Advertisement for orthosearch.org.uk
Results 1 - 8 of 8
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1590 - 1595
1 Dec 2017
Atrey A Ward SE Khoshbin A Hussain N Bogoch E Schemitsch EH Waddell JP

Aims

We present the ten-year data of a cohort of patients, aged between 18 and 65 years (mean age 52.7 years; 19 to 64), who underwent total hip arthroplasty. Patients were randomised to be treated with a cobalt-chrome (CoCr) femoral head with an ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE) or ceramic-on-ceramic (CoC) bearing surface.

Patients and Methods

A total of 102 hips (91 patients) were randomised into the three groups. At ten years, 97 hips were available for radiological and functional follow-up. Two hips (two patients) had been revised (one with deep infection and one for periprosthetic fracture) and three were lost to follow-up. Radiological analysis was performed using a validated digital assessment programme to give linear, directional and volumetric wear of the two polyethylene groups.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1573 - 1581
1 Dec 2016
Lewis PM Waddell JP

Fractures of the hip are common, often occurring in frail elderly patients, but also in younger fit healthy patients following trauma. They have a significant associated mortality and major social and financial implications to patients and health care providers. Many guidelines are available for the management of these patients, mostly recommending early surgery for the best outcomes. As a result, healthcare authorities now put pressure on surgical teams to ‘fast track’ patients with a fracture of the hip, often misquoting the available literature, which in itself can be confusing and even conflicting.

This paper has been written following an extensive review of the available literature. An attempt is made to clarify what is meant by early surgery (expeditious versus emergency), and we conclude with a personal view for the practical management of these patients of variable age, fitness and type of surgery performed within services that are often under considerable pressure of finance and available operating theatres and qualified staff.

Cite this article: Bone Joint J 2016;98-B:1573–81.


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1175 - 1179
1 Sep 2016
Olsen M Lewis PM Morrison Z McKee MD Waddell JP Schemitsch EH

Aims

One method of femoral head preservation following avascular necrosis (AVN) is core decompression and insertion of a tantalum rod. However, there may be a high failure rate associated with this procedure. The purpose of this study was to document the clinical and radiological outcomes following total hip arthroplasty (THA) subsequent to failed tantalum rod insertion.

Patients and Methods

A total of 37 failed tantalum rods requiring total hip arthroplasty were identified from a prospective database. There were 21 hips in 21 patients (12 men and nine women, mean age 37 years, 18 to 53) meeting minimum two year clinical and radiographic follow-up whose THAs were carried out between November 2002 and April 2013 (mean time between tantalum rod implantation and conversion to a THA was 26 months, 6 to 72). These were matched by age and gender to individuals (12 men, nine women, mean age 40 years, 18 to 58) receiving THA for AVN without prior tantalum rod insertion.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 459 - 465
1 Apr 2012
Nikolaou VS Edwards MR Bogoch E Schemitsch EH Waddell JP

The ideal bearing surface for young patients undergoing total hip replacement (THR) remains controversial. We report the five-year results of a randomised controlled trial comparing the clinical and radiological outcomes of 102 THRs in 91 patients who were <  65 years of age. These patients were randomised to receive a cobalt–chrome on ultra-high-molecular-weight polyethylene, cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic bearing. In all, 97 hip replacements in 87 patients were available for review at five years. Two hips had been revised, one for infection and one for peri-prosthetic fracture.

At the final follow-up there were no significant differences between the groups for the mean Western Ontario and McMaster Universities osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness, p = 0.99), Short Form-12 (physical component, p = 0.878; mental component, p = 0.818) or Harris hip scores (p = 0.22). Radiological outcomes revealed no significant wear in the ceramic group.

Comparison of standard and highly cross-linked polyethylene, however, revealed an almost threefold difference in the mean annual linear wear rates (0.151 mm/year versus 0.059 mm/year, respectively) (p < 0.001).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 310 - 315
1 Mar 2009
Olsen M Davis ET Waddell JP Schemitsch EH

We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip.

The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° (sd 2.6). The stem-shaft angle after operation differed from that planned by a mean of 2.8° (sd 2.0) and in 86% of cases the final angle measured within ± 5° of that planned. We had no instances of notching of the neck or varus alignment of the implant in our series. A learning curve was observed in the time taken for navigation, but not for accurate placement of the implant.

Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1522 - 1527
1 Nov 2008
Davis ET Olsen M Zdero R Waddell JP Schemitsch EH

A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in load to failure. Placing the component in 10° of valgus had no appreciable effect on increasing the load to failure except in the composite femurs with varus native femoral necks. Specimens in 10° of varus were significantly weaker than the neutrally-aligned specimens.

The results suggest that retention of the intact proximal femoral strength occurs at an implant angulation of ≥ 142°. However, the benefit of extreme valgus alignment may be outweighed in clinical practice by the risk of superior femoral neck notching, which was avoided in this study.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 557 - 560
1 Apr 2007
Davis ET Gallie P Macgroarty K Waddell JP Schemitsch E

A cadaver study using six pairs of lower limbs was conducted to investigate the accuracy of computer navigation and standard instrumentation for the placement of the Birmingham Hip Resurfacing femoral component. The aim was to place all the femoral components with a stem-shaft angle of 135°.

The mean stem-shaft angle obtained in the standard instrumentation group was 127.7° (120° to 132°), compared with 133.3° (131° to 139°) in the computer navigation group (p = 0.03). The scatter obtained with computer-assisted navigation was approximately half that found using the conventional jig.

Computer navigation was more accurate and more consistent in its placement of the femoral component than standard instrumentation. We suggest that image-free computer-assisted navigation may have an application in aligning the femoral component during hip resurfacing.


The Journal of Bone & Joint Surgery British Volume
Vol. 53-B, Issue 3 | Pages 392 - 405
1 Aug 1971
Schatzker J Rorabeck CH Waddell JP

1. Thirty-seven cases of fracture of the dens have been studied.

2. The incidence of non-union was high: 64 per cent after apparently adequate closed treatment.

3. Possible causes of the high incidence of non-union have been studied : attention is drawn to the effect of displacement and to that of posterior displacement in particular.

4. Non-union of the dens with potential instability at the atlanto-axial joint is not acceptable in a patient who expects to lead a normal active life.

5. Atlanto-axial fusion is the method of choice in the treatment of instability ; once that has been secured, pseudarthrosis of the dens is no longer significant.