Revision surgery of the hip was performed on 114 hips using an extensively porous-coated femoral component. Of these, 95 hips (94 patients) had a mean follow-up of 10.2 years (5 to 17). No cortical struts were used and the cortical index and the femoral cortical width were measured at different levels. There were two revisions for aseptic loosening. Survivorship at 12 years for all causes of failure was 96.9% (95% confidence interval 93.5 to 100) in the best-case scenario. Fibrous or unstable fixation was associated with major bone defects. The cortical index (p = 0.045) and the lateral cortical thickness (p = 0.008) decreased at the proximal level over time while the medial cortex increased (p = 0.001) at the proximal and distal levels. An increase in the proximal medial cortex was found in patients with an extended transtrochanteric osteotomy (p = 0.026) and in those with components shorter than 25 cm (p = 0.008). The use of the extensively porous-coated femoral component can provide a solution for difficult cases in revision surgery. Radiological bony ingrowth is common. Although without clinical relevance at the end of follow-up, the thickness of the medial femoral cortex often increased while that of the lateral cortex decreased. In cases in which a shorter component was used and in those undertaken using an extended trochanteric osteotomy, there was a greater increase in thickness of the femoral cortex over time.
Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal
We enrolled 98 patients (107 hips) with a mean age of 47 years (SD 8.6) into a prospective study of the Madreporic Lord THR; 34 hips had primary and 73 secondary osteoarthritis. After ten years, the survival rate using revision as the endpoint for failure was 70% (±9) for the cup and 98% (±0.3) for the stem. The combined clinical and radiological survival rates were 46% (±11) and 81% (±10), respectively. Osteoporosis due to
We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.Aims
Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods
The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique. Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.Aims
Patients and Methods
The aim of this study was to determine the stability of a new
short femoral stem compared with a conventional femoral stem in
patients undergoing cementless total hip arthroplasty (THA), in
a prospective randomized controlled trial using radiostereometric
analysis (RSA). A total of 53 patients were randomized to receive cementless
THA with either a short femoral stem (MiniHip, 26 patients, mean
age: 52 years, nine male) or a conventional length femoral stem
(MetaFix, 23 patients, mean age: 53 years, 11 male). All patients
received the same cementless acetabular component. Two-year follow-up
was available on 38 patients. Stability was assessed through migration
and dynamically inducible micromotion. Radiographs for RSA were
taken postoperatively and at three, six, 12, 18, and 24 months.Aims
Patients and Methods
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
We report on the outcome of the Synergy cementless femoral stem
with a minimum follow-up of 15 years (15 to 17). A retrospective review was undertaken of a consecutive series
of 112 routine primary cementless total hip arthroplasties (THAs)
in 102 patients (112 hips). There were 60 female and 42 male patients
with a mean age of 61 years (18 to 82) at the time of surgery. A
total of 78 hips in the 69 patients remain Aims
Patients and Methods
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of failure and modes of failure. We subsequently investigated
the factors that may affect the potential for the restoration of bone
stock and the durability of the re-revision reconstruction using
multivariate analysis. At the time of re-revision, there were ten host acetabula with
no significant defects, 14 with contained defects, nine with minor-column,
seven with major-column defects and ten with pelvic discontinuity.
When bone defects at re-revision were compared with those at the
previous revision, there was restoration of bone stock in 31 hips, deterioration
of bone stock in nine and remained unchanged in ten. This was a
significant improvement (p <
0.001). Morselised allografting
at the index revision was not associated with the restoration of
bone stock. In 17 hips (34%), re-revision was possible using a simple acetabular
component without allograft, augments, rings or cages. There were
47 patients with a mean follow-up of 70 months (6 to 146) available
for survival analysis. Within this group, the successful cases had
a minimum follow-up of two years after re-revision. There were 22 clinical
or radiological failures (46.7%), 18 of which were due to aseptic
loosening. The five and ten year Kaplan–Meier survival rate was
75% (95% CI, 60 to 86) and 56% (95% CI, 40 to 70) respectively with
aseptic loosening as the endpoint. The rate of aseptic loosening
was higher for hips with pelvic discontinuity (p = 0.049) and less
when the allograft had been in place for longer periods (p = 0.040). The use of a cage or ring over structural allograft bone for
massive uncontained defects in acetabular revision can restore host
bone stock and facilitate subsequent re-revision surgery to a certain
extent. Cite this article:
Bone preservation and physiological distribution of forces on the proximal femur are key elements in introducing a successful uncemented total hip replacement. In order to achieve this, in the mid 1990s, we developed an ultra short proximal loading custom-made component with a lateral flare, a high femoral neck osteotomy and without a diaphyseal stem. We report the outcome of 129 custom-made hydroxyapatite-coated uncemented short femoral components inserted into 109 patients between June 1995 and May 2004. The mean age of the patients was 51 years (21 to 71) and the mean follow-up was eight years (4.9 to 14.1). Bone behaviour around the implant was studied on the post-operative radiographs. The mean Harris hip score improved from 44 (8 to 66) pre-operatively to 95 (76 to 100) at final follow-up. The Western Ontario MacMaster University Osteoarthritis index was 93 of 100 at final review. None of the patients reported thigh pain. A total of five hips were revised, three for polyethylene liner exchange and two for complete revision of the acetabular component. No femoral components were revised. The radiological changes in the proximal femur were generally good, as evidenced by spot welds both on the medial and lateral aspects of the femur. No component migrated. The presence of a lateral flare and use of a high osteotomy of the femoral neck provided good clinical and radiological results. The absence of a diaphyseal portion of the stem did not impair stability.
Revision total hip arthroplasty (THA) is challenging
when there is severe loss of bone in the proximal femur. The purpose
of this study was to evaluate the clinical and radiographic outcomes
of revision THA in patients with severe proximal femoral bone loss
treated with a fluted, tapered, modular femoral component. Between
January 1998 and December 2004, 92 revision THAs were performed
in 92 patients using a single fluted, tapered, modular femoral stem
design. Pre-operative diagnoses included aseptic loosening, infection
and peri-prosthetic fracture. Bone loss was categorised pre-operatively
as Paprosky types III-IV, or Vancouver B3 in patients with a peri-prosthetic
fracture. The mean clinical follow-up was 6.4 years (2 to 12). A
total of 47 patients had peri-operative complications, 27 of whom
required further surgery. However, most of these further operations
involved retention of a well-fixed femoral stem, and 88/92 femoral
components (97%) remained Revision THA in patients with extensive proximal femoral bone
loss using the Link MP fluted, tapered, modular stem led to a high
rate of osseointegration of the stem at mid-term follow-up. Cite this article:
We undertook a randomised prospective follow-up study of changes in peri-prosthetic bone mineral density (BMD) after hip resurfacing and compared them with the results after total hip replacement. A total of 59 patients were allocated to receive a hip resurfacing (n = 29) or an uncemented distally fixed total hip replacement (n = 30). The BMD was prospectively determined in four separate regions of interest of the femoral neck and in the calcar region corresponding to Gruen zone 7 for the hip resurfacing group and compared only to the calcar region in the total hip replacement group. Standardised measurements were performed pre-operatively and after three, six and 12 months. The groups were well matched in terms of gender distribution and mean age. The mean BMD in the calcar region increased after one year to 105.2% of baseline levels in the resurfaced group compared with a significant decrease to 82.1% in the total hip replacement group (p <
0.001) by 12 months. For the resurfaced group, there was a decrease in bone density in all four regions of the femoral neck at three months which did not reach statistical significance and was followed by recovery to baseline levels after 12 months. Hip resurfacing did indeed preserve BMD in the inferior femoral neck. In contrast, a decrease in the mean BMD in Gruen zone 7 followed uncemented distally fixed total hip replacement. Long term follow-up studies are necessary to see whether this benefit in preservation of BMD will be clinically relevant at future revision surgery.
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84). Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the trunnion after fracture of a modular ceramic head, and in another two, removal of the femoral component was because of the preference of the surgeon. In all cases the femoral component was well fixed, but could be extracted at the time of acetabular revision. In one case both components were revised for deep infection. There has been one case of aseptic loosening of the stem which occurred at 14 years. This stem had migrated distally by 7.6 mm in ten years and 8.4 mm at the time of revision at which stage it was found to be rotationally loose. With hindsight this component had been undersized at implantation. The survivorship for the stem at 17 years with aseptic loosening as the endpoint was 98.6% (95% confidence interval 95.9 to 100) when 62 hips were at risk. All remaining stems had a satisfactory clinical and radiological outcome. The Freeman proximally hydroxyapatite-coated femoral component is therefore a dependable implant and its continued use can be recommended.
Cementless femoral stems are currently preferred
for total hip replacement (THR) in the United States. Improvements
in stem design, instrumentation and surgical technique have made
this technology highly successful, reproducible, and applicable
to the vast majority of patients requiring a THR. However, there
are ongoing developments in some aspects of stem design that influence
clinical results, the incidence of complications and their inherent adaptability
in accommodating the needs of individual patients. Here we examine
some of these design features. Cite this article:
This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading. A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12). The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur. This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.
Trabecular metal (TM) augments are a relatively
new option for reconstructing segmental bone loss during acetabular
revision. We studied 34 failed hip replacements in 34 patients that
were revised between October 2003 and March 2010 using a TM acetabular
shell and one or two augments. The mean age of the patients at the
time of surgery was 69.3 years (46 to 86) and the mean follow-up
was 64.5 months (27 to 107). In all, 18 patients had a minor column
defect, 14 had a major column defect, and two were associated with
pelvic discontinuity. The hip centre of rotation was restored in
27 patients (79.4%). The Oxford hip score increased from a mean
of 15.4 points (6 to 25) before revision to a mean of 37.7 (29 to
47) at the final follow-up. There were three aseptic loosenings
of the construct, two of them in the patients with pelvic discontinuity.
One septic loosening also occurred in a patient who had previously
had an infected hip replacement. The augments remained stable in
two of the failed hips. Whenever there was a loose acetabular component
in contact with a stable augment, progressive metal debris shedding
was evident on the serial radiographs. Complications included another
deep infection treated without revision surgery. Good clinical and
radiological results can be expected for bone-deficient acetabula
treated by a TM cup and augment, but for pelvic discontinuities
this might not be a reliable option. Cite this article:
Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements. We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant. At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 ( The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.
In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
We report the use of an allograft prosthetic composite for reconstruction of the skeletal defect in complex revision total hip replacement for severe proximal femoral bone loss. Between 1986 and 1999, 72 patients (20 men, 52 women) with a mean age of 59.9 years (38 to 78) underwent reconstruction using this technique. At a mean follow-up of 12 years (8 to 20) 57 patients were alive, 14 had died and one was lost to follow-up. Further revision was performed in 19 hips at a mean of 44.5 months (11 to 153) post-operatively. Causes of failure were aseptic loosening in four, allograft resorption in three, allograft nonunion in two, allograft fracture in four, fracture of the stem in one, and deep infection in five. The survivorship of the allograft-prosthesis composite at ten years was 69.0% (95% confidence interval 67.7 to 70.3) with 26 patients remaining at risk. Survivorship was statistically significantly affected by the severity of the pre-operative bone loss (Paprosky type IV; p = 0.019), the number of previous hip revisions exceeding two (p = 0.047), and the length of the allograft used (p = 0.005).