Advertisement for orthosearch.org.uk
Results 1 - 20 of 32
Results per page:
The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 485 - 492
1 Apr 2018
Gauci MO Bonnevialle N Moineau G Baba M Walch G Boileau P

Aims. Controversy about the use of an anatomical total shoulder arthroplasty (aTSA) in young arthritic patients relates to which is the ideal form of fixation for the glenoid component: cemented or cementless. This study aimed to evaluate implant survival of aTSA when used in patients aged < 60 years with primary glenohumeral osteoarthritis (OA), and to compare the survival of cemented all-polyethylene and cementless metal-backed glenoid components. Materials and Methods. A total of 69 consecutive aTSAs were performed in 67 patients aged < 60 years with primary glenohumeral OA. Their mean age at the time of surgery was 54 years (35 to 60). Of these aTSAs, 46 were undertaken using a cemented polyethylene component and 23 were undertaken using a cementless metal-backed component. The age, gender, preoperative function, mobility, premorbid glenoid erosion, and length of follow-up were comparable in the two groups. The patients were reviewed clinically and radiographically at a mean of 10.3 years (5 to 12, . sd. 26) postoperatively. Kaplan–Meier survivorship analysis was performed with revision as the endpoint. Results. A total of 26 shoulders (38%) underwent revision surgery: ten (22%) in the polyethylene group and 16 (70%) in the metal-backed group (p < 0.0001). At 12 years’ follow-up, the rate of implant survival was 74% (. sd.  0.09) for polyethylene components and 24% (. sd.  0.10) for metal-backed components (p < 0.0002). Glenoid loosening or failure was the indication for revision in the polyethylene group, whereas polyethylene wear with metal-on-metal contact, instability, and insufficiency of the rotator cuff were the indications for revision in the metal-backed group. Preoperative posterior subluxation of the humeral head with a biconcave/retroverted glenoid (Walch B2) had an adverse effect on the survival of a metal-backed component. Conclusion. The survival of a cemented polyethylene glenoid component is three times higher than that of a cementless metal-backed glenoid component ten years after aTSA in patients aged < 60 years with primary glenohumeral OA. Patients with a biconcave (B2) glenoid have the highest risk of failure. Cite this article: Bone Joint J 2018;100-B:485–92


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 482 - 491
1 May 2024
Davies A Sabharwal S Liddle AD Zamora Talaya MB Rangan A Reilly P

Aims

Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey.

Methods

NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 182 - 188
1 Feb 2024
Gallego JA Rotman D Watts AC

Aims

Acute and chronic injuries of the interosseus membrane can result in longitudinal instability of the forearm. Reconstruction of the central band of the interosseus membrane can help to restore biomechanical stability. Different methods have been used to reconstruct the central band, including tendon grafts, bone-ligament-bone grafts, and synthetic grafts. This Idea, Development, Exploration, Assessment, and Long-term (IDEAL) phase 1 study aims to review the clinical results of reconstruction using a synthetic braided cross-linked graft secured at either end with an Endobutton to restore the force balance between the bones of the forearm.

Methods

An independent retrospective review was conducted of a consecutive series of 21 patients with longitudinal instability injuries treated with anatomical central band reconstruction between February 2011 and July 2019. Patients with less than 12 months’ follow-up or who were treated acutely were excluded, leaving 18 patients in total. Preoperative clinical and radiological assessments were compared with prospectively gathered data using range of motion and the abbreviated version of the Disabilities of the Arm, Shoulder and Hand questionnaire (QuickDASH) functional outcome score.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 1000 - 1006
1 Sep 2023
Macken AA Haagmans-Suman A Spekenbrink-Spooren A van Noort A van den Bekerom MPJ Eygendaal D Buijze GA

Aims

The current evidence comparing the two most common approaches for reverse total shoulder arthroplasty (rTSA), the deltopectoral and anterosuperior approach, is limited. This study aims to compare the rate of loosening, instability, and implant survival between the two approaches for rTSA using data from the Dutch National Arthroplasty Registry with a minimum follow-up of five years.

Methods

All patients in the registry who underwent a primary rTSA between January 2014 and December 2016 using an anterosuperior or deltopectoral approach were included, with a minimum follow-up of five years. Cox and logistic regression models were used to assess the association between the approach and the implant survival, instability, and glenoid loosening, independent of confounders.


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1189 - 1195
1 Nov 2023
Kim JS Kim SH Kim SC Park JH Kim HG Lee SM Yoo JC

Aims

The aim of this study was to compare the clinical and radiological outcomes of reverse shoulder arthroplasty (RSA) using small and standard baseplates in Asian patients, and to investigate the impact of a mismatch in the sizes of the glenoid and the baseplate on the outcomes.

Methods

This was retrospective analysis of 50 and 33 RSAs using a standard (33.8 mm, ST group) and a small (29.5 mm, SM group) baseplate of the Equinoxe reverse shoulder system, which were undertaken between January 2017 and March 2021. Radiological evaluations included the size of the glenoid, the β-angle, the inclination of the glenoid component, inferior overhang, scapular notching, the location of the central cage in the baseplate within the vault and the mismatch in size between the glenoid and baseplate. Clinical evaluations included the range of motion (ROM) and functional scores. In subgroup analysis, comparisons were performed between those in whom the vault of the glenoid was perforated (VP group) and those in whom it was not perforated (VNP group).


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 76 - 82
1 Jan 2022
ten Brinke B Hesseling B Eygendaal D Hoelen MA Mathijssen NMC

Aims

Stemless humeral implants have been developed to overcome stem-related complications in total shoulder arthroplasty (TSA). However, stemless implant designs may hypothetically result in less stable initial fixation, potentially affecting long-term survival. The aim of this study is to investigate early fixation and migration patterns of the stemless humeral component of the Simpliciti Shoulder System and to evaluate clinical outcomes.

Methods

In this prospective cohort study, radiostereometric analysis (RSA) radiographs were obtained in 24 patients at one day, six weeks, six months, one year, and two years postoperatively. Migration was calculated using model-based RSA. Clinical outcomes were evaluated using the visual analogue scale (VAS), the Oxford Shoulder Score (OSS), the Constant-Murley Score (CMS), and the Disabilities of the Arm, Shoulder and Hand (DASH) score.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1292 - 1300
1 Jul 2021
Märtens N Heinze M Awiszus F Bertrand J Lohmann CH Berth A

Aims

The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up.

Methods

A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1619 - 1626
1 Oct 2021
Bi M Zhou K Gan K Ding W Zhang T Ding S Li J

Aims

The aim of this study is to provide a detailed description of cases combining bridging patch repair with artificial ligament “internal brace” reinforcement to treat irreparable massive rotator cuff tears, and report the preliminary results.

Methods

This is a retrospective review of patients with irreparable massive rotator cuff tears undergoing fascia lata autograft bridging repair with artificial ligament “internal brace” reinforcement technique between January 2017 and May 2018. Inclusion criteria were: patients treated arthroscopically for an incompletely reparable massive rotator cuff tear (dimension > 5 cm or two tendons fully torn), stage 0 to 4 supraspinatus fatty degeneration on MRI according to the Goutallier grading system, and an intact or reparable infraspinatus and/or subscapularis tendon of radiological classification Hamada 0 to 4. The surgical technique comprised two components: first, superior capsular reconstruction using an artificial ligament as an “internal brace” protective device for a fascia lata patch. The second was fascia lata autograft bridging repair for the torn supraspinatus. In all, 26 patients with a mean age 63.4 years (SD 6.2) were included.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1333 - 1338
2 Aug 2021
Kankanalu P Borton ZM Morgan ML Cresswell T Espag MP Tambe AA Clark DI

Aims

Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes.

Methods

All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 360 - 365
1 Feb 2021
Mahendraraj KA Shields MV Grubhofer F Golenbock SW Jawa A

Aims

Existing literature indicates that inferiorly inclined glenoid baseplates following reverse total shoulder arthroplasty (RSA) produce better outcomes compared to superiorly inclined baseplates. We aim to compare clinical outcomes for RSAs with superiorly and neutrally/inferiorly inclined lateralized glenospheres.

Methods

We retrospectively reviewed 154 consecutive patients undergoing RSA between July 2015 and July 2017 by one single-fellowship trained surgeon (AJ). Two raters (KAM and MVS) independently measured glenoid inclination in preoperative and minimum two year follow-up radiographs (anteroposterior/Grashey) using the RSA angle. Inclination was then compared to patient-reported outcomes, range of motion (ROM), and independently assessed degree of scapular notching and staging of heterotopic ossification at two year follow-up.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1107 - 1114
1 Sep 2019
Uy M Wang J Horner NS Bedi A Leroux T Alolabi B Khan M

Aims

The aim of this study was to evaluate the differences in revision and complication rates, functional outcomes, and radiological outcomes between cemented and press-fit humeral stems in primary anatomical total shoulder arthroplasty (TSA).

Materials and Methods

A comprehensive systematic review and meta-analysis was conducted searching for studies that included patients who underwent primary anatomical TSA for primary osteoarthritis or rheumatoid arthritis.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1066 - 1073
1 Aug 2018
Nishida K Hashizume K Nasu Y Ozawa M Fujiwara K Inoue H Ozaki T

Aims

The aim of this study was to report the mid-term clinical outcome of cemented unlinked J-alumina ceramic elbow (JACE) arthroplasties when used in patients with rheumatoid arthritis (RA).

Patients and Methods

We retrospectively reviewed 87 elbows, in 75 patients with RA, which was replaced using a cemented JACE total elbow arthroplasty (TEA) between August 2003 and December 2012, with a follow-up of 96%. There were 72 women and three men, with a mean age of 62 years (35 to 79). The mean follow-up was nine years (2 to 14). The clinical condition of each elbow before and after surgery was assessed using the Mayo Elbow Performance Index (MEPI, 0 to 100 points). Radiographic loosening was defined as a progressive radiolucent line of >1 mm that was completely circumferential around the prosthesis.


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1493 - 1498
1 Nov 2018
Wagner ER Hevesi M Houdek MT Cofield RH Sperling JW Sanchez-Sotelo J

Aims

Patients with a failed reverse shoulder arthroplasty (RSA) have limited salvage options. The aim of this study was to determine the outcome of revision RSA when used as a salvage procedure for a failed primary RSA.

Patients and Methods

We reviewed all revision RSAs performed for a failed primary RSA between 2006 and 2012, excluding patients with a follow-up of less than two years. A total of 27 revision RSAs were included in the study. The mean age of the patients at the time of revision was 70 years (58 to 82). Of the 27 patients, 14 (52% were female). The mean follow-up was 4.4 years (2 to 10).


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 461 - 469
1 Apr 2019
Lädermann A Schwitzguebel AJ Edwards TB Godeneche A Favard L Walch G Sirveaux F Boileau P Gerber C

Aims

The aim of this study was to report the outcomes of different treatment options for glenoid loosening following reverse shoulder arthroplasty (RSA) at a minimum follow-up of two years.

Patients and Methods

We retrospectively studied the records of 79 patients (19 men, 60 women; 84 shoulders) aged 70.4 years (21 to 87) treated for aseptic loosening of the glenosphere following RSA. Clinical evaluation included pre- and post-treatment active anterior elevation (AAE), external rotation, and Constant score.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1074 - 1079
1 Aug 2018
Paul R Knowles N Chaoui J Gauci M Ferreira L Walch G Athwal GS

Aims

The Walch Type C dysplastic glenoid is characterized by excessive retroversion. This anatomical study describes its morphology.

Patients and Methods

A total of 29 shoulders with a dysplastic glenoid were analyzed. CT was used to measure retroversion, inclination, height, width, radius-of-curvature, surface area, depth, subluxation of the humeral head and the Goutallier classification of fatty infiltration. The severity of dysplasia and deficiency of the posterior rim of the glenoid were recorded.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1609 - 1617
1 Dec 2018
Malhas AM Granville-Chapman J Robinson PM Brookes-Fazakerley S Walton M Monga P Bale S Trail I

Aims

We present our experience of using a metal-backed prosthesis and autologous bone graft to treat gross glenoid bone deficiency.

Patients and Methods

A prospective cohort study of the first 45 shoulder arthroplasties using the SMR Axioma Trabecular Titanium (TT) metal-backed glenoid with autologous bone graft. Between May 2013 and December 2014, 45 shoulder arthroplasties were carried out in 44 patients with a mean age of 64 years (35 to 89). The indications were 23 complex primary arthroplasties, 12 to revise a hemiarthroplasty or resurfacing, five for aseptic loosening of the glenoid, and five for infection.


The Bone & Joint Journal
Vol. 100-B, Issue 9 | Pages 1182 - 1186
1 Sep 2018
Werner BS Chaoui J Walch G

Aims

Scapular notching is a frequently observed radiographic phenomenon in reverse shoulder arthroplasty (RSA), signifying impingement of components. The purposes of this study were to evaluate the effect of glenoid component size and glenosphere type on impingement-free range of movement (ROM) for extension and internal and external rotation in a virtual RSA model, and to determine the optimal configuration to reduce the incidence of friction-type scapular notching.

Materials and Methods

Preoperative CT scans obtained in 21 patients (three male, 18 female) with primary osteoarthritis were analyzed using modelling software. Two concurrent factors were tested for impingement-free ROM and translation of the centre of rotation: glenosphere diameter (36 mm vs 39 mm) and type (centred, 2 mm inferior eccentric offset, 10° inferior tilt).


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1086 - 1092
1 Aug 2016
de Vos MJ Wagener ML Hannink G van der Pluijm M Verdonschot N Eygendaal D

Aims

Revision total elbow arthroplasty (TEA) is often challenging. The aim of this study was to report on the clinical and radiological results of revision arthroplasty of the elbow with the Latitude TEA.

Patients and Methods

Between 2006 and 2010 we used the Latitude TEA for revision in 18 consecutive elbows (17 patients); mean age 53 years (28 to 80); 14 women. A Kudo TEA was revised in 15 elbows and a Souter-Strathclyde TEA in three.

Stability, range of movement (ROM), visual analogue score (VAS) for pain and functional scores, Elbow Functional Assessment Scale (EFAS), the Functional Rating Index of Broberg and Morrey (FRIBM) and the Modified Andrews’ Elbow Scoring System (MAESS) were assessed pre-operatively and at each post-operative follow-up visit (six, 12 months and biennially thereafter). Radiographs were analysed for loosening, fractures and dislocation. The mean follow-up was 59 months (26 to 89).


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 969 - 975
1 Jul 2016
Theivendran K Varghese M Large R Bateman M Morgan M Tambe A Espag M Cresswell T Clark DI

Aim

We present the medium-term clinical results of a reverse total shoulder arthroplasty with a trabecular metal glenoid base plate.

Patients and Methods

We reviewed 125 consecutive primary reverse total shoulder arthroplasties (RTSA) implanted in 124 patients for rotator cuff arthropathy. There were 100 women and 24 men in the study group with a mean age of 76 years (58 to 89). The mean follow-up was 32 months (24 to 60). No patient was lost to follow-up.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 939 - 943
1 Jul 2017
Sowa B Bochenek M Bülhoff M Zeifang F Loew M Bruckner T Raiss P

Aims

Promising medium-term results from total shoulder arthroplasty (TSA) have been reported for the treatment of primary osteoarthritis in young and middle-aged patients. The aim of this study was to evaluate the long-term functional and radiological outcome of TSA in the middle-aged patient.

Patients and Methods

The data of all patients from the previous medium-term study were available. At a mean follow-up of 13 years (8 to 17), we reviewed 21 patients (12 men, nine women, 21 shoulders) with a mean age of 55 years (37 to 60). The Constant-Murley score (CS) with its subgroups and subjective satisfaction were measured. Radiological signs of implant loosening were analysed.