Objective. This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Methods. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm. 3. and 0.45 g/cm. 3. ). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests. Results. There was no statistically significant difference in seating
force (p = 0.104) or primary stability (pull-out p = 0.171, lever-out
p = 0.087) of the two components in the low-density substrate. Similarly,
in the high-density substrate, there was no statistically significant
difference in the peak