Advertisement for orthosearch.org.uk
Results 1 - 12 of 12
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 14 - 18
1 Nov 2012
Lombardi, Jr AV Barrack RL Berend KR Cuckler JM Jacobs JJ Mont MA Schmalzried TP

Since 1996 more than one million metal-on-metal articulations have been implanted worldwide. Adverse reactions to metal debris are escalating. Here we present an algorithmic approach to patient management. The general approach to all arthroplasty patients returning for follow-up begins with a detailed history, querying for pain, discomfort or compromise of function. Symptomatic patients should be evaluated for intra-articular and extra-articular causes of pain. In large head MoM arthroplasty, aseptic loosening may be the source of pain and is frequently difficult to diagnose. Sepsis should be ruled out as a source of pain. Plain radiographs are evaluated to rule out loosening and osteolysis, and assess component position. Laboratory evaluation commences with erythrocyte sedimentation rate and C-reactive protein, which may be elevated. Serum metal ions should be assessed by an approved facility. Aspiration, with manual cell count and culture/sensitivity should be performed, with cloudy to creamy fluid with predominance of monocytes often indicative of failure. Imaging should include ultrasound or metal artifact reduction sequence MRI, specifically evaluating for fluid collections and/or masses about the hip. If adverse reaction to metal debris is suspected then revision to metal or ceramic-on-polyethylene is indicated and can be successful. Delay may be associated with extensive soft-tissue damage and hence poor clinical outcome


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 11 - 13
1 Nov 2012
Cuckler JM

Hip implant retrieval analysis is the most important source of insight into the performance of new materials and designs of hip arthroplasties. Even the most rigorous in vitro testing will not accurately simulate the behavior of implant materials and new designs of prosthetic arthroplasties. Retrieval analysis has revealed such factors as the effects of gamma-in-air sterilisation of polyethylene, fatigue failure mechanisms of polymethylmethacrylate bone cement, fretting corrosion of Morse taper junctions, third body wear effects of both hard-on-hard and hard-on-soft bearing couples, and the effects of impingement of components on the full spectrum of bearing surfaces, none of which was predicted by pre-implantation in vitro testing of these materials and combinations. The temporal sequence of the retrieval process is approximately six years from first implantation through retrieval analysis, laboratory investigation, and publication of results, and thus, in addition to rigorous clinical evaluation, represents the true development and insight cycle for new designs and materials.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1409 - 1418
1 Nov 2006
Scheerlinck T Casteleyn P

We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from in vitro studies and finite element modelling were examined with reference to the survivorship results for each of the concepts of fixation.

It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 265 - 271
1 Mar 2008
Mandalia V Eyres K Schranz P Toms AD

Evaluation of patients with painful total knee replacement requires a thorough clinical examination and relevant investigations in order to reach a diagnosis. Awareness of the common and uncommon problems leading to painful total knee replacement is useful in the diagnostic approach. This review article aims to act as a guide to the evaluation of patients with painful total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 405 - 410
1 Apr 2008
Dattani R Patnaik S Kantak A Srikanth B Selvan TP

The management of injury to the distal tibiofibular syndesmosis remains controversial in the treatment of ankle fractures. Operative fixation usually involves the insertion of a metallic diastasis screw. There are a variety of options for the position and characterisation of the screw, the type of cortical fixation, and whether the screw should be removed prior to weight-bearing. This paper reviews the relevant anatomy, the clinical and radiological diagnosis and the mechanism of trauma and alternative methods of treatment for injuries to the syndesmosis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 567 - 573
1 May 2007
Keegan GM Learmonth ID Case CP

The long-term effects of metal-on-metal arthroplasty are currently under scrutiny because of the potential biological effects of metal wear debris. This review summarises data describing the release, dissemination, uptake, biological activity, and potential toxicity of metal wear debris released from alloys currently used in modern orthopaedics. The introduction of risk assessment for the evaluation of metal alloys and their use in arthroplasty patients is discussed and this should include potential harmful effects on immunity, reproduction, the kidney, developmental toxicity, the nervous system and carcinogenesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 469 - 476
1 Apr 2010
Shimmin AJ Walter WL Esposito C

The survivorship of contemporary resurfacing arthroplasty of the hip using metal-on-metal bearings is better than that of first generation designs, but short-term failures still occur. The most common reasons for failure are fracture of the femoral neck, loosening of the component, osteonecrosis of the femoral head, reaction to metal debris and malpositioning of the component. In 2008 the Australian National Joint Registry reported an inverse relationship between the size of the head component and the risk of revision in resurfacing hip arthroplasty. Hips with a femoral component size of ≤ 44 mm have a fivefold increased risk of revision than those with femoral components of ≥ 55 mm irrespective of gender. We have reviewed the literature to explore this observation and to identify possible reasons including the design of the implant, loading of the femoral neck, the orientation of the component, the production of wear debris and the effects of metal ions, penetration of cement and vascularity of the femoral head. Our conclusion is that although multifactorial, the most important contributors to failure in resurfacing arthroplasty of the hip are likely to be the design and geometry of the component and the orientation of the acetabular component.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 1 - 7
1 Jan 2009
Robinson AHN Pasapula C Brodsky JW

A comprehensive review of the literature relating to the pathology and management of the diabetic foot is presented. This should provide a guide for the treatment of ulcers, Charcot neuro-arthropathy and fractures involving the foot and ankle in diabetic patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 851 - 857
1 Jul 2007
Webb JCJ Spencer RF

Polymethylmethacrylate remains one of the most enduring materials in orthopaedic surgery. It has a central role in the success of total joint replacement and is also used in newer techniques such as percutaneous vertebroplasty and kyphoplasty.

This article describes the current uses and limitations of polymethylmethacrylate in orthopaedic surgery. It focuses on its mechanical and chemical properties and links these to its clinical performance. The behaviour of antibiotic-loaded bone cement are discussed, together with areas of research that are now shedding light upon the behaviour of this unique biomaterial.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1171 - 1177
1 Sep 2005
Trieb K


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1029 - 1037
1 Aug 2005
Mayer HM


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 562 - 575
1 May 2006
Boileau P Sinnerton RJ Chuinard C Walch G