The aim of this study was to evaluate the effects of using a
portable, accelerometer-based surgical navigation system (KneeAlign2)
in total knee arthroplasty (TKA) on the alignment of the femoral
component, and blood loss. A total of 241 consecutive patients with primary osteoarthritis
of the knee were enrolled in this prospective, randomised controlled
study. There were 207 women and 34 men. The mean age of the patients
was 74.0 years (57 to 89). The KneeAlign2 system was used for distal
femoral resection in 121 patients (KA2 group) and a conventional intramedullary
femoral guide was used in 120 patients (IM group).Aims
Patients and Methods
The shape of the flexion gap in 20 normal knees was evaluated by axial radiography of the distal femur, and the results compared with those obtained in a previous study by MRI. The observed asymmetry was reduced by 29% using radiography, with a mean value of 3.6° (1.5° to 6.3°) compared with that obtained by MRI of 5.1° (2.6° to 9.5°), a mean discrepancy of 1.49°. The results obtained by radiography and MRI showed a strong correlation (r = 0.78). Axial radiography is acceptable for the evaluation of the flexion gap and is less expensive and more comfortable to perform than MRI. Additionally, no metallic artefact occurs when the radiological method is used for assessment after arthroplasty.
Varus and valgus joint laxity of the normal living knee in flexion was assessed using MRI. Twenty knees were flexed to 90° and were imaged in neutral and under a varus-valgus stress in an open MRI system. The configuration of the tibiofemoral joint gap was studied in slices which crossed the epicondyles of the femur. When a varus stress was applied, the lateral joint gap opened by 6.7 ± 1.9 mm (mean ±