We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.Aims
Methods
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
In skeletally immature patients, resection of
bone tumours and reconstruction of the lower limb often results
in leg-length discrepancy. The Stanmore non-invasive extendible
endoprosthesis, which uses electromagnetic induction, allows post-operative
lengthening without anaesthesia. Between 2002 and 2009, 55 children
with a mean age of 11.4 years (5 to 16) underwent reconstruction
with this prosthesis; ten patients (18.2%) died of disseminated
disease and one child underwent amputation due to infection. We
reviewed 44 patients after a mean follow-up of 41.2 months (22 to
104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to
30) and the Toronto Extremity Salvage score was 92.3% (55.2% to
99.0%). There was no local recurrence of tumour. Complications developed
in 16 patients (29.1%) and ten (18.2%) underwent revision. The mean length gained per patient was 38.6 mm (3.5 to 161.5),
requiring a mean of 11.3 extensions (1 to 40), and ten component
exchanges were performed in nine patients (16.4%) after attaining
the maximum lengthening capacity of the implant. There were 11 patients
(20%) who were skeletally mature at follow-up, ten of whom had equal
leg lengths and nine had a full range of movement of the hip and
knee. This is the largest reported series using non-invasive extendible
endoprostheses after excision of primary bone tumours in skeletally
immature patients. The technique produces a good functional outcome,
with prevention of limb-length discrepancy at skeletal maturity.
Endoprosthetic replacement of the distal tibia and ankle joint for a primary bone tumour is a rarely attempted and technically challenging procedure. We report the outcome of six patients treated between 1981 and 2007. There were four males and two females, with a mean age of 43.5 years (15 to 75), and a mean follow-up of 9.6 years (1 to 27). No patient developed a local recurrence or metastasis. Two of the six went on to have a below-knee amputation for persistent infection after a mean 16 months (1 to 31). The four patients who retained their endoprosthesis had a mean musculoskeletal tumour society score of 70% and a mean Toronto extremity salvage score of 71%. All were pain free and able to perform most activities of daily living in comfort. A custom-made endoprosthetic replacement of the distal tibia and ankle joint is a viable treatment option for carefully selected patients with a primary bone tumour. Patients should, however, be informed of the risk of infection and the potential need for amputation if this cannot be controlled.