten years following surgery.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 33 - 41
1 Jan 2018
Matharu GS Hunt LP Murray DW Howard P Pandit HG Blom AW Bolland B Judge A

Aims

The aim of this study was to determine whether the rates of revision for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle components varied according to the year of the initial operation, and compare these with the rates of revision for other designs of MoM THA.

Patients and Methods

Data from the National Joint Registry for England and Wales included 36 mm MoM THAs with Pinnacle acetabular components which were undertaken between 2003 and 2012 with follow-up for at least five years (n = 10 776) and a control group of other MoM THAs (n = 13 817). The effect of the year of the primary operation on all-cause rates of revision was assessed using Cox regression and interrupted time-series analysis.


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 44 - 49
1 Jan 2018
Berstock JR Whitehouse MR Duncan CP

Aims

To present a surgically relevant update of trunnionosis.

Materials and Methods

Systematic review performed April 2017.


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1590 - 1595
1 Dec 2017
Atrey A Ward SE Khoshbin A Hussain N Bogoch E Schemitsch EH Waddell JP

Aims

We present the ten-year data of a cohort of patients, aged between 18 and 65 years (mean age 52.7 years; 19 to 64), who underwent total hip arthroplasty. Patients were randomised to be treated with a cobalt-chrome (CoCr) femoral head with an ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE) or ceramic-on-ceramic (CoC) bearing surface.

Patients and Methods

A total of 102 hips (91 patients) were randomised into the three groups. At ten years, 97 hips were available for radiological and functional follow-up. Two hips (two patients) had been revised (one with deep infection and one for periprosthetic fracture) and three were lost to follow-up. Radiological analysis was performed using a validated digital assessment programme to give linear, directional and volumetric wear of the two polyethylene groups.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 766 - 773
1 Jun 2017
Graves SE de Steiger R Davidson D Donnelly W Rainbird S Lorimer MF Cashman KS Vial RJ

Aims

Femoral stems with exchangeable (modular) necks were introduced to offer surgeons an increased choice when determining the version, offset and length of the femoral neck during total hip arthroplasty (THA). It was hoped that this would improve outcomes and reduce complications, particularly dislocation. In 2010, the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) first reported an increased rate of revision after primary THA using femoral stems with an exchangeable neck. The aim of this study was to provide a more comprehensive up-to-date analysis of primary THA using femoral stems with exchangeable and fixed necks.

Materials and Methods

The data included all primary THA procedures performed for osteoarthritis (OA), reported to the AOANJRR between 01 September 1999 and 31 December 2014. There were 9289 femoral stems with an exchangeable neck and 253 165 femoral stems with a fixed neck. The characteristics of the patients and prostheses including the bearing surface and stem/neck metal combinations were examined using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 325 - 329
1 Mar 2017
Viste A Perry KI Taunton MJ Hanssen AD Abdel MP

Aims

Loss or absence of proximal femoral bone in revision total hip arthroplasty (THA) remains a significant challenge. While the main indication for the use of proximal femoral replacements (PFRs) is in the treatment of malignant disease, they have a valuable role in revision THA for loosening, fracture and infection in patients with bone loss. Our aim was to determine the clinical outcomes, implant survivorship, and complications of PFRs used in revision THA for indications other than malignancy.

Patients and Methods

A retrospective review of 44 patients who underwent revision THA using a PFR between 2000 and 2013 was undertaken. Their mean age was 79 years (53 to 97); 31 (70%) were women. The bone loss was classified as Paprosky IIIB or IV in all patients. The mean follow-up was six years (2 to 12), at which time 22 patients had died and five were lost to follow-up.


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 29 - 36
1 Jan 2017
Martino ID Santis VD Apolito RD Sculco PK Cross MB Gasparini G

Aims

We report on the outcome of the Synergy cementless femoral stem with a minimum follow-up of 15 years (15 to 17).

Patients and Methods

A retrospective review was undertaken of a consecutive series of 112 routine primary cementless total hip arthroplasties (THAs) in 102 patients (112 hips). There were 60 female and 42 male patients with a mean age of 61 years (18 to 82) at the time of surgery. A total of 78 hips in the 69 patients remain in situ; nine hips in eight patients died before 15 years, and 16 hips in 16 patients were revised. Clinical outcome scores and radiographs were available for 94 hips in 85 patients.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 767 - 771
1 Jun 2016
Konan S Duncan CP Masri BA Garbuz DS

Aims

Reconstruction of the acetabulum after failed total hip arthroplasty (THA) can be a surgical challenge in the presence of severe bone loss. We report the long-term survival of a porous tantalum revision acetabular component, its radiological appearance and quality of life outcomes.

Patients and Methods

We reviewed the results of 46 patients who had undergone revision of a failed acetabular component with a Paprosky II or III bone defect and reconstruction with a hemispherical, tantalum acetabular component, supplementary screws and a cemented polyethylene liner.


The Bone & Joint Journal
Vol. 98-B, Issue 2 | Pages 187 - 193
1 Feb 2016
Lash NJ Whitehouse MR Greidanus NV Garbuz DS Masri BA Duncan CP

Aims

We present a case series of ten metal-on-polyethylene total hip arthroplasties (MoP THAs) with delayed dislocation associated with unrecognised adverse local tissue reaction due to corrosion at the trunnion and pseudotumour formation.

Methods

The diagnosis was not suspected in nine of the ten patients (six female/four male; mean age 66 years), despite treatment in a specialist unit (mean time from index surgery to revision was 58 months, 36 to 84). It was identified at revision surgery and subsequently confirmed by histological examination of resected tissue. Pre-operative assessment and culture results ruled out infection. A variety of treatment strategies were used, including resection of the pseudotumour and efforts to avoid recurrent dislocation.