We scanned 25 left knees in healthy human subjects
using MRI. Multiplanar reconstruction software was used to take
measurements of the inferior and posterior facets of the femoral
condyles and the trochlea. A ‘basic circle’ can be defined which, in the sagittal plane,
fits the posterior and inferior facets of the lateral condyle, the
posterior facet of the medial condyle and the floor of the groove
of the trochlea. It also approximately fits both condyles in the
coronal plane (inferior facets) and the axial plane (posterior facets).
The circle fitting the inferior facet of the medial condyle in the
sagittal plane was consistently 35% larger than the other circles
and was termed the ‘medial inferior circle’. There were strong correlations
between the radii of the circles, the relative positions of the
centres of the condyles, the width of the condyles, the total knee
width and skeletal measurements including height. There was poor
correlation between the radii of the circles and the position of
the trochlea relative to the condyles. In summary, the condyles are approximately spherical except for
the inferior facet medially, which has a larger radius in the sagittal
plane. The size and position of the condyles are consistent and
change with the size of the person. However, the position of the
trochlea is variable even though its radius is similar to that of
the condyles. This information has implications for understanding
anterior knee pain and for the design of knee replacements. Cite this article:
Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements. We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year. In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.
Retrieval studies have shown that the use of fully congruent meniscal bearings reduces wear in knee replacements. We report the outcome of 143 knees with anteromedial osteoarthritis and normal anterior cruciate ligaments treated by unicompartmental arthroplasty using fully congruous mobile polyethylene bearings. At review, 34 knees were in patients who had died and 109 were in those who were still living. The mean elapsed time since operation was 7.6 years (maximum 13.8). We established the status of all but one knee. There had been five revision operations giving a cumulative prosthetic survival rate at ten years (33 knees at risk) of 98% (95% CI 93% to 100%). Considering the knee lost to follow-up as a failure, the ‘worst-case’ survival rate was 97%. No failures were due to polyethylene wear or aseptic loosening of the tibial component. One bearing which dislocated at four years was reduced by closed manipulation. The ten-year survival rate is the best of those reported for unicompartmental arthroplasty and not significantly different from the best rates for total knee replacement.