Aims. To investigate
Aims. We present a case series of five patients who had revision surgery
following magnetic controlled growing rods (MGCR) for early onset
scoliosis. Metallosis was found during revision in four out of five
patients and we postulated a mechanism for rod failure based on
retrieval analysis. Patients and Methods. Retrieval analysis was performed on the seven explanted rods.
The mean duration of MCGR from implantation to revision was 35 months
(17 to 46). The mean age at revision was 12 years (7 to 15; four
boys, one girl). Results. A total of six out of seven rods had tissue
The aim of this study was to compare the clinical and radiological outcomes of patients with early-onset scoliosis (EOS), who had undergone spinal fusion after distraction-based spinal growth modulation using either traditional growing rods (TGRs) or magnetically controlled growing rods (MCGRs). We undertook a retrospective review of skeletally mature patients who had undergone fusion for an EOS, which had been previously treated using either TGRs or MCGRs. Measured outcomes included sequential coronal T1 to S1 height and major curve (Cobb) angle on plain radiographs and any complications requiring unplanned surgery before final fusion.Aims
Methods
To report the mid-term results of a modified self-growing rod (SGR) technique for the treatment of idiopathic and neuromuscular early-onset scoliosis (EOS). We carried out a retrospective analysis of 16 consecutive patients with EOS treated with an SGR construct at a single hospital between September 2008 and December 2014. General demographics and deformity variables (i.e. major Cobb angle, T1 to T12 length, T1 to S1 length, pelvic obliquity, shoulder obliquity, and C7 plumb line) were recorded preoperatively, and postoperatively at yearly follow-up. Complications and revision procedures were also recorded. Only patients with a minimum follow-up of five years after surgery were included.Aims
Methods
Magnetically controlled growing rod (MCGR) systems use non-invasive
spinal lengthening for the surgical treatment of early-onset scoliosis
(EOS). The primary aim of this study was to evaluate the performance
of these devices in the prevention of progression of the deformity.
A secondary aim was to record the rate of complications. An observational study of 31 consecutive children with EOS, of
whom 15 were male, who were treated between December 2011 and October
2017 was undertaken. Their mean age was 7.7 years (2 to 14). The
mean follow-up was 47 months (24 to 69). Distractions were completed
using the tailgating technique. The primary outcome measure was
correction of the radiographic deformity. Secondary outcomes were
growth, functional outcomes and complication rates.Aims
Patients and Methods
The primary aim of this study was to evaluate the performance
and safety of magnetically controlled growth rods in the treatment
of early onset scoliosis. Secondary aims were to evaluate the clinical
outcome, the rate of further surgery, the rate of complications,
and the durability of correction. We undertook an observational prospective cohort study of children
with early onset scoliosis, who were recruited over a one-year period
and followed up for a minimum of two years. Magnetically controlled
rods were introduced in a standardized manner with distractions
performed three-monthly thereafter. Adverse events which were both related
and unrelated to the device were recorded. Ten children, for whom
relevant key data points (such as demographic information, growth
parameters, Cobb angles, and functional outcomes) were available,
were recruited and followed up over the period of the study. There
were five boys and five girls. Their mean age was 6.2 years (2.5
to 10).Aims