Aims. Following the resection of an extensive amount of bone in the
treatment of a tumour, the residual segment may be insufficient
to accept a standard length intramedullary cemented stem. Short-stemmed
endoprostheses conceivably have an increased risk of aseptic
Aims. Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. Methods. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively. Results. Overall, 21 patients (16%) encountered complications, and the rate of aseptic
Aims. The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases. Methods. This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for
Aims. The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection. Methods. From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients. Results. Preparation of the prosthesis was a mean of 8.0 days (SD 1.5), during which time no patients experienced tumour progression. The mean operating time and intraoperative blood loss were 158.1 minutes (SD 67.6) and 176.9 ml (SD 187.8), respectively. All of the prostheses were implanted successfully. During a mean follow-up of 25.7 months (SD 7.8), no patients died, but four had complications (two superficial wound problems, one temporary palsy of radial nerve, and one dislocation). No aseptic
Aims. Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. Methods. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation. Results. Multivariate Fine-Gray regression analyses revealed different hazards for each Henderson failure mode: percentage of femoral resection (p = 0.001) and extent of quadriceps muscle resection (p = 0.005) for overall prosthetic failure; extent of quadriceps muscle resection (p = 0.002) and fixation of femoral component (p = 0.011) for type 2 failure (aseptic loosening); age (p = 0.009) and percentage of femoral resection (p = 0.019) for type 3 failure (mechanical failure); and type of joint resection (p = 0.037) for type 4 (infection) were independent predictors. A bone stem ratio of > 2.5 reliably predicted aseptic
Aims. Aseptic
Aims. The aim of this study was to present the long-term surgical outcomes, complications, implant survival, and causes of implant failure in patients treated with the modified Harrington procedure using antegrade large diameter pins. Patients and Methods. A cohort of 50 consecutive patients who underwent the modified Harrington procedure for periacetabular metastasis or haematological malignancy between January 1996 and April 2018 were studied. The median follow-up time for all survivors was 3.2 years (interquartile range 0.9 to 7.6 years). Results. The five-year overall survival rate was 33% for all the patients. However, implant survival rates were 100% and 46% at five and ten years, respectively. Eight patients survived beyond five years. There was no immediate perioperative mortality or complications. A total of 15 late complications occurred in 11 patients (22%). Five patients (10%) required further surgery to treat complications. The most frequent complication was pin breakage without evidence of acetabular
Osteoarticular reconstruction of the distal femur in childhood has the advantage of preserving the tibial physis. However, due to the small size of the distal femur, matching the host bone with an osteoarticular allograft is challenging. In this study, we compared the outcomes and complications of a resurfaced allograft-prosthesis composite (rAPC) with those of an osteoarticular allograft to reconstruct the distal femur in children. A retrospective analysis of 33 skeletally immature children with a malignant tumour of the distal femur, who underwent resection and reconstruction with a rAPC (n = 15) or osteoarticular allograft (n = 18), was conducted. The median age of the patients was ten years (interquartile range (IQR) 9 to 11) in the osteoarticular allograft group and nine years (IQR 8 to 10) in the rAPC group (p = 0.781). The median follow-up of the patients was seven years (IQR 4 to 8) in the osteoarticular allograft group and six years (IQR 3 to 7) in the rAPC group (p = 0.483). Limb function was evaluated using the Musculoskeletal Tumor Society (MSTS) score.Aims
Methods
Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation. A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.Aims
Methods
Aims. The purpose of this study was to report the long-term results of extendable endoprostheses of the humerus in children after the resection of a bone sarcoma. Methods. A total of 35 consecutive patients treated with extendable endoprosthetic replacement of the humerus in children were included. There were 17 boys and 18 girls in the series with a median age at the time of initial surgery of nine years (interquartile range (IQR) 7 to 11). Results. The median follow-up time was 10.6 years (IQR 3.9 to 20.4). The overall implant survival at ten years was 75%. Complications occurred in 13 patients (37%). Subluxation at the proximal humerus occurred in 19 patients (54%) but only six (17%) were symptomatic. Subluxation was seen more commonly in children under the age of nine years (86%) than in those aged nine years or more (33%) (p = 0.002). Implant failure occurred in nine patients (26%): the most common cause was aseptic
Aims. The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children. Patients and Methods. Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one. Results. The five-year overall survival rate was 78%. No patient developed local recurrence. The five-year implant survival rate was 79%. In all, 11 patients (61%) developed a complication. Seven patients (39%) required further surgery to treat the complications. Implant failures occurred in three patients (17%) including one patient with aseptic
Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).Aims
Methods
Aims. The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR). Patients and Methods. In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors. Results. The revision-free survival of the implant was 71% at five years and 63.3% at ten years. Three prostheses reached 15 years without revision. The mean Musculoskeletal Tumor Society score in the group was 26 (23 to 28). The mechanisms of failure were infection in 18%, structural failures in 6%, tumour progression in 5%, aseptic
We have previously reported cryoablation-assisted joint-sparing surgery for osteosarcoma with epiphyseal involvement. However, it is not clear whether this is a comparable alternative to conventional joint arthroplasty in terms of oncological and functional outcomes. A total of 22 patients who had localized osteosarcoma with epiphyseal involvement around the knee and underwent limb salvage surgery were allocated to joint preservation (JP) group and joint arthroplasty (JA) group. Subjects were followed with radiographs, Musculoskeletal Tumor Society (MSTS) score, and clinical evaluations at one, three, and five years postoperatively.Aims
Methods
Limb salvage for pelvic sarcomas involving the acetabulum is a major surgical challenge. There remains no consensus about what is the optimum type of reconstruction after resection of the tumour. The aim of this study was to evaluate the surgical outcomes in these patients according to the methods of periacetabular reconstruction. The study involved a consecutive series of 122 patients with a periacetabular bone sarcoma who underwent limb-salvage surgery involving a custom-made prosthesis in 65 (53%), an ice-cream cone prosthesis in 21 (17%), an extracorporeal irradiated autograft in 18 (15%), and nonskeletal reconstruction in 18 (15%).Aims
Methods
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
Hip reconstruction after resection of a periacetabular chondrosarcoma is complex and associated with a high rate of complications. Previous reports have compared no reconstruction with historical techniques that are no longer used. The aim of this study was to compare the results of tantalum acetabular reconstruction to both historical techniques and no reconstruction. We reviewed 66 patients (45 males and 21 females) with a mean age of 53 years (24 to 81) who had undergone acetabular resection for chondrosarcoma. A total of 36 patients (54%) underwent acetabular reconstruction, most commonly with a saddle prosthesis (n = 13; 36%) or a tantalum total hip arthroplasty (THA) (n = 10; 28%). Mean follow-up was nine years (SD 4).Aims
Methods
Aims. Pelvic reconstruction after the resection of a tumour around
the acetabulum is a challenging procedure due to the complex anatomy
and biomechanics. Several pelvic endoprostheses have been introduced,
but the rates of complication remain high. Our aim was to review
the use of a stemmed acetabular pedestal cup in the management of
these patients. Patients and Methods. The study involved 48 patients who underwent periacetabular reconstruction
using a stemmed pedestal cup (Schoellner cup; Zimmer Biomet Inc.,
Warsaw, Indiana) between 2000 and 2013. The indications for treatment included
a primary bone tumour in 27 patients and metastatic disease in 21
patients. The mean age of the patients at the time of surgery was
52 years (16 to 83). Results. At a median follow-up of 6.6 years (95% confidence interval 4.6
to 8.2), local control was achieved in all patients; 19 patients
had died (16 of disease). Complications occurred in 19 patients
(40%), of which deep infection was the most common, affecting eight
patients (17%). Seven patients (15%) had a dislocation of the hip.
Aseptic
We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database. A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic
We determined the efficacy of a devitalised autograft
(n = 13) and allograft (n = 16) cortical strut bone graft combined
with long-stem endoprosthetic reconstruction in the treatment of
massive tumours of the lower limb. A total of 29 patients (18 men:11
women, mean age 20.1 years (12 to 45) with a ratio of length of
resection to that of the whole prosthesis of >
50% were treated
between May 2003 and May 2012. The mean follow-up was 47 months
(15 to 132). The stem of the prosthesis was introduced through bone
graft struts filled with cement, then cemented into the residual
bone. Bone healing was achieved in 23 patients (86%). The mean Musculoskeletal
Tumour Society functional score was 85% (57 to 97). The five-year
survival rate of the endoprostheses was 81% (95% confidence intervals
67.3 to 92.3). The mean length of devitalised autografts and allografts
was 8.6 cm (5 to 15), which increased the ratio of the the length
of the stem of the prosthesis to that of the whole length of the
prosthesis from a theoretical 35% to an actual 55%. . Cortical strut bone grafting and long-stem endoprosthetic reconstruction
is an option for treating massive segmental defects following resection
of a tumour in the lower limb. Patients can regain good function
with a low incidence of aseptic