The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.Aims
Methods
Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies. Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment.Aims
Methods
Instability is a common cause of failure after total hip arthroplasty. A novel reverse total hip has been developed, with a femoral cup and acetabular ball, creating enhanced mechanical stability. The purpose of this study was to assess the implant fixation using radiostereometric analysis (RSA), and the clinical safety and efficacy of this novel design. Patients with end-stage osteoarthritis were enrolled in a prospective cohort at a single centre. The cohort consisted of 11 females and 11 males with mean age of 70.6 years (SD 3.5) and BMI of 31.0 kg/m2 (SD 5.7). Implant fixation was evaluated using RSA as well as Western Ontario and McMaster Universities Osteoarthritis Index, Harris Hip Score, Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, 38-item Short Form survey, and EuroQol five-dimension health questionnaire scores at two-year follow-up. At least one acetabular screw was used in all cases. RSA markers were inserted into the innominate bone and proximal femur with imaging at six weeks (baseline) and six, 12, and 24 months. Independent-samples Aims
Methods
The aim of this study was to compare open reduction and internal fixation (ORIF) with revision surgery for the surgical management of Unified Classification System (UCS) type B periprosthetic femoral fractures around cemented polished taper-slip femoral components following primary total hip arthroplasty (THA). Data were collected for patients admitted to five UK centres. The primary outcome measure was the two-year reoperation rate. Secondary outcomes were time to surgery, transfusion requirements, critical care requirements, length of stay, two-year local complication rates, six-month systemic complication rates, and mortality rates. Comparisons were made by the form of treatment (ORIF vs revision) and UCS type (B1 vs B2/B3). Kaplan-Meier survival analysis was performed with two-year reoperation for any reason as the endpoint.Aims
Methods
A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.Aims
Methods
We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive. We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score.Aims
Methods
We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.Aims
Methods
To compare the in vivo long-term fixation achieved by two acetabular components with different porous ingrowth surfaces using radiostereometric analysis (RSA). This was a minimum ten-year follow-up of a prospective randomized trial of 62 hips with two different porous ingrowth acetabular components. RSA exams had previously been acquired through two years of follow-up. Patients returned for RSA examination at a minimum of ten years. In addition, radiological appearance of these acetabular components was analyzed, and patient-reported outcome measures (PROMs) obtained.Aims
Methods
To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation.Aims
Methods
Complex total hip arthroplasty (THA) with subtrochanteric shortening osteotomy is necessary in conditions other than developmental dysplasia of the hip (DDH) and septic arthritis sequelae with significant proximal femur migration. Our aim was to evaluate the hip centre restoration with THAs in these hips. In all, 27 THAs in 25 patients requiring THA with femoral shortening between 2012 and 2019 were assessed. Bilateral shortening was required in two patients. Subtrochanteric shortening was required in 14 out of 27 hips (51.9%) with aetiology other than DDH or septic arthritis. Vertical centre of rotation (VCOR), horizontal centre of rotation, offset, and functional outcome was calculated. The mean followup was 24.4 months (5 to 92 months).Aims
Methods
In the 1990s, a bioactive bone cement (BABC) containing apatite-wollastonite glass-ceramic (AW-GC) powder and bisphenol-a-glycidyl methacrylate resin was developed at our hospital. In 1996, we used BABC to fix the acetabular component in primary total hip arthroplasty (THA) in 20 patients as part of a clinical trial. The purpose of this study was to investigate the long-term results of primary THA using BABC. A total of 20 patients (three men and 17 women) with a mean age of 57.4 years (40 to 71), a mean body weight of 52.3 kg (39 to 64), and a mean body mass index (BMI) of 23.0 kg/m2 (19.8 to 28.6) were evaluated clinically and radiologically. Survival analyses were undertaken, and wear analyses were carried out using a computer-aided method.Aims
Patients and Methods
We analyzed the acetabular morphology of Crowe type IV hips using CT data to identify a landmark for the ideal placement of the centre of the acetabular component, as assessed by morphometric geometrical analysis, and its reliability. A total of 52 Crowe IV hips (42 patients; seven male, 35 female; mean age 68.5 years (32 to 82)) and 50 normal hips (50 patients; eight male, 42 female; mean age 60.7 years (34 to 86)) undergoing total hip arthroplasty were retrospectively identified. In this CT-based simulation study, the acetabular component was positioned at the true acetabulum with a radiological inclination of 40° and anteversion of 20°. Acetabular shape and the position of the centre of the acetabular component were analyzed by morphometric geometrical analysis using the generalized Procrustes analysis.Aims
Patients and Methods
Aims. Few reconstructive techniques are available for patients requiring
complex acetabular revisions such as those involving Paprosky type
2C, 3A and 3B deficiencies and pelvic discontinuity. Our aim was
to describe the development of the patient specific Triflange acetabular
component for use in these patients, the surgical technique and
mid-term results. We include a description of the pre-operative
CT scanning, the construction of a model, operative planning, and
surgical technique. All implants were coated with porous plasma
spray and hydroxyapatite if desired. Patients and Methods. A multicentre, retrospective review of 95 complex acetabular
reconstructions in 94 patients was performed. A total of 61 (64.2%)
were female. The mean age of the patients was 66 (38 to 85). The
mean body mass index was 29 kg/m. 2. (18 to 51). Outcome
was reported using the Harris Hip Score (HHS), complications, failures
and survival. Results. The mean follow-up was 3.5 years (1 to 11). The mean HHS improved
from 46 (15 to 90)
pre-operatively to 75 (14 to 100). A total of 21 hips (22%) had
at least one complication with some having more than one; including
dislocation (6%), infection (6%), and femoral complications (2%).
The implant was subsequently removed in five hips (5%), only one
for suspected aseptic loosening. Conclusion. The Triflange patient specific
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
Aims. To determine the effect of a change in design of a cementless
ceramic
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
Our aim was to report survivorship data and lessons learned with
the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Between August 2005 and March 2015, a total of 4802 primary cementless
Corail/Pinnacle THAs were performed in 4309 patients. In March 2016,
we reviewed these hips from a prospectively maintained database.Aims
Patients and Methods
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented
Data on early morbidity and complications after
revision total hip replacement (THR) are limited. The aim of this nationwide
study was to describe and quantify early morbidity after aseptic
revision THR and relate the morbidity to the extent of the revision
surgical procedure. We analysed all aseptic revision THRs from 1st
October 2009 to 30th September 2011 using the Danish National Patient
Registry, with additional information from the Danish Hip Arthroplasty
Registry. There were 1553 procedures (1490 patients) performed in
40 centres and we divided them into total revisions, acetabular
component revisions, femoral stem revisions and partial revisions.
The mean age of the patients was 70.4 years (25 to 98) and the median
hospital stay was five days (interquartile range 3 to 7). Within 90
days of surgery, the readmission rate was 18.3%, mortality rate
1.4%, re-operation rate 6.1%, dislocation rate 7.0% and infection
rate 3.0%. There were no differences in these outcomes between high-
and low-volume centres. Of all readmissions, 255 (63.9%) were due
to ‘surgical’ complications Cite this article: