Aims. The primary aim of this study was to determine the ten-year outcome following surgical treatment for femoroacetabular impingement (FAI). We assessed whether the evolution of practice from open to arthroscopic techniques influenced outcomes and tested whether any patient, radiological, or surgical factors were associated with outcome. Methods. Prospectively collected data of a consecutive single-surgeon cohort, operated for FAI between January 2005 and January 2015, were retrospectively studied. The cohort comprised 393 hips (365 patients; 71% male (n = 278)), with a mean age of 34.5 years (SD 10.0). Over the study period, techniques evolved from open surgical dislocation (n = 94) to a combined arthroscopy-Hueter technique (HA + Hueter; n = 61) to a pure arthroscopic technique (HA; n = 238). Outcome measures of interest included
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods