We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone
Injury to the lateral femoral cutaneous nerve (LFCN) is one of the known complications after periacetabular osteotomy (PAO) performed using the anterior approach, reported to occur in between 1.5% and 65% of cases. In this study, we performed a prospective study on the incidence of LFCN injury as well as its clinical outcomes based on the Harris Hip Score (HHS), Short-Form 36 Health Survey (SF-36), and Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ). The study included 42 consecutive hips in 42 patients (three male and 39 female) who underwent PAO from May 2016 to July 2018. We prospectively evaluated the incidence of LFCN injury at ten days, three months, six months, and one year postoperatively. We also evaluated the clinical scores, including the HHS, SF-36, and JHEQ scores, at one year postoperatively.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.Aims
Methods
This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an uncemented component in acetabular revisions performed in a standardized manner since 1993. Between 1993 and 2013, 370 patients with a median age of 72 years (interquartile range (IQR) 63 to 79 years) underwent acetabular revision surgery. Of these, 229 were more than ten years following surgery and 137 were more than 15 years. All revisions were performed with extensive use of morcellized allograft firmly impacted into the entire acetabular cavity, followed by insertion of an uncemented component with supplementary screw fixation. All types of reoperation were captured using review of radiographs and medical charts, combined with data from the local surgical register and the Swedish Hip Arthroplasty Register.Aims
Methods
The aim of this study was to report the long-term results of rotational acetabular osteotomy (RAO) for symptomatic hip dysplasia in patients aged younger than 21 years at the time of surgery. We evaluated 31 patients (37 hips) aged younger than 21 years at the time of surgery retrospectively. There were 29 female and two male patients. Their mean age at the time of surgery was 17.4 years (12 to 21). The mean follow-up was 17.9 years (7 to 30). The RAO was combined with a varus or valgus femoral osteotomy or a greater trochanteric displacement in eight hips, as instability or congruence of the hip could not be corrected adequately using RAO alone.Aims
Patients and Methods
Femoral impaction bone grafting was first developed in 1987 using
morselised cancellous bone graft impacted into the femoral canal
in combination with a cemented, tapered, polished stem. We describe
the evolution of this technique and instrumentation since that time. Between 1987 and 2005, 705 revision total hip arthroplasties
(56 bilateral) were performed with femoral impaction grafting using
a cemented femoral stem. All surviving patients were prospectively
followed for a mean of 14.7 years (9.8 to 28.3) with no loss to
follow-up. By the time of the final review, 404 patients had died.Aims
Patients and Methods
The repair of chondral lesions associated with
femoroacetabular impingement requires specific treatment in addition
to that of the impingement. In this single-centre retrospective
analysis of a consecutive series of patients we compared treatment
with microfracture (MFx) with a technique of enhanced microfracture
autologous matrix-induced chondrogenesis (AMIC). Acetabular grade III and IV chondral lesions measuring between
2 cm2 and 8 cm2 in 147 patients were treated
by MFx in 77 and AMIC in 70. The outcome was assessed using the
modified Harris hip score at six months and one, two, three, four
and five years post-operatively. The outcome in both groups was
significantly improved at six months and one year post-operatively.
During the subsequent four years the outcome in the MFx group slowly deteriorated,
whereas that in the AMIC group remained stable. Six patients in
the MFx group subsequently required total hip arthroplasty, compared
with none in the AMIC group We conclude that the short-term clinical outcome improves in
patients with acetabular chondral damage following both MFx and
AMIC. However, the AMIC group had better and more durable improvement,
particularly in patients with large (≥ 4 cm2) lesions. Cite this article:
Until now, there has been no consensus as to whether stemmed
acetabular components are appropriate for use in patients undergoing
revision total hip arthroplasty (THA) who have major acetabular
defects or pelvic discontinuity. We wished to address this deficiency
in the literature. We carried out a retrospective study of 35 patients (six men
and 29 women) with a mean age of 68 years (37 to 87), with major
acetabular defects who underwent revision THA between 2000 and 2012.Aim
Patients and Methods
Sciatic nerve palsy following total hip arthroplasty
(THA) is a relatively rare yet potentially devastating complication.
The purpose of this case series was to report the results of patients
with a sciatic nerve palsy who presented between 2000 and 2010,
following primary and revision THA and were treated with neurolysis.
A retrospective review was made of 12 patients (eight women and
four men), with sciatic nerve palsy following THA. The mean age
of the patients was 62.7 years (50 to 72; standard deviation 6.9).
They underwent interfascicular neurolysis for sciatic nerve palsy,
after failing a trial of non-operative treatment for a minimum of
six months. Following surgery, a statistically and clinically significant
improvement in motor function was seen in all patients. The mean
peroneal nerve score function improved from 0.42 (0 to 3) to 3 (1
to 5) (p <
0.001). The mean tibial nerve motor function score
improved from 1.75 (1 to 4) to 3.92 (3 to 5) (p = 0.02).The mean
improvement in sensory function was a clinically negligible 1 out
of 5 in all patients. In total, 11 patients reported improvement
in their pain following surgery. We conclude that neurolysis of the sciatic nerve has a favourable
prognosis in patients with a sciatic nerve palsy following THA.
Our findings suggest that surgery should not be delayed for >
12
months following injury. Cite this article:
We identified 1305 femoral impaction bone grafting revisions using the Exeter stem performed between 1989 and 2002 in 30 hospitals throughout Sweden. There were 1188 patients with a mean age of 71 years (29 to 94) followed up for between five and 18 years. The participating departments reported 70 further revisions in total, of which 57 could also be identified on the Swedish National Arthroplasty Registry. Kaplan-Meier survivorship for all causes of failure was 94.0% (95% confidence interval (CI) 92 to 96) for women and 94.7% (95% CI, 92 to 96) for men at 15 years. Survivorship at 15 years for aseptic loosening was 99.1% (95% CI 98.4 to 99.5), for infection 98.6% (95% CI 97.6 to 99.2), for subsidence 99.0% (95% CI 98.2 to 99.4) and for fracture 98.7% (95% CI 97.9 to 99.2) Statistically significant predictors of failure were the year in which revision was conducted (p <
0.001). The number of previous revisions was slightly above the level of signifance (p = 0.056). Age, gender, the length of the stem and previous septic loosening were not predictors of failure (p = 0.213, p = 0.399, p = 0.337, p = 0.687, respectively). The difference in survivorship between high- and low-volume departments was only 3% at ten years. We conclude that impaction bone grafting with the Exeter stem has an excellent long-term survivorship following revision arthroplasty. The technique of impaction grafting appears to be reliable, can be learned rapidly and produces a predictably low incidence of aseptic loosening.
We undertook a randomised controlled trial to
compare the piriformis-sparing approach with the standard posterior approach
used for total hip replacement (THR). We recruited 100 patients
awaiting THR and randomly allocated them to either the piriformis-sparing
approach or the standard posterior approach. Pre- and post-operative
care programmes and rehabilitation regimes were identical for both
groups. Observers were blinded to the allocation throughout; patients
were blinded until the two-week assessment. Follow-up was at six
weeks, three months, one year and two years. In all 11 patients
died or were lost to follow-up. There was no significant difference between groups for any of
the functional outcomes. However, for patients in the piriformis-sparing
group there was a trend towards a better six-minute walk test at
two weeks and greater patient satisfaction at six weeks. The acetabular
components were less anteverted (p = 0.005) and had a lower mean
inclination angle (p = 0.02) in the piriformis-sparing group. However,
in both groups the mean component positions were within Lewinnek’s
safe zone. Surgeons perceived the piriformis-sparing approach to
be significantly more difficult than the standard approach (p =
0.03), particularly in obese patients. In conclusion, performing THR through a shorter incision involving
sparing piriformis is more difficult and only provides short-term
benefits compared with the standard posterior approach.
We prospectively evaluated the long-term outcome of 158 consecutive patients who underwent revision total hip replacement using uncemented computer-assisted design-computer-assisted manufacture femoral components. There were 97 men and 61 women. Their mean age was 63.1 years (34.6 to 85.9). The mean follow-up was 10.8 years (10 to 12). The mean Oxford, Harris and Western Ontario and McMaster hip scores improved from 41.1, 44.2 and 52.4 pre-operatively to 18.2, 89.3 and 12.3, respectively (p <
0.0001, for each). Six patients required further surgery. The overall survival of the femoral component was 97% (95% confidence interval 94.5 to 99.7). These results are comparable to those of previously published reports for revision total hip replacement using either cemented or uncemented components.
Revision surgery of the hip was performed on 114 hips using an extensively porous-coated femoral component. Of these, 95 hips (94 patients) had a mean follow-up of 10.2 years (5 to 17). No cortical struts were used and the cortical index and the femoral cortical width were measured at different levels. There were two revisions for aseptic loosening. Survivorship at 12 years for all causes of failure was 96.9% (95% confidence interval 93.5 to 100) in the best-case scenario. Fibrous or unstable fixation was associated with major bone defects. The cortical index (p = 0.045) and the lateral cortical thickness (p = 0.008) decreased at the proximal level over time while the medial cortex increased (p = 0.001) at the proximal and distal levels. An increase in the proximal medial cortex was found in patients with an extended transtrochanteric osteotomy (p = 0.026) and in those with components shorter than 25 cm (p = 0.008). The use of the extensively porous-coated femoral component can provide a solution for difficult cases in revision surgery. Radiological bony ingrowth is common. Although without clinical relevance at the end of follow-up, the thickness of the medial femoral cortex often increased while that of the lateral cortex decreased. In cases in which a shorter component was used and in those undertaken using an extended trochanteric osteotomy, there was a greater increase in thickness of the femoral cortex over time.
We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30. There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.
We reviewed the results of 71 revisions of the acetabular component in total hip replacement, using impaction of bone allograft. The mean follow-up was 7.2 years (1.6 to 9.7). All patients were assessed according to the American Academy of Orthopedic Surgeons (AAOS) classification of bone loss, the amount of bone graft required, thickness of the graft layer, signs of graft incorporation and use of augmentation. A total of 20 acetabular components required re-revision for aseptic loosening, giving an overall survival of 72% (95% CI, 54.4 to 80.5). Of these failures, 14 (70%) had an AAOS type III or IV bone defect. In the failed group, poor radiological and histological graft incorporation was seen. These results suggest that impaction allografting in acetabular revision with severe bone defects may have poorer results than have previously been reported.
We retrospectively reviewed, ten years after surgery, 100 consecutive total hip replacements in which the Duraloc 300 cup had been used. Post-operative radiographs were analysed for placement of the cup and interface gaps and follow-up radiographs for lucent lines, osteolysis, wear and migration. All the components were found to be stable with no evidence of loosening. The mean rate of wear was 0.12 mm/year. Three hips developed acetabular osteolysis at the level of the apex hole. Two have successfully undergone bone grafting without removal of the implants and one patient is awaiting surgery. The Duraloc 300 cup has a survival of 100% at ten years with no aseptic loosening and a low incidence of pelvic osteolysis.
We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups. Uncemented, press-fit acetabular components were used for both. No significant difference was found in the mean diameter of acetabular implant inserted in the groups (54.74 mm for total hip replacement and 54.90 mm for resurfacing arthroplasty). In seven resurfacing procedures (6.8%), the surgeon used a larger size of component in order to match the corresponding diameter of the femoral component. With resurfacing arthroplasty, conservation of bone is clearly advantageous on the femoral side. Our study has shown that, with a specific design of acetabular implant and by following a careful surgical technique, removal of bone on the acetabular side is comparable with that of total hip replacement.