Nonagenarians (aged 90 to 99 years) have experienced the fastest percent decile population growth in the USA recently, with a consequent increase in the prevalence of nonagenarians living with joint arthroplasties. As such, the number of revision total hip arthroplasties (THAs) and total knee arthroplasties (TKAs) in nonagenarians is expected to increase. We aimed to determine the mortality rate, implant survivorship, and complications of nonagenarians undergoing aseptic revision THAs and revision TKAs. Our institutional total joint registry was used to identify 96 nonagenarians who underwent 97 aseptic revisions (78 hips and 19 knees) between 1997 and 2018. The most common indications were aseptic loosening and periprosthetic fracture for both revision THAs and revision TKAs. Mean age at revision was 92 years (90 to 98), mean BMI was 27 kg/m2 (16 to 47), and 67% (n = 65) were female. Mean time between primary and revision was 18 years (SD 9). Kaplan-Meier survival was used for patient mortality, and compared to age- and sex-matched control populations. Reoperation risk was assessed using cumulative incidence with death as a competing risk. Mean follow-up was five years.Aims
Methods
The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre. All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery.Aims
Methods
Currently, the US Center for Medicaid and Medicare Services (CMS) has been testing bundled payments for revision total joint arthroplasty (TJA) through the Bundled Payment for Care Improvement (BPCI) programme. Under the BPCI, bundled payments for revision TJAs are defined on the basis of diagnosis-related groups (DRGs). However, these DRG-based bundled payment models may not be adequate to account appropriately for the varying case-complexity seen in revision TJAs. The 2008-2014 Medicare 5% Standard Analytical Files (SAF5) were used to identify patients undergoing revision TJA under DRG codes 466, 467, or 468. Generalized linear regression models were built to assess the independent marginal cost-impact of patient, procedural, and geographic characteristics on 90-day costs.Aims
Methods
The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check. Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).Aims
Materials and Methods
Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis. At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p <
0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate. There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.
We followed 138 patients (145 hips) who had had uncemented total hip arthroplasty using the Taperloc femoral component for a mean of ten years (8 to 12.5). No patient was lost to follow-up; 31 (31 hips) died before the minimum time of eight years for inclusion in the study, and 30 of these still had their femoral component in place. One well-fixed prosthesis had been exchanged at the time of