Advertisement for orthosearch.org.uk
Results 1 - 20 of 44
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 699 - 708
1 Jun 2020
Nieuwenhuijse MJ Vehmeijer SBW Mathijsen NMC Keizer SB

Aims. Short, bone-conserving femoral components are increasingly used in total hip arthroplasty (THA). They are expected to allow tissue-conserving implantation and to render future revision surgery more straightforward but the long-term data on such components is limited. One such component is the global tissue-sparing (GTS) stem. Following the model for stepwise introduction of new orthopaedic implants, we evaluated early implant fixation and clinical outcome of this novel short-stem THA and compared it to that of a component with established good long-term clinical outcome. Methods. In total, 50 consecutive patients ≤ 70 years old with end-stage symptomatic osteo-arthritis were randomized to receive THA with the GTS stem or the conventional Taperloc stem using the anterior supine intermuscular approach by two experienced hip surgeons in two hospitals in the Netherlands. Primary outcome was implant migration. Patients were followed using routine clinical examination, patient reported outcome using Harris Hip Score (HHS), Hip Disability And Osteoarthritis Outcome Score (HOOS), EuroQol five-dimension questionnaire (EQ5D), and Roentgen Stereophotogrammetric Analysis (RSA) at three, six, 12, and 24 months. This study evaluated the two-year follow-up results. Results. In addition to the initial migration pattern of distal migration (subsidence, Y-translation) and retroversion (Y-rotation) also exhibited by the Taperloc stem, the GTS stem showed an initial migration pattern of varization (X-translation combined with Z-rotation) and posterior translation (Z-translation). However, all components stabilized aside from one Taperloc stem which became loose secondary to malposition and was later revised. Clinical outcomes and complications were not statistically significantly different with the numbers available. Conclusion. A substantially different and more extensive initial migration pattern was seen for the GTS stem compared to the Taperloc stem. Although implant stabilization was achieved, excellent long-term survival similar to that of the Taperloc stem should not be inferred. Especially in the absence of clinically proven relevant improvement, widespread usage should be postponed until long-term safety has been established. Cite this article: Bone Joint J 2020;102-B(6):699–708


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 768 - 774
1 Jul 2023
Wooster BM Kennedy NI Dugdale EM Sierra RJ Perry KI Berry DJ Abdel MP

Aims

Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis.

Methods

We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 359 - 364
1 Apr 2024
Özdemir E de Lange B Buckens CFM Rijnen WHC Visser J

Aims

To investigate the extent of bone development around the scaffold of custom triflange acetabular components (CTACs) over time.

Methods

We performed a single-centre historical prospective cohort study, including all patients with revision THA using the aMace CTAC between January 2017 and March 2021. A total of 18 patients (18 CTACs) were included. Models of the hemipelvis and the scaffold component of the CTACs were created by segmentation of CT scans. The CT scans were performed immediately postoperatively and at least one year after surgery. The amount of bone in contact with the scaffold was analyzed at both times, and the difference was calculated.


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1045 - 1051
1 Oct 2023
Turgeon TR Righolt CH Burnell CD Gascoyne TC Hedden DR Bohm ER

Aims

The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components.

Methods

Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1032 - 1038
1 Sep 2022
García-Rey E Cruz-Pardos A Saldaña L

Aims

A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade.

Methods

Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 647 - 656
1 Jun 2022
Knudsen MB Thillemann JK Jørgensen PB Jakobsen SS Daugaard H Søballe K Stilling M

Aims

BoneMaster is a thin electrochemically applied hydroxyapatite (HA) coating for orthopaedic implants that is quickly resorbed during osseointegration. Early stabilization is a surrogacy marker of good survival of femoral stems. The hypothesis of this study was that a BoneMaster coating yields a fast early and lasting fixation of stems.

Methods

A total of 53 patients were randomized to be treated using Bi-Metric cementless femoral stems with either only a porous titanium plasma-sprayed coating (P group) or a porous titanium plasma-sprayed coating with an additional BoneMaster coating (PBM group). The patients were examined with radiostereometry until five years after surgery.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 206 - 211
1 Feb 2022
Bloch BV White JJE Matar HE Berber R Manktelow ARJ

Aims

Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost.

Methods

In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 221 - 226
1 Feb 2022
Edwards NM Varnum C Nelissen RGHH Overgaard S Pedersen AB

Aims

The aim of this study was to examine whether socioeconomic status (SES) is associated with a higher risk of infections following total hip arthroplasty (THA) at 30 and 90 days.

Methods

We obtained individual-based information on SES markers (cohabitation, education, income, and savings) on 103,901 THA patients from Danish health registries between 1 January 1995 and 31 December 2017. The primary outcome measure was any hospital-treated infection (i.e. all infections). The secondary outcomes were further specified to specific hospital-treated infections (pneumonia, urinary tract infection, and periprosthetic joint infection). The primary timepoint was within 90 days. In addition, the outcomes were further evaluated within 30 days. We calculated the cumulative incidence, and used the pseudo-observation method and generalized linear regression to estimate adjusted risk ratios (RRs) with 95% confidence intervals (CIs) for each marker.


Aims. To investigate the longevity of uncemented fixation of a femoral component in total hip arthroplasty (THA) in patients with Dorr type C proximal femoral morphology. Patients and Methods. A total of 350 consecutive uncemented THA in 320 patients were performed between 1983 and 1987, by a single surgeon using the Taperloc femoral component. The 63 patients (68 hips) with Dorr type C proximal femoral morphology were the focus of this review. The mean age of the patients was 69 years (24 to 88) and mean follow-up was 16.6 years (ten to 29). Survival analysis included eight patients (eight hips) who died without undergoing revision surgery prior to obtaining ten years follow-up. All 55 surviving patients (60 hips) were available for clinical assessment and radiographic review. As a comparator group, the survival and implant fixation in the remaining 282 THAs (257 patients) with Dorr type A and B morphology were evaluated. The mean age of these patients was 52 years (20 to 82). Results. In the Dorr C patient group the mean Harris hip score improved from 51 points (21 to 69 points) pre-operatively to 89 (74 to 100) at final follow-up. No femoral component was loose by radiological criteria and osteolysis was only identified around two stems (3.3%). There was one revision (1.6%) of a well-fixed femoral component for sepsis at 11 years. The survival of the Taperloc femoral component at 20 years with revision for any reason as the endpoint was 98% (95% confidence interval; 87 to 99). A total of ten (3.5%) of the Dorr A and B patient group of 282 THAs required revision surgery. Only one (0.4%) for aseptic loosening. A total of two hips (1%) were loose by radiographic criteria and osteolysis occurred around 12 hips (4%). Conclusion. This study demonstrates that excellent fixation can be achieved using the Taperloc stem in patients with Dorr type A and B, and Dorr type C bone. Take home message: The Taperloc stem demonstrated equivalent results in Dorr type A and B and Dorr type C bone. Cite this article: Bone Joint J 2016;98-B:595–600


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1184 - 1191
1 Sep 2013
Gordon M Stark A Sköldenberg OG Kärrholm J Garellick G

While an increasing amount of arthroplasty articles report comorbidity measures, none have been validated for outcomes. In this study, we compared commonly used International Classification of Diseases-based comorbidity measures with re-operation rates after total hip replacement (THR). Scores used included the Charlson, the Royal College of Surgeons Charlson, and the Elixhauser comorbidity score. We identified a nationwide cohort of 134 423 THRs from the Swedish Hip Arthroplasty Register. Re-operations were registered post-operatively for up to 12 years. The hazard ratio was estimated by Cox’s proportional hazards regression, and we used C-statistics to assess each measure’s ability to predict re-operation. Confounding variables were age, gender, type of implant fixation, hospital category, hospital implant volume and year of surgery. In the first two years only the Elixhauser score showed any significant relationship with increased risk of re-operation, with increased scores for both one to two and three or more comorbidities. However, the predictive C-statistic in this period for the Elixhauser score was poor (0.52). None of the measures proved to be of any value between two and 12 years. They might be of value in large cohort or registry studies, but not for the individual patient. Cite this article: Bone Joint J 2013;95-B:1184–91


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims

The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes.

Methods

A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 486 - 491
1 Mar 2021
Di Martino A Bordini B Ancarani C Viceconti M Faldini C

Aims

Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017).

Methods

A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 650 - 658
1 Apr 2021
Konow T Baetz J Melsheimer O Grimberg A Morlock M

Aims

Periprosthetic femoral fractures (PPF) are a serious complication of total hip arthroplasty (THA) and are becoming an increasingly common indication for revision arthroplasty with the ageing population. This study aimed to identify potential risk factors for PPF based on an analysis of registry data.

Methods

Cases recorded with PPF as the primary indication for revision arthroplasty in the German Arthroplasty Registry (Endoprothesenregister Deutschland (EPRD)), as well as those classified as having a PPF according to the International Classification of Diseases (ICD) codes in patients’ insurance records were identified from the complete datasets of 249,639 registered primary hip arthroplasties in the EPRD and included in the analysis.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 305 - 308
1 Feb 2021
Howell M Rae FJ Khan A Holt G

Aims

Iliopsoas pathology is a relatively uncommon cause of pain following total hip arthroplasty (THA), typically presenting with symptoms of groin pain on active flexion and/or extension of the hip. A variety of conservative and surgical treatment options have been reported. In this retrospective cohort study, we report the incidence of iliopsoas pathology and treatment outcomes.

Methods

A retrospective review of 1,000 patients who underwent THA over a five-year period was conducted, to determine the incidence of patients diagnosed with iliopsoas pathology. Outcome following non-surgical and surgical management was assessed.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 616 - 621
1 May 2011
O’Neill F Condon F McGloughlin T Lenehan B Coffey JC Walsh M

We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1662 - 1669
1 Dec 2020
Pollmann CT Gjertsen J Dale H Straume-Næsheim TM Dybvik E Hallan G

Aims

To compare the functional outcome, health-related quality of life (HRQoL), and satisfaction of patients who underwent primary total hip arthroplasty (THA) and a single debridement, antibiotics and implant retention (DAIR) procedure for deep infection, using either the transgluteal or the posterior surgical approach for both procedures.

Methods

The study was registered at clinicaltrials.gov (ID: NCT03161990) on 15 May 2017. Patients treated with a single DAIR procedure for deep infection through the same operative approach as their primary THA (either the transgluteal or the posterior approach) were identified in the Norwegian Arthroplasty Register and given a questionnaire. Median follow-up after DAIR by questionnaire was 5.5 years in the transgluteal group (n = 87) and 2.5 years in the posterior approach group (n = 102).


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 198 - 204
1 Feb 2020
Perlbach R Palm L Mohaddes M Ivarsson I Schilcher J

Aims

This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an uncemented component in acetabular revisions performed in a standardized manner since 1993.

Methods

Between 1993 and 2013, 370 patients with a median age of 72 years (interquartile range (IQR) 63 to 79 years) underwent acetabular revision surgery. Of these, 229 were more than ten years following surgery and 137 were more than 15 years. All revisions were performed with extensive use of morcellized allograft firmly impacted into the entire acetabular cavity, followed by insertion of an uncemented component with supplementary screw fixation. All types of reoperation were captured using review of radiographs and medical charts, combined with data from the local surgical register and the Swedish Hip Arthroplasty Register.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 414 - 422
1 Apr 2020
Gwynne-Jones DP Gray AR

Aims

To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem.

Methods

Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 62 - 67
1 Jun 2019
Tanzer M Chuang PJ Ngo CG Song L TenHuisen KS

Aims

The purpose of this study was to evaluate the biological fixation of a 3D printed porous implant, with and without different hydroxyapatite (HA) coatings, in a canine model.

Materials and Methods

A canine transcortical model was used to evaluate the characteristics of bone ingrowth of Ti6Al4V cylindrical implants fabricated using laser rapid manufacturing (LRM). At four and 12 weeks post-implantation, we performed histological analysis and mechanical push-out testing on three groups of implants: a HA-free control (LRM), LRM with precipitated HA (LRM-PA), and LRM with plasma-sprayed HA (LRM-PSHA).


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 396 - 402
1 Apr 2019
Ji B Wahafu T Li G Zhang X Wang Y Momin M Cao L

Aims

Single-stage revision is not widely pursued due to restrictive inclusion criteria. In this study, we evaluated the results of single-stage revision of chronically infected total hip arthroplasty (THA) using broad inclusion criteria and cementless implants.

Patients and Methods

Between 2010 and 2016, 126 patients underwent routine single-stage revision with cementless reconstruction with powdered vancomycin or imipenem poured into the medullary cavity and re-implantation of cementless components. For patients with a culture-negative hip, fungal infections, and multidrug-resistant organisms, a direct intra-articular infusion of pathogen-sensitive antibiotics was performed postoperatively. Recurrence of infection and clinical outcomes were evaluated. Three patients died and 12 patients (none with known recurrent infection) were lost to follow-up. There were 111 remaining patients (60 male, 51 female) with a mean age of 58.7 (sd 12.7; 20 to 79).