Varus and valgus joint laxity of the normal living knee in flexion was assessed using MRI. Twenty knees were flexed to 90° and were imaged in neutral and under a varus-valgus stress in an open MRI system. The configuration of the tibiofemoral joint gap was studied in slices which crossed the epicondyles of the femur. When a varus stress was applied, the lateral joint gap opened by 6.7 ± 1.9 mm (mean ± . sd. ; 2.1 to 9.2) whereas the medial joint gap opened by only by a mean of 2.1 ± 1.1 mm (0.2 to 4.2). These discrepancies indicate that the tibiofemoral flexion gap in the normal knee is not rectangular and that the lateral joint gap is significantly lax. These results may be useful for adequate soft-tissue balancing and bone resection in total knee arthroplasty and
This retrospective study evaluated the midterm clinical and radiographic outcomes of a second-generation total knee replacement system. In a multicentre consecutive series of 1512 patients, 1970 knees were treated with the PFC Sigma knee system (Depuy, Warsaw, Indiana). The patients were reviewed for functional outcome, and underwent independent radiographic evaluation at a mean follow-up of 7.3 years (5 to 10). A total of 40 knees (2%) required revision, 17 (0.9%) for infection. The incidence of osteolysis was 2.2%. The ten-year survival with revision for any cause other than infection as the endpoint was 97.2% (95% CI 95.4 to 99.1). The PFC Sigma knee system appears to provide excellent results in the medium term.
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.